← Product Code [BZQ](/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ) · K190734

# Linshom Respiratory Monitoring Device (LRMD) (K190734)

_Linshom Management, LLC · BZQ · Jan 30, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ/K190734

## Device Facts

- **Applicant:** Linshom Management, LLC
- **Product Code:** [BZQ](/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ.md)
- **Decision Date:** Jan 30, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

Linshom Respiratory Monitoring Device (LRMD) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult (at least 22 years of age) patients. LRMD is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume. LRMD measurements are used as an adjunct to other clinical information sources.

## Device Story

LRMD is a thermistor-based respiratory monitor used in healthcare facilities by professionals. It utilizes a Thermistor Sensor Assembly (TSA) to detect temperature changes associated with the breathing cycle; inputs are processed by the Integrated Linshom Module (ILM) Core. The device outputs respiratory rate, seconds since last breath, and a trend of tidal volume via a GUI. It serves as an adjunct to other clinical information sources, assisting clinicians in monitoring patient respiratory status. By providing real-time respiratory data, the device aids in clinical decision-making regarding patient stability, potentially benefiting patients by enabling earlier detection of respiratory distress.

## Clinical Evidence

Evidence includes bench testing (lifetime 24hr, mechanical lung, movement simulation) and clinical studies. Clinical data involved 38 volunteers and 12 endoscopy patients for respiratory rate, and 40 volunteers for tidal volume. Performance was compared against capnography and ventilators. Results showed respiratory rate accuracy of ± 1 BPM and tidal volume trend correlation (r²) of 0.97 to a ventilator, supporting substantial equivalence.

## Technological Characteristics

Thermistor-based sensing; includes Thermistor Sensor Assembly (TSA) and Integrated Linshom Module (ILM) Core. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC). Biocompatibility per ISO 10993 and ISO 18562. GUI interface. Operating range: 18.3°C–29.4°C. Respiration range: 5-60 BPM; Tidal volume range: 120-1000mL.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- ExSpiron ([K120087](/device/K120087.md))

## Reference Devices

- ReDe Mask ([K161953](/device/K161953.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

January 30, 2020

Linshom Management LLC Ronen Feldman President 2922 Excelsior Springs Court Ellicott City, Maryland 21042

Re: K190734

Trade/Device Name: Linshom Respiratory Monitoring Device (LRMD) Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BZQ Dated: January 1, 2020 Received: January 2, 2020

#### Dear Ronen Feldman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Michael Ryan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190734

Device Name

Linshom Respiratory Monitoring Device (LRMD)

Indications for Use (Describe)

Linshom Respiratory Monitoring Device (LRMD) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult (at least 22 years of age) patients.

LRMD is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume.

LRMD measurements are used as an adjunct to other clinical information sources.

| Type of Use (Select one or both, as applicable)                                |                                                                              |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span>     Prescription Use (Part 21 CFR 801 Subpart D)   </span> </div> | <div> <span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summarv – K190734

(per 21 CFR 807.92)

#### Submitter Information

| Name           | Linshom Management LLC                                  |
|----------------|---------------------------------------------------------|
| Address        | 2922 Excelsior Springs Court<br>Ellicott City, MD 21042 |
| Phone Number   | (410) 480-2700                                          |
| Contact Person | Ronen Feldman<br>President                              |
| Date Prepared  | January 30, 2020                                        |

## Device Information

| Trade Name     | Linshom Respiratory Monitoring Device (LRMD)                      |
|----------------|-------------------------------------------------------------------|
| Common Name    | Respiratory Monitor                                               |
| Classification | Breathing frequency monitor<br>21 CFR 878.2375 (Product Code BZQ) |

#### Predicate Device Information

| Primary Predicate | ExSpiron, K120087  |
|-------------------|--------------------|
| Reference Device  | ReDe Mask, K161953 |

#### Device Description

The Linshom Respiratory Monitoring Device (LRMD) is a medical device designed to monitor a patient's respiratory rate, tidal volume trend and seconds since last breath within a healthcare setting. The LRMD is a thermistor based respiratory monitor which includes the Thermistor Sensor Assembly (TSA), Integrated Linshom Module (ILM) Core, the power supply and software.

## Indications for Use

Linshom Respiratory Monitoring Device (LRMD) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult (at least 22 years of age) patients.

LRMD is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume.

LRMD measurements are used as an adjunct to other clinical information sources.

