Dynamic Gas Scavenging System 2 (DGSS - 2)

K193646 · Anesthetic Gas Reclamation, Inc. · CBN · Dec 4, 2020 · Anesthesiology

Device Facts

Record IDK193646
Device NameDynamic Gas Scavenging System 2 (DGSS - 2)
ApplicantAnesthetic Gas Reclamation, Inc.
Product CodeCBN · Anesthesiology
Decision DateDec 4, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5430
Device ClassClass 2
AttributesPediatric

Indications for Use

The Dynamic Gas Scavenging System 2 (DGSS-2) is intended to be used for the scavenging of waste anesthetic gases from anesthesia machines during the provision of general anesthesia to adults and children.

Device Story

The DGSS-2 is a waste anesthetic scavenging interface installed between an anesthesia workstation and a surgical facility's waste gas evacuation vacuum system. It utilizes a pressure transducer to detect waste anesthetic gas accumulation; upon detection, it triggers a solenoid to open the vacuum line, allowing gas exhaust. When no waste gas is present, the system interrupts exhaust flow. The device prevents positive and negative pressure on the patient breathing circuit. It is designed for use with low-flow (<50 L/min) waste gas disposal systems. The system is mounted on the anesthesia machine. The solenoid is normally open, ensuring a 'reversion to safety' full-flow mode during power loss. The device benefits patients by safely managing waste anesthetic gases, reducing environmental exposure in the surgical suite. It is operated by clinical staff in surgical care facilities.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical performance testing, including verification of specifications, materials testing (VOC), electrical safety, and electromagnetic compatibility testing per IEC 60601-1-2:2014 and ISO 80601-2-13.

Technological Characteristics

Waste anesthetic scavenging interface; digital control circuitry; solid-state pressure sensor; co-axial high-flow solenoid; 12 VDC power input with 6 VDC holding voltage; silicone rubber reservoir bag; RoHS compliant lead-free circuit board; external mount form factor; integrated positive/negative pressure relief.

Indications for Use

Indicated for scavenging waste anesthetic gases from anesthesia machines during general anesthesia for adult and pediatric patients.

