← Product Code [BTR](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTR) · K141939

# TELEFELX ISIS HVT TRACHEAL TUBE, CUFFED WITH SUBGLOTTIC SECRETION SUCTION PORT; WITH STYLET (K141939)

_Teleflexmedical, Inc. · BTR · Nov 24, 2014 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTR/K141939

## Device Facts

- **Applicant:** Teleflexmedical, Inc.
- **Product Code:** [BTR](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTR.md)
- **Decision Date:** Nov 24, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The TELEFLEX ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Port is indicated for airway management by oral intubation during mechanical ventilation and anesthesia including the capability to aid in the removal of subglottic secretions.

## Device Story

Single-use, sterile tracheal tube for adult airway management. Device features polyvinyl chloride tube body, cuff, inflation line, pilot balloon, one-way valve, and radiopaque line. Includes dorsal lumen with opening above cuff and male suction connector for subglottic secretion removal via suction accessory pack. Used in clinical environments where endotracheal intubation is performed. Physician or qualified clinician inserts tube orally to maintain airway during mechanical ventilation or anesthesia. Suction port allows removal of secretions accumulating above cuff, potentially reducing risk of ventilator-associated pneumonia. Output is physical airway access and secretion drainage capability.

## Clinical Evidence

Bench testing only. Testing included ink adherence (ISO 5361/ASTM D3002), cytotoxicity (ISO 10993-5), irritation and sensitization (ISO 10993-10), and extractables/leachables analysis (GC/MS, LC/MS, LC/UV, ICP, FTIR).

## Technological Characteristics

Polyvinyl chloride tube body; cuff inflation line; pilot balloon; one-way valve; 15mm connector; Murphy eye style. Biocompatibility per ISO 10993-1. Sterilization via Ethylene Oxide. Mechanical device; no software or energy source.

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port ([K091761](/device/K091761.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 24,2014

Teleflex Medical, Inc. James Cochie Sr. Regulatory Affairs Specialist 2917 Weck Dr. Research Triangle Park, NC 27709

Re: K141939

> Trade/Device Name: Teleflex ISIS HVT Tracheal Tube, Cuffed With Subglottic Secretion Suction Port Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: October 28, 2014 Received: October 29, 2014

Dear Mr. Cochie,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Runne DDS, mA

Erin Keith Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K141939

Device Name

TELEFLEX ISIS™ HVTM Tracheal Tube, Cuffed with Subglottic Secretion Suction Port

Indications for Use (Describe)

The TELEFLEX ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Port is indicated for airway management by oral intubation during mechanical ventilation and anesthesia including the capability to aid in the removal of subglottic secretions.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port

#### 510(k) Summary

## 510(k) SUMMARY

#### TELEFLEX ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port

#### A. Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8083 Fax: 919-433-4996

#### B. Contact Person

James Cochie Sr. Regulatory Affairs Specialist

#### C. Date Prepared

November 24, 2014

### D. Device Name

Trade Name: Teleflex ISIS™ HVTTM Tracheal Tube, Cuffed with Subglottic Secretion Suction Port

Common Name: Tracheal Tube

Classification Name: Tracheal Tube (W/Wo Connector) (21 CFR 868.5730, Product Code BTR)

### E. Device Description

The Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is a single-use, sterile tracheal tube made of a polyvinyl chloride tube body, cuff inflation line and a compatible pilot balloon and one-way valve. A radiopaque line is incorporated into the full length of the tracheal tube. Each tracheal tube is supplied with an appropriately sized 15mm connector. The Teleflex ISIS™ HVTTM Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is sold in Murphy eye style. Each tube has a dorsal lumen with an opening above the cuff and a male suction connector port attached to the tube, which is close to the machine end of the tube. Access to the suction lumen is accomplished via a connection of the suction accessory pack (P/N 5-23000) directly to the male suction connector.

{4}------------------------------------------------

## Section 29 - 510(k) Summary

#### F. Indications for Use

The Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is indicated for airway management by oral intubation during mechanical ventilation and anesthesia including the capability to aid in the removal of subglottic secretions.

#### G. Target Population

This device is intended for use on adult patients

#### H. Intended Environment of use

This device is intended for locations where endotracheal intubation may be performed.

This product is single use only.

#### I. Contraindications

There are no known contraindications.

