← Product Code [BSN](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSN) · K133004

# VENTED MILLEX -GV, CATHIVEX -GV (K133004)

_Merck Millipore , Ltd. · BSN · Mar 27, 2014 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSN/K133004

## Device Facts

- **Applicant:** Merck Millipore , Ltd.
- **Product Code:** [BSN](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSN.md)
- **Decision Date:** Mar 27, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5130
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

Vented Millex®-GV filter units are syringe filters for sterilization of low volume aqueous solutions for direct patient injection, filtration of laboratory solutions, and filtration of clinical solutions where venting and low protein binding membrane are required or desired. Cathivex®-GV filter units are in-line gravity fed filters for the sterilization of aqueous solutions for administration via the neuraxial route, filtration of laboratory solutions, and filtration of clinical solutions where venting and low protein binding membrane are required or desired.

## Device Story

Vented Millex®-GV and Cathivex®-GV are sterile, single-use, non-pyrogenic filter units. Vented Millex®-GV functions as a syringe filter; Cathivex®-GV functions as an in-line gravity-fed filter. Both utilize a 0.22um hydrophilic PVDF membrane to remove particles, microorganisms, and microprecipitates, and a 0.03um hydrophobic PTFE vent membrane to prevent air locks by automatically venting upstream air. Devices are used in clinical settings for sterile filtration of aqueous solutions. Healthcare providers (clinicians) attach the device to a syringe or administration line to ensure solution sterility before patient injection or neuraxial administration. The low protein binding property of the PVDF membrane preserves solution integrity. Benefits include reduced risk of particulate/microbial contamination and improved flow consistency via air venting.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including filter integrity, burst, flow rate, bubble point, water intrusion, endotoxin (LAL), particle count, bacterial retention, and biocompatibility testing per ISO 10993-1.

## Technological Characteristics

Filter units consist of a 0.22um hydrophilic PVDF membrane and a 0.03um hydrophobic PTFE vent membrane. Housing is molded from PVC. Vented Millex®-GV features Female Luer Lok™ inlet/Male Luer slip outlet; Cathivex®-GV features Female Luer Lok™ inlet/Male Luer Lok™ outlet. Sterilized via ethylene oxide (EO) to SAL 10^-6. Biocompatibility per ISO 10993-1.

## Regulatory Identification

An anesthesia conduction filter is a microporous filter used while administering to a patient injections of local anesthetics to minimize particulate (foreign material) contamination of the injected fluid.

## Predicate Devices

- Millex®-GV, 33mm ([K023892](/device/K023892.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for Merck Millipore. The logo consists of a large, stylized letter "M" with a break in the upper left portion of the letter. To the right of the "M", the words "MERCK MILLIPORE" are printed vertically, with the letters stacked on top of each other. The logo is black and white.

| 510(k) Summary                                        | K133004                                                                                                                 |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                                         | March 14, 2014                                                                                                          |
| Name, address and<br>telephone number of<br>submitter | Merck Millipore Ltd.<br>Tullagreen, Carrigtwohill<br>Co. Cork, Ireland<br>Phone: +353.21.4532308 / Fax: +353.21.4883048 |
| Contact Person                                        | Chris Parr, Regulatory Affairs Specialist III                                                                           |
| Device Trade Names                                    | Vented Millex®-GV, Cathivex®-GV                                                                                         |
| Device Common Name                                    | Filter Units                                                                                                            |
| Model Numbers                                         | SLGVV255F, SLGVN250S                                                                                                    |
| Device Class                                          | II                                                                                                                      |
| Product Code                                          | BSN                                                                                                                     |
| Regulation Number                                     | 868.5130                                                                                                                |

Anesthesia conduction filter

# Device Description

Regulation Name

Vented Millex®-GV and Cathivex®-GV filter units are sterile, non-pyrogenic, single-use filter devices intended for sterilizing aqueous solutions for medical applications. Vented Millex®-GV is designed with a Female Luer Lok™ inlet and a Male Luer slip outlet, and Cathivex®-GV is designed with a Female Luer Lok™ inlet and a Male Luer Lok™ Both devices contain a 0.22um Durapore® hydrophilic filter membrane outlet. constructed from polyvinylidene fluoride (PVDF) and a 0.03um hydrophobic vent membrane constructed from polytetrafluoroethylene (PTFE). The filter membrane is designed to remove particles, microorganisms, microprecipitates and undissolved powders which are larger than 0.22 um. The vent membrane is designed to prevent air locks by automatically venting air introduced upstream. The filter housing material is molded from PVC.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Merck Millipore. The logo consists of a stylized letter "M" with a small "i" to the left of it. To the right of the "M" is the text "MERCK MILLIPORE" in a vertical arrangement. The logo is in black and white.

