← Product Code [BZQ](/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ) · K220111

# RespiraSense (K220111)

_Pmd Solutions · BZQ · Oct 7, 2022 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ/K220111

## Device Facts

- **Applicant:** Pmd Solutions
- **Product Code:** [BZQ](/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ.md)
- **Decision Date:** Oct 7, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The RespiraSense is indicated for continuous, non-invasive, and real-time monitoring of respiratory rate. RespiraSense is indicated for patients 18 years and older in hospital-type facilities and while patients are mobile (e.g., walking). RespiraSense is not intended to be an apnoea monitor.

## Device Story

RespiraSense (RS) is a wireless, body-worn, non-invasive respiratory rate (RR) monitor. Input: thoracic and abdominal displacement signals measured via piezoelectric film elements. Operation: piezoelectric sensor detects chest/abdomen movement; integrated accelerometer corrects for motion artifacts; dsPIC microcontroller processes signals to derive RR. Output: RR data transmitted via Bluetooth Low Energy (BLE) to an iPad running the RS App. Usage: hospital-type facilities; operated by medical staff for continuous monitoring (up to 4 days). RS App allows configuration, alert limit setting, and optional display of SpO2/Pulse rate from connected Nonin devices. Benefits: enables reliable remote RR monitoring for mobile patients, assisting clinical decision-making regarding patient respiratory status.

## Clinical Evidence

Multi-site comparative clinical study (n=56) in intended population (adults 18+). Compared RespiraSense to gold standard EtCO2 monitor. Observers blinded to EtCO2. Primary endpoint: RR accuracy. Results demonstrated substantial equivalence to gold standard with accuracy of ±3 bpm at 95% confidence level.

## Technological Characteristics

Materials: single-use biocompatible adhesive (ISO 10993-5, ISO 10993-10). Sensing: piezoelectric film elements, accelerometer. Energy: rechargeable Li-Ion battery. Form factor: body-worn Lobe (57x98x18mm) + adhesive sensor. Connectivity: Bluetooth 4.2. Standards: AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ANSI IEEE C63.27, AIM 7351731.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- Philips Intellivue CL Respiration Pod ([K122223](/device/K122223.md))

## Reference Devices

- Masimo Rainbow SET Acoustic Monitoring sensor ([K120984](/device/K120984.md))
- Biobeat ([K212153](/device/K212153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

October 7, 2022

PMD Solutions % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE St. Petersburg, Florida 33704

Re: K220111

Trade/Device Name: RespiraSense Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BOZ Dated: August 30, 2022 Received: August 30, 2022

Dear Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

icies. You must comply with all the Act's

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220111

Device Name RespiraSense

Indications for Use (Describe)

The RespiraSense is indicated for continuous, non-invasive, and real-time monitoring of respiratory rate. RespiraSense is indicated for patients 18 years and older in hospital-type facilities and while patients are mobile (e.g., walking). RespiraSense is not intended to be an apnoea monitor.

| Type of Use (Select one or both, as applicable)                                                                                            |                                                                                                                                           |
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right: 5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right: 5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

# PMD Solutions RespiraSense

#### Date Summary was prepared: October 6, 2022

#### I. SPONSOR

PMD Solutions Bishoptown House Model Farm Road Cork, Ireland T12 T922 Tel - +353 (0) 212428760

| Sponsor Contact: | Myles Murray - President |
|------------------|--------------------------|
|------------------|--------------------------|

| Submission Correspondent: | Paul Dryden              |
|---------------------------|--------------------------|
|                           | ProMedic, LLC            |
|                           | 131 Bay Point Dr NE      |
|                           | St. Petersburg, FL 33704 |

#### II. DEVICE

| Name of the device:  | RespiraSense                 |
|----------------------|------------------------------|
| Common Name:         | Breathing Frequency Monitor  |
| Classification CFR:  | 21 CFR 868.2375              |
| Classification Name: | Monitor, Breathing Frequency |
| Product Code:        | BZQ                          |

