Microstream CO2 NanoPod

K213911 · Covidien, LLC · CCK · Jul 29, 2022 · Anesthesiology

Device Facts

Record IDK213911
Device NameMicrostream CO2 NanoPod
ApplicantCovidien, LLC
Product CodeCCK · Anesthesiology
Decision DateJul 29, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1400
Device ClassClass 2
AttributesPediatric

Indications for Use

The Microstream™ CO2 NanoPod, when connected to the host monitor, is intended to provide continuous, non-invasive measurement and monitoring of carbon dioxide (CO2) during the respiration cycle, EtCO2, FiCO2, and CO2-based respiration rate for adult, pediatric and neonatal patients, utilizing Microstream™ Advance CO2 sampling lines. The Microstream™ CO2 NanoPod also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/ hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for age 22 and up. IPI and ODI values can be calculated and displayed only if the host monitor collects SpO2 data and provides this SPO2 data to the NanoPod in order to calculate these parameters. The Microstream™ CO2 NanoPod is intended for use in hospitals, hospital-type facilities, during intra-hospital transport between and within areas of care.

Device Story

Microstream CO2 NanoPod is an external capnography module connecting to host monitors via LEMO cable; provides continuous, non-invasive CO2 measurement (EtCO2, FiCO2, respiration rate) using NDIR spectroscopy. When host monitor provides SpO2 data, device calculates Integrated Pulmonary Index (IPI), apnea per hour (A/hr), and oxygen desaturation index (ODI). Used in hospitals and during intra-hospital transport by trained healthcare providers. Device encloses NanoMediCO2 module; powered by host monitor. Output displayed on host monitor to assist clinicians in respiratory status assessment; aids in monitoring patient ventilation and oxygenation status. Benefits include continuous respiratory monitoring to support clinical decision-making.

Clinical Evidence

No clinical performance testing was required. Substantial equivalence supported by bench performance, safety, electromagnetic, and environmental testing, alongside software validation and ISO 14971:2019 hazard analysis.

Technological Characteristics

External capnography module; NDIR spectroscopy for CO2; powered by host monitor (DC). Includes NanoMediCO2 module, interface board with status LED, and mounting cradle. Connectivity via LEMO connector to host monitor. Standards: ISO 80601-2-55, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 80601-2-49.

Indications for Use

Indicated for continuous, non-invasive CO2 monitoring (EtCO2, FiCO2, respiration rate) in adult, pediatric, and neonatal patients. Integrated Pulmonary Index (IPI) indicated for pediatric and adult patients. Apnea per hour (A/hr) and Oxygen Desaturation Index (ODI) indicated for patients age 22 and up. Contraindicated for use as an apnea monitor.

