K Number
K993551

Validate with FDA (Live)

Manufacturer
Date Cleared
2000-01-13

(86 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SOCON® Spinal System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the SOCON® Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Device Description

The SOCON® Spinal System is a multiple component system comprised of a variety of single-use, non-sterile devices that allow the surgeon to build a spinal implant construct in order to provide stabilization and promote spinal fusion. The implants are manufactured from titanium alloy, Ti6A14V, in accordance to ISO 5832/3 and include pedicle bone screws, straight and pre-bent rods, cross bars and self-locking clamps. Specialized instrumentation made from surgical grade stainless steel in accordance to ISO 7153/1 is available for the application and removal of the SOCON implants.

AI/ML Overview

The provided text is a 510(k) summary for the SOCON® Spinal System Expanded Pedicle Screw Indications. It focuses on regulatory clearance for expanded indications based on FDA's final rule for pedicle screw systems and establishes substantial equivalence to a predicate device.

There is no study described in this document that proves the device meets acceptance criteria.

The document states: "There have been no design modifications or technological changes to the current SOCON® Spinal System since receiving FDA clearance under #K970285, and therefore it is substantially equivalent to itself." This explicitly indicates that no new performance testing was conducted for this specific 510(k) submission.

Therefore, I cannot provide information for the following points as they are not present in the provided text:

  1. A table of acceptance criteria and the reported device performance: No acceptance criteria or performance data are provided for this submission. The device is cleared based on substantial equivalence to a previously cleared device and updated indications for use.
  2. Sample size used for the test set and the data provenance: No new test set data was used.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method: Not applicable.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable (this is not an AI device).
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable (this is not an AI device).
  9. How the ground truth for the training set was established: Not applicable (this is not an AI device).

Summary of available information from the document:

This 510(k) submission (K993551) for the SOCON® Spinal System Expanded Pedicle Screw Indications is a regulatory submission for expanding the "Indications for Use" of an already cleared device (K970285). The expansion is based on an FDA final rule for pedicle screw systems. The device is deemed substantially equivalent to its previous clearance (K970285) because there have been no design modifications or technological changes. It is also found substantially equivalent to the VSP System (Pedicle Screw Indications) by DePuy AcroMed, Inc. (K984350) for the purpose of the expanded indications.

The document does not contain any information about new performance studies, acceptance criteria, or testing data for the device.

{0}------------------------------------------------

JAN 1 3 2000

K443551
Page 1 of 2

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

in Accordance with SMDA of 1990

SOCON® SPINAL SYSTEM EXPANDED PEDICLE SCREW INDICATIONS

October 19, 1999

  • Aesculap®, Inc. COMPANY: 1000 Gateway Blvd. So. San Francisco, CA 94080
  • Lia S. Jones, Requlatory Associate CONTACT: 650-624-5073 (phone) 650-589-3007 (fax) lia.iones@aesculap.com (email)
  • SOCON® Spinal System TRADE NAME:
  • COMMON NAME: Pedicle Screw Spinal Fixation System
  • DEVICE CLASS: Class II

PRODUCT CODE: MNH, MNI

  • CLASSIFICATION: 888.3070 Spondylolisthesis Spinal Fixation Device System 888.3070 - Pedicle Screw Spinal System
  • Orthopedic Devices Branch REVIEW PANEL: Division of General and Restorative Devices

DEVICE DESCRIPTION

The SOCON® Spinal System is a multiple component system comprised of a variety of single-use, non-sterile devices that allow the surgeon to build a spinal implant construct in order to provide stabilization and promote spinal fusion. The implants are manufactured from titanium alloy, Ti6A14V, in accordance to ISO 5832/3 and include pedicle bone screws, straight and pre-bent rods, cross bars and self-locking clamps. Specialized instrumentation made from surgical grade stainless steel in accordance to ISO 7153/1 is available for the application and removal of the SOCON implants.

PURPOSE FOR PREMARKET NOTIFICATION

The sole purpose of this 510(k) is to expand the Indications for Use for Aesculap's current SOCON® Spinal System (#K970285) based on the FDA's publication of the final rule for pedicle screw systems (Federal Register, Vol. 63, No. 143) which became effective on August 26, 1998.

)

{1}------------------------------------------------

K493551
Page 2 of 2

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS in Accordance with SMDA of 1990

SOCON® SPINAL SYSTEM EXPANDED PEDICLE SCREW INDICATIONS

October 19, 1999

INTENDED USE

The SOCON® Spinal System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the SOCON® Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

SUBSTANTIAL EQUIVALENCE

There have been no design modifications or technological changes to the current SOCON® Spinal System since receiving FDA clearance under #K970285, and therefore it is substantially equivalent to itself.

For the purpose of this 510(k), however, the expanded Indications for Use are based on the FDA's publication of the final rule for pedicle screw systems (Federal Register. Vol. 63, No. 143) and is substantially equivalent to the recently cleared VSP System (Pedicle Screw Indications) by DePuy AcroMed, Inc. under #K984350.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles three human profiles facing right, with flowing lines beneath them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

GAN 1 3 2000

Ms. Lia S. Jones Regulatory Associate Aesculap . Inc. 1000 Gateway Boulevard South San Francisco, California 94080-7028

Re: K993551

Trade Name: SOCON® Spinal System Regulatory Class: II Product Code: MNH and MNI Dated: October 19, 1999 Received: October 20, 1999

Dear Ms. Jones:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 - Ms. Lia S. Jones

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Arnell J. Ayer

James E. Dillard III So Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

K493-551

Device Name:

SOCON® Spinal System Expanded Pedicle Screw Indications

Indication for Use:

. "

The SOCON® Spinal System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the SOCON® Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Thunell Taylor 8-720

Division Sign-Off) . vision of General Restorative Devices 1593551 510(k) Number.

Prescription UseX
or Over-the-Counter Use
(per 21 CFR 801.109)

(Optional Format 3-10-98)

N/A