### Indications for Use

Vented Millex®-GV filter units are syringe filters for sterilization of low volume aqueous solutions for direct patient injection, filtration of laboratory solutions, and filtration of clinical solutions where venting and low protein binding membrane are required or desired. Cathivex®-GV filter units are in-line gravity fed filters for the sterilization of aqueous solutions for administration via the neuraxial route, filtration of laboratory solutions, and filtration of clinical solutions where venting and low protein binding membrane are required or desired.

#### Predicate Device

Vented Millex®-GV and Cathivex®-GV are substantially equivalent to the following legally marketed predicate device, Millex®-GV.

| Applicant           | Millipore Corporation                                                                                                                                           |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name   | Millex®-GV, 33mm                                                                                                                                                |
| Model Number        | SLGVM33RS                                                                                                                                                       |
| 510(k) Number       | K023892                                                                                                                                                         |
| Clearance date      | 1/17/2003                                                                                                                                                       |
| Device Class        | I I                                                                                                                                                             |
| Product Codes       | BSN, FPB                                                                                                                                                        |
| Regulation Numbers  | 868.5130, 880.5440                                                                                                                                              |
| Regulation Names    | Anesthesia conduction filter, Intravascular administration sets                                                                                                 |
| Indications for Use | Intended for use as a syringe filter to sterilize, ultraclean or<br>clarify low volume solutions in direct patient care and<br>pharmacy admixture applications. |

# Predicate Device Description

Millex®-GV 33mm syringe filter units are sterile, non-pyrogenic, single use syringe filter units intended for sterilizing and/or clarifying small volumes of aqueous solutions, alcohols and proteinaceous solutions. Millex@-GV 33mm is designed with a Female Luer Lok™ inlet and a Male Luer slip outlet. They are designed to remove by filtration, particles, microorganisms, microprecipitates and undissolved powders larger than the

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image is a logo for Merck Millipore. The logo consists of a stylized letter "M" on the left and the words "MERCK MILLIPORE" written vertically on the right. The letter "M" is bold and black, and the text is also black.

rated pore size of the device. The filter membrane is Durapore® (PVDF) with a pore size of 0.22um (micron). The filter housing material is molded from modified acrylic. Typical applications include the sterile filtration and/or clarification of protein pharmaceuticals, diagnostic imaging agents, chemotherapeutics and associated low volume solutions in direct patient care and pharmacy admixture applications.

#### Summary of Technological Characteristics

The Intended Use for Vented Millex®-GV and Cathivex®-GV is the same, and is similar to the predicate device in regards to the sterilization of solutions for medical applications. The predicate device includes additional applications in the Intended Use for ultracleaning and clarification. These applications are not part of the Intended Use for Vented Millex®-GV and Cathivex®-GV, and do not alter the intended therapeutic effect when compared to the predicate.

Vented Millex®-GV and Cathivex®-GV include an application in the Indications for Use The introduction of a vent is considered an incremental product for venting. enhancement designed to prevent air locks caused by air introduced upstream. The introduction of a vent membrane is not considered to adversely affect safety and effectiveness. Vented Millex®-GV and Cathivex®-GV also include an application in the Indications for Use for low protein binding. This application is making use of the existing properties of the Durapore® PVDF filter membrane used in the predicate device. Durapore® PVDF is a low protein binding material, and as such, the predicate device would also be considered to be low protein binding.

From a design perspective there are similarities and differences between the predicate device and the subject devices. The key similarity is the filter membrane type and pore size which is the same. The key difference is the introduction of a vent membrane for venting of air introduced upstream. The materials used in the molding of the filter housing of the predicate device are different to those of the subject devices. PVC is considered a well established material offering suitable mechanical properties for the application. Biocompatibility testing has been performed to address the differences between materials of the predicate and subject devices.

The performance of the subject devices has been verified through bench testing and the filter units have been demonstrated to perform as intended. The subject devices do not introduce new risks or adversely affect safety and effectiveness.

{3}------------------------------------------------

# Summary of Nonclinical Testing

The following design verification and validation tests were conducted on Vented All specified performance Millex®-GV and Cathivex®-GV post-sterilization. requirements were met.

## Device Testing

- Visual Inspection .
- Filter Integrity Test .
- Burst Test .
- Flow Rate Test (Vented Millex®-GV only) .
- Gravity Flow Test (Cathivex®-GV only) .
- Housing Test (Vented Millex®-GV only) .
- Bubble Point Test .
- Water Intrusion Test .
- Endotoxin LAL Test .
- Particle Count Downstream Test .
- Gravimetric Test .
- Luer Insertion Test .
- Bacterial Retention Test .
- Simulated Use Test .
- Packaging Test .
- Hold Up Volume Test .
- Mouse Safety Test .

