← Product Code [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD) · K241216

# myAir (K241216)

_Resmed Corp · BZD · Sep 26, 2024 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K241216

## Device Facts

- **Applicant:** Resmed Corp
- **Product Code:** [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD.md)
- **Decision Date:** Sep 26, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The myAir app is indicated for patients: • prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings. It is an optional software accessory to allow patients to acclimate to their therapy device. • prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process. The device is intended for home and hospital use for: • new and existing patients of ResMed Air10 and Air11 PAP therapy devices and • new users who are prescribed a compatible NightOwl home sleep test (HST).

## Device Story

myAir is a mobile medical application serving as a companion to ResMed Air11 PAP therapy devices and NightOwl home sleep test (HST) hardware. The app connects via Bluetooth to these devices to provide a user interface for temporary control (on/off) and facilitates data transfer between the hardware and cloud backend servers. It acts as a patient engagement and pre-therapy acclimation tool. The app does not perform diagnostic analysis or display diagnostic results; these functions remain with the NightOwl backend server. Used in home and hospital settings by patients, the app enables self-tracking of usage data and aids in the home sleep testing process. By providing a simplified interface for device operation and data transmission, it supports patient adherence and the diagnostic workflow.

## Clinical Evidence

Bench testing only. Non-clinical verification and validation testing confirmed that the addition of NightOwl HST interoperability (Stella functions) met all performance requirements. Testing included software verification, validation, non-functional requirements, and end-to-end functional testing.

## Technological Characteristics

Software-only mobile application. Connectivity via Bluetooth and HTTPS (cellular/wireless internet). Operates as a companion to compatible hardware (ResMed Air11, NightOwl HST). Software functions include device control and data transfer (MDDS). Complies with FDA guidance for software functions and cybersecurity.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Galapagos ([K200565](/device/K200565.md))

## Reference Devices

- NightOwl Companion App ([K220028](/device/K220028.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 26, 2024

ResMed Corp Rose Malonzo Senior Specialist, Regulatory Affairs 9001 Spectrum Center Blvd. San Diego, California 92123

Re: K241216

Trade/Device Name: myAir Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: August 28, 2024 Received: August 28, 2024

Dear Rose Malonzo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241216

Device Name myAir

Indications for Use (Describe)

The myAir app is indicated for patients:

• prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings. It is an optional software accessory to allow patients to acclimate to their therapy device. · prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process.

The device is intended for home and hospital use for:

• new and existing patients of ResMed Air11 PAP therapy devices and

· new users who are prescribed a compatible NightOwl home sleep test (HST).

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for ResMed. The logo features a curved line made of blue and red dots, with the blue dots on the left and the red dots on the right. Below the line is the word "ResMed" in a bold, sans-serif font.

# 510(k) SUMMARY

[As required by 21 CFR 807.92(c)]

| Date of Submission:                        | 25 September 2024                                                                                                                                                                                |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name/Owner:                        | ResMed Corp<br>9001 Spectrum Center Blvd<br>San Diego, CA 92123<br>USA                                                                                                                           |
| Official Contact:                          | Mrs. Rose Malonzo<br>Senior Specialist, Global Product Regulatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Blvd<br>San Diego, CA 92123 USA<br>Tel: 858-285-5670<br>rose.malonzo@resmed.com |
| Device Trade Name:                         | myAir                                                                                                                                                                                            |
| Device Common Name:                        | Ventilator, Non-Continuous (Respirator) [primary]                                                                                                                                                |
| Classification and<br>Classification Name: | II<br>Noncontinuous ventilator (IPPB) [primary]                                                                                                                                                  |
| Product Code:                              | BZD                                                                                                                                                                                              |
| Primary Predicate Device:                  | Galapagos<br>K200565                                                                                                                                                                             |
| Reference Device:                          | NightOwl Companion App (patient mobile application in the<br>NightOwl System)<br>K220028                                                                                                         |
| Submission Reason:                         | Modified device                                                                                                                                                                                  |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the ResMed logo. The logo features a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font. The logo is set against a white background.

# Device Description

myAir is a companion mobile medical application ("app") for self-monitoring use of the compatible devices that serve as a patient Home Sleep Test (HST), pre-therapy, and engagement platform for positive airway pressure (PAP) therapy. The app allows the patient to connect via Bluetooth to a compatible hardware device for temporary control of their prescribed HST or PAP device and to allow self-tracking of device usage data. myAir can also be used as a communication pathway using the Bluetooth connection with the compatible device in order to send or receive data. Analysis of patient diagnostic data or display of diagnostic results are not in scope of myAir features.

