K Number
K992587
Date Cleared
1999-12-22

(142 days)

Product Code
Regulation Number
866.3235
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for Epstein-Barr Early Antigen (EBV-EA) IgG Antibody is intended for the qualitative and semi-quantitative detection of IgG (Immunoglobulin G) antibody to the early antigen complex of the Epstein-Barr Virus in human serum. The test detects IgG antibodies to both the diffuse (D) and restricted (R) components of the EBV early antigen complex. The EBV-EA IFA Test Kit should be used in combination with other Epstein-Barr serologies (Viral Capsid Antigen (VCA) IgG and IgM, Epstein-Barr Nuclear Antigen (EBNA), and heterophile antibody) as an aid in the diagnosis of infectious mononucleosis (IM).

Device Description

The Epstein-Barr Virus Early Antigen (EBV-EA) IgG IFA is an Indirect Fluorescence Assay (IFA) for the detection of IgG antibodies to EBV-EA in human serum.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for Epstein-Barr Virus Early Antigen (EBV-EA) IgG Antibody, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria are not explicitly stated as numerical thresholds for sensitivity and specificity in the provided text. Instead, the study reports the device's performance relative to a commercially available predicate EBV-EA IFA. The implication is that comparable performance to the predicate device would be a key acceptance criterion.

Table 1: Relative Sensitivity and Specificity of the Stellar EBV-EA IFA Kit Relative to Alternate IFA

MetricAcceptance Criteria (Implied: Comparable to Alternate IFA)Reported Device Performance (Stellar EBV-EA IFA)
Relative SensitivityNot explicitly stated, likely expected to be high42/43 = 97.7% (with the alternate IFA as reference)
Relative SpecificityNot explicitly stated, likely expected to be high26/32 = 81.3% (with the alternate IFA as reference)
Overall AgreementNot explicitly stated, likely expected to be high68/75 = 90.7% (95% CI: 81.7% - 96.2%)
Titer Agreement (Identical)Not explicitly stated, likely expected to be high9/25 = 36%
Titer Agreement (+/- one two-fold dilution)Not explicitly stated, likely expected to be high12/25 = 48%
Titer Agreement (+/- two two-fold dilutions)Not explicitly stated, likely expected to be high4/25 = 16%
Reproducibility (Identical End Point Titer)Not explicitly stated, likely expected to be high131/180 = 72.8%
Reproducibility (Within +/- one two-fold dilution)Not explicitly stated, likely expected to be high49/180 = 27.2%
Specificity (Cross-reactivity)No cross-reactivity with CMV, VZV, HSV 1, HSV 2, HHV 6, HHV 8Found to be negative for EBV-EA IFA in these cases

Note on Acceptance Criteria: The document explicitly states: "Please be advised that 'relative' refers to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease." This highlights that the acceptance criteria are based on agreement with a predicate device, not on direct clinical truth or diagnostic accuracy.

Study Details

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size for Relative Sensitivity and Specificity: 75 samples (43 positive, 32 negative by the alternate IFA).
  • Test Set Sample Size for Titer Agreement: 25 positive sera.
  • Test Set Sample Size for Reproducibility: 180 (3 positive sera with various titers, 1 negative sera, each assayed 5 times on 3 different assays at 3 different sites).
  • Test Set Sample Size for Specificity (Cross-reactivity): At least 9 samples (CMV+, VZV+, HSV 1&2+, HHV 6+, HHV 8+; note multiple HSV 1&2+ and HHV 6+ samples are listed).
  • Data Provenance: The samples were selected from frozen retrospective sera tested at a commercial clinical laboratory. The document doesn't explicitly state the country of origin, but given the applicant's address (Columbia, MD 21046, USA), it's highly probable the data is from the USA.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for the primary relative sensitivity and specificity study was established by the results of a commercially available EBV-EA IFA (the "Alternate IFA"). There is no mention of human experts directly establishing the ground truth for this comparison.

