(75 days)
LiquiBand® Dental n-butyl cyanoacrylate is intended for use as an adhesive when bonding of caps and crowns to teeth is desired.
The LiquiBand® Dental adhesive formula consists of n-butyl cyanoacrylate monomer containing small amounts of hydroquinone (a free radical inhibitor) and sulfur dioxide (an anionic inhibitor).
LiquiBand® Dental, n-butyl cyanoacrylate adhesive, is a clear, colorless, free-flowing liquid packaged in a single use, 0.5 gram high density polyethylene ampoule with an attachable polypropylene applicator tip. It is applied as a liquid monomer to the bonding surface of the cap/crown to be cemented to a tooth. Upon seating the cap/crown and contact with weak bases LiguiBand® Dental polymerizes as to form a strong adhesive bond between the cap/crown and tooth.
The provided text describes a 510(k) submission for the LiquiBand® Dental adhesive, comparing it to a predicate device, OCTYLDENT® Dental Adhesive. The submission focuses on demonstrating substantial equivalence rather than establishing new acceptance criteria or conducting extensive clinical trials as would be common for novel devices.
Therefore, many of the typical elements of acceptance criteria and study design (like sample sizes for test and training sets, expert qualifications, MRMC studies, or standalone performance) are not detailed because they were not the focus of this type of regulatory submission.
Here's an analysis based on the information provided, with an emphasis on what is present and what is absent:
Acceptance Criteria and Device Performance
The core acceptance criterion for a 510(k) submission is substantial equivalence to a predicate device. This is demonstrated by showing similar intended use, materials, application methods, and performance in nonclinical testing.
| Acceptance Criterion (Implicit for 510(k)) | Reported Device Performance (LiquiBand® Dental adhesive) |
|---|---|
| Intended Use: Must be substantially similar to predicate device. | "Intended for use as a permanent cementation when bonding of crown and caps to teeth is desired." This is stated to be the "same intended use" as the predicate. |
| Material Composition: Must be similar to predicate device. | "n-butyl cyanoacrylate monomer containing small amounts of hydroquinone (a free radical inhibitor) and sulfur dioxide (an anionic inhibitor)." Stated as "similar materials" to the predicate. |
| Application Method: Must be similar to predicate device. | "Applied as a liquid monomer to the bonding surface of the cap/crown to be cemented to a tooth. Upon seating the cap/crown and contact with weak bases LiguiBand® Dental polymerizes..." Stated as "similar means of application" to the predicate. |
| Chemical, Physical, and Biocompatibility Properties: Test results must demonstrate similarity to predicate device. | "LiquiBand Dental adhesive was subjected to chemical, physical and biocompatibility testing... All of the testing supported the substantial equivalence by the similarity of the test results." Specific numerical results or thresholds are not provided, but the conclusion is that they were similar enough. |
| Effectiveness as Dental Cement: Must be effective as a dental cement, comparable to predicate. | "Both products are effective as dental cements." This is a general statement based on the nonclinical testing. |
Study Details for Substantial Equivalence
This is not a study to prove acceptance criteria for a novel device, but rather a submission to prove substantial equivalence to a legally marketed predicate device.
-
Sample sizes used for the test set and the data provenance: Not specified. The "test set" in this context refers to the samples used in the nonclinical chemical, physical, and biocompatibility testing. The document does not provide details on the number of samples tested for each assay or their provenance beyond being part of the manufacturer's testing. It is retrospective testing of the developed device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for a 510(k) nonclinical testing submission. No experts were used to establish "ground truth" in the clinical sense, as no clinical human data is presented or implied for the substantial equivalence claim. The "ground truth" or reference for the comparative nonclinical tests would be the established properties of the predicate device.
-
Adjudication method for the test set: Not applicable. There was no clinical trial or expert review panel described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/medical image analysis device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device (dental adhesive), not an algorithm. The nonclinical tests conducted can be considered "standalone" in that they assess the device's inherent physical, chemical, and biological properties without human intervention during the "performance" phase of the test itself.