Regulatory Classification

Identification

A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 29, 2022 Covidien LLC Andrew Berkeland Senior Regulatory Affairs Specialist 6135 Gunbarrel Ave Boulder, Colorado 80301 Re: K213911 Trade/Device Name: Microstream CO2 NanoPod Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK, BZQ Dated: June 27, 2022 Received: June 28, 2022 Dear Andrew Berkeland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K213911 Device Name Microstream™ CO2 NanoPod #### Indications for Use (Describe) The Microstream™ CO2 NanoPod, when connected to the host monitor, is intended to provide continuous, non-invasive measurement and monitoring of carbon dioxide (CO2) during the respiration cycle, EtCO2, FiCO2, and CO2-based respiration rate for adult, pediatric and neonatal patients, utilizing Microstream™ Advance CO2 sampling lines. The Microstream™ CO2 NanoPod also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/ hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for age 22 and up. IPI and ODI values can be calculated and displayed only if the host monitor collects SpO2 data and provides this SPO2 data to the NanoPod in order to calculate these parameters. The Microstream™ CO2 NanoPod is intended for use in hospitals, hospital-type facilities, during intra-hospital transport between and within areas of care. Contraindications: The device is not to be used as an apnea monitor. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Microstream™ CO2 NanoPod 510(k) Summary This summary of 510(k) safety and effectiveness information for the Microstream™ CO2 NanoPod is submitted in accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with the requirements of 21 CFR §807.92. ### SUBMITTER INFORMATION Submitted By: Covidien, IIc 6135 Gunbarrel Avenue Boulder, CO 80301 Date Prepared: December 14, 2021 ### Primary Contact Person: Andrew Berkeland Senior Requlatory Affairs Specialist Phone: (720) 501-7884 Email: andy.j.berkeland@medtronic.com ### Secondary Contact Person: Greeshma Kayala Senior Requlatory Affairs Manager Phone: (303) 579-1718 Email: greeshma.kayala@medtronic.com ### DEVICE | Trade Name: | Microstream™ CO <sub>2</sub> NanoPod | |----------------------|--------------------------------------| | Classification Name: | Capnography - NanoPod | | Regulation: | 21 CFR 868.1400, 21 CRF 868.2377 | | Regulatory Class: | Class II | | Product Code: | CCK, BZQ | | Review Panel: | Anesthesiology | ### PREDICATE DEVICE | Predicate Manufacturer: | Oridion Medical 1987 LTD | |-------------------------|----------------------------------------------| | Predicate Trade Name: | CapnostreamTM35 Portable Respiratory Monitor | | Predicate 510(k): | K150272 | {4}------------------------------------------------ ### DEVICE DESCRIPTION The subject device of this premarket 510(k) notification is referred to as the Microstream™ CO2 NanoPod throughout this submission. The Microstream™ CO2 NanoPod is a finished medical device that when connected to a host monitor is intended to provide continuous, non-invasive measurement and monitoring of carbon dioxide (CO2) during the respiration cycle, including End-tidal CO2 (EtCO2), Fractional Concentration of Inspired CO2 (FiCO2), and CO2 based respiration rate, for adult, pediatric, and neonatal patients utilizing Microstream™CO2 or Microstream™ Advance CO2 sampling lines. The Microstream™ CO2 NanoPod also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr), and oxygen desaturation index (ODI) values. The IPI measurement is intended for use in pediatric and adult patients only, and A/hr and ODI measurements are intended for age 22 and up. Both IPI and ODI values can be calculated and displayed only if the host monitor collects SpO2 data and provides this SpO2 data to the NanoPod to calculate these parameters. The Microstream™ CO2 NanoPod encloses a NanoMediCO2 capnography module, which measures inspired/expired carbon dioxide and respiration rate. The NanoPod is powered by the host monitor on a DC supply. The Microstream™ CO2 NanoPod system includes the following sub-systems: - . NanoPod is the medical device enclosure that encloses the NanoMediCO2 module. The NanoPod has a connected active cable that has a LEMO connector end, which connects to a host monitoring system. - . NanoMediCO2 capnography module is enclosed in the NanoPod. The NanoMediCO2 module measures inspired and expired carbon dioxide and respiration rate. The active cable is what provides bidirectional communication between the Host monitor and the NanoMediCO2 module enclosed inside the NanoPod. - . Interface Board is enclosed inside NanoPod which consists of an LED exposed on the outside of the NanoPod to provide the functional status of the subject device to the user. - l Cradle is a plastic holder used to connect the NanoPod with a GCX clamp or other mounting solution, which can then be used to mount the NanoPod on a pole or bedrail. There are no sharp edges on the Microstream™ CO2 NanoPod, which enables the medical device to be held in the hand at times when it is removed from its mounting cradle. The Microstream™ and or Microstream™ Advance CO2 sampling lines can attach to the NanoPod CO2 port on the NanoPod enclosure. Microstream™ Capnography parameters are displayed the Microstream™ and or Microstream™ Advance CO2 sampling line is connected to the NanoPod CO2 Port and the Microstream™ CO2 NanoPod is connected via LEMO connector to a host monitoring system. Please note that the Microstream™ and Microstream™ Advance sampling lines have been cleared in their own respective 510K submissions. ### INTENDED USE / INDICATIONS FOR USE The Microstream™CO2 NanoPod, when connected to the host monitor, is intended to provide continuous, non-invasive measurement and monitoring of carbon dioxide (CO2) during the {5}------------------------------------------------ respiration cycle, EtCO2, FiCO2, and