Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain, Express Dry Seal Chest Drain
Applicant
Atrium Medical Corporation
Product Code
CAC · Anesthesiology
Decision Date
Sep 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Auto transfusion (ATS) Chest Drains are intended to evacuate air and/or fluid from the chest cavity (pleural spaces or mediastinum).
Device Story
Sterile, single-use, disposable closed thoracic drainage systems; mimic traditional 3-bottle systems. Devices collect air/fluid from pleural/mediastinal spaces; ATS models facilitate autologous blood collection for reinfusion. Used in surgical suites, ICUs, or ERs by surgeons, physicians, or nurses. Operation involves connecting chest drain/catheter to patient; system uses water seals or dry suction to manage pressure and evacuate air/fluid. Output is visual monitoring of drainage and patient pressure via integrated chambers/scales. Benefits include restored lung expansion and breathing dynamics, plus blood salvage capability. Clinical decision-making relies on visual assessment of fluid levels and air leaks by healthcare providers.
Clinical Evidence
No clinical studies performed. Safety and performance demonstrated via bench testing (specification review, performance, biocompatibility, sterility, stability, design validation) and historical post-market data (low complaint rates/reportable events over 5 years).
Technological Characteristics
Closed thoracic drainage system; materials include ABS, PVC, TPE/TPR, HDPE, LDPE, Polysulfone, Polyisoprene, and stainless steel. Features include air leak monitor, collection chamber, dry suction regulator, manual high negativity vent, and positive pressure release valve. Standalone mechanical device; no software or energy source.
Indications for Use
Indicated for patients requiring evacuation of air and/or fluid from the chest cavity or mediastinum to re-establish lung expansion and restore breathing dynamics. Also indicated for collection of autologous blood from the pleural cavity or mediastinal area for reinfusion in postoperative and trauma blood loss management.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
K130131 — EDRAIN CHEST DRAINAGE SYSTEM · Atrium Medical Corp. · Feb 28, 2013
K043140 — EXPRESS AND OASIS CHEST DRAIN · Atrium Medical Corp. · Dec 8, 2004
K120953 — PLEUR-EVAC SAHARA PLUS CONTINOUS REINFUSION AUTOTRANSFUSION SYSTEM · Tlelflex Medical, Inc. · Dec 10, 2012
K984496 — ATRIUM MEDICAL CORPORATION EXPRESS CHEST DRAIN · Atrium Medical Corp. · May 5, 1999
K140197 — PLEUR-EVAC PLUS CONTINUOUS REINFUSION AND AUTOTRANSFUSION SYSTEM · Teleflexmedical, Inc. · Oct 21, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 3, 2021
Atrium Medical Corporation Robin Mulvey Senior Regulatory Affairs Specialist 40 Continental Blvd. Merrimack, New Hampshire 03054
#### Re: K201305
Trade/Device Name: Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain, Express Dry Seal Chest Drain Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: July 30, 2021 Received: August 4, 2021
#### Dear Robin Mulvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Nicole Gillette Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201305
#### Device Name
Auto transfusion (ATS) Chest Drains:
Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain, Express Dry Seal Chest Drain
Indications for Use (Describe)
· Evacuate air and/or fluid from the chest cavity or mediastinum.
· Help re-establish lung expansion and restore breathing dynamics.
Chest Drain Auto transfusion (ATS)
To facilitate collection of autologous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in postoperative and trauma blood loss management.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
区 Prescription Use (Part 21
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
### Section 5 510(K) Summary
Auto transfusion (ATS) Chest Drains 510(k) summary prepared in accordance with 21 CFR part 807.92
### General information:
| Submitter's name and address: | Atrium Medical Corporation<br>40 Continental Drive<br>Merrimack, NH 03054 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number: | 3011175548 |
| Contact Person: | Robin Mulvey<br>Principal Regulatory Affairs Specialist<br>Tel: 603.864.5424<br>Fax: 973.629.1325<br>email: robin.mulvey@getinge.com |
| Date Prepared: | May 13, 2020 |
| Device Information: | |
| Trade Name(s): | Auto transfusion (ATS) Chest Drains:<br>Ocean Water Seal Chest Drain<br>Oasis Dry Suction Water Seal Chest Drain<br>Express Dry Seal Chest Drain |
| Common/Generic Name: | apparatus, Auto transfusion |
| Classification Name: | Auto transfusion apparatus |
| Regulation Number: | 21 CFR 868.5830 |
| Product Code: | CAC (Anesthesiology Device Panel) |
| Device Class: | The Auto transfusion (ATS) Chest Drains are<br>classified as Class II devices in the US<br>according to 21 CFR 868.5830 |
{4}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
#### Predicate Device Information:
Atrium's Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drains, Oasis Dry Suction Water Seal Chest Drains and Express Dry Seal Chest Drains are substantially equivalent in function and intended use to the Express Chest Drain and Oasis Chest Drain (K043140) and Ocean Water Seal Chest Drain (K043582).