-
The type of ground truth used:
- For the nonclinical testing (chemical, physical, biocompatibility): The "ground truth" was implicitly established by comparing the results of LiquiBand® Dental adhesive to the known, established properties and performance of the legally marketed predicate device, OCTYLDENT® Dental Adhesive. The predicate device's existing regulatory status and performance data serve as the benchmark.
- For the intended use, material, and application method comparison: These were directly compared to the stated characteristics of the predicate device.
-
The sample size for the training set: Not applicable. This device is not an AI/machine learning model, so there is no concept of a "training set."
-
How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a sequence of characters, specifically 'K93556'. The characters are written in a bold, rounded font, giving them a distinct and somewhat playful appearance. The black color of the characters contrasts sharply with the white background, making them easily readable. The characters are arranged horizontally, with each character clearly separated from the others.
JAN - 3 2000
510(k) Summary
Submitter
MedLogic Global Corporation 4815 List Drive, Suite 111 Colorado Springs, Colorado 80919
Contact:
Stephen J. Tamsett Manager Regulatory Affairs and Quality Assurance
Phone: 1-800-625-6442 Fax: 1-719-535-2930
Date summary was prepared:
October 19, 1999
Name(s) of the device
LiquiBand® Dental adhesive
Identification of predicate device(s)
The predicate device for LiquiBand® Dental adhesive is OCTYLDENT® Dental Adhesive by Closure Medical Corporation
Description of the device
The LiquiBand® Dental adhesive formula consists of n-butyl cyanoacrylate monomer containing small amounts of hydroquinone (a free radical inhibitor) and sulfur dioxide (an anionic inhibitor).
LiquiBand® Dental, n-butyl cyanoacrylate adhesive, is a clear, colorless, free-flowing liquid packaged in a single use, 0.5 gram high density polyethylene ampoule with an attachable polypropylene applicator tip. It is applied as a liquid monomer to the bonding surface of the cap/crown to be cemented to a tooth. Upon seating the cap/crown and contact with weak bases LiguiBand® Dental polymerizes as to form a strong adhesive bond between the cap/crown and tooth.
Intended Use
LiquiBand® Dental n-butyl cyanoacrylate is intended for use as a permanent cementation when bonding of crown and caps to teeth is desired.
Comparison of device characteristics to predicate
LiquiBand® Dental adhesive is substantially equivalent to OCTYLDENT® Dental Adhesive. This is based on the same intended use, similar materials and similar means of application.
{1}------------------------------------------------
510(k) Notification
Nonclinical testing
LiquiBand Dental adhesive was subjected to chemical, physical and biocompatibility testing to demonstrate its equivalence to OCTYLDENT® Dental Adhesive. All of the testing supported the substantial equivalence by the similarity of the test results. Both products are effective as dental cements. Based on this data, it was concluded that LiquiBand® Dental adhesive is substantially equivalent to OCTYLDENT® Dental Adhesive.
Conclusion
Based on the descriptive information and the performance data provided in this premarket notification, it is concluded that the LiquiBand® Dental adhesive is substantially equivalent to OCTYLDENT® Dental Adhesive by Closure Medical Corporation, a legally marketed dental adhesive.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a symbol consisting of three stylized human figures or profiles, arranged in a way that they appear to be interconnected or overlapping.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 3 2000
Dr. Walter Fabisiak, Director Product Development and Regulatory Affairs and Quality Assurance MedLoqic Global Corporation 4815 List Drive Colorado Springs, Colorado 80919
K993556 Re :
Liquiband® Dental Trade Name: Requlatory Class: II Product Code: EMA December 14, 1999 Dated: Received: December 15, 1999
Dear Mr. Fabisiak:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in
{3}------------------------------------------------
Page 2 - Mr. Fabisiak
the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.ffa.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use: LiquiBand® Dental Adhesive
LiquiBand® Dental n-butyl cyanoacrylate is intended for use as an adhesive when bonding of caps and crowns to teeth is desired.
Prescription Use
(Per 21 CFR 801.109)
Susan Russo
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _
00021
MEDLOGIC GLOBAL CORPORATION - LiquiBand® Dental adhesive
§ 872.3275 Dental cement.
(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.