EC50 TOXCO+ BREATH CARBON MONOXIDE MONITOR

K070259 · Bedfont Scientific, Ltd. · CCJ · Feb 21, 2008 · Anesthesiology

Device Facts

Record IDK070259
Device NameEC50 TOXCO+ BREATH CARBON MONOXIDE MONITOR
ApplicantBedfont Scientific, Ltd.
Product CodeCCJ · Anesthesiology
Decision DateFeb 21, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1430
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The EC50 ToxCO+ Breath Carbon Monoxide Monitor and accessories are used by healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning.

Device Story

Handheld, microprocessor-controlled breath CO monitor; uses electrochemical gas sensor to detect CO in exhaled breath; provides non-invasive estimation of CO poisoning and Carboxyhaemoglobin (COHb) levels. Operated by healthcare professionals in clinical settings or triage; device features three-button interface (ON, CAL, RECALL) and alphanumeric LCD. Input is patient exhaled breath; output is CO concentration (0-500ppm) and COHb percentage (0-50%). Assists clinicians in identifying suspected CO poisoning cases when blood gas analysis is unavailable. Powered by 2 AA batteries; requires 6-month calibration intervals.

Clinical Evidence

Bench testing only. Performance metrics include sensor accuracy of ±5% of reading, concentration range of 0-50% COHb/0-500ppm CO, and H2 cross-interference <40% at 20°C. Compliance with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC) verified. Mouthpieces and facemasks meet ISO 10993 biocompatibility requirements.

Technological Characteristics

Handheld electrochemical gas analyzer; ABS case with synthetic rubber boot; 2 AA battery power. Sensor sensitivity 1 ppm; warm-up <10s. Operating range 0-30°C, 10-95% humidity. Microprocessor-controlled. Biocompatible mouthpieces/facemasks (ISO 10993).

Indications for Use

Indicated for healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning in patients where blood testing is unavailable or during triage.

Regulatory Classification

Identification

A carbon monoxide gas analyzer is a device intended to measure the concentration of carbon monoxide in a gas mixture to aid in determining the patient's ventilatory status. The device may use techniques such as infrared absorption or gas chromatography.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) # Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. | Submitter's Identification: | Bedfont Scientific Ltd<br>105 Laker Road<br>Rochester, Kent, United Kingdom ME 1 3QX | |-----------------------------|--------------------------------------------------------------------------------------| | Contact Person: | Mark Sandwell | | Telephone:<br>Fax: | 011 44 8700 844050<br>011 44 8700 844051 | | Date of Summary: | March 30, 2007 | | Device Name: | EC50 ToxCo + | | Classification Name: | Analyzer, Gas, Carbon Monoxide, Gaseous-Phase | | Product Code: | CCJ | | Product Class: | II | | Code of Federal Regulation: | 21 CFR 868.1430 | Substantial Equivalence: | Manufacturer | Trade Name | 510(k) Number | |--------------------|------------------|---------------| | Bedfont Scientific | Smokerlyzer | K892841 | | Bedfont Scientific | piCO Smokerlyzer | K000962 | #### 7.2 Substantial Equivalence The Bedfont EC50 ToxCO+, Bedfont piCO and Bedfont Smokerlyzer products are all part of the EC50 family, sharing similar electrochemical gas sensors for the detection of carbon monoxide. They share similar indications for use, accuracy, accessories and operating temperature ranges. All devices have similar power sources (Alkaline batteries) and meet general electrical standards IEC 60601-1. Bench testing has been carried out to ensure that the performance of the EC50 ToxCO+ is comparable with the predicate devices. Due to these similarities the ToxCO+ does not raise any new issues concerning safety and effectiveness. FEB 21 2008 {1}------------------------------------------------ #### 1.1 Description The EC50 ToxCO+ Breath Carbon Monoxide Monitor (figure 1 & 2) is a handheld device, which provides a non-invasive means of determining Carbon Monoxide (CO) poisoning and Carboxyhaemoglobin (COHb) when blood testing is not available. The EC50 ToxCO+ Breath Carbon Monoxide Monitor is also ideal for use during triage to ensure suspected cases of CO poisoning are not missed. The EC50 ToxCO+ Breath Carbon Monoxide Monitor is a microprocessor-controlled device powered by 2 AA batteries with an Alphanumerice LCD display housed in a sturdy ABS case with a synthetic rubber boot. The EC50 ToxCO+ Breath Carbon Monoxide Monitor is designed to have an operating temperature between 0-30° C, operating humidity 10-95% non-condensing and has storage temperature requirements of 0-30° C. The sensor sensitivity is 1 pmm, has a warm up time of <10 seconds-90% FSD, an accuracy of ±5% of reading, a concentration range of 0-50% COHb/0-500ppm CO and an H2 Cross Interference of <40% @ 20° C. The sensor operating life is 2-3 years with a company warranty of 6 months and requires calibrations at 6-month intervals. The EC50 ToxCO+ Breath Carbon Monoxide Monitor is an easy to use handheld device (figure 3) that has three clearly marked buttons: - . ON button, - . CAL calibration button - RECALL button . #### Indications for Use: 1.0 The EC50 ToxCO+ Breath Carbon Monoxide Monitor and accessories are used by healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning. #### 1.2 Standards Met The EC50 ToxCO+ Breath Carbon Monoxide Monitor has met IEC 60601-1 Electrical Safety for General Electrical Safety Standard and IEC 60601-1-2 Electromagnetic Compatibility - requirements and test. The mouthpiece used for the EC50 ToxCO+ Breath Carbon Monoxide Monitor is the same as the mouthpieces used for Bedfont EC50 Smokerlyzer Carbon Monoxide Monitor K892841 and Bedfont PICO Smokerlyzer Model EC-50 K000962 satisfying the ISO 10993 requirements. The facemasks used are commercially available for medical use and meet the requirements of ISO 10993. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES DEPARTMENT OF HEALTH & HUMAN SERVICES. USA FEB 21 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Bedfont Scientific Limited C/O Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087 Re: K070259 Trade/Device Name: EC50 ToxCO+ Breath Carbon Monoxide Monitor Regulation Number: 21 CFR 868.1430 Regulation Name: Carbon Monoxide Gas Analyzer Regulatory Class: II Product Code: CCJ Dated: February 7, 2008 Received: February 8, 2008 Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Mr. Lehtonen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice i requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clus Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: EC50 ToxCO+ Breath Carbon Monoxide Monitor Indications For Use: The EC50 ToxCO+ Breath Carbon Monoxide Monitor and accessories are used by healthcare professionals to determine levels of Carbon Monoxide (CO) poi soning. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mang Whal (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices V 0 Page 1 of 1 510(k) Number: 59
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...