(14 days)
DiaSorin EBV EA (D) IgG Clin-ELISA™, Gull Laboratories EBV EA IgG IFA
Gull Laboratories EBV EA IgG IFA
No
The device description and performance studies focus on standard ELISA technology and statistical analysis of results, with no mention of AI or ML algorithms.
No.
This device is an in vitro diagnostic (IVD) tool designed to detect antibodies to the EBV Early Antigen Diffuse, which aids in diagnosing EBV associated infectious mononucleosis. It does not directly treat or prevent a disease or condition.
Yes
The device uses ELISA technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse, and the presence of these antibodies is used as an aid in the diagnosis of EBV-associated infectious mononucleosis. This directly indicates its role in contributing to a medical diagnosis.
No
The device description clearly outlines a physical kit utilizing ELISA technology with reagents, microtiter wells, and a photometer for detection, indicating it is a hardware-based diagnostic assay, not software only.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states that the kit is for the "qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse" in "human serum." It also states that the results are used as an "aid in the diagnosis of EBV associated infectious mononucleosis." This clearly indicates the device is intended for use on biological specimens from humans to provide information for diagnostic purposes.
- Device Description: The description details an ELISA method, which is a common technique used in in vitro diagnostic testing to detect the presence of specific substances (in this case, antibodies) in a sample.
- Clinical Laboratory Setting: The intended user is a "clinical laboratory," which is where in vitro diagnostic tests are typically performed.
- Performance Studies: The document describes clinical trials and performance studies conducted to evaluate the device's performance in detecting the target antibodies in human samples from different patient populations and disease states. This is a standard requirement for IVD devices to demonstrate their accuracy and reliability.
The information provided aligns perfectly with the definition and characteristics of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The DiaSorin ETI-EA-G kit uses Enzyme Linked Immunosorbent Assay (ELISA) technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse. The assay is designed for human serum. The presence of EA antibodies is used as an aid in the diagnosis of EBV associated infectious mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion or significant change in antibody titer as evidence of recent infection. Both specimens should be tested simultaneously.
Product codes
LSE
Device Description
The method for the determination of specific anti-EA (D) IgG utilizes the enzyme-linked immunosorbent assay (ELISA) technique. Polystyrene microtiter wells are coated with recombinant EA (D) antigen. Diluted patient serum is incubated with the purified antigen bound to the solid surface of a microtiter well. The EA (D) IgG antibodies present in a patient's serum will be captured by the solid phase. After washing, affinity purified polyclonal goat anti-human IgG (Fc) antibodies coniugated to horseradish peroxidase are added to the well. After this incubation, chromogen containing tetramethylbenzidine is added. Enzyme action on the chromogen results in a color reaction. The color can be detected with a photometer at a wavelength of 450 nm. The measured enzyme activity is directly proportional to the concentration of specific anti-EA (D) IgG bound to the solid phase.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
pediatric, adult
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Clinical trials were conducted at 1 clinical laboratory to evaluate the performance of ETI-EA-G in detecting IgG antibodies to EBV EA (D). Assay performance was compared to the Gull Laboratories, Inc. EA IgG IFA test. Patients from the disease states defined below were tested. The screening population is a group of samples from patients suspected of disease. The transplant recipients were patients who had received, or were awaiting, solid organ or bone marrow transplants. Patient samples excluded if they failed to fit a recognized EBV marker pattern determined by EBV VCA (IgG, IgM) and EBNA testing.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Comparative Clinical Trials:
