MERCURY MEDICAL NEGATIVE INSPIRATORY FORCE (NIF) DISPOSABLE MONOMETER
K032593 · Mercury Medical · BXR · Feb 3, 2004 · Anesthesiology
Device Facts
Record ID
K032593
Device Name
MERCURY MEDICAL NEGATIVE INSPIRATORY FORCE (NIF) DISPOSABLE MONOMETER
Applicant
Mercury Medical
Product Code
BXR · Anesthesiology
Decision Date
Feb 3, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1780
Device Class
Class 2
Indications for Use
The Mercury Medical Negative Inspiratory Force (NIF) Disposable Manometer is used to indicate inspiratory force. Federal law (USA) restricts this device to sale by or on the order of a physician.
Device Story
Disposable manometer for measuring negative inspiratory force; connects to endotracheal tube. Clinician holds occlusion knob during patient inhalation to record maximum inspiratory pressure; memory pointer retains reading up to -60cm H2O. Knob rotation resets pointer. Used in hospital or transport settings by clinicians. Provides objective pressure data to assist in evaluating patient respiratory effort and clinical decision-making regarding weaning or respiratory support.
Clinical Evidence
Bench testing only. Performance and specifications verified against ISO 5356-1:1987 and ASTM F1054 standards for conical connectors.
Technological Characteristics
Mechanical manometer; 15mm and 22mm conical fittings per ASTM F1054; ISO 5356-1 compliant. Measures negative pressure via manual occlusion knob and memory pointer mechanism. No software or electronic components.
Indications for Use
Indicated for patients requiring measurement of inspiratory force to assess respiratory status.
Regulatory Classification
Identification
An inspiratory airway pressure meter is a device used to measure the amount of pressure produced in a patient's airway during maximal inspiration.
Predicate Devices
DHD Medical Products Negative Inspiratory Force Monitoring Kit
K143148 — Disposable Pressure Manometer (DPM) · Westmed, Inc. · Aug 11, 2015
K081693 — NS SERIES NIF (NEGATIVE INSPIRATORY FORCE) METERS · Instrumentation Industries, Inc. · Sep 11, 2008
K140370 — DISPOSABLE PRESSURE MANOMETER · Besmed Health Business Corp · Jun 19, 2014
K022402 — DISPOSABLE PRESSURE MANOMETER · Vital Signs, Inc. · Aug 22, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
#### ENCLOSURE L FEB - 3 2004
Premarket Notification (510(k) - Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer
#### Summary of Safety and Effectiveness
#### Non-Confidential Summary of Safety and Effectiveness
Page 1 of 2 January 29, 2004
Mercury Enterprises, Inc./Mercury Medical 11300 49th St. N. Clearwater, Fl. 33762 Tel: (800) 237-6418 Fax: (727) 572-4501
Official Contact: Karen Seltzer QA/RA Manager
#### Proprietary or Trade Name: Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer
Common/Usual Name: NIF Manometer
Classification: Class II, BXR, 21 CFR 868.1780
Classification Name: Airway Pressure Meter (Inspiratory Force)
Device: Negative Inspiratory Force (NIF) Disposable Manometer
Predicate Devices: DHD Medical Products Negative Inspiratory Force Monitoring Kit
Device Description:
The Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer is designed to connect to an endotracheal tube. To record the maximum inspiratory pressure, the occlusion knob is held down during patient inhalation. The knob is released immediately after measurement. A memory pointer identifies and remains at the maximum inspiratory pressure to -60cm H2O. Rotating the memory pointer knob counterclockwise resets the memory pointer.
{1}------------------------------------------------
## ENCLOSURE L
Premarket Notification (510(k) - Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer
#### Non-Confidential Summary of Safety and Effectiveness
Page 2 of 2 January 29, 2004
Indicated Use:
The Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer is used to indicate inspiratory force. Federal law (USA) restricts this device to sale by or on the order of a physician.
Technical Characteristics: The device has the same technical characteristics for measuring negative pressure as the DHD Medical Products NIF Monitoring kit predicate device.
Non-Clinical Data: Performance and specifications of the device are consistent with all requirements for this device type specified by ISO 5356-1: 1987 - Anesthetic and Respiratory Equipment-Conical connectors-Part 1: Cones and Sockets. ASTM F1054 - Standard Specification for Conical Fittings of 15mm and 22mm sizes.
Environment of Use: Hospital/Transport
Conclusions: The comparison to the predicate devices demonstrates that the proposed device is safe and effective and is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of three abstract shapes resembling waves or stylized human figures, stacked on top of each other.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 3 2004
Mr. Wayne Glover QA/RA Engineer Mercury Medical Inc. 11300 49th St. N. Clearwater, FL 33762
Re: K032593
Trade/Device Name: Mercury Medical Negative Inspiratory Force (NIF) Disposable Manometer Regulation Number: 21 CFR 868.1780 Regulation Name: Meter, Airway Pressure (Inspiratory Force) Regulatory Class: Class II Product Code: 73 BXR Dated: December 18, 2003 Received: December 19, 2003
Dear Mr. Glover:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{3}------------------------------------------------
Page 2 - Mr. Wayne Glover
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Chris Lins, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for Use
| 510(k) Number (if known): | K032593 |
|---------------------------|------------------------------------------------------------------------------------------------------------------|
| Device Name: | Negative Inspiratory Force (NIF) Disposable Nanometer |
| Intended Use: | The Mercury Medical Negative Inspiratory Force (NIF) Disposable Manometer is used to indicate inspiratory force. |
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L.A. Winterhausen
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
2 510(k) Number:
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.