Regulatory Classification

Identification

A gas-scavenging apparatus is a device intended to collect excess anesthetic, analgesic, or trace gases or vapors from a patient's breathing system, ventilator, or extracorporeal pump-oxygenator, and to conduct these gases out of the area by means of an exhaust system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 4, 2020 Anesthetic Gas Reclamation, Inc. % Laurel Arrigona Regulatory Ceutical Laboratories, Inc. 2300 Valley View Lane Ste. 230 Farmers Branch, Texas 75234 Re: K193646 Trade/Device Name: Dynamic Gas Scavenging System 2 (DGSS - 2) Regulation Number: 21 CFR 868.5430 Regulation Name: Gas-Scavenging Apparatus Regulatory Class: Class II Product Code: CBN Dated: November 6, 2020 Received: November 6, 2020 Dear Laurel Arrigona: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ence determination does not mean that FDA Page 2 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193646 Device Name Dynamic Gas Scavenging System 2 (DGSS - 2) Indications for Use (Describe) The Dynamic Gas Scavenging System is intended to be used for the scavenging of waste anesthesia machines during the provision of general anesthesia to adults and children. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with the letters "ATR" in a stylized font. The letters are green and gray, with a recycling symbol inside the letter "A". The logo has a modern and clean design. The letters are bold and easy to read. Section 7 – 510(k) Summary {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo with the letters 'AGR' in a stylized font. The 'A' is shaped like a triangle with a recycling symbol inside. The letters are colored in a gradient from green to gray. The logo has a modern and eco-friendly appearance. Email: larrigona@ceuticallabs.com Phone Number: 972-241-8374 Fax Number: 972-241-0619 | Date of Submission: | December 23, 2019 | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submission Sponsor: | James Berry, MD<br>Anesthetic Gas Reclamation, Inc.<br>10606 Shady Trail #22<br>Dallas, Texas 75220<br>Email: admin@gasrecycler.com<br>Phone Number: 972-707-7321<br>Fax Number: 972-707-7343 | | Submission Completed By: | Laurel Arrigona<br>Regulatory<br>Ceutical Laboratories, Inc.<br>2300 Valley View Lane, Suite 230<br>Farmers Branch, Texas 75234 | # Device Identification | Trade/Proprietary Name: | Dynamic Gas Scavenging System 2 (DGSS-2) | |----------------------------|------------------------------------------| | Common/Usual Name: | Waste Anesthetic Scavenging Interface | | Classification Name: | Apparatus, Gas-Scavenging | | Classification Regulation: | 868.5430 | | Product Code: | CBN | | Device Class: | Class II | | Classification Panel: | Anesthesiology | # Predicate Device | Trade/Proprietary Name: | Dynamic Gas Scavenging System | |----------------------------|-----------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Waste Anesthetic Scavenging Interface | | Classification Name: | Apparatus, Gas-Scavenging | | Classification Regulation: | 868.5430 | | Product Code: | CBN | | Device Class: | Class II | | Classification Panel: | Anesthesiology | | 510(k) Number: | K063519 | | Manufacturer: | Anesthetic Gas Reclamation, Inc | | Intended Use: | This device is intended to be used for the scavenging of | | | waste anesthetic gases from anesthesia machines during the<br>provision of general anesthesia to adults and children. | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo with the letters "AGR" in a stylized font. The letters are green and gray, with a recycling symbol inside the "A". The logo has a modern and clean design. The letters are bold and easy to read. # Device Description The Dynamic Gas Scavenging System 2 (DGSS-2) is a waste anesthetic scavenging interface placed between the individual anesthetic workstation and the waste gas evacuation vacuum system in a surgical care facility. Through a sensor and solenoid combination, it allows waste gas exhaust flow to the waste gas vacuum line only in the presence of waste anesthetic gas, and interrupts all exhaust flow when no waste gas is present. The system effectively prevents both positive and negative pressure on the patient breathing circuit, and it is usable over a wide range of anesthetic gas flows. It is designed for use in conjunction with low-flow (<50 liters/min) waste gas disposal (WAGD) systems. # Intended Use The Dynamic Gas Scavenging System 2 (DGSS-2) is intended to be used for the scavenging of waste anesthetic gases from anesthesia machines during the provision of general anesthesia to adults and children. # Comparison of Technological Characteristics The Dynamic Gas Scavenging System 2 (DGSS-2) shares the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate device, Dynamic Gas Scavenging System (DGSS). The Dynamic Gas Scavenging System 2 is similar in design and function to the predicate device for the modes of operation and use. The Dynamic Gas Scavenging System 2 was created as an update to the predicate device, Dynamic Gas Scavenging System. The result is both an aesthetically more pleasing, as well as more efficient, scavenger interface. The Dynamic Gas Scavenging System 2 differs from the predicate device, Dynamic Gas Scavenging System, in three ways: - 1. Electronic sensing of pressure and control of vacuum solenoid. - 2. Smaller, simpler design. - 3. Reduced power consumption and heat production. The Dynamic Gas Scavenging System 2 has no different functionality or effect on the anesthesia workstation than any other anesthetic gas-scavenging accessory attached to any currently-marketed anesthetic workstation. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a logo with the letters "AGR" in a stylized font. The letters are green and gray, with a recycling symbol inside the "A". The logo has a modern and clean design. The letters are bold and easy to read. Differences in Technology: Mode of Operation: The Dynamic Gas Scavenging System 2 is identical to the predicate in its functional operation of sensing the accumulation of waste gas via a sensitive pressure transducer and translating this signal into a command to provide vacuum flow for a set period of time. Because the vacuum control solenoid is normally open, the single failure mode for both the Dynamic Gas Scavenging System 2 and the Dynamic Gas Scavenging System is full-open flow. This full flow mode is identical to the current standard Anesthetic Gas Scavenging System (AGSS) mode of operation, providing a "reversion to safety" failure mode in case of power loss. # Engineering: Like the Dynamic Gas Scavenging System, the Dynamic Gas Scavenging System 2 uses an electrically-actuated solenoid to control gas flow. Because the solenoid is normally open and spends 90% of its time closed, it is energized more often than not and thus operates at above-ambient temperature. Unlike the predicate, though, the Dynamic Gas Scavenging System 2 also has circuitry minimizing the "on" voltage to the solenoid, reducing heat accumulation and potential for any hazard. This feature actuates the solenoid with full 12 VDC and within 100 ms, drops the voltage holding the solenoid closed to 6 VDC (adequate to maintain displacement) to minimize electrical use and waste heat. ### Materials: There is no change in the housing and fittings of the Dynamic Gas Scavenging System 2 from the predicate. ### Control Circuitry: Pressure sensing and flow control in the Dynamic Gas Scavenging System 2 is digital rather than analog, as was the predicate device. The pressure sensor has increased sensitivity and predicted longevity, while the solid-state construction reduces electrical load, weight and bulk. This allows the complete system to be mounted directly onto the anesthesia machine without external hoses or connections. The circuit board is lead-free and manufactured from Restriction of Hazardous Substances (RoHS) compliant components. It is also isolated from the gas pathway as an additional control against degradation from moisture or oxygen. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows a logo with the letters "AGR" in a stylized font. The "A" is shaped like a triangle with a recycling symbol inside. The letters are colored in a gradient from green to gray. The logo appears to be for a recycling or environmental company. Lifespan/Shelf Life: Shelf life for the Dynamic Gas Scavenging System 2 is not applicable, since there is a very low likelihood of time-dependent product degradation. The predicate, Dynamic Gas Scavenging System, after 9 years of service in some cases, has no established lifespan. The nitrile reservoir bag used in the predicate device was found to be subject to some environmental degradation over 6 - 12 months and the labelling now indicates that it was to be examined every 6 months and replaced as needed. The sensors and electrical components had no determined life expectancy and, in our current experience of 2 million patient hours, have yet to fail. The only moving part, the solenoid, is rated at 1,000,000 cycles; but, we have yet to see a failure. The Dynamic Gas Scavenging System 2 is estimated to have a lifespan of 10 years, calculated primarily to anticipate solenoid failure (> 1,000,000 cycles), but no definitive determination has been made. The reservoir bag is now composed of silicone rubber and resists environmental degradation. It should not be a factor although semi-annual inspection is recommended. ### Substantial Equivalence Discussion The following table compares the Dynamic Gas Scavenging System 2 to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the bases for the determination of substantial equivalence. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows a logo with the letters "AGR" in a stylized font. The letters are green and gray, with the green portion of the letters appearing to be in front of the gray portion. The letter "A" has a recycling symbol inside of it. The logo is simple and modern, and the colors suggest that the company is environmentally friendly. | Table 7A – Comparison of Characteristics | | | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Anesthetic Gas<br>Reclamation LLC | Anesthetic Gas<br>Reclamation LLC | | Trade Name | New Device<br>Dynamic Gas Scavenging<br>System – 2 (DGSS-2) | Dynamic Gas Scavenging<br>System (DGSS) | | 510(k) Number | Not Assigned | K063519 | | Intended Use | The device is intended to<br>be used for the<br>scavenging of waste<br>anesthetic gases from<br>anesthesia machines used<br>during the provision of<br>general anesthesia to<br>adults and children. | The device is intended to<br>be used for the<br>scavenging of waste<br>anesthetic gases from<br>anesthesia machines<br>during the provision of<br>general anesthesia to<br>adults and children. | | Anesthesia Machine –<br>Specific | No | No | | Control Circuit | Digital | Analog | | Pressure Sensing | Solid – State | Mechanical Diaphragm | | Solenoid | Co – Axial High Flow | 5 mm Orifice | | Power Requirement | 0.4 A | 0.8 A | | Status Indicator | Green LED | Green LED | | Positive Pressure Relief | Integrated | Integrated | | Negative Pressure Relief | -2 cm H2O | -2 cm H2O | | WAGD Maximum Flow Rate | 50 Ipm | 50 Ipm | | Machine Location | External Mount | External Mount | | Waste Anesthetic Ingress | Via 19mm Hose | Via 19mm Hose | | Physical Attachment | GCX or Other Rail | GCX or Other Rail | Table 7A – Comparison of Characteristics {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows a logo with the letters "ATR" in a stylized font. The letters are green and gray, with a recycling symbol inside the "A". The logo has a modern and clean design. ### Non-Clinical Performance Data The Dynamic Gas Scavenging System 2 has been thoroughly tested through verification of specifications and validation, including materials testing including volatile organic compounds, to ensure safe use of the device in its intended use environment. Verification of compliance with applicable standards has also been completed. The following quality assurance measures were applied during the development of the Dynamic Gas Scavenging System 2: | Quality Assurance Measure | Standard(s) | |----------------------------------------------------------------------------------------------------|---------------------------------------| | Risk Analysis | ISO 80601-2-13 | | Requirements/Specification Reviews | ISO 80601-2-13 | | Design Reviews | ISO 80601-2-13 | | Testing on Unit Level (Module Verification) | ISO 80601-2-13 | | Integration Testing (System Verification) | ISO 80601-2-13 | | Performance Testing (Verification) | IEC 60601-1-2:2014 and ISO 80601-2-13 | | Materials Testing including Volatile Organic Compounds (VOC) | ISO 80601-2-13 | | Verification Testing including Electrical Safety Testing and Electromagnetic Compatibility Testing | IEC 60601-1-2:2014 | | Simulated Use/User Requirements Testing (Validation) | ISO 80601-2-13 | ### Clinical Testing There was no clinical testing required to support the Dynamic Gas Scavenging System 2 as the indications for use is equivalent to the predicate device. The safety and efficacy of the Dynamic Gas Scavenging System 2 is supported by the non-clinical testing performed. The verification and validation testing of the Dynamic Gas Scavenging System 2 was found to be acceptable and support the claims of substantial equivalence. ### Statement of Substantial Equivalence It has been shown in this 510(k) submission that the difference between the Dynamic Gas Scavenging System 2 and the predicate device does not raise any questions regarding its safety and effectiveness. The Dynamic Gas Scavenging System 2, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows a logo with the letters "AGR" in a stylized font. The "A" has a recycling symbol inside of it. The letters are colored in a gradient, transitioning from gray on the left to green on the right. The logo has a modern and eco-friendly design. # Conclusion The Dynamic Gas Scavenging System 2 has similar intended use and technological characteristics as the predicate device. The Dynamic Gas Scavenging System 2 is both safe and effective for the scavenging of waste anesthetic gases from anesthesia workstations during the provision of general anesthesia to adults and children.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...