For the categorization of sera by serological status (Acute, Seropositive, Seronegative), the criteria were based on ELISA results for VCA IgM/IgG and EBNA. Again, this points to laboratory assay results, not human expert consensus, as the ground truth.

4. Adjudication method for the test set

No adjudication method involving multiple human readers is mentioned. The ground truth was determined by the results of established laboratory assays (the Alternate IFA and various EBV ELISAs).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or reported. This device is a diagnostic assay (IFA), not an AI-assisted interpretation tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the performance data presented is for the standalone device (the Stellar EBV-EA IFA kit). The "human-in-the-loop" equivalent for an IFA assay would typically be a lab technician reading the fluorescence under a microscope. The study design tests the performance of the kit itself in producing results compared to other similar kits.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used was primarily the results from a commercially available predicate EBV-EA IFA kit or other serological assays (VCA IgG/IgM ELISA, EBNA ELISA, heterophile antibody assay), rather than expert consensus, pathology, or outcomes data. The study explicitly states there was no attempt to correlate assay results with disease presence or absence.

8. The sample size for the training set

The document does not provide information on a specific training set or its size. As this is a traditional in-vitro diagnostic (IVD) kit based on an established IFA methodology, it's unlikely to have a "training set" in the machine learning sense. The assay development would involve optimizing reagents and protocols during its design phase, but not typically a labeled dataset for algorithm training.

9. How the ground truth for the training set was established

Since no training set for an algorithm is mentioned, the method for establishing its ground truth is not applicable/not provided.

{0}------------------------------------------------

DEC 2 2 1999

K992587

510k Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

Applicant Information:

Date Prepared:July 20, 1999
Name:Stellar Bio Systems, Inc
Address:9075 Guilford RoadColumbia, MD 21046
Contact Person:John Brewer
Contact Person:John Brewer
PhoneNumber.410-381-8550
Fax Number.410-381-8984

Device Information:

Trade Name:Indirect Fluorescence Assay for Epstein-Barr Virus Early Antigen (EBV-EA) IgGAntibody
Common Name:EBV-EA IgG IFA Test
Classification Name:EBV-EA IgG Serological Reagent

Equivalent Device: IFA

Device Description: The Epstein-Barr Virus Early Antigen (EBV-EA) IgG IFA is an Indirect Fluorescence Assay (IFA) for the detection of IgG antibodies to EBV-EA in human serum.

Intended Use: The Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for Epstein-Barr Virus Early Antigen (EBV-EA) IgG Antibody is intended for the qualitative and semi-quantitative detection of IgG (Immunoglobulin G) antibody to the early antigen complex of the Epstein-Barr Virus in human serum. The test detects IgG antibodies to EBV antigens in chemically induced Raji cells. The EBV-EA IFA Test Kit should be used in combination with other Epstein-Barr serologies [Viral Capsid Antigen (VCA) IgG and IgM, Epstein-Barr Nuclear Antigen (EBNA), and heterophile antibody] as an aid in the diagnosis of infectious mononucleosis (IM).

Principle of Procedure:

Stellar Bio Systems' fluorescent antibody assays use the indirect method of antibody detection and titer determination. Diluted patient serum sampled to fixed antigens provided on paint delineated wells on glass microscope slides. During a 30 minute incubation, antibody specific for EA forms an antigen/antibody complex with the fixed antigens. In a brief washing step, nonspecific antibody and other unreacted serum proteins are eliminated. Fluorescein-conjugated goat anti-human IgG is then applied to the glass slide. The anti-IgG conjugate combines with human IgG, if present, during a 30 minute incubation. After a brief wash to remove unreacted conjugate, the slides are viewed by fluorescence microscopy. A positive antibody reaction is denoted by bright green fluorescence at the antigen sites.

