← Product Code [BWF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF) · K192000

# D R Burton OxyPAP (K192000)

_D R Burton Healthcare, LLC · BWF · Feb 19, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K192000

## Device Facts

- **Applicant:** D R Burton Healthcare, LLC
- **Product Code:** [BWF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF.md)
- **Decision Date:** Feb 19, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5690
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The D R Burton® OxyPAP™ device is indicated for the treatment and prevention of atelectasis. It also has the ability to provide supplemental oxygen when used with compressed oxygen. It includes the option of a pressure manometer. The device is for patients (ages 5 years and above) who are capable of following directions for positive airway pressure (PAP) therapy. Normal use of the D R Burton OxyPAP™ should not be more than 24 hours total combined time. The environment of use is for hospital and clinical setting.

## Device Story

Handheld respiratory therapy device; creates positive airway pressure (PAP) via Venturi principle using compressed gas source. Used in hospital/clinical settings under healthcare professional supervision. Features supplemental oxygen port and optional pressure manometer (0-20 cmH2O) for monitoring airway pressure. Patient exhales against adjustable PEEP-type valve to facilitate airway opening. Output is visual pressure reading; assists in treating/preventing atelectasis. Single-patient use; cumulative use < 24 hours.

## Clinical Evidence

Bench testing only. Comparative performance testing conducted against predicate and reference devices across operational ranges. Testing included pressure manometer performance, drop testing, and effects of aging. Biocompatibility confirmed via material certification identical to legally marketed devices.

## Technological Characteristics

Handheld respiratory device; Venturi-based PAP generation; adjustable PEEP-type valve; optional spring-based pressure manometer (0-20 cmH2O range). Materials are externally communicating/mucosal contact, limited duration (<24 hours). Single-patient use. No software or electronic components.

## Regulatory Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

## Predicate Devices

- DHD Healthcare Corp. – Boeing, Positive Airway Pressure (PAP) Therapy Device ([K991300](/device/K991300.md))

## Reference Devices

- Trudell VersaPAP Device ([K173819](/device/K173819.md))
- Ambu Disposable Pressure Manometer ([K040991](/device/K040991.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 19, 2020

D R Burton Healthcare, LLC Paul Dryden Consultant 3936 S Fields St Farmville, North Carolina 27828

Re: K192000

Trade/Device Name: D R Burton OxyPAP Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: February 8, 2020 Received: February 11, 2020

Dear Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Michael Ryan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

## K192000

Device Name

# D R Burton® OxyPAP™M

Indications for Use (Describe)

The D R Burton® OxyPAP™ device is indicated for the treatment and prevention of atelectasis. It also has the ability to provide supplemental oxygen when used with compressed oxygen. It includes the option of a pressure manometer. The device is for patients (ages 5 years and above) who are capable of following directions for positive airway pressure (PAP) therapy.

Normal use of the D R Burton OxyPAP™ should not be more than 24 hours total combined time.

The environment of use is for hospital and clinical setting.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (7/17)

{3}------------------------------------------------

### 510(k) Summary

| Date Prepared                                                                               | 19-Feb-20                                                                                 |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| D R Burton Healthcare, LLC<br>3936 S Fields St<br>Farmville, NC 27828<br>Tel - 252-228-7038 |                                                                                           |
| Official Contact:                                                                           | Dennis Cook - President                                                                   |
| Proprietary or Trade Name:                                                                  | D R Burton OxyPAP                                                                         |
| Common/Usual Name:                                                                          | Incentive Spirometer                                                                      |
| Classification Code/Name:                                                                   | BWF – Incentive Spirometer<br>21 CFR 868.5690<br>Class II                                 |
| Predicate Device:                                                                           | K991300 – DHD Healthcare Corp. – Boeing, Positive Airway<br>Pressure (PAP) Therapy Device |
| Reference Devices:                                                                          | K173819 – Trudell VersaPAPTM Device<br>K040991 – Ambu Disposable Pressure Manometer       |

### Device Description:

The OxyPAP device is a handheld respiratory therapy device that creates a positive airway pressure. The OxyPAP device is intended to be used by pediatric (ages 5 years and above) and adult patients in the hospital environment, under the supervision of a healthcare professional. The OxyPAP device is a single patient use device.

### Indications for Use:

The D R Burton® OxyPAP™ device is indicated for the treatment and prevention of atelectasis. It also has the ability to provide supplemental oxygen when used with compressed oxygen. It includes the option of a pressure manometer. The device is for patients (ages 5 years and above) who are capable of following directions for positive airway pressure (PAP) therapy.

Normal use of the D R Burton OxyPAP™ should not be more than 24 hours total combined time.

