Gas Module 3

K180788 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · BZK · Oct 24, 2018 · Anesthesiology

Device Facts

Record IDK180788
Device NameGas Module 3
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeBZK · Anesthesiology
Decision DateOct 24, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1850
Device ClassClass 2
AttributesPediatric

Indications for Use

The indications for use for the Gas Module 3 include monitoring of airway gases during anesthesia and/or assisted respiration. The intended environment of use is the anesthesia department, including the Operating Room (OR) and post anesthesia care units (PACU), etc.

Device Story

Gas Module 3 is an independently powered respiratory gas monitor; interfaces with Mindray patient monitors via RS232 using proprietary communication protocol. Device samples airway gases (CO2, O2, N2O, anesthetic agents: Halothane, Isoflurane, Sevoflurane, Desflurane, Enflurane) in real-time; utilizes AION Multigas Analyzer and Servomex Paramagnetic Oxygen sensor. System draws gas samples via water trap and sampling line at 200ml/min (adult/pediatric) or 120ml/min (neonate). Data displays on connected patient monitor; user commands entered via monitor. Provides breath-by-breath concentration and respiration rate monitoring. Used in OR/PACU by clinicians to manage patient ventilation and anesthesia delivery. Benefits include improved gas management for smaller patients via lower flow rates and faster rise times for gas detection compared to predicate.

Clinical Evidence

Bench testing only. Performance evaluated via functional and system-level testing, including EMC and electrical safety per AAMI ANSI ES 60601-1 and IEC 60601-1-2. Gas measurement performance validated against ISO 80601-2-55:2011. No clinical data presented.

Technological Characteristics

Components: AION Multigas Analyzer, Servomex Paramagnetic Oxygen sensor, power supply board, communication board. Connectivity: RS232 to patient monitor. Power: 100-240VAC, 50-60Hz. Standards: AAMI ANSI ES 60601-1, IEC 60601-1-2, ISO 80601-2-55. Sampling: 200ml/min or 120ml/min flow rates.

Indications for Use

Indicated for monitoring airway gases (CO2, O2, N2O, anesthetic agents) and respiration rate during anesthesia or assisted respiration in OR and PACU settings for adult, pediatric, and neonatal patients.

