K160735 · Fresenius Kabi AG · CAC · Sep 16, 2016 · Anesthesiology
Device Facts
Record ID
K160735
Device Name
CATSmart
Applicant
Fresenius Kabi AG
Product Code
CAC · Anesthesiology
Decision Date
Sep 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CATSmart (Continuous Autotransfusion System) device by Fresenius Kabi is an autotransfusion device indicated for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed red blood cells for reinfusion.
Device Story
CATSmart is an intraoperative/postoperative autotransfusion system; processes anticoagulated shed blood collected in a sterile reservoir. Operates via continuous flow centrifuge; removes plasmatic components, non-erythrocytic cells, activated coagulation factors, fibrinolysis products, cell trauma, and anticoagulants. Produces washed packed red cells for reinfusion. Used in clinical settings; operated by healthcare professionals. Features include hematocrit monitoring via integrated cuvette in AT3 disposable tubing; touchscreen interface. Output provides hematocrit values for process monitoring; not for diagnostic/quality control. Benefits include recovery of autologous blood for patient reinfusion.
Clinical Evidence
Bench testing only. In vitro blood quality studies compared CATSmart to the predicate C.A.T.S. across three wash programs (emergency, low volume, smart). Results confirmed successful validation of all wash programs using the AT3 tubing set. Separate evaluation demonstrated effective removal of fat and heparin. Electrical safety tested per IEC 60601-1-2:2007; additional testing performed for vibration, shock, and durability.
Technological Characteristics
Continuous flow centrifuge system. Includes AT3 disposable tubing set with integrated cuvette for hematocrit sensing. Touchscreen control interface. Electrical safety per IEC 60601-1-2:2007. System supports multiple wash programs (emergency, low volume, smart).
Indications for Use
Indicated for processing autologous shed blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion in patients undergoing surgery or trauma.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
C.A.T.S. Continuous Autotransfusion System (K960006)
FRESENIUS ATR40 AND ATR120 AUTOTRANSFUSION RESERVOIRS (K984586)
Related Devices
K210089 — CATSmart, Automated Blood Processing Autotransfusion System · Fresenius Kabi AG · Feb 11, 2021
K192368 — CATSmart · Fresenius Kabi AG · Sep 27, 2019
K180831 — CATSmart · Fresenius Kabi AG · Dec 10, 2018
K973378 — C.A.T.S CONTINUOUS AUTOTRANSFUSION SYSTEM MODEL 900500 1, AT1 AUTOTRANSFUSION SET MODEL 900 510 1, PSQ SET (DD) MODEL 61 · Fresenius USA, Inc. · Nov 3, 1997
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all capital letters. The text is in a bold, serif font. The words are arranged on a single line and centered horizontally.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 16, 2016
Fresenius Kabi USA, LLC % Barry Hicks Manager, Regulatory Affairs Three Corporate Drive Lake Zurich, Illinois 60047
Re: K160735
Trade/Device Name: CATSmart Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: September 14, 2016 Received: September 16, 2016
Dear Barry Hicks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Willhemen
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K160735
Device Name CATSmart
Indications for Use (Describe)
The CATSmart (Continuous Autotransfusion System) device by Fresenius is an autotransfusion device indicated for the processing of autologous shed blood collected intraoperatively to obtain washed packed red blood cells for reinfusion.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Fresenius Kabi. The logo consists of three blue, curved lines on the left, followed by the text "FRESENIUS" in blue, and "KABI" in blue below that. The logo is simple and modern, and the blue color gives it a sense of trustworthiness.