# Technological Characteristics

As shown in Table 1, many technological similarities exist between the subject and predicate devices. The operating mechanism utilized in the LRMD is thermistor-based, similar to the cited reference predicate. Other technological characteristics of LRMD, including software, EMC and ES, and usability aspects, are similar to predicate devices.

{4}------------------------------------------------

| Characteristic                                  | Subject Device<br>LRMD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>(K120087)<br>ExSpiron                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Reference Device<br>(K161953)<br>ReDe Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                                                                                                                                                                                                        |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                                      | Linshom Respiratory<br>Monitoring Device<br>(LRMD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ExSpiron                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ReDe Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                   |
| Common Name                                     | Respiratory Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Respiratory Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Breathing Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                   |
| 510(K) Number                                   | K190734                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K120087                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K161953                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                   |
| Regulation Classification<br>(Product Code)     | 21 CFR 868.2375 (BZQ)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 21 CFR 868.2375 (BZQ)<br>AND<br>21 CFR 868.1850 (BZK)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR 879.2375 (PRK)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                              |
| Intended Use                                    | Non-invasive monitoring<br>of respiration and tidal<br>volume trends for adults<br>in healthcare settings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Non-invasive monitoring<br>of respiration and tidal<br>volume for adults in<br>healthcare settings                                                                                                                                                                                                                                                                                                                                                                                                                                                | Non-invasive monitoring<br>of respiration for adults<br>in healthcare settings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same (ExSpiron)                                                                                                                                                                                                                                   |
| Indications for Use                             | Linshom Respiratory<br>Monitoring Device (LRMD)<br>is indicated for use by<br>healthcare professionals in<br>healthcare facilities, such<br>as procedural areas and<br>recovery rooms, to<br>monitor breathing in adult<br>(at least 22 years of age)<br>patients.<br>LRMD is a non-invasive<br>system that graphically<br>displays temperature<br>changes against time and<br>reports values of<br>respiratory rate and<br>seconds since last breath,<br>along with a trend of tidal<br>volume.<br>LRMD measurements are<br>used as an adjunct to other<br>clinical information<br>sources. | ExSpiron is indicated for<br>use by healthcare<br>professionals in healthcare<br>facilities, such as post-<br>operative care and critical<br>care units, to monitor<br>breathing in adult (at least<br>21 years of age) patients.<br>ExSpiron is a non-invasive<br>system that graphically<br>displays lung volume<br>against time and reports<br>an approximate value of:<br>· Tidal volume,<br>· Respiratory rate, and<br>· Minute ventilation.<br>ExSpiron measurements<br>are used as an adjunct to<br>other clinical information<br>sources. | The ReDe Mark is indicated<br>for use by healthcare<br>professionals in healthcare<br>facility procedural areas<br>and recovery rooms as an<br>adjunct to monitor<br>breathing in adult patients<br>who are sedated for a<br>diagnostic or therapeutic<br>procedure. The ReDe Mask<br>measures the time period<br>between the current and<br>previous exhalation and<br>illuminates a colored light<br>during the exhalation that<br>reflects the interval of time<br>between breaths. If the<br>interval is less than 7.5<br>seconds, the green light<br>illuminates during<br>exhalation; if the interval is<br>greater than 7.5 seconds<br>but up to 20 seconds, the<br>yellow light illuminates<br>during exhalation; and if<br>the interval between<br>breaths is 20 seconds or<br>longer, the red light flashes<br>continuously. The ReDe<br>Mask is only to be used<br>when supplemental<br>oxygen is provided by the<br>facemask. The ReDe Mask<br>is not a standalone device<br>and is only to be used as<br>an adjunct to pulse<br>oximetry. | Equivalent (ExSpiron)<br>Minor text differences to not suggest<br>a new or difference intended use.<br>Justification: minor differences reflect minor<br>differences in devices but does not change<br>the use, users, or conditions of use       |
| Mechanism (General)                             | Thermistor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Thoracic Bioimpedence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Thermistor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same (ReDe Mask)                                                                                                                                                                                                                                  |
| Measurements                                    | Respiratory Rate<br>Seconds Since Last<br>Breath<br>Tidal Volume Trend                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Respiratory Rate<br>Tidal Volume<br>Minute Volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Seconds Since Last<br>Breath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Equivalent (All)<br>Justification: these parameters are all related<br>to the breathing cycle                                                                                                                                                     |
| Characteristic                                  | Subject Device<br>LRMD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>(K120087)<br>ExSpiron                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Reference Device<br>(K161953)<br>ReDe Mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                                                                                                                                                                                                        |