### III. PREDICATE DEVICE AND REFERENCE DEVICE

| Predicate Device:    | Philips Intellivue CL Respiration Pod (K122223)                                                                                                                    |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:   | Breathing Frequency Monitor                                                                                                                                        |
| Classification CFR:  | 21 CFR 868.2375                                                                                                                                                    |
| Classification Code: | BZQ / DRG / MSX                                                                                                                                                    |
| Classification Name: | Monitor, Breathing Frequency / Transmitters and Receivers,<br>Physiological Signal, Radio Frequency / System, Network and<br>Communication, Physiological Monitors |
| Reference Device:    | Masimo Rainbow SET Acoustic Monitoring sensor (K120984)                                                                                                            |
| Common/Usual Name:   | Oximeter                                                                                                                                                           |
| Classification CFR:  | 21 CFR 870.2700                                                                                                                                                    |
| Classification Name: | Oximeter / Monitor, Breathing Frequency                                                                                                                            |
| Product Code:        | DQA / BZQ                                                                                                                                                          |

#### IV. DEVICE DESCRIPTION

The RespiraSense (RS) monitors respiratory rate by directly measuring displacements of the chest and abdomen that occur during breathing which are analogous of respiratory rate. This is measured using piezoelectric film elements that output a varying voltage when displaced.

{4}------------------------------------------------

The RS device is a non-invasive, wireless, respiratory rate (RR) monitor that is worn on the body of the patient. It is internally-powered, and enables continuous RR monitoring from a single application for up to four days. It is designed to enhance the ability of medical staff to reliably measure respiratory rate remotely.

The RS Device is comprised of the RespiraSense Lobe (hereafter the Lobe), the reusable component that houses the RS Device's electronics, a single use adhesive RespiraSense Sensor (hereafter the Sensor) and Cradle. The Lobe and Sensor connect via a secured flat cable and are mechanically fastened together using a plastic Cradle. The Lobe and Sensor are placed on the left-hand side of a patient's torso.

The Lobe transmits wirelessly via Bluetooth to an iPad running PMDs RS App for display, control and alert functionality. The RS App allows for

- Configuring the Lobe ●
- Displaying respiration rate ●
- Display SpO2 and Pulse rate from a connected Nonin device (optional)
- Configuring alert limits .

The RS application can also connect to and display information from multiple Lobes simultaneously.

The Lobe is rechargeable. It is designed to be charged using the supplied Charging Station. Charging must be done outside of the immediate patient environment. Charging cannot occur while the device is in operation. Up to six Lobes can be charged at one time.

The Sensor is the only patient-contacting device and is secured to the patient with medical grade adhesive.

## V. INDICATIONS FOR USE

The RespiraSense is indicated for continuous, non-invasive, and real-time monitoring of respiratory rate. RespiraSense is indicated for patients 18 years and older in hospital-type facilities and while patients are mobile (e.g., walking). RespiraSense is not intended to be an apnoea monitor.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE AND REFERENCE DEVICE

The table below outlines the features of the Respiratory Rate Monitor and compares it to the predicate devices to establish substantial equivalence.

