K201418 · Sunset Healthcare Solutions, Inc. · BZE · Jan 29, 2021 · Anesthesiology
Device Facts
Record ID
K201418
Device Name
Sunset Heated CPAP Tube
Applicant
Sunset Healthcare Solutions, Inc.
Product Code
BZE · Anesthesiology
Decision Date
Jan 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5270
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sunset Heated CPAP Tube is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with non-invasive ventilation for adult patients. It is compatible with the Philips Respironics System One Heated Humidifier and Philips Respironics Dreamstation Heated Humidifier.
Device Story
Heated wire breathing tube; warms/humidifies breathing gases for non-invasive ventilation; reduces condensation in breathing circuit. Input: electrical power from compatible humidifier; Output: warmed air delivered to patient airway. Operated by clinicians or patients in home/clinical settings. Integrated NTC thermistor at mask-end senses air temperature; provides feedback to humidifier to regulate current through heating wires. Benefits patient by maintaining gas temperature, preventing condensation/pooling, and improving therapy comfort.
Clinical Evidence
Bench testing only. Testing included performance of sleep apnea breathing therapy equipment, respiratory humidifying equipment interaction, breathing sets/connectors, EMC, electrical safety, and mechanical safety. Biocompatibility testing performed per ISO 10993-1 and ISO 18562 series (particulate matter, VOCs, leachables). Usability evaluation conducted per FDA human factors guidance.
Technological Characteristics
Heated wire breathing tube; 1.83m length, 15mm ID. Polymeric materials. Powered by host humidifier (3.3 VDC, max 5A). Integrated NTC thermistor for temperature sensing/regulation. Non-sterile. Standards: ISO 5367, ISO 80601-2-70, ISO 80601-2-74, ANSI/AAMI/ES 60601-1, IEC 60601-1-2, IEC 60601-1-11.
Indications for Use
Indicated for adult patients requiring non-invasive ventilation in home or clinical settings (hospitals, institutions, sleep labs, sub-acute care). Compatible with specific Philips Respironics heated humidifiers.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 29, 2021
Sunset Healthcare Solutions, Inc. % Thomas Kroenke Application Correspondent Speed To Market, Inc. 2235 East Flamingo Road, Suite 201G, Las Vegas, NV 89119 USA
Re: K201418
Trade/Device Name: Sunset Heated CPAP Tube Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE Dated: December 22, 2020 Received: December 28, 2020
Dear Thomas Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201418
Device Name Sunset Heated CPAP Tube
#### Indications for Use (Describe)
The Sunset Heated CPAP Tube is a heated wire breathing tube intended to provide warmed and/or humidffed breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with non-invasive ventilation for adult patients. It is compatible with the Philips Respironics System One Heated Humidifier and Philips Respironics Dreamstation Heated Humidifier.
Type of Use (Select one or both, as applicable)
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| I SUBMITTER | |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name: | Sunset Healthcare Solutions. |
| Submitter Address: | 180 N. Michigan Ave, Suite 2000<br>Chicago, IL 60601, US |
| Phone Number: | +1 312-533-2449 |
| Contact Person: | Greg Marler |
| Date Prepared: | January 28, 2021 |
| II. DEVICE | |
| Device Trade Name: | Sunset Heated CPAP Tube |
| Common Name | Breathing system, heater |
| Classification | 21 CFR 868.5270 |
| Product Code: | BZE |
| Review Panel: | Anesthesiology |
| III. PREDICATE DEVICE | K140424, Philips Respironics Reusable Heated Tubing,<br>Respironics, Inc. |
| IV. DEVICE DESCRIPTION | |
| Description | The Sunset Heated CPAP Tube warms air, or breathable<br>gas, as it travels to and from the respiratory patient along the<br>breathing circuit of a respiratory system. It reduces<br>condensation that can form in breathing circuit.<br><br>The Sunset Heated CPAP Tube has a proprietary connector<br>with two locking tabs that makes it compatible with the<br>Respironics System One and with the Respironics<br>Dreamstation series. |
| Physical Description: | Sunset Heated CPAP Tube: length: 1.83 m, internal diameter 15mm<br>Operating voltage: 3.3 VDC<br>Power consumption: max 5A |
| V. INTENDED USE /<br>INDICATION FOR USE<br>STATEMENT | The Sunset Heated CPAP Tube is a heated wire breathing<br>tube intended to provide warmed and/or humidified breathing<br>gases before they enter a patient's airway. It is indicated for<br>single-patient reuse in the home and in clinical settings, such<br>as hospitals, institutions, sleep laboratories, and sub-acute<br>care facilities. It may be used with non-invasive ventilation for<br>adult patients. It is compatible with the Philips Respironics<br>System One Heated Humidifier and Philips Respironics<br>Dreamstation Heated Humidifier. |
{4}------------------------------------------------
- VI. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO PREDICATE DEVICES
