K Number
K993687

Validate with FDA (Live)

Device Name
UROVIEW 2800
Date Cleared
1999-12-21

(50 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UroView 2800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical applications may include but are not limited to urologic and endoscopic procedures. The system may be used for other imaging applications at the physician's discretion.

Device Description

The UroView 2800 is a fluoroscopic x-ray system including a tilting patient-support table with overhead x-ray tube assembly, high-voltage x-ray generator, image intensifier, and video image display. A separate mobile workstation provides digital image processing, image storage capability, and additional video display.

AI/ML Overview

The provided document K993687, concerning the UroView 2800, is a 510(k) summary and not a comprehensive study report. Therefore, it primarily focuses on establishing substantial equivalence to predicate devices and adherence to relevant standards rather than presenting detailed experimental data on acceptance criteria and device performance as one might find in a clinical trial or performance study.

Based on the provided text, a detailed table of acceptance criteria and reported device performance, information on sample sizes, expert qualifications, and adjudication methods for a test set, and specific details about MRMC studies, standalone algorithm performance, or training set ground truth cannot be fully extracted.

The document mainly highlights the device's adherence to various safety and performance standards without explicitly stating "acceptance criteria" through a defined study with quantifiable outcomes against a benchmark. The "performance" described is in the context of meeting general functional requirements and safety standards.

Here's a breakdown of what can be inferred or is missing based on your request:


1. A table of acceptance criteria and the reported device performance

The document does not present a table of specific, quantifiable acceptance criteria alongside reported device performance derived from a dedicated study. Instead, the device's "performance" is implicitly deemed acceptable by its compliance with various recognized industry standards.

Acceptance Criteria (Inferred from Standards and Device Description)Reported Device Performance (Inferred from Compliance)
Fluoroscopic and Spot-Film Imaging Capability:Provides fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures.
Clinical Application Versatility:Usable for urologic and endoscopic procedures, and other imaging applications at the physician's discretion.
General Safety and Performance:Designed in accordance with product safety and performance requirements established in standards such as: - 21 CFR 1020.30-32 (Federal Performance Standard for Diagnostic X-ray Systems) - ANSI/NFPA 70 & 99 (National Electrical Code and Standard for Health Care Facilities) - UL 2601 (Medical Electrical Equipment) - CSA-C22.2 No. 601.1-M90 (Medical Electrical Equipment) - IEC 60601-1 (Medical Electrical Equipment, General Requirements for Safety) - IEC 60601-1-2 (Medical Electrical Equipment, Electromagnetic Compatibility) - IEC 60601-1-3 (Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray) - IEC 60601-1-4 (Medical Electrical Equipment, Programmable Electrical Medical Systems) - IEC 60601-2-7 (Medical Electrical Equipment, Safety of HV/X-ray Generators) - IEC 60601-2-28 (Medical Electrical Equipment, X-ray Tubes and X-ray Source Assemblies) - IEC 60601-2-32 (Medical Electrical Equipment, Safety of Associated X-ray Equipment) - IEC 60601-2-46 (Medical Electrical Equipment, Safety of Operating Tables) - 93/42/EEC - Annex 1 (Essential Requirements of the Medical Devices Directive)
Environmental Parameters (EMC):Complies with IEC 60601-1-2 (Electromagnetic Compatibility).
Radiation Protection:Complies with IEC 60601-1-3 (Radiation Protection in Diagnostic X-ray).
Electrical Safety:Complies with UL 2601, CSA-C22.2 No. 601.1-M90, IEC 60601-1, and relevant parts of IEC 60601-2 series for X-ray generators, tubes, and associated equipment.
Software/Programmability:Complies with IEC 60601-1-4 (Programmable Electrical Medical Systems).
Patient Support Table Safety:Complies with IEC 60601-2-46 (Safety of Operating Tables).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Missing. The document does not describe a "test set" in the context of an AI/algorithm evaluation. This is a medical device (fluoroscopic x-ray system), not an AI algorithm. Its performance is assessed through compliance with engineering and safety standards, not typically by analyzing a dataset of medical images.
  • Data Provenance: Not applicable as it's not an AI/data-driven device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Missing/Not Applicable. The concept of "ground truth" established by experts for a test set is not relevant for this type of device submission. Its safety and effectiveness are established through adherence to engineering and performance standards, and comparisons to predicate devices, rather than through diagnosis or interpretation of clinical images by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Missing/Not Applicable. Adjudication methods are typically used in clinical studies involving interpretation tasks and expert consensus, which is not the nature of this 510(k) submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. The UroView 2800 is a physical medical imaging system, not an AI or computer-aided detection/diagnosis (CAD) system. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not have been performed or reported in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable in the traditional sense. The "ground truth" for this device's acceptance is its compliance with established regulatory and industry standards for X-ray systems (e.g., radiation output limits, electrical safety, image quality specifications as defined by the standards).

