← Product Code [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD) · K241939

# EasyCare Tx 2 (K241939)

_Resmed Corp · BZD · Sep 27, 2024 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K241939

## Device Facts

- **Applicant:** Resmed Corp
- **Product Code:** [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD.md)
- **Decision Date:** Sep 27, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The Easy Care Tx 2 software is intended to be used with ResMed compatible therapy devices in a clinical environment. EasyCare Tx 2 provides therapy device setting changes and displays real-time data and treatment settings from compatible ResMed therapy devices when used together with the TxLink 2 module.

## Device Story

EasyCare Tx 2 is a PC-based software application for clinicians in sleep labs or hospitals. It interfaces with ResMed therapy devices via the TxLink 2 module to monitor real-time patient data and adjust treatment settings remotely. The software receives data packets from compatible flow generators (e.g., AirSense 11), allowing clinicians to modify therapy modes, pressure, and other parameters. By enabling remote control and real-time visualization of therapy performance, the device assists clinicians in optimizing patient treatment settings during titration. It enhances usability through updated auto-mode features and expanded parameter ranges compared to the predicate. The system incorporates secure communication protocols, including CSR-based initialization and code-signing, to ensure data integrity and secure operation within the clinical network.

## Clinical Evidence

Bench testing only. Software verification and validation were performed against non-functional requirements and end-to-end functional testing. No clinical data was required to demonstrate safety and effectiveness.

## Technological Characteristics

PC-based software application running on Microsoft Windows. Connectivity via TxLink 2 module to ResMed therapy devices. Implements secure communication via CSR and code-signing. Supports various therapy modes (CPAP, AutoSet, S, ST, TVAuto, ASV, ASVAuto, PAC).

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- EasyCare Tx System ([K113780](/device/K113780.md))

## Reference Devices

- AirSense 11 ([K203126](/device/K203126.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

September 27, 2024

ResMed Corp Prerana Gurubasavaraj Senior Regulatory Affairs Specialist 9001 Spectrum Center Blvd San Diego, California 92123

Re: K241939

Trade/Device Name: EasyCare Tx 2 Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: July 1, 2024 Received: July 2, 2024

Dear Prerana Gurubasavaraj:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241939

Device Name EasyCare Tx 2

Indications for Use (Describe)

The Easy Care Tx 2 software is intended to be used with ResMed compatible therapy devices in a clinical environment. EasyCare Tx 2 provides therapy device setting changes and displays real-time data and treatment settings from compatible ResMed therapy devices when used together with the TxLink 2 module.

| Type of Use (Select one or both, as applicable)                                 |                                                                                |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the ResMed logo. The logo features a curved line made of dots that transition in color from blue to red. Below the line of dots, the text "ResMed" is written in a bold, sans-serif font.

#### 510(k) Summary 1

[As required by 21 CFR 807.921(c)]

| Date of Submission:                        | July 1, 2024                                                                                                                                                                  |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name/Owner:                        | ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123, US                                                                                                      |
| Official Contact:                          | Prerana Gurubasavaraj<br>Senior Regulatory Affairs Specialist<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123, US<br>prerana.gurubasavaraj@resmed.com |
| Device Trade Name:                         | EasyCare Tx 2                                                                                                                                                                 |
| Device Common Name:                        | Therapy titration software                                                                                                                                                    |
| Classification and<br>Classification Name: | II<br>Noncontinuous Ventilator (IPPB)                                                                                                                                         |
| Product Code:                              | BZD                                                                                                                                                                           |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the ResMed logo. The logo features a curved line made of dots that transition in color from blue to red. Below the line of dots, the text "ResMed" is written in a bold, sans-serif font.