{5}------------------------------------------------

| Parameter                    | Subject Device RespiraSense<br>(K220111)                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>Philips Intellivue CL Respiration Pod (K122223)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Reference Device<br>Masimo Rainbow SET Acoustic<br>Monitoring sensor (K120984)                                                                                                                                                                                                                                                                                                                                               | Comparison                                                                                                                                                                                                                    |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Name       | Breathing Frequency Monitor                                                                                                                                                                                                                                                                                                                                                                                             | Breathing Frequency Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Breathing Frequency Monitor                                                                                                                                                                                                                                                                                                                                                                                                  | Similar                                                                                                                                                                                                                       |
| Indications for<br>Use       | The RespiraSense is indicated for<br>continuous, non-invasive, and real-time<br>monitoring of respiratory rate.<br>RespiraSense is indicated for patients<br>18 years and older in hospitals,<br>hospital-type facilities and while patients<br>are mobile (e.g., walking). RespiraSense<br>is not intended to be an apnoea monitor.                                                                                    | The IntelliVue CL Respiration Pod is indicated for<br>use by health care professionals whenever there is a<br>need for intermittent or spot-check acquisition and<br>monitoring of physiological patient parameters<br>respiration rate and pulse rate wirelessly in specific<br>hospital areas. The IntelliVue CL Respiration Pod<br>is mainly indicated for use in general medical and<br>surgery wards and in waiting areas of emergency<br>rooms. It is not indicated for use in hospital areas in<br>which continuous patient monitoring is needed,<br>such as intensive care units or operating rooms.<br>The intended use of the IntelliVue CL Respiration<br>Pod when used together with a patient monitor is<br>for intermittent or spot-check monitoring and<br>recording of, and to generate alarms for,<br>respiration rate and pulse rate of adult patients.<br>The IntelliVue CL Respiration Pod is also<br>intended for acquisition of respiration rate and<br>pulse rate data of adult patients for a clinical<br>information management system.<br>The Intellivue CL Respiration Pod is intended for<br>use by health care professionals. It is not intended<br>for home use. It is not a therapeutic device.<br>The IntelliVue CL Respiration Pod is not intended<br>for use on patients with extremely high values for<br>respiration rate (above 60 rpm).<br>The IntelliVue CL Respiration Pod is not intended<br>for use on acutely ill cardiac patients with the<br>potential to develop life threatening arrhythmias,<br>e.g. very fast atrial fibrillation or ventricular<br>tachycardia (rapid irregular pulse rate). For<br>monitoring of these patients, a device for<br>continuous ECG monitoring is necessary. The<br>IntelliVue CL Respiration Pod is not a substitute<br>for an ECG monitor. | The Rainbow® Acoustic<br>Monitoring sensors and cables are<br>indicated for continuous,<br>noninvasive monitoring of<br>respiratory rate (RRa).<br>The RAS- 125 is indicated for adult<br>patients in hospitals, hospital-type<br>facilities, mobile and home<br>environments.<br>The RAS- 125c is indicated for<br>adult and pediatric patients in<br>hospitals, hospital-type facilities,<br>mobile and home environments. | Similar to both<br>predicate and<br>reference for<br>measuring respiratory<br>rate in the intended<br>population and<br>environment.                                                                                          |
| Parameter                    | Subject Device RespiraSense<br>(K220111)                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>Philips Intellivue CL Respiration Pod<br>(K122223)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Reference Device<br>Masimo Rainbow SET Acoustic<br>Monitoring sensor (K120984)                                                                                                                                                                                                                                                                                                                                               | Comparison                                                                                                                                                                                                                    |
| Measurement<br>Principle     | Thoracic and abdominal movements<br>from respiratory effort. Bending of the<br>piezo sensor   caused by chest and<br>abdomen movements during breathing<br>produce a voltage signal, from which<br>respiration rate is derived. Integrated<br>accelerometer provides means to<br>correct artefacts.                                                                                                                     | Thoracic movements from respiratory effort -<br>Inclination changes of the incorporated<br>accelerometer sensor, caused by chest and<br>abdomen movements during breathing and heart<br>contraction, produce a voltage signal, from which<br>respiration and pulse rate signals are derived.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Use of piezoelectric sensor element<br>for the purpose of non-invasive<br>respiratory rate monitoring. As air<br>flows, the acoustic vibrations are<br>measured by the sensors and<br>processed to determined respiratory<br>rate.                                                                                                                                                                                           | Similar principle of<br>operations, sensor<br>technology of subject<br>device is different<br>from predicate but<br>similar to reference<br>device.                                                                           |
| Patient Interface            | Lobe attached to sensor. Sensor<br>adhesively attached to patient's chest.                                                                                                                                                                                                                                                                                                                                              | Pod attached to strap. Strap adhesively attached to<br>patient.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Sensor adhered to the neck of the<br>patient, close to the windpipe. A<br>cable connects the sensor to<br>a nearby monitor.                                                                                                                                                                                                                                                                                                  | Similar                                                                                                                                                                                                                       |
| Patient<br>Population        | Patients aged 18 or over                                                                                                                                                                                                                                                                                                                                                                                                | Adult                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Adult and pediatric patients                                                                                                                                                                                                                                                                                                                                                                                                 | Similar                                                                                                                                                                                                                       |
| Prescriptive                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar                                                                                                                                                                                                                       |
| Environment                  | Hospital, hospital-type facilities, and<br>mobile                                                                                                                                                                                                                                                                                                                                                                       | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Hospitals, hospital-type facilities,<br>mobile and home environments.                                                                                                                                                                                                                                                                                                                                                        | Similar                                                                                                                                                                                                                       |