### J. Substantial Equivalence

The modified Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is substantially equivalent to the predicate device:

| Comparative<br>Characteristics | Proposed Device,<br>Teleflex ISIS™ HVT™<br>Tracheal Tube, Cuffed with<br>Subglottic Secretion<br>Suction Port | Predicate Device,<br>Teleflex ISIS™ HVT™<br>Tracheal Tube, Cuffed with<br>Subglottic Secretion<br>Suction Port                                                                                                                                                                              |
|--------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                   | Teleflex Medical, Inc.                                                                                        | Teleflex Medical, Inc.                                                                                                                                                                                                                                                                      |
| 510(k) Number                  | N/A                                                                                                           | K091761                                                                                                                                                                                                                                                                                     |
| Indications for Use            | Identical                                                                                                     | The Teleflex ISIS™ HVT™<br>Tracheal Tube, Cuffed with<br>Subglottic Secretion Suction<br>Port is indicated for airway<br>management by oral<br>intubation during mechanical<br>ventilation and anesthesia<br>including the capability to aid<br>in the removal of subglottic<br>secretions. |
| Main Tube                      | Identical                                                                                                     | Yes                                                                                                                                                                                                                                                                                         |
| Inflation Line                 | Identical                                                                                                     | Yes                                                                                                                                                                                                                                                                                         |

{5}------------------------------------------------

## Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port

| Pilot Balloon                   | Identical                                                                           | Yes                                                                                     |
|---------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Bi-directional Valve            | Identical                                                                           | Yes                                                                                     |
| Cuff Balloon                    | Identical                                                                           | Yes                                                                                     |
| 15mm Connector                  | Identical                                                                           | Yes                                                                                     |
| Male Suction Lumen<br>Connector | Identical                                                                           | Yes                                                                                     |
| Black Ink                       | Printing Ink – 2405 Black Ink, Markem-Imaja Solvent – Medical Ink Reducer, Colorcon | Black Ink, Nazdar Thinner, Nazdar Retarder, Nazdar                                      |
| Descriptive Size (ID)           | Identical                                                                           | 6.0mm, 6.5mm, 7.0mm,<br>7.5mm, 8.0mm, 8.5mm,<br>9.0mm                                   |
| Biocompatibility                | Identical                                                                           | Meets ISO 10993-1                                                                       |
| Sterilization                   | Ethylene Oxide                                                                      | Ethylene Oxide                                                                          |
| Single Use                      | Yes                                                                                 | Yes                                                                                     |
| Shelf Life                      | 12 Months (extending to 60<br>months as data becomes<br>available)                  | 12 Months (extending to 60<br>months as data becomes<br>available)                      |
| Packaging                       | Identical                                                                           | 10 individually packaged in<br>either banana pack or Tyvek<br>pouches per dispenser box |

## Section 29 – 510(k) Summary

## K. Non-clinical Comparative Performance Testing

Bench testing has been performed to verify that the performance of the proposed Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is substantially equivalent to the predicate device, and that the Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port will perform as intended.

| Test          | Reference to Standard (if         | Principle of Test                            |
|---------------|-----------------------------------|----------------------------------------------|
| Performed     | applicable)                       |                                              |
| Ink adherence | ISO 5361, Section 5.3.2 (check    | Ink adherence is tested by application       |
|               | compliance by inspection, as      | and removal of tape to the inked surface     |
|               | indicated in 6.4.1 of ASTM        | according to the test parameters and         |
|               | D3002.                            | inspected to a rating scale.                 |
| Cytotoxicity  | ISO 10993-5, Biological           | Tests the potential biological reactivity of |
|               | Evaluation of Medical Devices -   | a mammalian cell culture in response to      |
|               | Part 5: Tests for In Vitro        | exposure to the extract of the test article. |
|               | Cytotoxicity                      |                                              |
| Irritation    | ISO 10993-10, Biological          | Test the potential of the test article to    |
|               | Evaluation of Medical Devices -   | produce primary skin irritation after        |
|               | Part 10: Tests for Irritation and | topical application when extracted in        |
|               | Skin Sensitization                | sodium chloride for injections and           |
|               |                                   | cottonseed oil.                              |

{6}------------------------------------------------

## Teleflex ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port

| Sensitization                | ISO 10993-10, Biological<br>Evaluation of Medical Devices –<br>Part 10: Tests for Irritation and<br>Skin Sensitization | Test the potential of the test article to<br>produce allergenic reaction when<br>extracted in sodium chloride and<br>cottonseed oil.          |
|------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Extractables<br>& Leachables | N/A                                                                                                                    | Assesses the ET tube relative to<br>leachable/extractable profiles with<br>GC/MS, LC/MS, LC/UV, ICP and FTIR<br>under clinical use conditions |

## Section 29 – 510(k) Summary

## L. Substantial Equivalence

The modified Teleflex ISIS™ HVTTM Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is substantially equivalent in intended use, design, performance and principles of operation to the identified predicate device cleared under 510(k) K091761. The differences between the ISIS Tube and the predicated device are minor and raise no new issues of safety and efficacy. The ISIS Tube is substantially equivalent to the currently marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTR/K141939](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTR/K141939)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
