MINISCAV

K160950 · Sedation Systems, LLC · CBN · May 5, 2017 · Anesthesiology

Device Facts

Record IDK160950
Device NameMINISCAV
ApplicantSedation Systems, LLC
Product CodeCBN · Anesthesiology
Decision DateMay 5, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5430
Device ClassClass 2

Indications for Use

The MINISCAV™ waste gas evacuation apparatus is intended to remove patients' exhaled waste gases during procedures where analgesia is administered to a patient via inspiration of mixtures oxide and oxygen from a nitrous oxide / oxygen delivery device. Not intended for use with flammable anesthetic gases. This device is intended for professional use only in healthcare facilities, clinics, and physician and dentist offices.

Device Story

MINISCAV™ is a portable, diaphragm-based vacuum pump designed to evacuate exhaled waste gases from nitrous oxide/oxygen analgesia delivery systems. It connects via a standard 19mm flexible vacuum hose to the exhaust port of analgesia equipment. Operated by healthcare professionals in clinics and offices, the device provides a constant vacuum (≤90 mmHg) and flow rate (42 Lpm ± 5 Lpm) to ensure effective gas scavenging. By removing waste gases from the patient circuit, the device helps maintain a safe environment for both patients and staff. It is compatible with specific analgesia systems (e.g., Porter, Accutron, Nitronox). The device is not intended for use with flammable anesthetic gases.

Clinical Evidence

Bench testing only. Testing included electrical safety (AAMI/ANSI/ES60601-1), EMC (IEC 60601-1-2), durability (10,000 hours continuous operation), vacuum/flow performance (≤90 mmHg at 42 Lpm ± 5 Lpm), leakage testing, and oxygen-rich environment safety (IEC 60601-1 section 11.2.2). Compatibility was verified with Porter, Accutron, and Nitronox systems.

Technological Characteristics

Diaphragm-type vacuum pump; hermetically sealed; portable form factor. Connects via 19mm hose barb. Fixed vacuum output (≤90 mmHg) and flow rate (42 Lpm ± 5 Lpm). Electrical power source. Complies with IEC 60601-1 and IEC 60601-1-2 standards.

Indications for Use

Indicated for patients undergoing procedures requiring analgesia via nitrous oxide/oxygen delivery. Intended for professional use in healthcare facilities, clinics, and physician/dentist offices. Not for use with flammable anesthetic gases.