The subject device modifies the predicate device by adding interoperability with the NightOwl HST (Stella functions) in addition to sustaining previously cleared medical device function of connecting to the Air 1 platform.

## Indications for Use

The myAir app is indicated for patients:

- prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their ● device with their prescribed settings. It is an optional software accessory to allow patients to acclimate to their therapy device.
- . prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process.

The device is intended for home and hospital use for:

- new and existing patients of ResMed Air10 and Air11 PAP therapy devices and ●
- new users who are prescribed a compatible NightOwl home sleep test (HST). ●

### Non-clinical testing

Non-clinical verification and validation testing completed for NightOwl HST interoperable (Stella) functions introduced with the modifications to the myAir app demonstrated that the device met all intended performance requirements. Testing included:

- Software verification and validation, including non-functional requirements and end-to-end ● functional testing.
The following FDA guidance were conformed to:

Public

- Content of Premarket Submissions for Device Software Functions: June 2023 ●
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket . Submissions: September 2023

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the ResMed logo. The logo consists of a curved line of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

## Substantial Equivalence

The subject device and predicate device have the same intended use, and similar technological characteristics and operating principle.

The technological differences between the predicate device and subject device are:

- The subject device adds interoperability with the NightOwl HST, allowing myAir to control the . hardware HST (on/off) and facilitate data transfer during the home sleep testing process (i.e., non-device MDDS function) to the EctoSense NightOwl backend server for analysis.
- . The subject device sustains the previously cleared medical device function of connecting to the Air11 platform device to allow the subject device app to control (on/off the hardware and to facilitate transfer of data (i.e., non-device MDDS function) between the hardware and cloud backend server.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for ResMed. The logo features a curved line of dots that transition in color from blue to red. Below the dots, the word "ResMed" is written in a bold, sans-serif font.