# Packaging Testing

- Peelability Test .
- Dye Test .
- Strength of Blister Seal and Burst Strength Test .
- Blister Seal Width .
- Unit Packaging including Print Inspection .
- Product Sterility Test .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Merck Millipore. The logo consists of a large, stylized letter "M" with a small "i" to the left of it. To the right of the "M", the words "MERCK MILLIPORE" are written vertically in a smaller font. The logo is black and white.

## Biocompatibility

- Biocompatibility testing was conducted in accordance with standard ISO 10993-. 1:2009 Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process, and the FDA Blue Book Memorandum #G95-1 Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and demonstrated acceptable results.
#### Sterilization

- Vented Millex®-GV and Cathivex®-GV are sterilized with ethylene oxide (EO) . gas using a validated sterilization cycle. Sterilization was conducted in accordance with ISO 11135-1:2007 Sterilization of health care products -Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and FDA Guidance Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.
- Vented Millex®-GV and Cathivex®-GV Filter Units are sterilized using Ethylene . Oxide (EO) at 100% gas concentration in a 3 hour 30 min exposure cycle at the target parameters for temperature, relative humidity and gas concentration. The sterilization cycle has been validated to provide a Sterility Assurance Level (SAL) of 10-6.

#### Shelf Life

- Vented Millex®-GV and Cathivex®-GV have a 1 year shelf life that is supported . by accelerated and real-time stability studies.
- The results demonstrate that the devices maintain their performance and sterility . throughout the duration of the study and support a 1 year shelf life.

#### Summary of Clinical Testing

This 510(k) premarket notification does not contain any clinical performance testing data obtained from clinical investigations or from literature sources with the predicate or subject devices for the purposes of demonstrating substantial equivalence or safety and effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Merck Millipore. The logo consists of a large, stylized letter "M" in black, with a smaller "i" placed to the left of the "M". To the right of the "M", the words "MERCK MILLIPORE" are written vertically in a sans-serif font.

# Summary and Conclusions from Non-Clinical and Clinical Testing

Vented Millex®-GV and Cathivex®-GV Filter Units are substantially equivalent to the predicate device Millex®-GV 33mm because they share the following similarities:

- Device Class (class II) .
- Product Code (BSN) .
- Intended Use (sterilization) ●
- Filter Membrane Type (Durapore® PVDF, 0.22 um) ●
- Packaging Design (Blister with medical paper lid) ●
- High degree of similarity in design and principle of operation .

Non-Clinical performance testing was conducted for Vented Millex®-GV and Cathivex®-GV in order to demonstrate that the filter units perform as intended and meet user needs and intended uses. All specified performance requirements were met for Device Testing and Packaging Testing. Biocompatibility testing was conducted in accordance with ISO 10993-1 and Blue Book Memorandum #G95-1, and demonstrated acceptable results. Sterilization Validation was completed in accordance with ISO 11135-1:2007 to achieve a Sterility Assurance Level (SAL) of 106. Shelf life testing demonstrated that the devices maintain sterility and performance throughout their 1 year shelf life.

Based on the data presented in this 510(k), the subject devices are sufficiently similar in design and intended use to the predicate device that they are considered to be substantially equivalent as supported by Non-Clinical performance testing. Any differences that do exist are deemed not to significantly affect the safety and effectiveness of the devices.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the branches of government. The logo is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is in all caps.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

#### March 27, 2014

Merck Millipore Limited Mr. Chris Parr Regulatory Affairs Specialist III Tullagreen. Carrigtwohill Co. Cork lreland

Re: K133004

Trade/Device Name: Vented Millex®-GV, Cathivex®-GV Regulation Number: 21 CFR 868.5130 Regulation Name: Anesthesia Conduction Filter Regulatory Class: Il Product Code: BSN Dated: January 13, 2014 Received: January 16. 2014

Dear Mr. Parr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. Iabeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not mislcading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

{7}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S. Runner -S

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

{8}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K133004

Device Name Vented Millex®-GV, Cathivex®-GV

Vented Millex®-GV filter units are syringe filters for sterilization of low volume aqueous solutions for direct patient injection, filtration of laboratory solutions, and filtration of clinical solutions where venting and low protein binding membrane are required or desired.

Cathivex®-GV filter units are in-line gravity fed filters for the sterilization of aqueous solutions for administration via the neuraxial route, filtration of laboratory solutions, and filtration of clinical solutions where venting and low protein binding membrane are required or desired.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY .

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

| ారామెడియా ప్రాచ్చిన్నీ                                                          |
|---------------------------------------------------------------------------------|
| 2 44 45 11 11 11                                                                |
| and the country of the same of the states<br>and and the first the stage on the |
|                                                                                 |

Digitally signed by Richard C. Chapman Date: 2014.03.26 19:26:38 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

. - - -

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSN/K133004](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSN/K133004)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