- Adult Population; Sensitivity: 45/46 = 97.8% (95% CI: 88.5 - 99.9%); Agreement: 45/46 = 97.8%; Prevalence ELISA: 45/46 = 97.8%; Prevalence IFA: 46/46 = 100%
- Pediatric Population; Sensitivity: 26/29 = 89.7% (95% CI: 72.6 – 97.8%); Agreement: 26/29 = 89.7%; Prevalence ELISA: 26/29 = 89.7%; Prevalence IFA: 29/29 = 100%
- Seronegative; Adult Population; Specificity: 7/7 = 100.0% (95% CI: 59.0 - 100.0%); Agreement: 7/7 = 100.0%; Prevalence ELISA: 0/7 = 0.0%; Prevalence IFA: 0/7 = 0.0%
- Seronegative; Pediatric Population; Specificity: 48/50 = 96.0% (95% CI: 86.3 - 99.5%); Agreement: 48/50 = 96.0%; Prevalence ELISA: 2/50 = 4.0%; Prevalence IFA: 0/50 = 0.0%
- Reactivated; Adult Population; Sensitivity: 30/31 = 96.8% (95% CI: 83.3-99.9%); Agreement: 30/31 = 96.8%; Prevalence ELISA: 30/31 = 96.8%; Prevalence IFA: 31/31 = 100%
- Reactivated; Pediatric Population; Sensitivity: 4/4 = 100.0% (95% CI: 39.8 - 100.0%); Agreement: 4/4 = 100.0%; Prevalence ELISA: 4/4 = 100.0%; Prevalence IFA: 4/4 = 100.0%
- Past Infection; Adult Population; Sensitivity: 12/33 = 36.4% (95% CI: 20.4 - 54.9%); Specificity: 3/4= 75.0% (95% CI: 19.4-99.4); Agreement: 15/37 = 40.5%; Prevalence ELISA: 13/37 = 32.4%; Prevalence IFA: 33/37 = 89.2%
- Past Infection; Pediatric Population; Sensitivity: 2/3 = 66.7% (95% CI: 9.0 - 99.9); Agreement: 2/3 = 66.7%; Prevalence ELISA: 2/3 = 66.7%; Prevalence IFA: 3/3 = 100.0%
- Transplant Recipient patients: Adult Population; Sensitivity: 10/21 = 47.6% (95% CI: 25.7 - 70.2%); Specificity: 1/3 = 33.3% (95% CI: 0.8% - 90.6%); Agreement: 11/24 = 45.8%; Prevalence ELISA: 12/24 = 50.0%; Prevalence IFA: 21/24 = 87.5%
- Transplant Recipient patients; Pediatric Population; Specificity: 6/7 = 85.7% (95% CI: 42.1 - 99.6%); Agreement: 6/8 = 75.0%; Prevalence ELISA: 1/8 = 12.5%; Prevalence IFA: 1/8 = 12.5%
The ETI-EA-G ELISA demonstrates good sensitivity and specificity in Primary. Seronegative and Reactivated populations. The prevalence rates were within the expected range for each of these populations.
Reproducibility: Reproducibility studies were performed at 3 sites using one lot of ETI-EA-G reagents. Assay reproducibility was determined using 6 samples that spanned the range of the assay. Samples were tested in triplicate once a day for 3 days.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
- Sensitivity
- Specificity
- Agreement
- Prevalence ELISA
- Prevalence IFA
- Within-run %CV
- Between day %CV
- Between site %CV
- Total %CV
Predicate Device(s)
DiaSorin EBV EA (D) IgG Clin-ELISA™, Gull Laboratories EBV EA IgG IFA
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 866.3235 Epstein-Barr virus serological reagents.
(a)
Identification. Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).(b)
Classification. Class I (general controls).
0
JUL 12 1560
K992191
510 (k) SUMMARY
SUBMITTED BY:
Judith J. Smith DiaSorin, Inc. 9175 Guilford Rd. Suite 100 Columbia, MD 21046
NAME OF DEVICES: Trade Name:
DiaSorin ETI-EA-G Epstein-Barr Viral Early Antigen Diffuse EA (D) IaG ELISA
Immunoassay for the detection of IgG Common Names/Descriptions: antibodies to the Epstein-Barr Viral (EBV) Early Antigen Diffuse antigen
Classification Names:
EBV Serology Test
PREDICATE DEVICES:
DiaSorin EBV EA (D) IgG Clin-ELISA™ Gull Laboratories EBV EA IgG IFA
DEVICE DESCRIPTION:
INTENDED USE: The DiaSorin ETI-EA-G kit uses Enzyme Linked Immunosorbent Assay (ELISA) technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse. The assay is designed for human serum. The presence of EA antibodies is used as an aid in the diagnosis of EBV associated infectious mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion or significant change in antibody titer as evidence of recent infection. Both specimens should be tested simultaneously.