| Characteristic                                    | Primary Predicate<br>Device:<br>Galapagos<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K200565                                                                                                                                                                                                              | Reference Device:<br>NightOwl Companion<br>App (patient mobile<br>application in the<br>NightOwl System)<br>Manufacturer:<br>EctoSense<br>510(k) Number:<br>K220028                                                                                                                                                                                                                                                           | Subject Device:<br>myAir with Stella functions<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K241216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                            | The Galapagos app is<br>intended for patients who<br>are prescribed a<br>compatible ResMed S10<br>platform device to<br>simulate therapy prior to<br>using their device with<br>their prescribed settings. It<br>is an optional software<br>accessory to allow<br>patients to acclimate to<br>their therapy device. | The NightOwl is a<br>wearable device<br>intended for use in the<br>recording, analysis,<br>displaying, exporting,<br>and storage of<br>biophysical parameters<br>to aid in the evaluation<br>of sleep-related<br>breathing disorders of<br>adult patients<br>suspected of sleep<br>apnea. The device is<br>intended for the clinical<br>and home setting use<br>under the direction of a<br>Healthcare<br>Professional (HCP). | The myAir app is indicated for<br>patients:<br>• prescribed with a<br>compatible ResMed<br>Air11 platform device to<br>simulate therapy prior to<br>using their device with<br>their prescribed settings.<br>It is an optional software<br>accessory to allow<br>patients to acclimate to<br>their therapy device.<br>prescribed a NightOwl<br>wearable device to<br>provide the user interface<br>to operate the connected<br>device and aid in the<br>home sleep testing<br>process.<br>The device is intended for<br>home and hospital use for:<br>• new and existing patients<br>of ResMed Air10 and Air11<br>PAP therapy devices and | The proposed myAir indications for use adding Stella<br>functions (i.e., compatibility with NightOwl HST<br>sensor) with this premarket submission is similar to<br>the predicate device in that the subject device mobile<br>app has software device functions which are<br>considered medical functions associated with<br>Bluetooth connection intended for limited device<br>control and non-device-MDDS data transfer.<br>The difference in the indications for use does not raise<br>new questions of safety or efficacy for the subject<br>device as the devices have the same intended use. |
| Characteristic                                    | Primary Predicate<br>Device:<br>Galapagos<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K200565                                                                                                                                                                                                              | Reference Device:<br>NightOwl Companion<br>App (patient mobile<br>application in the<br>NightOwl System)<br>Manufacturer:<br>EctoSense<br>510(k) Number:<br>K220028                                                                                                                                                                                                                                                           | Subject Device:<br>myAir with Stella functions<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K241216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                   |                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                               | • new users who are<br>prescribed a compatible<br>NightOwl home sleep test<br>(HST).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation<br>Number                              | 21 CFR §868.5905                                                                                                                                                                                                                                                                                                    | 21 CFR §868.2375                                                                                                                                                                                                                                                                                                                                                                                                              | Primary: 21 CFR §868.5905                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar. The modification to add device compatibility<br>does not raise new questions of safety or efficacy for<br>the subject device as the devices<br>have the same intended use.                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification<br>Name                            | Noncontinuous ventilator<br>(IPPB)                                                                                                                                                                                                                                                                                  | Ventilatory effort<br>recorder                                                                                                                                                                                                                                                                                                                                                                                                | Primary: Noncontinuous<br>ventilator (IPPB)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar. The subject device adds device compatibility<br>with the cleared NightOwl device (K220028). Data<br>analysis, processing, and results are out of scope from<br>the subject device functions.                                                                                                                                                                                                                                                                                                                                                                                               |
| FDA Product<br>Code                               | BZD                                                                                                                                                                                                                                                                                                                 | MNR                                                                                                                                                                                                                                                                                                                                                                                                                           | BZD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similar. The subject device adds compatibility with<br>cleared NightOwl device (K220028).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Prescription<br>Use                               | Yes                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Characteristic                                    | Primary Predicate<br>Device:<br>Galapagos<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K200565                                                                                                                                                                                                              | Reference Device:<br>NightOwl Companion<br>App (patient mobile<br>application in the<br>NightOwl System)<br>Manufacturer:<br>EctoSense<br>510(k) Number:<br>K220028                                                                                                                                                                                                                                                           | Subject Device:<br>myAir with Stella functions<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K241216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Patient<br>Population                             | Patients weighing more<br>than 66 lb (30kg)                                                                                                                                                                                                                                                                         | 22 years old and older                                                                                                                                                                                                                                                                                                                                                                                                        | Patients weighing more<br>than 66 lb (30kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical. The difference in patient population for the<br>reference device does not raise new questions of safety<br>or effectiveness for the subject device. Patients are<br>unable to use myAir as intended without connecting to<br>the intended compatible device. Due to the intended<br>compatibility of myAir with the Air11 or NightOwl<br>devices, the same patient population and user<br>environment for the respective compatible hardware<br>devices inherently applies to the subject myAir device.                                                                                  |
| Intended<br>Environment of<br>Use                 | Home                                                                                                                                                                                                                                                                                                                | Clinical and home                                                                                                                                                                                                                                                                                                                                                                                                             | Home                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Patient<br>Contacting                             | No, Galapagos is<br>software.                                                                                                                                                                                                                                                                                       | No, NightOwl<br>Companion App is<br>software.                                                                                                                                                                                                                                                                                                                                                                                 | No, myAir is software.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Display Type                                      | Smartphone display                                                                                                                                                                                                                                                                                                  | Smartphone display                                                                                                                                                                                                                                                                                                                                                                                                            | Smartphone display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device<br>Interoperability<br>and V&V             | ResMed Air11 platform<br>devices                                                                                                                                                                                                                                                                                    | NightOwl HST                                                                                                                                                                                                                                                                                                                                                                                                                  | ResMed Air11 platform<br>devices and EctoSense<br>NightOwl HST                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | The modification for additional compatibility with the<br>cleared NightOwl Companion App (patient mobile app<br>within the NightOwl System, K220028) does not raise<br>new questions of safety or effectiveness.<br>Furthermore, all non-clinical performance bench<br>testing required for the subject device myAir with Stella<br>functions support the intended design of the product,<br>and device safety and effectiveness.                                                                                                                                                                   |
| Characteristic                                    | Primary Predicate<br>Device:<br>Galapagos<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K200565                                                                                                                                                                                                              | Reference Device:<br>NightOwl Companion<br>App (patient mobile<br>application in the<br>NightOwl System)<br>Manufacturer:<br>EctoSense<br>510(k) Number:<br>K220028                                                                                                                                                                                                                                                           | Subject Device:<br>myAir with Stella functions<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K241216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device<br>Connection<br>Requirement               | Yes. Galapagos app must<br>be connected to a<br>compatible therapy device<br>for usage.                                                                                                                                                                                                                             | Yes. NightOwl<br>Companion App must<br>be connected to a<br>compatible HST device<br>for usage.                                                                                                                                                                                                                                                                                                                               | Yes. myAir app must be<br>connected to a<br>compatible HST to<br>operate Stella functions<br>as intended.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Difference in functionality does not raise new<br>questions of safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Control                                    | Yes. Galapagos can<br>control the connected<br>therapy device.                                                                                                                                                                                                                                                      | No. NightOwl<br>Companion App does<br>not allow control of the<br>connected therapy<br>device.                                                                                                                                                                                                                                                                                                                                | Yes. myAir can control<br>the connected therapy<br>or HST device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Difference in functionality from the NightOwl<br>companion app does not raise new questions of safety<br>or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Communication<br>Pathways<br>Available            | Bluetooth and HTTPS<br>(cellular or wireless<br>internet connection)                                                                                                                                                                                                                                                | Bluetooth and HTTPS<br>(cellular or wireless<br>internet connection)                                                                                                                                                                                                                                                                                                                                                          | Bluetooth and HTTPS<br>(cellular or wireless internet<br>connection)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device data to<br>app<br>communication<br>pathway | Pathways:<br>1) Device data transfers to<br>Machine Cloud Service<br>(MCS) then to Galapagos.<br>Therapy device data<br>transferred from device to<br>app to MCS.                                                                                                                                                   | Pathways:<br>HST device data<br>transfers from sensor<br>through the NightOwl<br>Companion App to the<br>EctoSense backend<br>server for analysis.                                                                                                                                                                                                                                                                            | Pathways:<br>1) Device data transfers to<br>Machine Cloud Service<br>(MCS) then to Galapagos.<br>Therapy device data<br>transferred from device to<br>app to MCS.                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Differences in types of data transferred and<br>communication pathways available does not raise new<br>questions of safety or effectiveness. Software functions<br>related to data analysis and display of<br>diagnostic results are out of scope of the subject<br>device myAir app as these medical device software<br>functions are within the NightOwl Backend Server and                                                                                                                                                                                                                       |
| Characteristic                                    | Primary Predicate<br>Device:<br>Galapagos<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K200565                                                                                                                                                                                                              | Reference Device:<br>NightOwl Companion<br>App (patient mobile<br>application in the<br>NightOwl System)<br>Manufacturer:<br>EctoSense<br>510(k) Number:<br>K220028                                                                                                                                                                                                                                                           | Subject Device:<br>myAir with Stella functions<br>Manufacturer:<br>ResMed Corp<br>510(k) Number:<br>K241216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Substantial Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                   |                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                               | 3) HST device data transfers<br>from sensor through the<br>myAir app to EctoSense<br>backend server for analysis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | HCP Dashboard components of the NightOwl System<br>(K220028).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| App Store<br>Availability                         | Google Play Store                                                                                                                                                                                                                                                                                                   | Google Play Store and<br>Apple App Store                                                                                                                                                                                                                                                                                                                                                                                      | Apple App Store                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Difference in mobile app store availability does not<br>raise new questions of safety or efficacy for the subject<br>device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Therapy Settings<br>Changes                       | None. The Galapagos app<br>does not have capabilities<br>of altering therapy settings.                                                                                                                                                                                                                              | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                                | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Adjust Comfort<br>Settings                        | None. The Galapagos app<br>does not have capabilities<br>of adjusting comfort<br>settings.                                                                                                                                                                                                                          | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                                | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Software Level of<br>Documentation                | Moderate                                                                                                                                                                                                                                                                                                            | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                      | Basic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Similar. The differences in level of software<br>documentation are determined in accordance with the<br>FDA guidance "Content of Premarket Submissions for<br>Device Software Functions". The subject device<br>follows FDA's risk-based approach for basic vs.<br>enhanced (2023), while the predicate devices followed<br>FDA's Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical<br>Devices issued on May 11, 2005.                                                                                                                                         |