The environment of use is for hospital and clinical setting.

{4}------------------------------------------------

| Table 1 – Comparison to Predicate and Reference Devices   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                |                                                                                                                                                                                                                                                                  |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                           | Proposed<br>OxyPAP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate<br>DHD EzPAP                                                                                                                                                                                                                                                                          | Reference<br>Trudell VersaPAP                                                                                                                                                                  | Reference<br>Ambu Pressure Manometer                                                                                                                                                                                                                             |
| 510(k)<br>CFR                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K991300                                                                                                                                                                                                                                                                                         | K173918                                                                                                                                                                                        | K040991                                                                                                                                                                                                                                                          |
| Classification                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 868.5690<br>BWF                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                | 868.2600<br>CAP                                                                                                                                                                                                                                                  |
| Classification name                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Incentive Spirometer                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                | Monitor, airway pressure (includes gauge and/or alarm                                                                                                                                                                                                            |
| Indications for Use                                       | D R Burton OxyPAP is<br>indicated for the treatment<br>and prevention of atelectasis.<br>It also has the ability to<br>provide supplemental oxygen<br>when used with compressed<br>oxygen. The device is for<br>patients (ages 5 years and<br>above) who are capable of<br>following directions for<br>positive airway pressure<br>(PAP) therapy. Normal use of<br>the D R Burton OxyPAPTM<br>should not be more than 24<br>hours total combined time.<br>The environment of use is for<br>hospital and clinical setting. | Boeing is indicated for the<br>treatment and prevention of<br>atelectasis. Boing facilitates<br>opening of airways in patients<br>requiring prevention or<br>treatment of atelectasis.<br>Boeing also has the ability to<br>provide supplemental oxygen<br>when used with compressed<br>oxygen. | VersaPAPTM is indicated<br>for the treatment and<br>prevention of atelectasis.<br>VersaPAPTM also has the<br>ability to provide<br>supplemental oxygen<br>when used with<br>compressed oxygen. | To provide visual indication of a<br>patient's airway pressure during<br>ventilation. It may be attached to the<br>manometer port or proximal port on<br>ventilation devices such as resuscitation<br>bags, hyperinflation bags, CPAP mask,<br>or CPAP circuits. |
| Patient Population                                        | Patients (ages 5 years and<br>above) who are capable of<br>following directions for<br>positive airway pressure<br>(PAP) therapy.                                                                                                                                                                                                                                                                                                                                                                                         | Patients requiring therapy for<br>treatment and prevention of<br>atelectasis who are capable of<br>following directions for PAP<br>therapy                                                                                                                                                      | Patients (ages 5 years and<br>above) who are capable of<br>following directions for<br>positive airway pressure<br>(PAP) therapy.                                                              | No population                                                                                                                                                                                                                                                    |
| Environments                                              | Hospital and clinical settings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Hospital and clinical settings                                                                                                                                                                                                                                                                  | Hospital and clinical<br>settings                                                                                                                                                              | Hospital and clinical settings                                                                                                                                                                                                                                   |
|                                                           | Proposed<br>OxyPAP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate<br>DHD EzPAP                                                                                                                                                                                                                                                                          | Reference<br>Trudell VersaPAP                                                                                                                                                                  | Reference<br>Ambu Pressure Manometer                                                                                                                                                                                                                             |
| Operating principles                                      | • Venturi principle created by<br>compressed gas source<br>• Supplemental oxygen port<br>• Measures airway pressure<br>with manometer<br>• Adjustable pressure                                                                                                                                                                                                                                                                                                                                                            | • Venturi principle created by<br>compressed gas source<br>• Supplemental oxygen port                                                                                                                                                                                                           | • Venturi principle created<br>by compressed gas<br>source<br>• Supplemental oxygen<br>port<br>• Measures airway<br>pressure with<br>manometer<br>• Adjustable pressure                        | • Measures airway pressure with<br>manometer                                                                                                                                                                                                                     |
| Single patient,<br>multi-use                              | Yes. Typical use is < 14 days, less than 24 hours cumulative use                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                |                                                                                                                                                                                                                                                                  |
| Ability to adjust flow<br>rate                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                              |
| Ability to adjust<br>pressure during<br>exhalation        | Yes, via a “PEEP” type valve                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes via an adjustable opening<br>and via flow rate                                                                                                                                                                                                                                              | Yes adjustable                                                                                                                                                                                 | N/A                                                                                                                                                                                                                                                              |
| Maximum<br>recommended pressure                           | < 20 cmH2O                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                              |
| Pressure manometer<br>range                               | 0-20 cmH2O                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | No manometer                                                                                                                                                                                                                                                                                    | 0-20 cmH2O                                                                                                                                                                                     | 0-20 cmH2O                                                                                                                                                                                                                                                       |
| PAP Flow rate                                             | 5-15 Lpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                              |
| Pressure                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                 | N/A                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                              |
| Mean PEEP<br>(cmH2O)<br>@800 ml TV                        | 2.81 @ 5 Lpm<br>12 @ 10 Lpm<br>12.75 @ 15 Lpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 2 @ 5 Lpm<br>8 @ 10 Lpm<br>16 @ 15 Lpm                                                                                                                                                                                                                                                          | N/A                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                              |
| Mean Peak Expiratory<br>Pressure<br>(cmH2O)<br>@800 ml TV | 15 Lpm / 15 cmH2O<br>20 (400 TV) / 21 (800 TV)                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 15 Lpm / 15 cmH2O<br>22 (400 TV) / 26 (800 TV)                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                              |
|                                                           | Proposed<br>OxyPAP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate<br>DHD EZPAP                                                                                                                                                                                                                                                                          | Reference<br>Trudell VersaPAP                                                                                                                                                                  | Reference<br>Ambu Pressure Manometer                                                                                                                                                                                                                             |
| Mean Inspiratory<br>Pressure<br>(cmH2O)<br>@800 ml TV     | -0.7 @ 5 Lpm<br>-4.7 @ 10 Lpm<br>-3.46 @ 15 Lpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Not known                                                                                                                                                                                                                                                                                       | Not known                                                                                                                                                                                      |                                                                                                                                                                                                                                                                  |
| Mean Expiratory<br>Pressure<br>(cm H2O)<br>@ 800 ml TV    | 3.1 @ 5 Lpm<br>12.1 @ 10 Lpm<br>13 @ 15 Lpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 5 @ 5 Lpm<br>11 @ 10 Lpm<br>20 @ 15 Lpm                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                            |                                                                                                                                                                                                                                                                  |
| Shelf-life                                                | 1 year                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | N/A                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                              |
| Biocompatibility                                          | Externally communicating, tissue and Surface Contact, mucosal contact, limited duration (<24 hours)<br>Use of material certification<br>confirmed materials and<br>processing were identical to a<br>legally marketed device.                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                              |