CO2-based respiration rate for adult, pediatric and neonatal patients, utilizing Microstream™ and Microstream™ Advance CO2 sampling lines. The Microstream™ CO2 NanoPod also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for age 22 and up. IPI and ODI values can be calculated and displayed only if the host monitor collects SpO2 data and provides this SpO2 data to the NanoPod in order to calculate these parameters. The Microstream™ CO2 NanoPod is intended for use in hospital-type facilities, during intra-hospital transport between and within areas of care. ### CONTRAINDICATIONS The device is not to be used as an apnea monitor. ### TECHNOLOGICAL CHARACTERISTICS The main difference between the subject device and the predicate, is the fact that the Microstream™ CO2 NanoPod is an external single parameter, capnography module, designed to be connected to a host monitor. The predicate Capnostream™35 Portable Respiratory Monitor (K150272), is a standalone, dual parameter monitor, equipped with MicroMediCO2 capnography module (product code CCK) and a pulse oximetry module (product code DQA). The subject device, the Microstream™ CO2 NanoPod, has the same intended population, principles of operation, and fundamental technology as the predicate device, the Capnostream™35 Portable Respiratory Monitor. The subject device is a derivative of the capnography module that is inside the predicate device with software and hardware modifications to enable the subject device to be used in conjunction with an Original Equipment Manufacturer (OEM) host monitor. Based on the results of the verification and validation studies. Covidien has established that the subject device, the Microstream™ CO2 NanoPod, is substantially equivalent to the predicate device. The following technological characteristics were compared between the subject device and the predicate device to demonstrate substantial equivalence. | Characteristic | Subject Device | Predicate Device K150272 | |-------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------| | Classification | II | II | | Device Classification<br>Name | Carbon Dioxide Gas Analyzer<br>and<br>Breathing Frequency Monitor | Carbon Dioxide Gas Analyzer<br>and<br>Breathing Frequency Monitor | | Product Code | CCK,<br>BZQ | CCK<br>DQA<br>MNR | | Purpose and Function | Continuous non-invasive<br>monitoring of expired and<br>inspired CO2, EtCO2, | Continuous non-invasive<br>monitoring of expired and<br>inspired CO2, EtCO2, | ### Table 1. Comparison of Technological Characteristics {6}------------------------------------------------ | Characteristic | Subject Device | Predicate Device K150272 | |---------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | capnography derived<br>respiration rate | capnography derived respiration<br>rate, SpO2 and pulse rate | | Energy Source | Powered by the host monitor | Power supply operating on AC<br>90-264VAC, 47 to 63 Hz;<br>alternatively, on a rechargeable<br>internal lithium-ion battery or on<br>a removable lithium-ion battery. | | Target population | Adult, pediatric, and neonatal<br>patients | Adult, pediatric, and neonatal<br>patients | | Intended user | Professionally trained health<br>care providers | Professionally trained health<br>care providers | | Where used | In hospitals, hospital-type<br>facilities, during intra-hospital<br>transport between and within<br>areas of care. | In hospitals, hospital-type<br>facilities, intra-hospital transport,<br>out-of-hospital Emergency<br>Medical Service applications<br>that include ground and air<br>transport. | | Fundamental<br>Technology | Non-dispersive infrared (NDIR)<br>(CO2) Spectroscopy<br>and Plethysmography | Non-dispersive infrared (NDIR)<br>(CO2) Spectroscopy<br>and Plethysmography | | Performance Standards | ISO 80601-2-55 | ISO 80601-2-55<br>ISO 80601-2-61 | | Main Safety Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 80601-2-49 (section<br>202.8.102, Figure 202 104) | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-49<br>IEC 60601-1-12<br>IEC 60601-1-6<br>IEC 60601-1-8 | ### PERFORMANCE DATA The following performance data were provided to support the substantial equivalence determination with the predicate device. ### Bench Performance Testing The subject device, the Microstream™ CO2 NanoPod, is substantially equivalent to the predicate device, the Capnostream™35 (Portable Respiratory Monitor). The functional features and the intended use and indications for use of the Microstream™ CO2 NanoPod are substantially equivalent to the predicate device. The Microstream™ CO2 NanoPod has successfully undergone performance, safety, electromagnetic and environmental testing to ensure compliance with expected performance standards. Software testing was performed to validate the performance of the Microstream™ CO2 NanoPod and its substantial equivalence to the predicate device. {7}------------------------------------------------ A hazard analysis was carried out on the Microstream™ CO2 NanoPod in compliance with ISO 14971:2019. This hazard analysis concluded that any residual risks were judged as acceptable when weighed against the intended benefits of use of the device. ### Animal Performance Testing Not Applicable. No animal performance testing was required to demonstrate device safety and effectiveness. ### Clinical Performance Testing Not Applicable. No clinical performance testing was required to demonstrate device safety and effectiveness. ### CONCLUSIONS The subject device, Microstream™ CO2 NanoPod is equivalent in intended use, technological characteristics, and performance to the existing legally marketed device. Testing did not raise any issue of safety and effectiveness; therefore, the device is substantially equivalent to the predicate device with respect to safety, effectiveness, indications for use and intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...