K Number
K994318

Validate with FDA (Live)

Date Cleared
2000-02-23

(63 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VISTAKON Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eves who may have 1.00 D or less of astigmatism.

The VISTAKON BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with nondiseased eyes who may have 0.75 D or less of astigmatism.

The VISTAKON TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

The VISTAKON TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.

VISTAKON Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

Device Description

The device description does not change from that cleared under K991163.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the VISTAKON contact lenses:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state specific quantitative acceptance criteria or detailed device performance metrics in the way typically seen for new medical device clearances. This 510(k) submission is for a labeling modification to an already cleared device (K991163), not for a new device or a significant change in its design or performance.

The "acceptance criteria" here are implicitly related to demonstrating that the revised labeling (specifically, the claim about reduced risk of giant papillary conjunctivitis) is supported by existing clinical evidence.

However, based on the information, we can infer the "performance" goal as demonstrating the established safety and efficacy for the original device, as the current submission relies on that prior clearance.

Acceptance Criterion (Inferred)Reported Device Performance
Safety and Efficacy for intended use (refractive error correction, UV protection)Established by prior clearance K991163 (details not provided in this document)
Support for revised labeling claim regarding reduced risk of giant papillary conjunctivitis (GPC) with specific wear schedulesSupported by clinical studies published in the July 1999 CLAO Journal. (Specific findings not detailed here, but the FDA's clearance implies the claim was adequately supported).

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated in this document. The submission references "clinical studies published in the July 1999 CLAO Journal" (Attachment A) as the basis for the labeling modification. The sample size would be found within those external publications, but is not provided here.
  • Data Provenance: The document does not specify the country of origin of the data. It refers to "clinical studies published in the July 1999 CLAO Journal," which could be from various locations. The studies are assumed to be prospective given they are clinical studies, likely trials, investigating the impact of wear schedules.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided in the submission. The "ground truth" for the claims about GPC risk reduction would have been established by the methods described in the referenced clinical studies (e.g., clinical examination, diagnosis by ophthalmologists/optometrists). The number and qualifications of the experts involved in those studies are not detailed in this 510(k).

4. Adjudication Method for the Test Set

This information is not provided. As the submission refers to external clinical studies, the adjudication methods (if any, for outcomes like GPC diagnosis) would be detailed within those publications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done for this submission. MRMC studies are typically used to evaluate the impact of AI-assisted diagnostic devices on human reader performance. This submission concerns a contact lens, not a diagnostic AI system.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone performance evaluation was not done. This is not an AI/algorithm-based device. The "performance" being evaluated is related to the clinical outcomes associated with the wear schedule of the contact lens.

7. The Type of Ground Truth Used

The ground truth used for the claims about the reduction of giant papillary conjunctivitis (GPC) would have been clinical diagnosis by eye care professionals. This would involve direct observation and examination of the patient's eyes for signs of GPC as per established diagnostic criteria. It is a form of "expert consensus" or "clinical diagnosis" based on observable medical conditions.

8. The Sample Size for the Training Set

This information is not applicable as this is not an AI/machine learning device that requires a training set. The clinical studies supporting the claims would have involved a patient cohort, but it's not a "training set" in the context of an algorithm.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for an AI/ML algorithm. For the clinical studies, the "ground truth" for GPC diagnosis would have been established by the clinical assessments and diagnoses of the investigators or study clinicians.

{0}------------------------------------------------

FEB 2 3 2000

K 9943/8

Summary of 510(k) Submission

Name andaddress ofsubmitterVistakon, Johnson & Johnson Vision Products, Inc.7500 Centurion ParkwayJacksonville, Florida 32256Contact: Sharon BriggsPhone: (904) 443-1471Date Prepared: December 21, 1999
Identification ofdevice• The trade name is VISTAKON (lenefilcon A) soft (hydrophilic) contactlenses, clear and visibility Tinteded, with UV blocker, for daily wear• The common or usual name is Soft (hydrophilic) Contact Lens (dailywear).• The FDA Classification is Class II.
PredicatedevicesThe predicate device is VISTAKON (lenefilcon A) soft (hydrophilic) contactlenses, clear and visibility Tinteded, with UV blocker, for daily wear coveredunder K991163.
Description ofdeviceThe device description does not change from that cleared under K991163.
Continued on next page