KIT DESCRIPTION: The method for the determination of specific anti-EA (D) IgG utilizes the enzyme-linked immunosorbent assay (ELISA) technique. Polystyrene microtiter wells are coated with recombinant EA (D) antigen. Diluted patient serum is incubated with the purified antigen bound to the solid surface of a microtiter well. The EA (D) IgG antibodies present in a patient's serum will be captured by the solid phase. After washing, affinity purified polyclonal goat anti-human IgG (Fc) antibodies coniugated to horseradish peroxidase are added to the well. After this incubation, chromogen containing tetramethylbenzidine is added. Enzyme action on the chromogen results in a color reaction. The color can be detected with a photometer at a wavelength of 450 nm. The measured enzyme activity is directly proportional to the concentration of specific anti-EA (D) IgG bound to the solid phase.
1
PERFORMANCE DATA:
Comparative Clinical Trials Clinical trials were conducted at 1 clinical laboratory to evaluate the performance of ETI-EA-G in detecting IgG antibodies to EBV EA (D). Assay performance was compared to the Gull Laboratories, Inc. EA IgG IFA test. Patients from the disease states defined below were tested. The screening population is a group of samples from patients suspected of disease. The transplant recipients were patients who had received, or were awaiting, solid organ or bone marrow transplants. Patient samples excluded if they failed to fit a recognized EBV marker pattern determined by EBV VCA (IgG, IgM) and EBNA testing. Results are shown below.
Expected Pattern | + |
---|---|
Sensitivity | 45/46 = 97.8% |
(95% CI) | (88.5 - 99.9%) |
Specificity | N/A |
(95% CI) | |
Agreement | 45/46 = 97.8% |
Prevalence ELISA | 45/46 = 97.8% |
Prevalence IFA | 46/46 = 100% |
Expected Prevalence | 80-100% |
Primary Disease State: Adult Population
Primary Disease State; Pediatric Population
Expected Pattern | + |
---|---|
Sensitivity | 26/29 = 89.7% |
(95% CI) | (72.6 – 97.8%) |
Specificity | N/A |
(95% CI) | |
Agreement | 26/29 = 89.7% |
Prevalence ELISA | 26/29 = 89.7% |
Prevalence IFA | 29/29 = 100% |
Expected Prevalence | 80-100% |
2
Seronegative; Adult Population
Expected Pattern | - |
---|---|
Sensitivity | N/A |
(95% CI) | |
Specificity | 7/7 = 100.0% |
(95% CI) | (59.0 - 100.0%) |
Agreement | 7/7 = 100.0% |
Prevalence ELISA | 0/7 = 0.0% |
Prevalence IFA | 0/7 = 0.0% |
Expected Prevalence | 0% |
:
:
Seronegative; Pediatric Population
Expected Pattern | - |
---|---|
Sensitivity | N/A |
(95% CI) | |
Specificity | 48/50 = 96.0% |
(95% CI) | (86.3 - 99.5%) |
Agreement | 48/50 = 96.0% |
Prevalence ELISA | 2/50 = 4.0% |
Prevalence IFA | 0/50 = 0.0% |
Expected Prevalence | 0% |
Reactivated; Adult Population
Expected Pattern | + |
---|---|
Sensitivity | 30/31 = 96.8% |
(95% CI) | (83.3-99.9%) |
Specificity | N/A |
(95% Cl) | N/A |
Agreement | 30/31 = 96.8% |
Prevalence ELISA | 30/31 = 96.8% |
Prevalence IFA | 31/31 = 100% |
Expected Prevalence | 90-100% |
3
Reactivated; Pediatric Population
Expected Pattern | + |
---|---|
Sensitivity | 4/4 = 100.0% |
(95% CI) | (39.8 - 100.0%) |
Specificity | N/A |
(95% CI) | |
Agreement | 4/4 = 100.0% |
Prevalence ELISA | 4/4 = 100.0% |
Prevalence IFA | 4/4 = 100.0% |
Expected Prevalence | 90-100% |
. .
