Accu O2 Oxygen Analyzer

K173807 · Precision Medical, Inc. · CCL · May 2, 2018 · Anesthesiology

Device Facts

Record IDK173807
Device NameAccu O2 Oxygen Analyzer
ApplicantPrecision Medical, Inc.
Product CodeCCL · Anesthesiology
Decision DateMay 2, 2018
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 868.1720
Device ClassClass 2

Indications for Use

The PM5950 Accu O2 Analyzer is intended as a tool for use by qualified personnel to spot-check or measure oxygen concentration of a delivered air/oxygen mixture. The PM5950 Accu O2 Analyzer is not intended for use in continuous monitoring of oxygen delivery to patient.

Device Story

Portable oxygen analyzer; measures O2 concentration in air/oxygen breathing circuits. Input: galvanic fuel cell sensor (amperometric electrochemical principle) connected via interface cable. Processing: microprocessor-based analyzer; converts sensor voltage to O2 percentage; manages user inputs (push buttons) and LCD output. Output: digital display of O2 concentration (0.0-100.0%), low battery status, and calibration indicators. Used by trained professionals in hospitals, home care, transport, and sub-acute settings. Healthcare providers use spot-check readings to verify delivered gas mixtures; output informs clinical decisions regarding oxygen delivery settings. Benefits: provides accurate, portable verification of oxygen concentration to ensure appropriate therapy delivery.

Clinical Evidence

Bench testing only. No clinical data. Performance validated via ISO 80601-2-55:2011 compliance, including measurement accuracy, response time (<12s), drift, and environmental testing (shock, vibration, EMC, temperature, humidity).

Technological Characteristics

Galvanic fuel cell sensor; microprocessor-based analyzer; 3.5" backlight LCD; 2 AA alkaline battery power; IPX1 ingress protection; non-sterile; 10ft extendible cable; 0.0-100% measurement range; 0.1% resolution; +/- 3.0% accuracy. Software is moderate level of concern.

Indications for Use

Indicated for use by trained professionals in hospital, home care, transport, and sub-acute institutions to spot-check or measure oxygen concentration of delivered air/oxygen mixtures. Not for continuous patient monitoring or use in MRI environments.