### Device Description
Atrium's Auto transfusion (ATS) Chest Drains are sterile, single use, disposable devices that mimic a traditional "3 Bottle Systems".
A closed thoracic drainage system (chest drain and catheter together) are used to restore the chest to a more normalized condition after surgery, trauma, or spontaneous need for air and/or fluid to be removed. The Auto transfusion (ATS) features of the drains facilitate postoperative collection and reinfusion of autologous blood from the patient's pleural cavity or mediastinal area.
#### Intended Use
The Auto transfusion (ATS) Chest Drains are intended to evacuate air and/or fluid from the chest cavity (pleural spaces or mediastinum).
{5}------------------------------------------------
# GENTIC F
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
#### Indications for Use
- · Evacuate air and/or fluid from the chest cavity or mediastinum.
- · Help re-establish lung expansion and restore breathing dynamics. Chest Drain Auto transfusion (ATS)
To facilitate collection of autologous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in postoperative and trauma blood loss management.
### Intended User/Operator
The primary intended users of the Auto transfusion Chest Drains; Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain are surgeons, physicians, or nurses. Specifically, the primary intended users are chest medicine physicians, including but not limited to surgeons (cardiothoracic, general trauma, hepatobiliary, urology and other trained and qualified to treat conditions in the thorax requiring drainage), interventional physicians (pulmonology, radiology, anesthesiology, emergency medicine, critical care medicine, pain medicine and others trained and qualified to treat conditions in the thorax requiring drainage), physicians' assistants and certified nurse advanced practitioners or nurses who have been trained in the theoretical, technical, and clinical aspects of the use the devices.
### Intended Use Environment
Auto transfusion Chest Drains; Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain are used in a sterile environment (e.g. the surgical suite) or in a non-sterile environment (e.g. the Intensive Care Unit (ICU) or the Emergency Room (ER)) where a local sterile environment for procedures can be created.
{6}------------------------------------------------
# GENTING
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
#### Intended Patient
The target patient population is patients who need fluid and/or air evacuated from the chest cavity (pleural spaces or mediastinum). The actual choice of a specific drainage product is determined by the clinical indications and contraindications and the specific requirements of each patient's anatomy and physiologic condition as clearly described in the respective guides accompanying each
device. The choice of device is at the sole discretion of the physician in charge of that patient.
{7}------------------------------------------------
## GETINGE *
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
### Example Auto transfusion Drain
Image /page/7/Picture/3 description: The image shows an Atrium Ocean water seal chest drain. The device has multiple chambers labeled A, B, C, and D with numerical scales for measuring fluid levels. The scales range from 0 to 2000, with finer gradations for more precise measurements.
{8}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
### Technological characteristics:
Atrium's Auto transfusions (ATS) Chest Drains have the same technology and attributes as the predicate devices (K043140 and K043582).
| Auto transfusions (ATS)<br>Chest Drains Features | Ocean<br>Water Seal<br>Chest Drain | Oasis Dry<br>Suction<br>Water Seal<br>Chest<br>Drains | Express Dry<br>Seal Chest<br>Drain |
|--------------------------------------------------|------------------------------------|-------------------------------------------------------|------------------------------------|
| Air Leak Monitor | √ | √ | √ |
| Collection Chamber | √ | √ | √ |
| Dry Suction Regulator | | √ | √ |
| Easy-to-Grip Handle | √ | √ | √ |
| In-Line Connector | √ | √ | √ |
| Manual High Negativity Vent | √ | √ | √ |
| Multi-Position Hangers | √ | √ | √ |
| Needleless Access Port | √ | √ | √ |
| Patient Connector | √ | √ | √ |
| Patient Pressure Float Ball | √ | √ | √ |
| Patient Tube | √ | √ | √ |
| Patient Tube Clamp | √ | √ | √ |
| Positive Pressure Release Valve | √ | √ | √ |
| Suction Control Chamber | √ | | |
| Suction Monitor Bellows | | √ | √ |
| Suction Port | √ | √ | √ |
| Swing Out Floor Stand | √ | √ | √ |
| Vacuum Indicator | | | √ |
| Water Seal Chamber | √ | √ | |
| Ampoule, Water, 45ml | | √ | |
| Prefilled Syringe, 30ml | | | √ |
### Table 1: Technological Characteristics
{9}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
### Major Materials of Construction:
Atrium's Auto transfusions (ATS) Chest Drains have the same Major Materials of Construction as the predicate devices (K043140 and K043582).