# Section 5: 510(k) Summary
## 1. Submitter
This 510(k) summary was created on 14 March 2016.
| The 510(k) owner and contact person for this application is: | |
|--------------------------------------------------------------|-------------------------------------------------------|
| Name: | Barry G. Hicks |
| Position: | Manager, Regulatory Affairs |
| Company: | Fresenius Kabi USA, LLC. |
| Address | Three Corporate Drive<br>Lake Zurich, IL<br>60047 USA |
| Phone: | 847-550-7981 |
| Fax: | 847-550-2960 |
| Email: | barry.hicks@fresenius-kabi.com |
| Establishment registration number: | 3004548776 |
### 2. Device Name and Classification
| Proprietary / Trade Name: | CATSmart |
|---------------------------|---------------------------------------------------|
| Common Name: | Automated Blood Processing Autotransfusion System |
| Classification Name: | Apparatus, Autotransfusion |
| Classification: | Class II per 21 CFR § 868.5830 |
| Product Code: | CAC |
| Panel: | Anesthesiology |
## 3. Predicate Device
This application is intended to demonstrate that the CATSmart Autotransfusion System, including its associated disposables, is substantially equivalent to the following predicate device:
| 510(k) number: | K960006 |
|---------------------------|-----------------------------------------------------------------|
| Proprietary / Trade Name: | C.A.T.S. Continuous Autotransfusion System |
| Common Name: | Automated Blood Processing Autotranfusion System |
| Classification Name: | Apparatus, Autotransfusion |
| Classification: | Class II per 21 CFR § 868.5830 |
| Product Code: | CAC |
| Panel: | Cardiovascular |
| Manufacturer: | Fresenius AG, Frankfurter Str. 6-8, 66606 St. Wendel<br>Germany |
{4}------------------------------------------------
CATSmart
Image /page/4/Picture/2 description: The image contains the logo for Fresenius Kabi. The logo consists of three blue vertical lines that are curved at the top. To the right of the lines, the words "FRESENIUS" and "KABI" are written in blue, with "FRESENIUS" on top of "KABI". The logo is simple and modern.
The C.A.T.S. Autotransfusion System was modified with the addition of the following accessories under K984233 and K984586:
| 510(k) number: | K984233 |
|---------------------------|------------------------------------------------------------------------------------------------------------|
| Proprietary / Trade Name: | C.A.T.S AUTOTRANSFUSION ACCESSORIES,<br>MODELS ATS SUCTION LINE, ATY Y-ADAPTER,<br>AND ATO OXYGENATOR LINE |
| Common Name: | Automated Blood Processing Autotranfusion System |
| Classification Name: | Apparatus, Autotransfusion |
| Classification: | Class II per 21 CFR § 868.5830 |
| Product Code: | CAC |
| Panel: | Cardiovascular |
| Manufacturer: | Fresenius AG, Frankfurter Str. 6-8, 66606 St. Wendel,<br>Germany |
| 510(k) number: | K984586 |
| Proprietary / Trade Name: | FRESENIUS ATR40 AND ATR120<br>AUTOTRANSFUSION RESERVOIRS |
| Common Name: | Cardiovascular Surgical Devices |
| Classification Name: | Reservoir, Blood, Cardiopulmonary Bypass |
| Classification: | Class II per 21 CFR § 870.4400 |
| Product Code: | DTN |
| Panel: | Cardiovascular |
| Manufacturer: | Fresenius AG, Frankfurter Str. 6-8, 66606 St. Wendel,<br>Germany |
#### 4. Description
The product described in this document is the CATSmart (Continuous Autotransfusion System) including the disposables AT1, AT3, ATS, ATY, ATO, ATR40 and ATR120, ATV-70 and ATV-180, ATF40 and ATF120.
The Fresenius Kabi CATSmart device is an intraoperative autotransfusion system for intraand/or postoperative processing of blood lost through surgery or trauma. The CATSmart device operates on the principle of a continuous flow centrifyge, comparable to continuous systems for hemapheresis which, for decades, have been widely used in blood banks.
The shed blood, which is anticoagulated and collected in a sterile reservoir, is processed in a continuous washing process to obtain washed packed red cells for reinfusion to the patient. During this process all plasmatic and non-erythrocytic cellular components of the collected blood, and thus activated coagulation factors, products of fibrinolysis and cell trauma as well as the anticoagulant are removed. The packed red cells are collected in a reinfusion bag from which they can be reinfused to the patient via a transfusion set when needed.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of three blue, curved lines on the left side. To the right of the lines, the words "FRESENIUS" and "KABI" are stacked on top of each other, also in blue. The logo is simple and modern in design.