{1}------------------------------------------------

Performance Characteristics

    1. Relative Sensitivity and Specificity The Stellar EBV-EA IFA kit was evaluated relative to a commercially available EBV-EA IFA. The samples were selected frozen retrospective sera tested at a commercial clinical laboratory. The acute sera were selected by being positive in a heterophile assay. The seronegative sera were selected by being negative for VCA G and EBNA G by ELISA. The seropositive sera were selected by being positive for VCA G and EBNA G by ELISA.. The data in Table 1 summarizes the data.

Table 1 Relative Sensitivity and Specificity of the Stellar EBV-EA IFA Kit Relative to Alternate IFA

Stellar EBV-EA IFA
PositiveNegativeTotal
Positive42143
AlternateIFANegative6*2632
Total482775

95% Confidence Interval Overall Agreement = 68/75 = 90.7% 81.7% - 96.2%

  • The 6 Stellar positive/Alternate IFA negative samples were acute sera determined by ELISA VCA IgM.

Please be advised that 'relative' refers to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease.

Since the above studies were performed on a pre-selected retrospective population no calculations for the assay's positive and negative predictive value may be done or inferred.

{2}------------------------------------------------

The above sera were categorized relative to their serological status to EBV. The following table illustrates the results:

AcuteVCA IgM (+)EBNA (-)SeropositiveVCA IgG (+)EBNA (+) VCA IgM (-)SeronegativeVCA IgG (-)EBNA (-)
EBV EAIgG IFA
Positive20253
Negative10710
Total303213
    1. Titer Agreement: 25 positive sera were serially two-fold diluted and the endpoint titer was determined on the Stellar EBV-EA IFA and a commercially available EBV-EA IFA. The endpoint titer results are as follows.
Identical Titer9/25
+ one, two-fold dilution-12/25
+ two, two-fold dilutions------------------------------------------------------------------------------------------------------------------------------------------------------------------------------4/25
    1. Reproducibility: Three positive sera with various titers (1:80,1:160, 1:640) and one negative sera were serially diluted and assayed five times each on three different assays at three different sites. 131/180 of the end point titers were identical. 49/180 of the end point titers was within + one, two-fold dilution.

4. Specificity

The following sera that were positive for IgG antibodies to CMV, VZV, HSV 1, HSV 2, HHV 6 and HHV 8 by IFA were found to be negative for EBV-EA IFA, indicating a lack of cross-reactivity to these viruses.

SampleEA IFA Result
CMV +-
VZV +-
HSV 1&2 +-
HSV 1&2 +-
HSV 1&2 +-
HSV 1&2 +-
HHV 6+-
HHV 6+-
HHV 8+-

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 22 1999

Mr. John Brewer President Stellar Bio Systems, Inc. 9075 Guilford Road Columbia, Maryland 21046

Re: K992587

Trade Name: Indirect Fluorescence Assay (IFA) for Epstein-Barr Virus Early Antigen (EBV-EA) IgG Antibody Regulatory Class: I

:"... Product Code: LSE Dated: October 29, 1999 Received: October 29, 1999

Dear Mr. Brewer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K992587

Device Name: Indirect Fluorescence Assay (IFA) for Epstein-Barr Virus Early Antigen (EBV-EA) IgG Antibody

Indications For Use: The Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for Epstein-Barr Early Antigen (EBV-EA) IgG Antibody is intended for the qualitative and semi-quantitative detection of IgG (Immunoglobulin G) antibody to the early antigen complex of the Epstein-Barr Virus in human serum. The test detects IgG antibodies to both the diffuse (D) and restricted (R) components of the EBV early antigen complex. The EBV-EA IFA Test Kit should be used in combination with other Epstein-Barr serologies (Viral Capsid Antigen (VCA) IgG and IgM, Epstein-Barr Nuclear Antigen (EBNA), and heterophile antibody) as an aid in the diagnosis of infectious mononucleosis (IM).

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Duboe

(Division Sign Off)
Division of Clinical Laboratory Devices
510(k) Number K992587

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 866.3235 Epstein-Barr virus serological reagents.

(a)
Identification. Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).(b)
Classification. Class I (general controls).