Regulatory Classification

Identification

A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 24, 2018 Shenzhen Mindray Bio-medical Electronics Co., Ltd Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South High-tech Industrial Park, Nanshan Shenzhen, 518057 CN Re: K180788 Trade/Device Name: Gas Module 3 Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: Class II Product Code: BZK Dated: September 19, 2018 Received: September 24, 2018 #### Dear Yanhong Bai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K180788 Device Name Gas Module 3 Indications for Use (Describe) The indications for use for the Gas Module 3 include monitoring of airway gases during anesthesia and/or assisted respiration. The intended environment of use is the anesthesia department, including the Operating Room (OR) and post anesthesia care units (PACU), etc. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------| | <div> <span> × Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Gas Module 3 is provided below. | Device Common Name: | Spirometer, Monitoring (W/wo Alarm) | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Proprietary Name: | Gas Module 3 | | Submitter: | Shenzhen Mindray Bio-medical Electronics Co., LTD.<br>Mindray Building, Keji 12th Road South,<br>Hi-tech Industrial Park, Nanshan,<br>Shenzhen, 518057, P. R. China<br>Tel: +86 755 8188 8998<br>Fax: +86 755 2658 2680 | | Contact: | Ms. Yanhong Bai<br>Manager Regulatory Affairs<br>Shenzhen Mindray Bio-medical Electronics Co., LTD.<br>Mindray Building, Keji 12th Road South,<br>High-tech Industrial Park, Nanshan, Shenzhen<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com | | Date Prepared: | October 23, 2018 | | Classification Regulation: | 21 CFR 868.1850, Class II, Monitoring spirometer | | Panel: | Anesthesiology | {4}------------------------------------------------ | Primary | | | | | |--------------|-------------------|----------------|------------------------|-------------------------------------| | Product Code | Regulation Number | Panel | Regulation description | Device Common Name | | BZK | 21 CFR 868.1850 | Anesthesiology | Monitoring spirometer | Spirometer, Monitoring (W/wo Alarm) | #### Classification Regulation, Classification Name and Product Codes: #### Primary Predicate Device: K062754 - GAS MODULE SE; DATASCOPE CORP. #### Indications for Use: The indications for use for the Gas Module 3 include monitoring of airway gases during anesthesia and/or assisted respiration. The intended environment of use is the anesthesia department, including the Operating Room (OR) and post anesthesia care units (PACU), etc. #### Device Description: The Gas Module 3 is an independently powered unit capable of interfacing with Mindray DS USA, Inc. (hereafter MRDS) Patient Monitors using a proprietary communications protocol. The system connects to the patient monitor via a RS232 connector. All Gas Module 3 data display on the monitor screen. And all user commands are entered on the monitor and then electronically transmitted to the Gas Module 3. The Gas Module 3 measures in real-time, breath-by-breath CO2, O2, N2O gases. Additionally, the Gas Module 3 monitors the anesthetic agents Halothane (HAL), Isoflurane (ISO), Sevoflurane (SEV), Desflurane (DES) and Enflurane (ENF). The Gas Module 3 consists of AION Multigas Analyzer, Servomex Paramagnetic Oxygen sensor, Power supply Board and Communication Board. The subject device is a modified version of a previously cleared model, the Gas Module SE. The Gas Module SE was cleared under K062754, on September 13, 2006. #### Non-clinical testing: EMC and Electrical Safety - To establish the substantial equivalence of the subject Gas Module 3 to the previously cleared Gas Module SE (K062754), Mindray conducted additional Electromagnetic Compatibility and Electrical Safety testing on the subject device. The testing provides an evaluation of the EMC and electrical safety of the subject devices relevant to each of the modifications since clearance of the predicate Gas Module SE. The Electromagnetic Compatibility and Electrical Safety testing shows that the subject device continues to meet the specifications and the performance of the subject device is equivalent to the predicate device. AAMI ANSI ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and ● {5}------------------------------------------------ A2:2010/(R)2012* Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests Performance Testing - Bench - To establish the substantial equivalence of the subject Gas Module 3 to the previously cleared Gas Module SE (K062754). Mindray conducted functional and system level testing on the subject devices. The testing provides an evaluation of the performance of the subject devices relevant to each of the modifications since clearance of the Gas Module SE. The functional and system level testing shows that the subject device continues to meet the specifications and the performance of the subject device is equivalent to the predicate device. In addition, Mindray has conducted testing to ensure the subject device meets following consensus standards. - ISO 80601-2-55: 2011 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors ## Software Verification and Validation Testing Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the subject Gas Module 3 was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product. ## Substantial Equivalence: Comparison of