| Hardware                     | Small body-worn device with adhesive<br>attachment, data is sent to the<br>RespiraSense Software Application.<br>Lobe hardware consists of a battery and<br>an electronic module with the embedded<br>processor, accelerometer, dsPIC<br>microcontroller, Bluetooth Low Energy<br>(BLE) transceiver, Sounder and LED.<br>Encased in a plastic housing.                                                                  | Small body-worn pod with adhesive<br>attachment, data is sent to an assigned<br>compatible IntelliVue Patient Monitor or<br>IntelliVue Guardian Software.<br>Pod hardware consists of a battery and an<br>electronic module with the embedded processor,<br>accelerometer, SRR transceiver and speaker.<br>Encased in a plastic housing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | A small piezoelectric sensor strip<br>adhered to the neck of the patient<br>and connected via a Masimo RD<br>cables from the sensor to a<br>peripheral bedside patient monitor.                                                                                                                                                                                                                                              | Similar                                                                                                                                                                                                                       |
| Software                     | The RespiraSense Lobe provides and<br>communicates measurement values<br>and other information (e.g., battery<br>state) wirelessly as encrypted data via<br>an encrypted BLE wireless protocol to<br>the RespiraSense Software Application<br>on a iPad. The RespiraSense<br>Application manages bidirectional<br>communication between Lobe and the<br>RespiraSense Application installed on<br>a COTS Tablet Computer | The CL Respiration Pod provides and<br>communicates measurements values and other<br>information (e.g., battery state) wirelessly via Short<br>Range Radio (SRR) to an assigned compatible<br>IntelliVue Patient Monitor or IntelliVue Guardian<br>Software.<br>The CL Respiration Pod can be controlled from<br>the assigned IntelliVue Patient Monitor or<br>IntelliVue Guardian Software run on PC or<br>server.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | All software is managed by the<br>peripheral bedside patient monitor<br>to which the sensor is connected to.<br>The Bedside patient monitor<br>provides power to the sensor.                                                                                                                                                                                                                                                 | Similar                                                                                                                                                                                                                       |
| Parameter                    | Subject Device RespiraSense<br>(K220111)                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>Philips Intellivue CL Respiration Pod<br>(K122223)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Reference Device<br>Masimo Rainbow SET Acoustic<br>Monitoring sensor (K120984)                                                                                                                                                                                                                                                                                                                                               | Comparison                                                                                                                                                                                                                    |
| Performance<br>Specification | Range: 6 - 60 breaths per minute<br>Resolution: 1 breath per minute<br>Accuracy: $\pm$ 3 breaths per minute                                                                                                                                                                                                                                                                                                             | Range: 5-60 breaths per minute Resolution: 1 breath per minute Accuracy: $\pm$ 1 breath per minute                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Range: 4-70 breaths per minute<br>Resolution: 1 breath per minute<br>Accuracy $\pm$ 1 breaths per minute                                                                                                                                                                                                                                                                                                                     | Difference on the<br>lower range of<br>respiration has no<br>impact on safety or<br>efficacy. For accuracy<br>there is reference<br>Biobeat K212153<br>which has BPM<br>accuracy of $\pm$ 3 bpm<br>to support<br>equivalence. |
| Wireless<br>Connectivity     | Bluetooth 4.2                                                                                                                                                                                                                                                                                                                                                                                                           | Short Range Radio                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | N/A                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar principle<br>of operation                                                                                                                                                                                             |
| Power Source                 | Rechargeable Lithium-Ion battery                                                                                                                                                                                                                                                                                                                                                                                        | Rechargeable Lithium-Ion battery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Peripheral Bedside Monitor                                                                                                                                                                                                                                                                                                                                                                                                   | Similar                                                                                                                                                                                                                       |
| Patient<br>contacting parts  | Adhesive attaching device to<br>patient's chest.                                                                                                                                                                                                                                                                                                                                                                        | Adhesive attaching device to patient's chest.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Adhesive attaching device to the<br>patient's neck.                                                                                                                                                                                                                                                                                                                                                                          | Similar                                                                                                                                                                                                                       |
| Safety and EMC<br>Standards  | AAMI ANSI ES 60601-1 :2015<br>+A1:2012<br>IEC 60601-1-2:2014<br>IEC 60601-1-8:2012<br>ANSI 63.27-2017<br>AIM 735173                                                                                                                                                                                                                                                                                                     | AAMI ANSI ES 60601-1:2005<br>IEC 60601-1-2:2007<br>IEC 60601-1-8:2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | IEC 60601-1<br>IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                 | Subject device<br>complies with the<br>current versions of<br>the applicable<br>standards.                                                                                                                                    |
| Dimensions                   | 57 mm x 98 mm x 18 mm (without<br>sensor)                                                                                                                                                                                                                                                                                                                                                                               | Pod: 53.5 mm x 27 mm x 65 mm ±5% (2.1 in x<br>1.1 in x 2.6 in ±5%) (without cradle and sensor)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Sensor: 135 mm x 22mm x 3 mm<br>accompanied by a 3-3.5m cable                                                                                                                                                                                                                                                                                                                                                                | Similar                                                                                                                                                                                                                       |
| Weight                       | 57 g                                                                                                                                                                                                                                                                                                                                                                                                                    | 80 g ±10% (2.8 oz ±10%)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | <20g (excluding cable)                                                                                                                                                                                                                                                                                                                                                                                                       | Similar                                                                                                                                                                                                                       |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