Regulatory Classification

Identification

An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 2, 2018 Precision Medical Inc. James Parker Quality Assurance Manager 300 Held Drive Northampton, Pennsylvania 18067 Re: K173807 Trade/Device Name: PM5950 Accu O2 Oxygen Analyzer Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: Class II Product Code: CCL Dated: March 12, 2018 Received: March 29, 2018 Dear James Parker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ K173807 and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Geeta K. Pamidimukkala -ഗ for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173807 Device Name PM5950 Accu O2 Analyzer Indications for Use (Describe) The PM5950 Accu O2 Analyzer is intended as a tool for use by qualified personnel to spot-check or measure oxygen concentration of a delivered air/oxygen mixture. The PM5950 Accu O2 Analyzer is not intended for use in continuous monitoring of oxygen delivery to patient. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------|----------------------------------------------------------------| | <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Precision Medical. The words "PRECISION MEDICAL" are written in a serif font, with "PRECISION" in green and "MEDICAL" in black. To the right of the words is a green graphic that resembles a series of angled lines. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 # 510(k) Summary # Abbreviated 510K # Precision Medical, Inc. PM5950 Accu O2 Analyzer. Revised 5/1/2018 ## 1. Submitter Information Submitter: Precision Medical, Inc. 300 Held Drive Northampton, Pa. 18067 Facility Registration #: 2523148 | Contact: | James Parker | | |-------------------|--------------------------------|------------------------| | | Quality Assurance Manager | | | Tel: | (610)-262-6090 Extensions 2228 | | | Fax: | (610)-262-6080 | | | Preparation Date: | 11/17/2017 | Revised Date: 5/1/2018 | # 2. Device Name | Proprietary Name: | Accu O2 | |------------------------------|--------------------------------------| | Model number: | PM5950 | | Common Name: | Oxygen Analyzer | | Classification Name: | Analyzer, Gas, Oxygen, Gaseous-Phase | | Classification Product Code: | CCL | | Regulation number: | 868.1720 | | Regulatory Status: | Class II | {4}------------------------------------------------ ## 3. Predicate Device Equivalence Precision Medical, Inc. is claiming substantial equivalence; # Primary Predicate Device; Substantial Equivalent: | Proprietary Name: | MAXO2®+AE | |-------------------|--------------------------------------------------------------| | Model Number: | R217P72 | | Manufactured by: | Maxtec<br>2305 South 1070 West<br>Salt Lake City, Utah 84119 | Classification Name: Analyzer, Gas, Oxygen, Gaseous-Phase Premarket Notification #: K040484 --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- # Referenced Predicate Device; | Proprietary Name: | Oxygen Monitor | | |-------------------|------------------------|--| | Model Number: | PM5900 | | | Manufactured by: | Precision Medical Inc. | | | | 300 Held Drive | | | | Northampton, Pa 18067 | | Classification Name: Analyzer, Gas, Oxygen, Gaseous-Phase Premarket Notification #: K063096 Image /page/4/Picture/11 description: The image shows the logo for Precision Medical. The words "PRECISION MEDICAL" are written in a serif font, with "PRECISION" in green and "MEDICAL" in black. A green graphic resembling a stylized leaf or feather is placed between the two words. The logo has a clean and professional appearance. Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 {5}------------------------------------------------ 510(k): K173807 Image /page/5/Picture/2 description: The image shows the logo for Precision Medical. The words "PRECISION MEDICAL" are written in a serif font, with "PRECISION" in green and "MEDICAL" in black. A green graphic resembling a stylized leaf or feather is placed between the two words. The registered trademark symbol is placed after the word "MEDICAL". 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 #### 4. Device Description (per 21 CFR 807.92) #### General Description: The PM5950 Accu O2 Analyzer measures and displays the amount of oxygen present, percent (%), in an air/oxygen breathing circuit. #### Intended Use: The PM5950 Accu O2 Analyzer is intended as a tool for use by qualified personnel to spot-check or measure oxygen concentration of a delivered air/oxygen mixture. The PM5950 Accu O2 Analyzer is not intended for use in continuous monitoring of oxygen delivery to patient. #### Indications for Use: The PM5950 Accu O2 Analyzer is intended as a tool for use by qualified personnel to spot-check or measure oxygen concentration of a delivered air/oxygen mixture. The PM5950 Accu O2 Analyzer is not intended for use in continuous monitoring of oxygen delivery to