Past Infection; Adult Population
:
Expected Pattern | -/+ |
---|---|
Sensitivity | 12/33 = 36.4% |
(95% CI) | (20.4 - 54.9%) |
Specificity | 3/4= 75.0% |
(95% CI) | (19.4-99.4) |
Agreement | 15/37 = 40.5% |
Prevalence ELISA | 13/37 = 32.4% |
Prevalence IFA | 33/37 = 89.2% |
Expected Prevalence | 10-40% |
Past Infection; Pediatric Population
Expected Pattern | -/+ |
---|---|
Sensitivity | 2/3 = 66.7% |
(95% CI) | (9.0 - 99.9) |
Specificity | N/A |
(95% CI) | |
Agreement | 2/3 = 66.7% |
Prevalence ELISA | 2/3 = 66.7% |
Prevalence IFA | 3/3 = 100.0% |
Expected Prevalence | 10-40% |
4
ELISA Result | EA (D) IgG |
---|---|
Sensitivity | 10/21 = 47.6% |
(95% CI) | (25.7 - 70.2%) |
Specificity | 1/3 = 33.3% |
(95% CI) | (0.8% - 90.6%) |
Agreement | 11/24 = 45.8% |
Prevalence ELISA | 12/24 = 50.0% |
Prevalence IFA | 21/24 = 87.5% |
Transplant Recipient patients: Adult Population
Transplant Recipient patients; Pediatric Population
ELISA Result | EA (D) IgG |
---|---|
Sensitivity | |
(95% CI) | N/A |
Specificity | |
(95% CI) | 6/7 = 85.7% |
(42.1 - 99.6%) | |
Agreement | 6/8 = 75.0% |
Prevalence ELISA | 1/8 = 12.5% |
Prevalence IFA | 1/8 = 12.5% |
The ETI-EA-G ELISA demonstrates good sensitivity and specificity in Primary. Seronegative and Reactivated populations, The prevalence rates were within the expected range for each of these populations.
In the Past Infection population the prevalence rate for ETI-EA-G was consistent with published rates of 10-40%, while IFA rates were 90%. This appears to reflect the fact that the IFA detects both the Restricted and the Diffuse EA antigen, while ETI-EA-G detects only the Diffuse antigen. The Restricted antigen is only occasionally seen in acute IM cases, and then only late in the acute phase. Following IM, however, the Restricted antigen persists in a substantial number of patients for at least 10-104 months. In light of this, it is predictable that the Gull IFA would detect Early Antigen at a much higher rate in a Past Infection population than would the ETI-EA-G ELISA. The same situation appears to account for the results seen in the Transplant Recipients who included a high percentage of samples with EBV markers consistent with Past Infection. The results support the Transplant claim in the Intended Use.
Within each of the defined disease states, results were similar across the adult and pediatric populations, supporting the separate adult and pediatric claims in the Intended Use for ETI-EA-G.
Reproducibility: Reproducibility studies were performed at 3 sites using one lot of ETI-EA-G reagents. Assay reproducibility was determined using 6 samples that spanned
5
the range of the assay. Samples were tested in triplicate once a day for 3 days. Combined results are summarized below.
Within-run | Between day | Between site | Total | ||
---|---|---|---|---|---|
Sample | Mean (AU) | %CV | %CV | %CV | %CV |
Negative | 7.6 | 12.14 | 22.02 | 14.77 | 21.63 |
Negative | 9.2 | 14.24 | 24.19 | 20.01 | 19.61 |
low pos | 29.4 | 7.57 | 4.97 | 11.37 | 8.18 |
low pos | 26.0 | 13.47 | 12.32 | 9.59 | 15.88 |
mid pos | 74.5 | 9.71 | 2.68 | 3.36 | 10.12 |
high pos | 122.4 | 6.74 | 3.06 | 0.98 | 6.64 |
Reproducibility for ETI-EA-G based on Arbitrary Units (AU)
�
6
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1.2 1999
DiaSorin, Inc. c/o Ms. Carole Stamp TÜV Product Service, Inc. 1775 Old Highway 8 NW Suite 104 New Brighton, MN 55112-1891
Re: K992191 Trade Name: DiaSorin ETI-EA-G Regulatory Class: I Product Code: LSE Dated: June 25, 1999 Received: June 28, 1999
Dear Ms. Stamp:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
7
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
8
INDICATIONS FOR USE
510(k) Number (if known): Not known
Device Name: DiaSorin ETI-EA-G
Indications For Use:
The DiaSorin ETI-EA-G kit uses Enzyme Linked Immunosorbent Assay (ELISA) technology for the qualitative and/or semi-quantitative detection of IgG antibodies to the EBV Early Antigen Diffuse. The assay is designed for human serum. The presence of EA antibodies is used as an aid in the diagnosis of EBV associated infectious mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. When evaluating properly paired sera, the results of these assays are used to demonstrate seroconversion or significant change in antibody titer as evidence of recent infection. Both specimens should be tested simultaneously.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubois
ical I aboratory Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)