| Technological<br>Characteristics: | Characteristics | Philips Respironics<br>Reusable Heated Tubing<br>(K140424) | Sunset Healthcare<br>Solutions<br>Heated CPAP Tube<br>(proposed device) | Discussion of<br>Differences |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| | Intended Use | The Philips Respironics<br>Reusable Heated Tubing<br>is a heated wire breathing<br>tube intended to provide<br>warmed and/or humidified<br>breathing gases before<br>they enter a patient's<br>airway. It is indicated for<br>single-patient reuse in the<br>home and multi-patient<br>use in clinical settings,<br>such as hospitals,<br>institutions, sleep<br>laboratories, and sub-<br>acute care facilities. It may<br>be used with non-invasive<br>ventilation for patients<br>weighing over 10 kg<br>(22lbs).<br>It is compatible with the | The Sunset Heated CPAP<br>Tube is a heated wire<br>breathing tube intended to<br>provide warmed and/or<br>humidified breathing gases<br>before they enter a<br>patient's airway. It is<br>indicated for single-patient<br>reuse in the home and in<br>clinical settings, such as<br>hospitals, institutions,<br>sleep laboratories, and<br>sub-acute care facilities. It<br>may be used with non-<br>invasive ventilation for<br>adult patients.<br>It is compatible with the | Similar.<br>Proposed<br>device is<br>restricted to:<br>• Single<br>patient<br>reuse.<br>• Adult<br>patients. |
| | | Philips Respironics<br>System One Heated<br>Humidifier and Philips<br>Respironics A-Series<br>System One Heated<br>Humidifier. | Philips Respironics<br>System One Heated<br>Humidifier and Philips<br>Respironics Dreamstation<br>Heated Humidifier. | |
| | Compatibility<br>with Humidifiers,<br>Standard<br>Connectors and<br>Humidification<br>Chambers | The tubing has a<br>proprietary connector with<br>two locking tabs that<br>makes it compatible with<br>the Philips Respironics<br>System One Humidifier<br>(K113068) and Philips<br>Respironics A-Series<br>System One Heated<br>Humidifier (K121623). | The tubing has a<br>proprietary connector with<br>two locking tabs that<br>makes it compatible with<br>the Respironics System<br>one (K113068) and with<br>the Respironics<br>Dreamstation series. | Similar.<br>Compatibility<br>with a subset<br>of the devices<br>with which the<br>predicate is<br>compatible. |
| | Anatomical Site | Non-invasive | Non-invasive | Same. |
| | Patient<br>Population | Patients weighing over 10<br>kg (22lbs). | Adult patients | Similar.<br>Device is used<br>with a subset<br>of the<br>predicate<br>device's<br>population. |
| | Environment of<br>Use | Home and hospital | Home and hospital | Same. |
| | Operating<br>Principle | During use a voltage is<br>applied and a current flows<br>through the heating wires,<br>encapsulated in the tubing.<br>Due to the wire resistance,<br>heat is dissipated through<br>the wall of the tube<br>construction into the air<br>flow in the lumen of the<br>tubing. As a result, the air<br>passing through the tubing<br>is warmed reducing or<br>eliminating water<br>condensation and/or<br>pooling of water in the tubing. | During use a voltage is<br>applied and a current flows<br>through the heating wires,<br>encapsulated in the tubing.<br>Due to the wire resistance,<br>heat is dissipated through<br>the wall of the tube<br>construction into the air<br>flow in the lumen of the<br>tubing. As a result, the air<br>passing through the tubing<br>is warmed reducing or<br>eliminating water<br>condensation and/or<br>pooling of water in the tubing. | Same. |
| Characteristics<br>(continued) | Philips Respironics<br>Reusable Heated Tubing<br>(K140424) | Sunset Healthcare<br>Solutions<br>Heated CPAP Tube<br>(proposed device) | Discussion of<br>Differences | |
| Technology | The power is generated by<br>the humidifier and due to<br>the wire resistance in the<br>device, heat is dissipated<br>through the wall of the<br>tube construction into the<br>air flow in the lumen of the<br>tubing. An in-circuit<br>integrated Negative<br>Temperature Coefficient<br>thermistor (NTC) at the<br>mask-end senses the<br>temperature of the passing<br>air flow. The resistance<br>characteristics of the NTC<br>changes with temperature<br>and regulates the current<br>through the heated wires<br>and thereby regulates the<br>temperature of the<br>breathed air. | The power is generated by<br>the humidifier and due to<br>the wire resistance in the<br>device, heat is dissipated<br>through the wall of the<br>tube construction into the<br>air flow in the lumen of the<br>tubing. An in-circuit<br>integrated Negative<br>Temperature Coefficient<br>thermistor (NTC) at the<br>mask-end senses the<br>temperature of the passing<br>air flow. The resistance<br>characteristics of the NTC<br>changes with temperature<br>and regulates the current<br>through the heated wires<br>and thereby regulates the<br>temperature of the<br>breathed air. | Same. | |
| Materials | Standard breathing circuit<br>Polymeric materials | Standard breathing circuit<br>Polymeric materials | Similar. | |
| Device Design<br>and Physical<br>Properties | Use of heated tubing on<br>respiratory systems is a<br>proven technology and is<br>well accepted by the<br>medical community | Use of heated tubing on<br>respiratory systems is a<br>proven technology and is<br>well accepted by the<br>medical community | Similar. | |