Indications - The indication for use of the subject device (Gas Module 3) is identical to that of the predicate device (Gas Module SE), with only the device name changed. The minor change to the indication for use does not change the fundamental intended use of the Gas Module 3. Comparison of Technological Characteristics - The table below compares the key technological feature of the subject device to the primary predicate device (Gas Module SE, K062754). The features in gray are the main features that have been modified since their previous clearances. ## Device Comparison Table {6}------------------------------------------------ | Feature | Predicate Device<br>(K062754) | Subject Device | Comparison analysis | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Gas Module SE | Gas Module 3 | | | Indications for<br>Use | The indications for use<br>for the Gas Module SE<br>include monitoring of<br>airway gases, during<br>anesthesia and/or<br>assisted respiration. The<br>intended environment of<br>use is the anesthesia<br>department, including the<br>Operating Room (OR)<br>and post anesthesia care<br>units (PACU), etc. | The indications for use for<br>the Gas Module 3 include<br>monitoring of airway gases<br>during anesthesia and/or<br>assisted respiration. The<br>intended environment of<br>use is the anesthesia<br>department, including the<br>Operating Room (OR) and<br>post anesthesia care units<br>(PACU), etc. | Same<br>The indication for use of<br>the subject device (Gas<br>Module 3) is identical to<br>that of the predicate<br>device (Gas Module SE),<br>with only the device<br>name changed. | | Power supply | Multi-voltage power<br>supply, source 100-<br>240VAC, 50-60Hz. | Multi-voltage power<br>supply, source 100-<br>240VAC, 50-60Hz. | Same | | User interface | Operation is controlled<br>by a connected patient<br>monitor via Mindray DS<br>USA proprietary<br>communication protocol<br>by a RS232 connector. | Operation is controlled by a<br>connected patient monitor<br>via Mindray DS USA<br>proprietary communication<br>protocol by a RS232<br>connector. | Same | | Measured<br>gases | CO2, O2, N2O and<br>Anesthetic agents<br>(Halothane, Isoflurane,<br>Sevoflurane, Desflurane<br>and Enflurane). | CO2, O2, N2O and<br>Anesthetic agents<br>(Halothane, Isoflurane,<br>Sevoflurane, Desflurane<br>and Enflurane). | Same | | Measured<br>parameters | Patient inspired/expired<br>gas concentration and<br>respiration rate. | Patient inspired/expired gas<br>concentration and<br>respiration rate. | Same | | Warm-up time | 2 minutes to operation<br>with CO2, O2 and N2O. 5<br>minutes to operation of<br>anesthetic agents. 30<br>minutes for full<br>specification. | 45 seconds to operation<br>with CO2, O2, N2O and<br>anesthetic agents. 10<br>minutes for full<br>specification. | Similar, warm-up time<br>improved in Gas Module 3. | | Sampling rate | The sampling system<br>takes care of drawing a<br>gas sample to analyzers<br>at a fixed rate<br>(200 $\pm$ 20ml/min). | 200 $\pm$ 20ml/min with larger<br>size water trap.<br>120 $\pm$ 12ml/min with smaller<br>sizewater trap. | Same - The flow rate for<br>larger size water trap is the<br>same. The option to use a<br>new smaller size water trap<br>is introduced which support<br>a flow rate of 120ml/min,<br>which may provide<br>enhanced gas management<br>for smaller sized patients. | | Gas sampling<br>accessories | Water trap<br>Sampling line | Adult/Pediatric Water trap<br>Neonate Water trap<br>Adult/Pediatric Sampling<br>line<br>Neonate Sampling line | The water trap and<br>sampling line is changed<br>from Datex-Ohmeda to<br>Artema Technology. | {7}------------------------------------------------ {8}------------------------------------------------ | Adapter, Straight<br>Adapter, Elbow | Adapter, Straight<br>Adapter, Elbow | In Gas Module SE, there is<br>one water trap (PN: 0202-<br>00-0129) and one sample<br>line (PN: 0683-00-0451-10)<br>for all patient types (Adult,<br>Pediatric and Neonate) with<br>the sample line flow rate<br>setting 200ml/min.<br>In Gas Module 3, there is<br>one water trap (PN: 0202-<br>00-0182-10) and sample<br>line (PN: 0683-00-0525-25)<br>for adult and pediatric<br>patient with the sampling<br>flow rate setting<br>200ml/min. And a small<br>size water trap (PN: 0202-<br>00-0181-10) and sample<br>line (PN: 0683-00-0524-25)<br>is introduced for the flow<br>rate 120ml/min, which is an<br>enhancement for neonate<br>patients.<br>The adapter is the same. | |-------------------------------------|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| {9}------------------------------------------------ | Measurement<br>range | CO2:<br>O2:<br>100%<br>N2O:<br>DES:<br>SEV:<br>ENF:<br>ISO:<br>HAL:<br>awRR:<br>rpm | 0 to 15%<br>0 to 100%<br>0 to 100%<br>0 to 20%<br>0 to 8%<br>0 to 6%<br>0 to 6%<br>0 to 6%<br>4 to 60<br>rpm | CO2:<br>O2:<br>N2O:<br>DES:<br>SEV:<br>ENF:<br>ISO:<br>HAL:<br>awRR:<br>rpm | 0 to 10%<br>0 to 100%<br>0 to 100%<br>0 to 18%<br>0 to 8%<br>0 to 5%<br>0 to 5%<br>0 to 5%<br>2 to 100 | Although the measurement range of CO2, DES, ENF, ISO and HAL for subject device is small than the predicate device, the difference in the measurement range of these gases for the subject device is insignificant in for the clinical use.<br>The awRR measurement range for the subject device is wider than the predicate device.