Regulatory Classification

Identification

A gas-scavenging apparatus is a device intended to collect excess anesthetic, analgesic, or trace gases or vapors from a patient's breathing system, ventilator, or extracorporeal pump-oxygenator, and to conduct these gases out of the area by means of an exhaust system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or movement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 5, 2017 Sedation Systems LLC % Paul Dryden Consultant, Promedic LLC 2471 McMullen Booth Rd., Ste. 316 Clearwater, Florida 33759 Re: K160950 Trade/Device Name: MINISCAVTM Regulation Number: 21 CFR 868.5430 Regulation Name: Gas-Scavenging Apparatus Regulatory Class: Class II Product Code: CBN, BTA Dated: April 27, 2017 Received: May 1, 2017 Dear Paul Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K160950 Device Name MINISCAVTM Indications for Use (Describe) The MINISCAV™ waste gas evacuation apparatus is intended to remove patients' exhaled waste gases during procedures where analgesia is administered to a patient via inspiration of mixtures oxide and oxygen from a nitrous oxide / oxygen delivery device. Not intended for use with flammable anesthetic gases. This device is intended for professional use only in healthcare facilities, clinics, and physician and dentist offices. Type of Use (Select one or both, as applicable) XX Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) FORM FDA 3881 (8/14) Page 1 of 1 EF PSC Publishing Services (301) 443-6740 Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. {3}------------------------------------------------ Sedation Systems LLC 2471 McMullen Booth Rd., Suite 316 Clearwater, FL 33759 Tel – 888-959-5288 | Official Contact: | S. Michael Bender, Managing Member | |-------------------------------------------------|------------------------------------------------------------------------------------------------------| | Proprietary or Trade Name: | MINISCAV™ | | Common/Usual Name: | Apparatus, Gas Scavenging | | Classification Name /<br>Product Classification | Apparatus, Gas Scavenging<br>CBN, 21CFR 868.5430, Class II | | Predicate Devices: | G. Dundas, Active Waste Gas Scavenger System, K11093<br>Medela, Dominant Flex Suction Pumps, K150134 | #### Device Description: The MINISCAV™ is a small, portable, vacuum pump that generates a constant vacuum that can be connected via a standard flexible vacuum hose to nitrous oxide / oxygen delivery equipment. The vacuum hose is connected to the "vacuum / suction" nipple on exemplary equipment. #### Indications for Use: The MINISCAV™ waste gas evacuation apparatus is intended to remove patients' exhaled waste gases during procedures where analgesia is administered to a patient via inspiration of mixtures of nitrous oxide and oxygen from a nitrous oxide / oxygen delivery device. Not intended for use with flammable anesthetic gases. This device is intended for professional use only in healthcare facilities, clinics, and physician and dentist offices. | Feature | Predicate<br>G. Dundas<br>Active Waste Gas<br>Scavenger System<br>K110930 | Predicate<br>Medela<br>Dominant Flex<br>Suction Pumps<br>K150134 | Proposed<br>MINISCAVTM | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Classification | CBN<br>CFR 868.5430<br>Gas scavenging<br>apparatus<br>Class II | BTA<br>CFR 878.4780<br>Powered suction pump<br>Class II | CBN<br>CFR 868.5430<br>Gas scavenging apparatus<br>Class II | | Indications for Use | The Waste Gas<br>Scavenger is designed<br>for use with vacuum<br>(suction) waste gas<br>disposal systems with<br>anesthesia machines<br>and heart/lung bypass | The Dominant Flex<br>Suction Pump is<br>indicated for vacuum<br>extraction, aesthetic<br>body contouring,<br>aspiration during<br>flexible endoscopy, and | The MINISCAVTM waste<br>gas evacuation apparatus<br>is intended to remove<br>patients' exhaled waste<br>gases during procedures<br>where analgesia is<br>administered to a patient | #### Table 1 Substantial equivalence Comparison to Predicates {4}------------------------------------------------ #### 510(k) Summary 5-May-17 Page 2 of 4 | Page 2 of 4 | | | | |-----------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | machines. | aspiration and removal<br>of surgical fluids, tissue<br>(including bone), gases,<br>bodily fluids or<br>infectious materials<br>from wounds or from a<br>patient's airway or<br>respiratory support<br>system either during<br>surgery or at the beside. | via inspiration of<br>mixtures of nitrous oxide<br>and oxygen from a<br>nitrous oxide / oxygen<br>delivery device. Not<br>intended for use with<br>flammable anesthetic<br>gases.