| Legally Marketed Predicate<br>Device: | EasyCare Tx System (K113780)                                                                                                                                                                                                                                                                                                 |                                                                                                                                |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Device Description:                   | EasyCare Tx 2 is a personal computer application designed to allow<br>clinicians to monitor real-time data from the ResMed compatible<br>therapy device and adjust the therapy device settings from the<br>control room of the sleep lab.                                                                                    |                                                                                                                                |
| Indication For Use:                   | The EasyCare Tx 2 software is intended to be used with ResMed<br>compatible therapy devices in a clinical environment. EasyCare Tx 2<br>provides therapy device setting changes and displays real-time data<br>and treatment settings from compatible ResMed therapy devices<br>when used together with the TxLink 2 module. |                                                                                                                                |
| Intended Use Comparison               | EasyCare Tx                                                                                                                                                                                                                                                                                                                  | EasyCare Tx 2                                                                                                                  |
|                                       | (K113780)                                                                                                                                                                                                                                                                                                                    |                                                                                                                                |
|                                       | EasyCare Tx is intended to be                                                                                                                                                                                                                                                                                                | The EasyCare Tx 2 software is                                                                                                  |
|                                       | used with ResMed compatible<br>therapy devices via the Tx Link.                                                                                                                                                                                                                                                              | intended to be used with<br>ResMed compatible therapy                                                                          |
|                                       | EasyCare Tx provides real-time                                                                                                                                                                                                                                                                                               | devices in a clinical                                                                                                          |
|                                       | data and treatment settings<br>display, and can also provide                                                                                                                                                                                                                                                                 | environment. EasyCare Tx 2<br>provides therapy device setting<br>changes and displays real-time<br>data and treatment settings |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for ResMed. The logo features a curved line of colored dots that transition from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.

| therapy device setting changes<br>remotely.<br><br>EasyCare Tx is intended to be<br>used in a clinical environment. | from compatible ResMed<br>therapy devices when used<br>together with the TxLink 2<br>module. |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|