<br>The awRR measurement range of the subject device has been tested using the test apparatus in figure 201.101 of ISO80601-2-55, and the test result meets the specification. | |----------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Full accuracy | CO2 accuracy: | $\pm$ (0.2% + 2% of reading) | CO2 accuracy: | $\pm$ 0.1% @ 0 - 1%<br>$\pm$ 0.2% @ 1 - 5%<br>$\pm$ 0.3% @ 5 - 7%<br>$\pm$ 0.5% @ 7 - 10%<br>Unspecified @ > 10% | The CO2 measurement accuracy is now specified in different ranges. These accuracies in different ranges meet the requirements of ISO 80601-2-55. | | | O2 accuracy: | $\pm$ (1% + 2% of reading) | O2 accuracy: | $\pm$ 1% @ 0 - 25%<br>$\pm$ 2% @ 25 - 80%<br>$\pm$ 3% @ 80 - 100% | The O2 measurement accuracy is now specified in different ranges. These accuracies in different ranges continue to meet the requirements of ISO 80601-2-55 | {10}------------------------------------------------ | N2O accuracy: $\pm$ (1% +<br>2% of reading) | N2O accuracy:<br>$\pm$ 2% @ 0 – 20%<br>$\pm$ 3% @ 20 – 100% | The N2O measurement<br>accuracy is now specified in<br>different ranges. These<br>accuracies in different<br>ranges continue to meet the<br>requirements of ISO 80601-<br>2-55. | |------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | DES accuracy: $\pm$ (0.15% + 5% of<br>reading) | DES accuracy:<br>$\pm$ 0.1% @ 0 - 1%<br>$\pm$ 0.2% @ 1 - 5%<br>$\pm$ 0.4% @ 5 - 10%<br>$\pm$ 0.6% @ 10 - 15%<br>$\pm$ 1% @ 15 - 18%<br>Unspecified @ > 18% | The DES measurement<br>accuracy is now specified in<br>different ranges. These<br>accuracies in different<br>ranges still meet the<br>requirements of ISO 80601-<br>2-55. | | SEV accuracy: $\pm$ (0.15%<br>+ 5% of reading) | SEV accuracy:<br>$\pm$ 0.15% @ 0 - 1%<br>$\pm$ 0.2% @ 1 - 5%<br>$\pm$ 0.4% @ 5 - 8%<br>Unspecified @ > 8% | The SEV measurement<br>accuracy is now specified in<br>different ranges. These<br>accuracies in different<br>ranges continue to meet the<br>requirements of ISO 80601-<br>2-55. | | ENF accuracy: $\pm$ (0.15%<br>+ 5% of reading) | ENF accuracy:<br>$\pm$ 0.15% @ 0 - 1%<br>$\pm$ 0.2% @ 1 - 5%<br>Unspecified @ > 5% | The ENF measurement<br>accuracy is now specified in<br>different ranges. These<br>accuracies in different<br>ranges continue to meet the<br>requirements of ISO 80601-<br>2-55. | | ISO accuracy: $\pm$ (0.15% +<br>5% of reading) | ISO accuracy:<br>$\pm$ 0.15% @ 0 - 1%<br>$\pm$ 0.2% @ 1 - 5%<br>Unspecified @ > 5% | The ISO measurement<br>accuracy is now specified in<br>different ranges. These<br>accuracies in different<br>ranges continue to meet the<br>requirements of ISO 80601-<br>2-55. | l {11}------------------------------------------------ | | HAL accuracy: ±(0.15% + 5% of reading) | HAL accuracy:<br>±0.15 % @ 0 - 1%<br>±0.2 % @ 1 - 5%<br>Unspecified @ > 5% | The Hal measurement<br>accuracy is now specified in<br>different ranges. These<br>accuracies in different<br>ranges continue to meet the<br>requirements of ISO 80601-<br>2-55. | |------------|----------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Rise time | CO2 rise time: < 400ms | CO2 rise time: < 250 ms<br>for 200ml/min<br>CO2 rise time: < 250 ms<br>for 120ml/min | The CO2 rise time for<br>subject device is smaller<br>than the predicate device.<br>The smaller rise time means<br>the device can detect the<br>variety of the concentration<br>of the gas with greater<br>sensitively. | | | O2 rise time : < 400ms | O2 rise time: < 400 ms<br>for 200ml/min<br>O2 rise time: < 450 ms for<br>120ml/min | The difference of rise time<br>between the subject device<br>and the predicate device is<br>insignificant in clinical use.<br>So no clinical risks or<br>safety issue introduced. | | | N2O rise time: < 400ms | N2O rise time: < 250 ms<br>for 200ml/min<br>N2O rise time: < 250 ms<br>for 120ml/min | The N2O rise time for<br>subject device is smaller<br>than the predicate device. | | | DES rise time: <600ms | DES rise time: <300ms<br>for 200ml/min.<br>DES rise time: <300ms<br>for 120ml/min. | The DES rise time for<br>subject device is smaller<br>than the predicate device. | | | SEV rise time: <600ms | SEV rise time: <300ms for<br>200ml/min.<br>SEV rise time: <300ms for<br>120ml/min. | The SEV rise time for<br>subject device is smaller<br>than the predicate device. | | | ENF rise time: <600ms | ENF rise time: <350ms for<br>200ml/min.<br>ENF rise time: <350ms for<br>120ml/min | The ENF rise time for<br>subject device is smaller<br>than the predicate device. | | | ISO rise time: <600ms | ISO rise time: <300ms for 200ml/min.<br>ISO rise time: <300ms for 120ml/min. | The ISO rise time for subject device is smaller than the predicate device. | | | HAL rise time: <1000ms | HAL rise time: <300ms for 200ml/min.<br>HAL rise time: <300ms for 120ml/min. | The HAL rise time for subject device is smaller than the predicate device. | | Delay time | All gases delay time: <2.5 s | All gases delay time: <4 s for 200ml/min<br>All gases delay time: <4 s for 120ml/min | The difference of delay time between the subject device and the predicate device is insignificant in clinical use.<br>So no clinical risks or safety issue introduced. | {12}------------------------------------------------ ### Substantial Equivalence Conclusion: Based on the detailed comparison of specifications and modifications to the subject device (Gas Module 3), predicted device (Gas Module SE, K062754), the performance testing results and conformance with applicable standards the Gas Module 3 can be found substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...