| Auto transfusions (ATS)<br>Chest Drains<br>Major Materials | Ocean Auto<br>transfusion<br>Drain | Oasis Ocean<br>Auto transfusion<br>Drain | Express<br>Ocean Auto<br>transfusion<br>Drain |
|-----------------------------------------------------------------------------|------------------------------------|------------------------------------------|-----------------------------------------------|
| Acrylonitrile butadiene<br>styrene (ABS) | √ | √ | √ |
| Polyvinyl Chloride (PVC) | √ | √ | √ |
| TPE/TPR<br>Thermoplastic Elastomers<br>(TPE)/ Thermoplastic<br>Rubber (TPR) | √ | √ | √ |
| High Density Polyethylene<br>(HDPE) | √ | √ | √ |
| Low-Density Polyethylene<br>(LDPE) | √ | √ | √ |
### Table 2: Major Materials of Construction
{10}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
Table 3 below demonstrates the equivelancy of the Auto transfusion (ATS) Chest drains subject of this submission against the Auto transfusion (ATS) Chest Drains subject of the predicate submissions K043140 and K043582.
### Table 3: Summary Comparison to Predicates K043140 and K043582
| Device Component | PREDICATES:<br>K043140<br>K043582 | Comparison |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Filter, Screen, Blood Recovery Unit (BRU)<br>300CC<br>Acrylonitrile butadiene styrene and<br>Polysulfone | Filter, Screen, Blood Recovery Unit<br>(BRU) 300CC<br>Acrylonitrile butadiene styrene and<br>Polysulfone | Same as Predicate. |
| Sub Assembly, Drain, Blood Recover Unit<br>(BRU)<br>Polyisoprene, Low-Density Polyethylene,<br>Cryolite acrylic, Acrylonitrile butadiene<br>styrene, High Density Polyethylene | Sub Assembly, Drain, Blood Recover<br>Unit (BRU), Ocean<br>Polyisoprene, Low-Density<br>Polyethylene, Cryolite acrylic,<br>Acrylonitrile butadiene styrene, High<br>Density Polyethylene | Same as Predicate. |
| Hanger, Drain, Flexible<br>Nylon , white | Hanger, Drain, Flexible<br>Nylon, white | Same as Predicate. |
| Stand, Chest Drain, Universal<br>Acrylonitrile butadiene styrene (ABS), white | Stand, Chest Drain, Universal<br>Acrylonitrile butadiene styrene (ABS),<br>white | Same as Predicate. |
| Dye Ball - white W/blue dye<br>Polypropylene ball | Dye Ball - white W/blue dye<br>Polypropylene ball | Same as Predicate. |
| Spring, Manual Vent Valve<br>302 stainless steel | Spring, Manual Vent Valve<br>302 stainless steel | Same as Predicate. |
| Plug, Dye Carrier W/blue Dye<br>Porous Hydrophilic Polyethylene | Plug, Dye Carrier W/blue Dye<br>Porous Hydrophilic Polyethylene | Same as Predicate. |
| Seal, Manual Vent Valve | Seal, Manual Vent Valve | Same as Predicate. |
| Ball, 1/2" white<br>polypropylene<br>homopolymer resin | Ball, 1/2" white polypropylene<br>homopolymer resin | Same as Predicate. |
| Stopper, Tethered<br>Kraton, color: gray | Stopper, Tethered<br>Kraton, color: gray | Same as Predicate. |
| Drain Body/ Insert/ Back<br>Acrylonitrile butadiene styrene (ABS),<br>White | Drain Body/ Insert/ Back<br>Acrylonitrile butadiene styrene (ABS),<br>White | Same as Predicate. |
| Drain Cover (Front)<br>Acrylonitrile butadiene styrene (ABS) clear | Drain Cover (Front)<br>Cryo, Cryolite | Same as Predicate<br>with the exception of<br>a change to material<br>from Cryo, Cryolite<br>to Acrylonitrile<br>butadiene styrene<br>(ABS) clear |
| Device Component | PREDICATES:<br>K043140<br>K043582 | Comparison |
| Tubing Set, Auto Connect (A/C),69 | Tubing Set, Auto Connect (A/C),80 | Same as Predicate<br>with the exception of a<br>change in tubing<br>length. |
| Sub Assembly Stopcock Suction Line<br>Low-Density Polyethylene (LDPE). | Sub Assembly Stopcock Suction Line<br>302 stainless steel spring | The strain relief was<br>changed from a 302<br>stainless steel spring<br>to a Low-Density<br>Polyethylene (LDPE). |
| Strain Relief, Patient Tube 6<br>Low-Density Polyethylene (LDPE) | Strain Relief, Patient Tube 6<br>302 stainless steel spring | The strain relief was<br>changed from 302<br>stainless steel spring<br>to a Low-Density<br>Polyethylene (LDPE). |