{8}------------------------------------------------

Image /page/8/Figure/0 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

{9}------------------------------------------------

Image /page/9/Figure/0 description: The image shows the logo for ResMed. The logo features a curved line of dots that transition in color from blue to red. Below the dots, the word "ResMed" is written in a bold, italicized font.

{10}------------------------------------------------

Image /page/10/Figure/0 description: The image shows the logo for ResMed. The logo consists of the company name "ResMed" in a bold, italicized font. Above the name is a curved line made up of a series of circles that transition in color from blue to red.

Public

{11}------------------------------------------------

Image /page/11/Figure/0 description: The image shows the ResMed logo. The logo consists of the word "ResMed" in a bold, sans-serif font. Above the word is a curved line made up of a series of dots that transition in color from blue to red. The dots are arranged in a wave-like pattern.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the ResMed logo. The logo consists of the word "ResMed" in a bold, sans-serif font, with the "R" slightly larger than the other letters. Above the wordmark is a curved line made up of a series of colored dots, transitioning from blue to red. The dots are arranged in a wave-like pattern, creating a visual representation of airflow or breathing.

# Substantial Equivalence Conclusion

The subject device is substantially equivalent to the predicate K200656 because:

- They have the same intended use; ●
- They have similar technological characteristics;
- . They have similar performance characteristics;
- The differences do not raise any new questions of safety or effectiveness; and
- . It is as safe and effective as the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K241216](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K241216)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