| Dimensions | Length: 1.83 meter<br>Inner Diameter: 15 mm<br>version and 22 mm<br>version | Length: 1.83 meter<br>Inner Diameter: 15 mm | Proposed<br>device is the<br>similar to the<br>15 mm<br>predicate<br>version. | |
| Reusable | Single-patient reuse in the<br>home and multi-patient<br>use in clinical setting | Single patient reuse. | Similar.<br>Proposed<br>device is<br>labeled the<br>same for<br>home use and<br>for single use<br>only in a<br>clinical setting. | |
| Sterility | Non-sterile. | Non-sterile. | Same. | |
| Breathing<br>Gases Specified | Not specified | Not Specified | Same. | |
| Power Source | Humidifier controlled | Humidifier controlled | Same. | |
| Heating Wire | Encased | Encased | Similar. | |
| Active Controller | No, humidifier controlled | No, humidifier controlled | Same. | |
| Standards of<br>Conformity /<br>Performance | ISO 5367<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 8185 | ISO 5367<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-70<br>ISO 80601-2-74 | Similar.<br>Proposed<br>device has<br>applied<br>relevant<br>standards. | |
| Compliance<br>(mlpa)<br>Resistance to<br>Flow (mb)2 Tube<br>Volume | ISO 5367 compliant | ISO 5367 compliant | Same. | |
| Characteristics<br>(continued) | Philips Respironics<br>Reusable Heated Tubing<br>(K140424) | Sunset Healthcare<br>Solutions<br>Heated CPAP Tube<br>(proposed device) | Discussion of<br>Differences | |
| Biocompatibility | ISO 10993<br>tests for: Cytotoxicity,<br>sensitization, irritation,<br>genotoxicity, implantation,<br>extractables and<br>leachables. | ISO 10993-1<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3<br>ISO 18562-4<br>tests for: emissions of<br>particulate matter,<br>emissions of volatile<br>organic compounds<br>(VOCs) and leachables in<br>condensate. | Similar.<br>Proposed<br>device has<br>applied<br>relevant<br>standards. | |
{5}------------------------------------------------
{6}------------------------------------------------
#### Summary of Performance Data And Design Controls:
Bench testing was carried out on the following characteristics:
- . Performance of sleep apnea breathing therapy equipment
- . Performance with interaction of respiratory humidifying equipment
- Requirements for breathing sets and connectors .
- . Electromagnetic compatibility (EMC)
- Electrical safety testing
- Mechanical safety testing ●
## Usability Testing:
In addition to the above, usability evaluation was also conducted as per the Guidance for Industry and Food and Drug Administration Staff, Applying Human Factors and Usability Engineering to Medical Devices February 3, 2016.
## Referenced Standards and Performance Testing:
The Sunset Heated CPAP Tube was tested and meets the requirements of following performance standards.
- ISO 5367, edition 5.0. Anaesthetic and respiratory . equipment -- Breathing sets and connectors
- ISO 80601-2-70, edition 1.0 Medical Electrical . Equipment - Part 2-70: Particular Requirements for Basic Safety and Essential Performance of Sleep Apnea Breathing Therapy Equipment
- ISO 80601-2-74. edition 1.0 Medical Electrical . Equipment - Part 2-70: Particular Requirements for Basic Safety and Essential Performance of respiratory humidifying equipment
{7}------------------------------------------------
#### Summary of Performance Biocompatibility testing: The Sunset Heated CPAP Tube was tested on biological Data And Design Controls (continued): safety and meets the requirements of following biocompatibility standards:
- ISO 10993-1:2018: Biological Evaluation of Medical . Devices Part 1: Evaluation and testing within a risk management process
- . ISO 18562-1 First edition, Biocompatibility evaluation of breathing gas pathways in healthcare applications -Part 1: Evaluation and testing within a risk management process
- ISO 18562-2 First edition, Biocompatibility evaluation . of breathing gas pathways in healthcare applications -Part 2: Tests for emissions of particulate matter
- . ISO 18562-3 First edition, Biocompatibility evaluation of breathing gas pathways in healthcare applications -Part 3: Tests for emissions of volatile organic compounds
- ISO 18562-4 First edition, Biocompatibility evaluation . of breathing gas pathways in healthcare applications -Part 4: Tests for leachables in condensate
# Electrical & Mechanical safety and electromagnetic compatibility (EMC)
Electrical & Mechanical safety and EMC testing were conducted on the Sunset Heated CPAP Tube. The device complies with the following safety standards:
- . ANSI/AAMI/ES 60601-1:2005/(R)2012, A1:2012 edition 3.1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2, edition 4.0 2014, Medical electrical . equipment - Part 1-2 General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -Requirements and tests
- IEC 60601-1-11, edition 2.0 2015, Medical electrical . equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Comparison with the predicate device demonstrates that the VIII. CONCLUSION OF Sunset Heated CPAP Tube is substantially equivalent to the SUBSTANTIAL predicate device. The non-clinical data support the safety of EQUIVALENCE the device and the hardware verification and validation demonstrate that the Sunset Heated CPAP Tube should perform as intended in the specified use conditions.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.