| Ink Jet Printing | Silk Screen Printing Ink | Same as Predicate<br>with the exception of a<br>change from Silk<br>Screen printing to<br>Inkjet printing. |
| Tape, 1/2 Smooth Crepe Color: beige | Tape, 1/2 Smooth Crepe Color:<br>beige | Same as Predicate. |
| Shipper, Corrugated | Shipper, Corrugated | Same as Predicate. |
| Label, Blank, Color Stripe, Rub. Red<br>Transprint | Label, Blank, Color Stripe, Rub. Red<br>Transprint | Same as Predicate<br>with the exception of a<br>change to transprint<br>specification. |
| Instruction For Use (IFU)<br>40 pound offset | Instruction For Use (IFU)<br>50 pound offset | Same as Predicate with<br>the exception of<br>change to 40 pound<br>offset. |
| DEHP Label<br>Label, Blank, 6.5 X 3.75 | Label not present in Predicate device | DEHP label added to<br>device . |
| Iconic IFU<br>Label,Flat,Blank,WT,6.5X4.625 | Not present in Predicate device | The addition of an<br>Iconic IFU. |
| Pouch, Nylon/Tyvek | Bag, Breather | The change did not<br>impact the products<br>ability to meet sterile<br>barrier, presentation<br>into a sterile field,<br>sterilization of the<br>finished device, or<br>packaging and<br>handling requirements. |
{11}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: `,
Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest
Drain and Express Dry Seal Chest Drain
{12}------------------------------------------------
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
#### Non-clinical tests:
The Auto transfusion (ATS) Chest Drains comply with the voluntary standards identified in section 9 of this submission. Atrium Medical's development process required that the following activities be completed during the development of the current configuration of the Ocean Water Seal, Oasis Dry Suction Water Seal and Express Dry Seal Chest Drains.
- · specification review
- · performance testing
- · biocompatibility testing
- · sterility testing
- · stability testing
- design validation
### Clinical tests:
There were no clinical studies of the modified device.
The Auto transfusion (ATS) Chest Drain products are designed and manufactured in such a way that, when used under the conditions and for the purposes intended, they will not compromise the clinical condition or the safety of patients, or the safety and health of users. The risks associated with their intended use constitute acceptable risks when weighed against the benefits to the patient and are compatible with a high level of protection of health and safety. The clinical evaluation of the product is sufficient to demonstrate both clinical safety and performance of the Auto transfusion (ATS) Chest Drains based on the articles reviewed, and from the low complaint rates and reportable events over the last 5 years. Post-market experience indicates that complications rates in clinical use constitute risks commensurate with the state of the art as indicated by the clinical evaluation.
{13}------------------------------------------------
# GENTIC F
Atrium Medical Corporation Premarket Notification Traditional 510(k) Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain and Express Dry Seal Chest Drain
#### Conclusion:
The Ocean Water Seal, Oasis Dry Suction Water Seal and Express Dry Seal Chest Drains are similar to the predicate devices in the intended use and the fundamental scientific technology of the device.
The design verification and validation testing established that the Ocean Water Seal, Oasis Dry Suction Water Seal and Express Dry Seal Chest Drains are safe and effective and performs as well as the predicate devices.
Based upon the information submitted in this Traditional 510(k) premarket notification, Ocean Water Seal, Oasis Dry Suction Water Seal and Express Dry Seal Chest Drains are substantially equivalent to the currently marketed Express Chest Drain and Oasis Chest Drain (K043140) and Ocean Water Seal Chest Drain (K043582).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.