patient. #### Contraindication: The PM5950 Accu O2 Analyzer is not intended to actively monitor oxygen concentration or oxygen gas mixtures while being delivered to a patient. PM5950 Accu O2 Analyzer is not intended for use in a MRI environment. ### Intended population \ Environment \ Conditions to be used for: The PM5950 Accu O2 Analyzer is intended for use by trained professionals in Hospital, Home Care, Transport, and sub-acute institutions. ### Prescription: Caution! US. Federal Law restricts this device to sale by or on the order of a physician. ### Principals of Operation: Oxygen measurement is accomplished by placing an oxygen sensor into a breathing circuit. The sensor employs amperometric electrochemical measurement principles (e.g. galvanic fuel cell; lead oxygen battery). The sensor is a purchased and an existing legally marketed device. The oxygen sensor is connected to the analyzer unit via an interface cable. The analyzer unit is battery powered. The analyzer is microprocessor based and interfaces with user push buttons and a LCD display. User push buttons allow the user to turn the unit on/off and to initiate calibration. The LCD provides indication of the per-cent oxygen measured, low battery indication and calibration required. #### New Technologies/Features: Neither new technologies nor features are incorporated ### Configurations/Variations: The PM5950 Accu O2 Analyzer will be provided in the configuration submitted; use of the Tee Adaptor is optional. No variations exist. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Precision Medical. The logo is in green and black. The word "PRECISION" is in green, and the word "MEDICAL" is in black. There is a green graphic between the two words. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 ### Key Functional Elements: Image /page/6/Figure/5 description: The image shows a Precision Medical oxygen analyzer and its components. The oxygen analyzer has a digital display showing an oxygen concentration of 100.0%. The components include a coiled cable, locking ring, oxygen sensor, diverter fitting, and tee adapter. The device consists of four basic elements: - Base Device: The Base Device is a portable / mountable enclosure that houses the device electronic circuitry, a 3 ½" backlight display, membrane keypad, sensor interface jack and 2 AA Alkaline Batteries. - Extendible Cable: The extendible cable allows connection of the sensor the Base Device. The cable can be extended up to 10 feet from the side of the device. - Oxygen Sensor W/ Diverter Fitting: The Oxygen sensor is a galvanic fuel cell that generates a voltage output relative to the measured percent oxygen. The Oxygen Sensor (with diverter) is designed to fit industry standard, 15 mm I.D. "T" adaptors. - Tee Adaptor: The Tee adaptor is used to connect the Oxygen Sensor and Diverter Fitting to an oxygen pathway circuit. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Precision Medical. The logo is in green and black. The word "PRECISION" is in green, and the word "MEDICAL" is in black. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 # The device User interface | ITEM NAME | DESCRIPTION | |----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Power Key<br>Image: [Power Key Symbol] | The Power Key turns the Oxygen Analyzer ON and OFF. | | Calibration Key<br>Image: [Calibration Key Symbol] | Pressing the Calibration Key calibrates the Oxygen Analyzer with air or oxygen. | | Image: [Calibration Symbol] | Calibration Symbol - The calibration symbol is located on the display and is timed to activate when a calibration is necessary, or when a calibration is being conducted. | | Image: [Low Battery Symbol] | Low Battery Symbol - The low battery indicator is located on the top of the display and is only activated when the voltage on the batteries is below a normal operating level. | | Image: [Invalid Reading Symbol] | Invalid Reading Symbol - failed sensor, failed cable , invalid calibration | | 100.0 % | Oxygen Concentration Reading. (0.0 to 100.0 %) | # Technical Specifications # Base Device Specifications Dimensions: (Analyzer without cable and sensor attached): | Depth: | 1.30" (3.32 cm) | |---------------|---------------------------------| | Width: | 2.90" (7.30 cm) | | Height: | 4.00" (10.23 cm) | | Cable Length: | 10 ft. (3.05m) (fully extended) | # Weight: Device Weight: 0.35 lbs (5.60 oz / 0.15 kg) {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Precision Medical. The logo is in green and black. The word "PRECISION" is in green, and the word "MEDICAL" is in black. There is a green graphic between the two words. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 (Includes: analyzer, sensor, batteries and cable) | Operating Conditions: | | |-------------------------------|-----------------------------------------------------------------------| | Temperature: | 50°F - 113°F (10°C - 45°C) | | Altitude: | Sea Level to 8000 feet | | Humidity: | 0-95% non-condensing | | Storage Conditions: | | | Temperature: | 5°F - 122°F (0°C - 50°C) | | Humidity: | 0 - 95% non-condensing | | Mode of Operation: | Continuous | | Electrical Classification: | Internally powered Medical Electrical equipment | | Diverter Fitting: | Fits industry standard, 15 mm “T” adapter | | Measurement Range: | 0.0 - 100% Oxygen | | Resolution: | 0.1 % | | Total Accuracy: | $\pm$ 3.0% Actual Oxygen Level over full operating temperature range | | Drift of Measurement: | < +/-1% of full scale at constant temperature, pressure and humidity | | Response Time: | 90% of final value in less than 12 seconds at 77°F (25°C) | | Warm-up Time: | Not required | | Low Battery Indication: | Low battery icon displayed | | Patient Contact: | None | | Classifications: | | | | Protection against electric shock: Internally powered equipment | | | Protection against water: IPX1 (Drip Proof) | | Mode of Operation: | Continuous | | Sterilization: | Non-Sterile Device | | Flammable anesthetic mixture: | Not suitable for use in presence of a flammable anesthetic<br>mixture | | Sensor Specifications: | | | Sensor Type: | Galvanic oxygen sensor; Precision Medical PN 504877 | | Expected Sensor Life: | > 1,000,000 O2 % Hours | {9}------------------------------------------------ # 5. Difference in technology from MAXO2®+AE R217P72 (Predicate) to PM5950 Accu O2 Analyzer | Technical<br>difference | MAXO2®+AE<br>(Primary Predicate)<br>K040484 | PM5900 Oxygen<br>Monitor<br>(Reference Predicate)<br>K063096 | PM5950 Accu O2<br>Analyzer<br>K173807 | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The MAXO2+ is<br>intended as a tool for<br>use by qualified<br>personnel to spot-check<br>or measure oxygen<br>concentration of a<br>delivered air/oxygen<br>mixture. The MAXO 2+<br>is not intended for use<br>in continuous<br>monitoring of oxygen<br>delivery to a patient. | The oxygen monitor<br>provides continuous,<br>direct monitoring of<br>oxygen mixtures in a<br>wide variety of medical<br>applications such as<br>anesthesiology [e.g.<br>anesthesia machines],<br>respiratory devices [e.g.<br>respirators, ventilators,<br>pediatric incubators], and<br>oxygen therapy [e.g.,<br>oxygen tents]. The<br>oxygen monitor is to be<br>used by trained<br>healthcare professionals<br>under the supervision, or<br>on the order, of a<br>physician in a hospital [or<br>other clinical setting. The<br>Precision Medical, Inc.<br>oxygen monitor is not<br>intended for transport<br>use. This device is not an<br>oxygen supply source. | The PM5950 Accu O2<br>Analyzer is intended as a<br>tool for use by qualified<br>personnel to spot-check<br>or measure oxygen<br>concentration of a<br>delivered air/oxygen<br>mixture. The PM5950<br>Accu O2 Analyzer is not<br>intended for use in<br>continuous monitoring of<br>oxygen delivery to patient. | | Calibration | After the device is<br>turned on it will<br>automatically calibrate<br>to room air. The display<br>should be stable and<br>reading 20.9%. | The device requires<br>manual calibration | The device requires<br>manual calibration. | PRECISION® 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 {10}------------------------------------------------ Abbreviated 510k for PM5950 Accu O2 Analyzer Model # PM5950 510(k): K173807 Image /page/10/Picture/2 description: The image shows the logo for Precision Medical. The word "PRECISION" is in green, and the word "MEDICAL" is in black. There is a green graphic between the two words. The logo also includes the registered trademark symbol. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 # 6. Certifications The PM5950 Accu O2 Analyzer was tested in compliance with FDA recognized consensus standard ISO 80601-2-55:2011: Medical electrical equipment — Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors and meets the requirements of the standard. - IEC 60601-1:2005-12 Third Edition: Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance - IEC 60601-1-2: 2014 Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests - IEC 60601-1-6:2013-3.1 Medical Electrical Equipment Part 1 -6: General . Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability - IEC 61000-3-2 Edition 3.2:2009: Electromagnetic compatibility (EMC) Part 3-2: . Limits - Limits for harmonic current emissions (equipment input current < 16A per phase) - . IEC 61000-3-3 Edition 2.0: 2008: Electromagnetic compatibility (EMC) – Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in low-voltage supply systems, for equipment with rated current < 16A per phase and not subject to conditional connection - IEC 61000-4-2 Edition 2.0 2008-12: Electromagnetic compatibility (EMC) Part 4-2: . Testing and measurement techniques - Electrostatic discharge immunity test - IEC 61000-4-3 Edition 3.2: 2010-04: Electromagnetic compatibility (EMC) Part 4-3: . Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test - IEC 61000-4-4 Edition 2.1: 2011-03: Electromagnetic compatibility (EMC) Part 4-4: . Testing and measurement techniques – Electrical fast transient/burst immunity test - IEC 61000-4-5 Edition 2.0: 2005-11: Electromagnetic compatibility (EMC) Part 4-5: . Testing and measurement techniques - Surge immunity test - IEC 61000-4-6 Edition 4.0: 2013-10: Electromagnetic compatibility (EMC) Part 4-6: ● Testing and measurement techniques - Immunity to conducted disturbances, induced by radio frequency fields. - IEC 61000-4-8 Edition 2.0: 2009-09: Electromagnetic compatibility (EMC) Part 4-8: ● Testing and measurement techniques - Power frequency magnetic field immunity test - IEC 61000-4-11Edition 2.0: 2004-03: Electromagnetic compatibility (EMC) Part 4-11: . Testing and measurement techniques –Voltage dips, short interruptions and voltage variations immunity test {11}------------------------------------------------ Image /page/11/Picture/1 description: The image shows the logo for Precision Medical. The word "PRECISION" is in a bold, green, sans-serif font. To the right of the word "PRECISION" is a green graphic that resembles blades of grass. To the right of the graphic is the word "MEDICAL" in a black, serif font. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 #### Table of Comparisons to Predicate Device 7. | | R217P72<br>MAXO2®+AE<br>(Primary Predicate)<br>K040484 | PM5900<br>Oxygen Monitor<br>(Referenced Predicate)<br>K063096 | PM5950<br>Accu O2 Analyzer | |----------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------| | Specifications | | | | | Weight (device only) | 0.4 lbs. (170g) | 1.11 lbs (0.50 kg) | 0.33 Lbs. (150g) | | Dimensions (inches): | | | | | Length | 1.5" | 1.72" | 1.3" | | Height | 4.0" | 5.44" | 4.0" | | Width | 3.0" | 3.56" | 2.9" | | Electrical Requirements | 2, AA Alkaline<br>batteries (2 x 1.5<br>Volts) | 4, AA Alkaline batteries<br>(4 x 1.5 Volts) | 2, AA Alkaline batteries<br>(2 x 1.5 Volts) | | Operating temp range | 15°C - 40°C<br>(59°F - 104°F) | 10°C - 45°C<br>(50°F - 113°F) | 10°C - 45°C<br>(50°F- 113°F) | | Storage temp range | -15°C - 50°C<br>(5°F - 122°F) | -15°C- 50°C<br>(5°F - 122°F) | -0°C- 50°C<br>(32°F - 122°F) | | Humidity | 0 - 95% non-<br>condensing | 0 - 95% non-<br>condensing | 0 - 95% non-<br>condensing | | Mode of Operation: | Continuous | Continuous | Continuous | | Electrical Classification: | Internally powered<br>Medical Electrical<br>equipment | Internally powered<br>Medical Electrical<br>equipment | Internally powered<br>Medical Electrical<br>equipment | | Protection against water: | IPX1 | IPX1 | IPX1 | | Measurement Range: | 0.0 - 100.0% | 0.0 - 100.0% | 0.0 - 100.0% | | Resolution: | 0.1 Increments | 0.1 Increments | 0.1 Increments | | Total Accuracy: | $\pm$ 3% Actual Oxygen<br>Level over full<br>operating temperature<br>range | $\pm$ 3% Actual Oxygen<br>Level over full<br>operating temperature<br>range | $\pm$ 3% Actual Oxygen<br>Level over full<br>operating temperature<br>range | | Specifications | R217P72<br>MAXO2®+AE<br>(Primary Predicate)<br>K040484 | PM5900<br>Oxygen Monitor<br>(Referenced Predicate)<br>K063096 | PM5950<br>Accu O2 Analyzer | | Response Time: | 90% of final value in<br>approximately 15<br>seconds at 23°C | 90% of final value in<br>less than 12 seconds at<br>77°F (25°C) | 90% of final value in<br>less than 12 seconds at<br>77°F (25°C) | | Warm-up Time: | none required | none required | none required | | Low Battery Indication: | Low battery icon<br>displayed on graphics<br>screen | Low battery icon<br>displayed on graphics<br>screen | Low battery icon<br>displayed on graphics<br>screen | | Interface | Two button<br>On\Off, Cal | Two button<br>On\Off, Cal | Two button<br>On\Off, Cal | | Sensor Type | Maxtec MAX 13/CAG-13<br>Galvanic fuel sensor<br>(0-100%) | Maxtec MAX 13/CAG-13<br>Galvanic fuel sensor<br>(0-100%) | Maxtec MAX 13/CAG-13<br>Galvanic fuel sensor<br>(0-100%) | | Sensor Life | 2-years in typical<br>applications | 2-years in typical<br>applications | 2-years in typical<br>applications | | Drift of Measurement: | < +/-1% of full scale at<br>constant temperature,<br>pressure and humidity | < +/-1% of full scale at<br>constant temperature,<br>pressure and humidity | < +/-1% of full scale at<br>constant temperature,<br>pressure and humidity | | Alarms | None | high/low alarms,<br>respective flashing red<br>LEDs and graphics,<br>68db audible alarm @<br>1 meter | None | {12}------------------------------------------------ Image /page/12/Picture/1 description: The image shows the logo for Precision Medical. The word "PRECISION" is in green, block letters. To the right of the word "PRECISION" is a green graphic that looks like blades of grass. To the right of the graphic is the word "MEDICAL" in black, block letters. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 {13}------------------------------------------------ Image /page/13/Picture/1 description: The image shows the logo for Precision Medical. The logo is in green and black. The word "PRECISION" is in green, and the word "MEDICAL" is in black. There is a green graphic between the two words. Pa 18067 Tel: 610-262-6090 Fax: 610-262-6080 ## 8. Declarations of Conformity and Summary Reports The following list is high level summary of the tests that were used to demonstrate substantial equivalence. IEC 60601-1 3rd + A1 (60601_1k) Report CB_PM5950 IEC 60601-1-2 EMC Test Report_PM5950 IEC 60601-1-6 _ IEC 62366 COMPLIANCE Report_PM5950 ISO 80601-2-55 2011 Compliance Report PM5950 IEC 60068-2-27& 64 Shock & Vibration Test_PM5950 ISO 14971 Risk Management File PM5950_101617 The following list is a summary of the tests performed by Precision Medical Inc. that were used to demonstrate substantial equivalence. | 811-1 | Ingress of water ISO 80601-2-55 (201.11.6.5) | |--------|----------------------------------------------------------------------------------------------| | 811-2 | Oxygen Analyzer Measurement Accuracy ISO 80601-2-55:2011<br>(201.12.1.101,102,104) | | 811-3 | Shock & Vibration Testing for PM5950 Oxygen Analyzer ISO 80601-2-55:2011<br>(201.15.3.5.101) | | 811-4 | EMC Testing for PM5950 Oxygen Analyzer ISO 80601-2-55:2011 (202.6.2) | | 811-6 | Oxygen Analyzer - Leakage to Atmosphere ISO 80601-2-55:2011 (201.102) | | 811-7 | Operating and Storage Temp | | 811-7A | Operating and Storage Temperature of the PM5950 O2 Analyzer Review | | 811-8 | Software Validation of PM5950 | | 811-10 | Effects of Elevation-Barometric Pressure Change | | 811-11 | Drop Test of PM5950 in Shipping Box | | 811-12 | Performance Test | | 811-13 | Process Validation of the PM5950 | | 811-15 | Life Expectancy of the Keypad | | 811-16 | Label Durability | | 811-17 | Cleaning and disinfection of ME Equipment of ME Systems | {14}------------------------------------------------ Image /page/14/Picture/1 description: The image features the logo for "Precision Medical." The word "PRECISION" is written in a bold, green font, followed by a stylized green graphic that resembles a series of angled lines. To the right of the graphic, the word "MEDICAL" is written in a thinner, black font, with the registered trademark symbol (®) next to it. The overall design is clean and professional, suggesting a company in the healthcare or medical industry. 300 Held Drive Northampton, Pa 18067 USA Tel: 610-262-6090 Fax: 610-262-6080 #### 811-18 Compatibility with substances used with ME Equipment ## Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for PM5950 Oxygen Analyzer was considered to be "Moderate" because a failure of the Device Software could result in Minor Injury to a patient, and a malfunction of a latent design flaw in the Device Software may lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury. ### 9. Conclusions Precision Medical's PM5950 Accu O2 Analyzer operates in the same manner as the Primary Predicate Device: MAXO2+AE. The Referenced Predicate Device, Precision Medical PM5900, is only presented to indicate no new technologies or features have been developed or incorporated. Compliance testing indicates the technical specifications presented have been achieved. The PM5950 Accu O2 Analyzer has the same Intended/Indicated use as the Primary Predicate Device. Precision Medical's PM5950 Accu O2 Analyzer has the same therapeutic use. Precision Medical's PM5950 Accu O2 Analyzer is substantially equivalent to the Primary Predicate Device. The Precision Medical PM5900 Oxygen Monitor was used a reference predicate device as the same applied consensus standard was used for compliance testing (ISO 80601-2-55:2011). The significant difference between the Precision Medical, Inc PM5900 Oxygen Monitor and PM5950 Accu O2 Analyzer is that the Oxygen Monitor may be used to continuously monitor and contains alarms. The PM5950 Accu O2 Analyzer is intended as a tool for use by qualified personnel to spot-check or measure oxygen concentration of a delivered air/oxygen mixture. The PM5950 Accu O2 Analyzer is not intended for use in continuous monitoring of oxygen delivery to patient. In summary, Precision Medical, Inc had demonstrated that the Precision Medical, Inc PM5950 Accu O2 Analyzer is as safe and as effective as the predicate device. The combined non-clinical testing and analysis of results provides assurance that the device meets the specifications and is as safe, as effective, and performs as well as or better than the legally marketed predicate MAXO2+AE device..
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...