### Table 1 - Comparison to Predicate and Reference Devices

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

The OxyPAP device is substantially equivalent to the predicate device because:

## Indications -

The proposed indications for use for the treatment and prevention of pulmonary atelectasis and has the ability to provide supplemental oxygen when used with compressed oxygen are similar. Discussion -

# The subject device has similar indications for use as the predicate.

## Patient Population -

The patient population is similar, where we have specified the age, the predicate did not.

# Discussion -

The subject device has similar population as the predicate.

## Environment of Use -

Hospital and healthcare setting due to the requirement for compressed air / oxygen source.

## Discussion -

The subject device has similar environments of use as the predicate.

## Technology -

The technology for generating PAP is similar to the predicate. The ability to adjust pressure and measure airway pressure is similar to the reference K173918 - Trudell VersaPAP.

The technology of measuring and monitoring pressure via a spring pressure manometer is similar to the references K173918 and K040991.

### Discussion -

The subject device has similar technology and principle of operation, interface with the patient as the predicate.

The predicate was not cleared with the ability to attach a pressure manometer, yet the reference K173918 which has similar indications for use does have this feature.

## Performance –

The differences in performance features is compared to the predicate and reference for pressure monitoring and they were not significant and can be found to be substantially equivalent.

# Discussion -

The subject device has substantially equivalent performance to the predicate and applicable reference devices.

## Non-clinical Comparative Performance

### Biocompatibility -

The materials in patient contact are considered as having 2 types of patient contact:

- External communicating, tissue contacting, limited duration and
- Surface contact, mucosal contact, limited duration for the mouthpiece ●
- . Use of material certification confirmed materials and processing were identical to a legally marketed device.

### Bench Testing -

Performance testing included:

- Comparative performance testing to the predicate across the range of performance .

{8}------------------------------------------------

- Comparative performance of the pressure manometer to the reference ●
- . Drop testing
- . Effects of aging

## Discussion of Differences

The subject device incorporates a pressure manometer which the predicate does not. However, we provide a reference device. K173819, which has the same indications for use and incorporates a pressure manometer. We evaluated and compared the performance of the pressure manometer to another reference device, K040991, as K173819 did not include performance data and samples were not available to test.

There were no differences which would raise new or different concerns of safety and effectiveness.

# Substantial Equivalence Conclusion

Based upon the comparative performance testing we have demonstrated that the proposed device compared to the predicate and reference devices can be found to be substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K192000](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K192000)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