8. The sample size for the training set

  • Not Applicable. The UroView 2800 is a conventional medical device; there is no "training set" as it does not employ machine learning or AI algorithms that require data for training.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

K993687

DEC 21 1999

Image /page/0/Picture/2 description: The image shows the logo for OEC Medical Systems, Inc. The logo consists of the letters "OEC" stacked on top of the words "OEC MEDICAL SYSTEMS, INC". The letters "OEC" are in a bold, sans-serif font, and the words "OEC MEDICAL SYSTEMS, INC" are in a smaller, serif font.

384 WRIGHT BROTHERS DRIVE SALT LAKE CITY, UTAH 84116 801 - 328 - 9 300 FAX 801-328-4300

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

Date:October 25, 1999
Name of Submitter:OEC Medical Systems, Inc.384 Wright Brothers DriveSalt Lake City, UT 84116801-328-9300
Corresponding Official:Ted L. Parrot, Vice President,Quality Assurance & Regulatory Affairs
Device Proprietary Name:UroView 2800
Classification Name:Image Intensified Fluoroscopic X-ray System
Common/Usual Names:Fluoroscopic Imaging SystemUrology Table
Substantial Equivalence:The UroView 2800 is substantially equivalent to the followingdevices that are currently marketed:
• OEC UroView 2600(K940295)
• Liebel-Flarsheim Hydradjust IV(K943581)
• Liebel-Flarsheim HydraVision(K004145)

Device Description:

General Description

The UroView 2800 is a fluoroscopic x-ray system including a tilting patient-support table with overhead x-ray tube assembly, high-voltage x-ray generator, image intensifier, and video image display. A separate mobile workstation provides digital image processing, image storage capability, and additional video display.

{1}------------------------------------------------

Indications for Use

The UroView 2800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical applications may include but are not limited to urologic and endoscopic procedures. The system may be used for other imaging applications at the physician's discretion.

User Characteristics

The device is used by health care professionals such as medical doctors, surgeons, radiologists and technologists in a hospital or clinical environment. In addition to being qualified within their respective medical fields, users must be trained in the use of medical x-ray equipment. OEC applications specialists train the user in the proper use of this product. The device labeling stipulates that only properly trained persons operate this equipment.

Standards:

The UroView 2800 is designed in accordance with product safety and performance requirements established in the following standards:

21 CFR 1020.30-32Federal Performance Standard for Diagnostic X-ray Systems
ANSI/NFPA 70 & 99National Electrical Code and Standard for Health Care Facilities
UL 2601Medical Electrical Equipment
CSA-C22.2 No. 601.1-M90Medical Electrical Equipment
IEC 60601-1Medical Electrical Equipment, General Requirements for Safety
IEC 60601-1-2Medical Electrical Equipment, Electromagnetic Compatibility
IEC 60601-1-3Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray
IEC 60601-1-4Medical Electrical Equipment, Programmable Electrical Medical Systems
IEC 60601-2-7Medical Electrical Equipment, Safety of HV/X-ray Generators
IEC 60601-2-28Medical Electrical Equipment, X-ray Tubes and X-ray Source Assemblies
IEC 60601-2-32Medical Electrical Equipment, Safety of Associated X-ray Equipment
IEC 60601-2-46Medical Electrical Equipment, Safety of Operating Tables
93/42/EEC - Annex 1Essential Requirements of the Medical Devices Directive

This concludes this 510(k) Summary.

OEC, MEDICAL SYSTEMS, INC.

Hamo

Ted L. Parrot Vice President, Quality Assurance & Regulatory Affairs

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem depicting a stylized eagle or bird-like figure with three human profiles incorporated into its design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 21 1999

Ted L. Parrot Vice President, Quality Assurance/ Regulatory Affairs and Official Correspondent OEC Medical Systems, Inc. 384 Wright Brothers Drive Salt Lake City, Utah 84116

Re:

K993687 UroView 2800 Dated: October 25, 1999 Received: November 1, 1999 Regulatory class: II 21 CFR 892.1650/Procode: 90 JAA 21 CFR 876.4890/Procode: 78 KQS

Dear Mr. Parrot:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications For Use Statement

OEC Medical Systems, Inc. Applicant:

510(k) No. (if known):

UroView 2800 Device name:

  • The UroView 2800 is designed to provide fluoroscopic and spot-film Indications for use: imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical applications include but are not limited to urologic and endoscopic procedures. The system may be used for other imaging applications at the physician's discretion.
    (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use ﮐﮯ (Per 21 CFR 801.109)

OR

Over-The-Counter

(Optional Format 1-2-96)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.