<br>This device is intended<br>for professional use only<br>in healthcare facilities,<br>clinics, and physician and<br>dentist offices. | | Environment of Use | Healthcare facilities,<br>clinics and physician<br>and dentist offices. | Healthcare facilities<br>and clinics | | | Used with flammable<br>anesthetic gases | No | No | No | | Requires a vacuum<br>source | Yes, connects to any<br>vacuum source<br>Central supply / wall /<br>Portable pump | Portable vacuum source | Portable vacuum source<br>to connect to a waste gas<br>scavenging apparatus to<br>supply the vacuum source | | Means of connection | A flexible vacuum hose<br>from the device to the<br>vacuum source | Connects a flexible<br>hose from this pump to<br>any device requiring a<br>vacuum source, i.e., gas<br>scavenging apparatus | Connects a flexible hose<br>from this pump to a gas<br>scavenging apparatus | | Fittings | 19 mm<br>Hose barb for vacuum | Hose barb for vacuum | 19 mm<br>Hose barb for vacuum | | Vacuum pressures<br>Flow rate | Can accept a range of<br>applied vacuum as it no<br>range limit specified | Up to -700 mmHg | 90 mmHg<br>42 Lpm +/- 5 Lpm<br>Connects to a nitrous<br>oxide/ oxygen flowmeter<br>exhaust port | | Technology | Passive device which<br>requires connection to a<br>vacuum source | Piston / cylinder design<br>Portable | Diaphragm design<br>Portable | #### Substantial Equivalence Discussion - Table 1 above compares the key features of the proposed MINISCA V™ with the identified predicate and reference to demonstrate that the proposed device can be found to be substantially equivalent. #### Indications for Use - While the indications for use are not identical to that of the predicate device (K110930), the subject device and Dundas predicate are intended to remove waste gases from anesthesia gas machine system.. Discussion - The MINISCAV™ is one part of a waste gas scavenging system. The parts include: (1) a scavenger to titrate exhaust gas flow from the patient circuit, which is connected to (2) a vacuum source. This device is intended to scavenge gas, which the same as the intended use of the predicate Dundas scavenger. #### Technology, Construction, Performance - {5}------------------------------------------------ The MINISCAVIM is a diaphragm hermetically sealed pump with a fixed vacuum flow rate. The inlet line is connected to a flowmeter exhaust port of the nitrous oxide / oxygen delivery equipment which has an exhaust port for scavenging gases. The Dundas scavenger (K110930) is a flowmeter that titrates exhaust flow with wall-supplied vacuum sources. While the Dundas scavenger does not supply the vacuum, both the MINISCAV and Dundas devices validate the ability to expel gases at specified vacuum pressures, mitigating risks associated with failure to vacuum. Therefore, the MINISCAV™ does not raise different concerns of safety or effectiveness for substantial equivalence. The Medela Dominant Flex Suction Pump (K150134) is pump with an adjustable vacuum flow as high as -700 mmHg. The Medela includes indications to remove gases from respiratory support svstems, which the MINISCAV also performs by providing vacuum via a separate gas scavenger, Therefore, both devices are validated for vacuum specifications. Although different technologies (vacuum range -700mmHg compared to -90mmHg) exist, these differences do not raise different questions of safety and effectiveness. #### Environment of Use - The environments of use are similar to the predicate Medela Dominant Flex Suction Pump (K150134) and the Dundas Active Waste Gas Scavenger System (K110930). Discussion - As the environments of use are similar to the predicate they should be considered substantially equivalent. #### Non-Clinical Testing Summary - We performed testing which evaluated: - AAMI/ANSI/ES60601-1 for electrical safety ● - IEC 60601-1-2 for EMC ● - Durability ● - o Continuous running demonstrated that the vacuum pump stays within its specifications for at least 10,000 hours - Real-time testing of the complete unit supports that the device meets its O performance specifications after 2 years use - . Vacuum and Flow testing - Testing demonstrates that the device provides a constant vacuum of <=90 mmHg o at a constant Flow rate of 42 Lpm +/- 5 Lpm which has been deemed sufficient for scavenging waste gases from nitrous oxide / oxygen delivery equipment - Determination of maximum length of exhaust tubing о - Leakage - Testing demonstrates that after 2 years the unit has no leaks and meets its O performance specifications - Evaluation in an Oxygen Rich Environment per IEC 60601-1 section 11.2.2. . - Risk Analysis of critical components, oxygen build-up in housing, and o construction and separation of the housing and components support compliance to IEC 60601-1 section 11.2.2, clause 1.b(3). - . Compatibility - o with Porter, Accutron, and Nitronox analgesia systems. #### Discussion of Differences – The differences between the proposed device and the predicate are: {6}------------------------------------------------ - MINISCAV™ is intended to scavenge waste gases, which is similar to the Dundas ● Active Waste Gas Scavenger System (K110930). - . While the intended uses are similar, the Dundas device (K110930) must be connected to a vacuum source (such as the MINISCAVTM) to operate. - Limitation of vacuum. While the Medela Dominant Flex Suction Pump (K150134) has a ● higher vacuum, for connection to a waste gas scavenger, the vacuum pressure is much lower and within the performance specifications of the proposed device. - o This lower vacuum range is appropriate for the intended use of scavenging gases. These differences still allow one to find the subject device substantially equivalent to the predicate devices for the proposed indications for use. ### Substantial Equivalence Conclusion - Based upon the presented information the sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate device are substantially equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...