CATSmart is based on the predecessor products C.A.T.S. and C.A.T.SP."", which have been in clinical use for almost 20 years. CATS was originally cleared in 1996. Following minor modifications it has been marketed as C.A.T.Smits since 2005. CATSmart has as an additional technical feature: monitoring of the hematocrit value.
# 5. Intended Use
The CATSmart (Continuous Autotransfusion System) device by Fresenius Kabi is an autotransfusion device indicated for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed red blood cells for reinfusion.
# 6. Substantial Equivalence
Fresenius CATSmart is substantially equivalent to the predicate device system, with the following components.
#### a) The reusable autotransfusion device:
CATSmart is based on the predecessor products C.A.T.S. and C.A.T.SP.446, which have been in clinical use for almost 20 years. The main functions of CATSmart and the underlying methodology remained unchanged. In fact, the only functional difference of CATSmart compared to C.A.T.S./C.A.T.S./C.A.T.S./C.A.T.S. of the hematocrit value as an additional technical feature. Moreover, the main controller software was updated. The update includes the hematocrit function as a new item and usage of a touchscreen control instead of a keyboard. The software structure and the requirements remain unchanged compared to the C.A.T.S./C.A.T.Splus software.
#### b) Disposables:
The predicate device C.A.T.S./C.A.T.SPl46 system also includes the disposable blood processing chamber with all necessary tubing (called AT1) as described in the 510(k) submission K960006. The new disposable for the proposed device CATSmart is AT3. The only difference between AT3 and AT1 is the integration of a cuvette into the tubing for the hematocrit measurement. The AT1 disposable may also be used with CATSmart, but no hematocrit information will be displayed.
The following disposables have received prior 510(k) clearance, do not change in their usage with CATSmart, and are not further mentioned in this 510(k):
- ATS suction line (cleared under K984233), -
- ATY Y-Adapter (cleared under K984233). -
- -ATO Oxygenator Line (cleared under K984233).
- ATR40 Autotransfusion Reservoir (cleared under K984586), -
- ATR120 Autotransfusion Reservoir (cleared under K984586) -
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of three blue, curved lines on the left side. To the right of the lines, the text "FRESENIUS KABI" is written in blue, with "FRESENIUS" on top of "KABI".
The following additional disposables are exempted from 510(k) submission, do not change with their usage with CATSmart, and are not further mentioned in this 510(k):
- -ATV-70 and ATV-180 vacuum lines. These devices are classified as cardiopulmonary bypass accessory equipment and exempted according to 21 CFR & 870.4200.
#### Discussion of differences between CATSmart and the predicate device.
The only functional difference of CATSmart compared to C.A.T.S./C.A.T.Sn/W is the monitoring of the hematocrit value as an additional feature. The hematocrit monitoring function of CATSmart is only used to monitor the wash process for informational purposes and is not intended for diagnostic or quality control purpose. As stated in the Operator's Manual, the hematocrit values given by the device are not a substitute for the hematocrit check on the product before it is reinfused in the patient. The sensors are not calibrated measuring instruments.
#### Conclusion:
Based on the identified characteristics it is claimed that the device is substantially equivalent to the predicate devices.
### 7. Performance testing
The performance of the CATSmart device was tested by in vitro blood quality studies in direct comparison to the predicate device C.A.T.S. (currently marketed as C.A.T.Sº™). All three wash programs (emergency wash program, low volume wash, smart wash) were tested. Overall it was judged that all validation tests were passed for all wash programs on the CATSmart device with the AT3 tubing set. A separate in vitro evaluation demonstrated that fat and heparin were almost completely removed from the processed blood. The electrical safety of the CATSmart device was tested according to the general requirements for basic safety and essential performance directive of IEC 60601-1-2:2007. Furthermore, vibration and shock resistance was tested as well as the durability.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.