# VII. PERFORMANCE DATA:

Bench testing - Bench testing was performed that includes:

- Shelf-life / Aging ●
- Software Verification and Validation Guidance for the Content of Premarket Submissions . for Software Contained in Medical Devices
- AAMI ANSI ES 60601-1:2015+A1:2012 Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Collateral standard: Electromagnetic Disturbances Requirements ● and Tests
- IEC 60601-1-8 2012 Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- ANSI IEEE C63.27-2017 American National Standard for Evaluation of Wireless ● Coexistence
- AIM 7351731 Rev. 2.00 Medical Electrical Equipment and System Electromagnetic ● Immunity Test for Exposure to Radio Frequency Identification Readers
- Cybersecurity Content of Premarket Submissions for Management of Cybersecurity in Medical devices

Discussion - The subject device meets the applicable standards similar to the predicate device.

Biocompatibility - The adhesive of the sensor is applied to intact skin with a prolonged duration (up to 4 days). Testing was performed per:

- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro . cytotoxicity
- . ISO 10993-10:2010 - Biological evaluation of medical devices - Part 10: Tests for skin sensitization

Discussion - Both devices use single-use disposable biocompatible adhesive compliant with applicable parts of the ISO 10993-1 Biological Evaluation standard to adhere to intact skin.

# Clinical Testing

We performed a multi-site comparative clinical testing in the intended population and the environment comparing the subject device to the gold standard EtCO2. A total of 56 subjects were evaluated by manual counting and / or comparison to a FDA cleared end-tidal CO2 monitor with the observer blinded to the EtCO2 monitor. The subject covered the age range of > 18 years old and split by gender. The statistical analysis showed that the subject device are substantially equivalent to the gold standard of + 3 bpm at a 95% confidence level. The results supported the claims of substantial equivalence for the intended use.

# VII. CONCLUSION:

The testing performed and comparison to the predicate demonstrate that the subject device is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ/K220111](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ/K220111)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