{1}------------------------------------------------

Summary of 510(k) Submission, Continued

Indications forUseThe VISTAKON Contact Lens (spherical) is indicated for daily wearfor the correction of refractive ametropia (myopia and hyperopia) inaphakic or non-aphakic persons with non-diseased eyes who may have1.00 D or less of astigmatism.The VISTAKON BIFOCAL Contact Lens is indicated for daily wearfor the correction of distance and near vision in presbyopic, aphakic ornon-aphakic persons with non-diseased eyes who may have 0.75 D orless of astigmatism.The VISTAKON TORIC Contact Lens is indicated for daily wear forthe correction of visual acuity in aphakic or non-aphakic persons withnon-diseased eyes that are hyperopic or myopic and may have 10.00 Dor less of astigmatism.The VISTAKON TORIC BIFOCAL Contact Lens is indicated fordaily wear for the correction of distance and near vision in presbyopicaphakic or non-phakic persons with non-diseased eyes who may have10.00 D of astigmatism or less.VISTAKON Contact Lenses help protect against transmission ofharmful UV radiation to the cornea and into the eye.Eye care practitioners may prescribe the lens for either single-usedisposable wear or for frequent/planned replacement wear, withcleaning, disinfection and scheduled replacement. When prescribedfor frequent/planned replacement wear, the lens may be disinfectedusing a chemical disinfection system only.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Continued on next page

{2}------------------------------------------------

Summary of 510(k) Submission, Continued

:

Reason for510(k)The reason for the 510(k) is to revise the “Wear Schedule” in the DraftPackage Insert, and the “Introduction” statements in the Draft PatientInstruction Guides (Disposable and Frequent Replacement) for VISTAKON(lenefilcon A) soft (hydrophilic) contact lenses, clear and visibility tinted,with UV blocker, for daily wear to include the following statement:When VISTAKON UV Blocking Contact Lenses are replaced atintervals ranging from 1 day to 2 weeks, the risk of developing giantpapillary conjunctivitis may be reduced .
CharacteristicsThe characteristics do not change. They are the same as previously submittedin K991163.
Non-clinicalstudiesNon-clinical studies (chemistry, toxicology, microbiology, shelf-life, andleachability) on the lens material were not conducted because the lensmaterial, lenefilcon A, does not change.
Clinical studiesThis 510(k) describes a labeling modification which is supported by clinicalstudies published in the July 1999 CLAO Journal (see Attachment A). Thereis no change in lens material, the manufacturing process, nor the parametersand properties, therefore, the clinical data previously submitted in K991163supports the clinical safety of the subject device.
Conclusionsdrawn fromstudiesAdditional studies were not conducted, therefore, the conclusions drawn fromstudies previously submitted in K991163 support the non-clinical and clinicalsafety of the subject device. The clinical studies published in the July 1999CLAO Journal support the labeling modification that is the subject of this510(k).

・・・・・・・

·

For the disposable patient guide, the statement is revised to: "When VISTAKON UV Blocking Contact Leases are replaced daily, the risk of developing giant papillary conjunctivitis may be reduced."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 3 2000

Ms. Sharon A. Briggs Manager, Regulatory Submissions VISTAKONTM 7500 Centurion Parkway Jacksonville, FL 32256

Re: K994318

Trade Name: VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lens Clear and Visibility Tinted with UV blocker

Regulatory Class: II Product Code: 86 LPL Dated: December 21, 1999 Received: December 22, 1999

Dear Ms. Briggs:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{4}------------------------------------------------

Page 2 - Ms. Sharon A. Briggs

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications Statement

510(k) Number (if known):

VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lens Clear and Device Name: Visibility Tinted with UV blocker

Indication for Use:

The VISTAKON Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eves who may have 1.00 D or less of astigmatism.

The VISTAKON BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with nondiseased eyes who may have 0.75 D or less of astigmatism.

The VISTAKON TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

The VISTAKON TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.

VISTAKON Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseX
OROver the Counter

(Division Sign-Off)

Division of Ophthalmic Devices

510(k) Number_K 99 43 | 4

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.