| Technological<br>Characteristics Comparison                                                                                                                      | EasyCare Tx<br>(K113780)                                                                                                                                               | EasyCare Tx 2                                                                  | Discussion                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                   |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription<br>Use                                                                                                                                              | Yes                                                                                                                                                                    | Yes                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Intended<br>Environment of<br>Use                                                                                                                                | Hospital or<br>Sleep Lab                                                                                                                                               | Hospital or<br>Sleep Lab                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Display Type                                                                                                                                                     | PC                                                                                                                                                                     | PC                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Hardware<br>Requirements<br>(User PC)                                                                                                                            | Standard PC<br>platform with<br>microprocessor<br>and ancillary<br>controllers                                                                                         | Standard PC<br>platform with<br>microprocessor<br>and ancillary<br>controllers | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Operating<br>System (User<br>PC)                                                                                                                                 | Microsoft<br>Windows                                                                                                                                                   | Microsoft<br>Windows                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Accessories                                                                                                                                                      | CD for<br>downloading<br>to PC<br>Tx Link                                                                                                                              | CD for<br>downloading<br>to PC<br>Tx Link 2                                    | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Compatible<br>Therapy<br>Devices                                                                                                                                 | BZD<br>MNS<br>MNT                                                                                                                                                      | BZD                                                                            | Same<br>EasyCare Tx 2<br>does not<br>support<br>product<br>codes MNS<br>and MNT                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Therapy Device Adjustments (Settings)                                                                                                                            |                                                                                                                                                                        |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Adjust Device<br>Therapy Mode                                                                                                                                    | Yes                                                                                                                                                                    | Yes                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Adjust Device<br>Therapy<br>Settings                                                                                                                             | Yes                                                                                                                                                                    | Yes                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Available<br>Therapy Modes                                                                                                                                       | CPAP                                                                                                                                                                   | CPAP                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
|                                                                                                                                                                  | AutoSet                                                                                                                                                                | AutoSet                                                                        | The PAC and<br>¡VAPS modes<br>are no longer<br>used and<br>supported as<br>they are for<br>devices under                                                                                                                                                                                                                                                                            |                                                                                                                                                   |
|                                                                                                                                                                  | AutoSet for<br>Her                                                                                                                                                     | AutoSet for<br>Her                                                             |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | S                                                                                                                                                                      | S                                                                              |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | ST                                                                                                                                                                     | ST                                                                             |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | TVAuto                                                                                                                                                                 | TVAuto                                                                         | the MNS and<br>MNT product<br>codes, and                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                   |
|                                                                                                                                                                  | ASV                                                                                                                                                                    | ASV                                                                            | not in scope<br>for EasyCare<br>Tx 2 with BZD<br>devices.                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                   |
|                                                                                                                                                                  | ASVAuto                                                                                                                                                                | ASVAuto                                                                        |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | PAC<br>¡VAPS                                                                                                                                                           | PAC                                                                            |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Humidifier<br>Control                                                                                                                                            | Humidifier<br>on/off                                                                                                                                                   | Humidifier<br>on/off, level<br>control, auto                                   | Similar                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                   |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                | Auto mode in<br>the attached<br>ResMed<br>therapy<br>device<br>optimizes the<br>humidifier<br>temperature<br>to ensure<br>rainout does<br>not occur as<br>ambient<br>temperature<br>varies during<br>the night.<br>EasyCare Tx 2<br>controls the<br>humidifier<br>settings on<br>the ResMed<br>compatible<br>therapy<br>device.                                                     |                                                                                                                                                   |
| Tubing Control                                                                                                                                                   | Tube<br>Temperature<br>on/off, auto                                                                                                                                    | Tube<br>Temperature<br>on/off, auto                                            | Same<br>ResMed<br>therapy<br>device<br>optimizes the<br>tube<br>temperature<br>to ensure<br>rainout does<br>not occur as<br>ambient                                                                                                                                                                                                                                                 |                                                                                                                                                   |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                | temperature varies during the night.                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Ramp                                                                                                                                                             | 0 to 45 min                                                                                                                                                            | 5 to 45 min,<br>off, auto                                                      | Similar<br><br>Ramp is in<br>the attached<br>ResMed<br>therapy<br>device. In<br>auto mode,<br>the ramp<br>stops as soon<br>as sleep is<br>detected to<br>ensure the<br>patient<br>obtains the<br>best therapy<br>outcome.<br><br>Off mode was<br>added for<br>usability<br>purposes<br>instead of<br>ramping to 0<br>in the<br>predicate.                                           |                                                                                                                                                   |
| Output Parameters1                                                                                                                                               |                                                                                                                                                                        |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Pressure                                                                                                                                                         | -3 to 35 cm<br>H2O                                                                                                                                                     | 0 to 30 cm H2O                                                                 | Similar<br><br>Therapy<br>range has not<br>changed.                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                   |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                | -3cm H₂O was<br>previously<br>used to cover<br>an extremely<br>niche case,<br>which<br>presented<br>little clinical<br>benefit in<br>AirSense 10<br>and was<br>subsequently<br>removed in<br>AirSense 11.<br>The reduction<br>in pressure is<br>based on the<br>fact that that<br>EasyCare Tx 2<br>only supports<br>OSA devices<br>which require<br>a lower<br>therapy<br>pressure. |                                                                                                                                                   |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Flow                                                                                                                                                             | -120 to 120<br>L/min                                                                                                                                                   | -240 to 240<br>L/min                                                           | Similar<br>Risks<br>associated                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                   |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                | with the<br>increase in<br>flow as an<br>output<br>parameter<br>are ultimately<br>considered by<br>the parent<br>therapy<br>device.<br>EasyCare Tx 2<br>uses a larger<br>data model<br>with scalable<br>architecture<br>to achieve<br>the greater<br>range.                                                                                                                         |                                                                                                                                                   |
| Leak                                                                                                                                                             | -30 to 120<br>L/min                                                                                                                                                    | 0 to 2 L/s                                                                     | Similar<br>Leak<br>parameters<br>are sent from<br>the attached<br>ResMed<br>therapy<br>device in L/s.<br>EasyCare Tx 2<br>allows the<br>option to<br>toggle<br>between L/s<br>and L/min.                                                                                                                                                                                            |                                                                                                                                                   |
| Respiratory<br>Rate                                                                                                                                              | 6 to 35 BPM                                                                                                                                                            | 0 to 90 BPM                                                                    | Similar                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                   |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                | The subject<br>device's<br>higher range<br>supports<br>pediatric<br>patients using<br>the AirSense<br>11.                                                                                                                                                                                                                                                                           |                                                                                                                                                   |
| Tidal Volume                                                                                                                                                     | 0.01 to 3 L                                                                                                                                                            | 0 to 4 L                                                                       | Similar                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                   |
| Target<br>Ventilation                                                                                                                                            | 0 to 30 L/min                                                                                                                                                          | 0 to 30 L/min                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Minute<br>Ventilation                                                                                                                                            | 0 to 30 L/min                                                                                                                                                          | 0 to 30 L/min                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                   |
| Ti (Inspiration<br>Time)                                                                                                                                         | 0 to 9 seconds                                                                                                                                                         | 0.3 to 4<br>seconds                                                            | Similar                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                   |
| Pulse Rate                                                                                                                                                       | 18 to 300 BPM                                                                                                                                                          | 0 to 300 BPM                                                                   | Similar                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                   |
|                                                                                                                                                                  | SpO₂                                                                                                                                                                   | 40 to 100%                                                                     | 0 to 100%                                                                                                                                                                                                                                                                                                                                                                           | Similar                                                                                                                                           |
|                                                                                                                                                                  |                                                                                                                                                                        |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     | Expansion of range on pulse rate and SpO₂ – the new sensor that is utilised for EasyCare Tx 2 now supports an increased range of data monitoring. |
| ¹Easycare Tx 2 does not contain any PAP device information and receives data packets from AirSense 11 flow generators that contain the range for each parameter. |                                                                                                                                                                        |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Submission Reason:                                                                                                                                               | New Device                                                                                                                                                             |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
| Non-clinical testing:                                                                                                                                            | Non-clinical verification and validation testing completed for EasyCare Tx 2 demonstrated that the device met all intended performance requirements. Testing included: |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | Software verification and validation                                                                                                                                   |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | The following FDA guidance were conformed to:                                                                                                                          |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | Content of Premarket Submissions for Device Software Functions: June 2023                                                                                              |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |
|                                                                                                                                                                  | Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: September 2023                                                   |                                                                                |                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                   |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the ResMed logo. The logo consists of a series of blue to red dots arranged in a curved line above the word "ResMed" in a bold, sans-serif font. The dots transition in color from blue on the left to red on the right, creating a visual gradient.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the ResMed logo. The logo features a curved line made of connected circles that transition in color from blue to red. Below the line is the word "ResMed" in a bold, sans-serif font.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the text "EasyCare Tx 2". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "EasyCare" on the first line and "Tx 2" on the second line.