| Communication Method                            | GUI Interface                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | GUI Interface                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Light Display (Red /<br>Yellow / Green)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same (ExSpiron)                                                                                                                                                                                                                                   |
| ES / EMC Testing                                | IEC 60601-1 & 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | IEC 60601-1 & 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | IEC 60601-1 & 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                              |
| Material Biocompatibility                       | ISO 10993<br>ISO 18562                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Tested (mask<br>components)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Equivalent<br>Justification: all devices address<br>biocompatibility risks (albeit different<br>mechanisms)                                                                                                                                       |
| Mask Type                                       | Face mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | ExSpiron 1Xi PadSets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Face mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Equivalent (ReDe Mask)<br>Justification: LRMD utilizes a legally marketed<br>facemask while the ReDe Mask incorporates<br>its own mask; regardless, both devices attach<br>to the patient face via an oxygen mask                                 |
| Mounting Design                                 | Sensor attached to<br>separate electronics box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Electrodes attached to<br>separate electronics box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Face mask and<br>electronics housing<br>placed on patient's face                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Equivalent (ExSpiron)<br>Justification: these are similar concepts just<br>have different electronics                                                                                                                                             |
| Ambient Operating<br>Temperature                | 65°F -- 85°F<br>(18.3°C - 29.4°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 16°C to 31°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Equivalent (ReDe Mask)<br>Justification: both devices demonstrate safe<br>and effective performance at the extremes of<br>its stated operating range                                                                                              |
| Working Range                                   | Respiration: 5-60 BPM<br>Tidal Volume: 120-<br>1000mL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Respiration: 5-30 BPM<br>Tidal Volume: unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Equivalent (ExSpiron)<br>Justification: both devices have performance<br>testing that demonstrates safety and<br>effectiveness against a gold standard (e.g.,<br>capnography) for the entire working range                                        |
| Accuracy                                        | Respiration: ± 1 BPM<br>Tidal Volume Trend: 0.97<br>(r² correlation to<br>ventilator)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Respiration: 2%<br>Tidal Volume: 3.9%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Equivalent (ExSpiron)<br>Justification: although LRMD reports its<br>accuracy across the working range (via<br>correlation), performance data showed that<br>accuracy is equivalent across working range<br>for both respiration and tidal volume |
| Weight (at point of<br>measurement)             | ~15g (Thermistor Sensor<br>Assembly only)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 40g                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Within Range of Existing Products                                                                                                                                                                                                                 |
| Dimensions (of unit as<br>point of measurement) | 60mm H<br>254mm L<br>158mm W<br>(ILM Core)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 43.43mm H<br>30.48mm W<br>14.99mm D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Within Range of Existing Products                                                                                                                                                                                                                 |
| Performance Testing                             | Lifetime Test (24hr)<br>Breathing Rate Test<br>(Mechanical Lung)<br>Movement Test<br>38 volunteer subjects<br>(for respiratory rate)<br>12 endoscopy subjects<br>(for respiratory rate)<br>40 volunteer subjects<br>(for tidal volume)<br>Human Factors Test<br>Tidal Volume Trend<br>Bench Test                                                                                                                                                                                                                                                                                             | 20 subjects (for all 3<br>measures, compared<br>against spirometer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Lifetime Test (8hr)<br>Human Factors (for<br>overall use and for light<br>interpretation)<br>38 volunteer subjects<br>(for detection of<br>exhalation and low<br>respiratory rate<br>compared to<br>capnography)<br>50 volunteer subjects<br>(for detection of<br>exhalation and low<br>respiratory rate<br>compared to<br>capnography)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Demonstrates subject device is as<br>safe and as effective as predicate<br>device.                                                                                                                                                                |

{5}------------------------------------------------

{6}------------------------------------------------

## Performance Data

Bench and clinical performance data were collected to support a substantial equivalence determination. This testing included mechanical lung data, patient movement simulation, 24-hour testing, clinical comparison of respiratory rate to capnography, and clinical comparison of tidal volume trend to a ventilator. These data suggest the LRMD is as safe and as effective as the identified predicate devices.

## Conclusions

The LRMD and predicate devices have intended use and similar indications, both serving as medical devices to monitor respiration and tidal volume trend. As described above, the minor technological differences between the LRMD and its predicates do not present any new or different issues of safety or effectiveness. Based upon analysis and valid scientific evidence the LRMD is substantially equivalent to its predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ/K190734](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/BZQ/K190734)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