Image /page/9/Picture/3 description: The image shows the logo for ResMed. The logo consists of the company name "ResMed" in a bold, sans-serif font, positioned below a curved line of colored dots. The dots transition in color from blue on the left to red on the right, creating a gradient effect.

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the ResMed logo. The logo consists of a curved line of dots that transition in color from blue to red. Below the dots is the company name, "ResMed", in a bold, sans-serif font.

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.

{13}------------------------------------------------

Image /page/13/Picture/2 description: The image contains the ResMed logo. The logo features a curved line made up of a series of dots that transition in color from blue to purple to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the ResMed logo. The logo consists of the word "ResMed" in a bold, sans-serif font. Above the word, there is a curved line made up of a series of dots that transition in color from blue to red. The dots are arranged in a wave-like pattern.

| Clinical Testing        | The verification testing performed with EasyCare Tx 2 was software<br>verification and validation against the non-functional requirements<br>and end-to-end functional testing. The predicate testing conducted<br>with EasyCare Tx 2 to the predicate EasyCare Tx System device<br>(K113780) demonstrate substantial equivalence to the technology<br>andoperating principle of the devices.<br>Clinical tests were not required to demonstrate the safety and<br>effectiveness of EasyCare Tx 2.                                                                                                                                                                                                                                                                                                      |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence | The subject device has similar features and interfaces as the device<br>cleared in K113780. The main differences between ResMed's<br>EasyCare Tx System and the EasyCare Tx 2 include:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                         | 1. Compatible therapy devices are not hardcoded into the<br>EasyCare Tx 2 software, and EasyCare Tx 2 can receive data<br>packets from the compatible flow generators (e.g., AirSense<br>11, K203126).<br><br>2. EasyCare Tx 2 contains all the same therapy settings and<br>output parameters as the predicate, but has updated features<br>for auto modes, up-to-date therapy settings, and extended<br>ranges.<br><br>3. The EasyCare Tx 2 is only for use with compatible ResMed<br>devices, including the updated TxLink™ 2. EasyCare Tx 2 is not<br>reverse compatible with EasyCare Tx or TxLink™.<br><br>4. The EasyCare Tx 2 introduces secure communication<br>initialization via CSR, implements code-signing for the installer,<br>and enhances the certificate generation and use process. |
|                         | The subject device has the same intended use and similar<br>technological characteristics as the predicate device (K113780).<br>EasyCare Tx 2 has undergone performance testing to ensure the<br>device is as safe and effective as the predicate, and it can be<br>concluded that the subject device does not raise new issues of safety<br>or effectiveness compared to the predicate device. The similar<br>indications for use, technological characteristics, and performance<br>characteristics for the proposed EasyCare Tx 2 are assessed to be<br>substantially equivalent to the predicate device.                                                                                                                                                                                            |

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image contains the logo for ResMed. The logo features a curved line made up of a series of circles that transition in color from blue to purple to red. Below the curved line, the word "ResMed" is written in a bold, sans-serif font.

## Substantial Equivalence Conclusion:

 The EasyCare Tx 2 is substantially equivalent to the predicate EasyCare Tx System (K113780):

- It has similar intended use
- . It has similar technological characteristics
- . It has similar operating principles
- The differences do not raise any new questions of safety . or effectiveness
- . It is at least as safe and as effective as the predicate device

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K241939](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K241939)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
