K Number
K991688
Device Name
WRIST TYPE DIGITAL BLOOD PRESSURE MONITOR, MODEL HD-303S
Date Cleared
1999-06-01

(15 days)

Product Code
Regulation Number
870.1130
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Noninvasive measurement of systolic and diastolic blood pressure and heart rate in adult patients, i.e., age 18 and above, in a home care environment.
Device Description
The Jawon Model HD-303S is a compact, automatic sphygmomanometer intended for measurement of blood pressure at the wrist. The method of operation is the oscillometric method. The control unit and the cuff are physically integrated into a single wrist-mounted unit. The system is microprocessor controlled and includes an air pump; fuzzy logic to regulate inflation, deflation, and measurement operations; circuitry to detect and process minute pressure oscillations; a six-digit LCD display of systolic and diastolic pressure readings and heart rate; a memory function that stores the previous eight measurement results; and two pushbutton controls.
More Information

Not Found

Yes
The device description explicitly mentions the use of "fuzzy logic" to regulate inflation, deflation, and measurement operations, which is a form of AI.

No
The device is described as a noninvasive measurement device for blood pressure and heart rate, which are diagnostic or monitoring functions, not therapeutic ones.

Yes

The device measures blood pressure and heart rate, providing information that can be used to monitor and potentially diagnose cardiovascular conditions.

No

The device description explicitly states it is a "compact, automatic sphygmomanometer" with integrated hardware components like an air pump, circuitry, LCD display, and pushbutton controls. It is a physical device, not solely software.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD devices are used to examine specimens derived from the human body (like blood, urine, or tissue) to provide information for diagnosis, monitoring, or screening.
  • This device measures blood pressure and heart rate directly from the patient's wrist. It does not analyze any biological samples.

The description clearly indicates it's a noninvasive device for measuring physiological parameters directly from the body, which falls under the category of a medical device, but not specifically an IVD.

No
The letter does not mention any Predetermined Change Control Plan (PCCP) or that the FDA has reviewed and approved or cleared such a plan for this device.

Intended Use / Indications for Use

The Model HD-303S system is indicated for the noninvasive measurement of systolic and diastolic blood pressure and determination of heart rate in adult patients, age 18 and above. Because the device is recommended for use in a home care environment, use should be limited to patients capable of understanding written and/or oral directions.

Product codes (comma separated list FDA assigned to the subject device)

DXN

Device Description

General: The Jawon Model HD-303S is a compact, automatic sphygmomanometer intended for measurement of blood pressure at the wrist. The method of operation is the oscillometric method. The control unit and the cuff are physically integrated into a single wrist-mounted unit. The system is microprocessor controlled and includes an air pump; fuzzy logic to regulate inflation, deflation, and measurement operations; circuitry to detect and process minute pressure oscillations; a six-digit LCD display of systolic and diastolic pressure readings and heart rate; a memory function that stores the previous eight measurement results; and two pushbutton controls.

Operation: If occlusion of the systolic pulse is not achieved by initial pressurization, cuff pressure is automatically increased in 30 mmHg increments until a proper systolic measurement can be obtained. The device employs a pressure measurement algorithm designed to detect, filter, process, and store pressure readings. An error message is presented whenever results fail to satisfy preprogrammed accuracy criteria. If cuff pressure starts to exceed 320 mmHg, a high-speed exhaust valve is opened automatically.

Power: The Model HD-303S is powered by two AAA-size batteries. Power is shut down automatically if the unit remains idle for a period of approximately two minutes.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

fuzzy logic to regulate inflation, deflation, and measurement operations

Input Imaging Modality

Not Found

Anatomical Site

wrist

Indicated Patient Age Range

age 18 and above

Intended User / Care Setting

patients capable of understanding written and/or oral directions / home care environment.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The measurement performance of the Jawon system has been evaluated in clinical studies conducted in accordance with ANSI/AAMI Standard SP10-1992 and found to comply fully with the accuracy criteria established in the standard. Safety testing including electrical characteristics of the system and components, life testing over 10,005 operational cycles, intra-device variability, environmental integrity under various operating and storage conditions including high and low altitude extremes, and resistance to vibration and shock has been conducted with satisfactory results. Similarly, electromagnetic compatibility and FCC Part 15 compliance studies have been conducted by ONETECH Testing & Evaluation Laboratories, and the device was found to comply with all relevant standards. Software verification and validation have been performed. It is concluded that the subject device complies with all applicable safety and performance standards.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K952494

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).

0

510(k) SUMMARY

MODEL HD-303S NONINVASIVE BLOOD PRESSURE MEASUREMENT SYSTEM

    1. COMPANY INFORMATION. Name: Jawon Medical Co., Ltd. Address: 7F Jeong Ju Bldg., #1451-38, Seocho-Dong, Seocho-Ku, Seoul 137-070, Korea (011) 82-2-587-4056 Contact: Mr. J. N. Kim, Manager Phone: -
    1. DEVICE IDENTIFICATION. Trade Name: Model HD-303S Wrist Type Digital Blood Pressure Monitor Common Name and Classification Name: Noninvasive Blood Pressure Measurement System, 74 DXN
    1. PREDICATE DEVICE. Model WS-200 Automatic Digital Electronic Wrist Blood Pressure Monitor, Nihon Seimitsu Sokki Co., Ltd. - K952494, SE decision 12/28/95.
    1. DEVICE DESCRIPTION. General: The Jawon Model HD-303S is a compact, automatic sphygmomanometer intended for measurement of blood pressure at the wrist. The method of operation is the oscillometric method. The control unit and the cuff are physically integrated into a single wrist-mounted unit. The system is microprocessor controlled and includes an air pump; fuzzy logic to regulate inflation, deflation, and measurement operations; circuitry to detect and process minute pressure oscillations; a six-digit LCD display of systolic and diastolic pressure readings and heart rate; a memory function that stores the previous eight measurement results; and two pushbutton controls.

Operation: If occlusion of the systolic pulse is not achieved by initial pressurization, cuff pressure is automatically increased in 30 mmHg increments until a proper systolic measurement can be obtained. The device employs a pressure measurement algorithm designed to detect, filter, process, and store pressure readings. An error message is presented whenever results fail to satisfy preprogrammed accuracy criteria. If cuff pressure starts to exceed 320 mmHg, a high-speed exhaust valve is opened automatically.

Power: The Model HD-303S is powered by two AAA-size batteries. Power is shut down automatically if the unit remains idle for a period of approximately two minutes.

    1. INTENDED USES. The Model HD-303S system is indicated for the noninvasive measurement of systolic and diastolic blood pressure and determination of heart rate in adult patients, age 18 and above. Because the device is recommended for use in a home care environment, use should be limited to patients capable of understanding written and/or oral directions.
      00018

1

    1. COMPARISON WITH PREDICATE DEVICE. The Jawon device has been compared with the Nihon Seimitsu Sokki Co. Model WS-200 Wrist Blood Pressure Monitor. The intended use of the two systems is the same. The principle of operation (oscillometric measurement) and many operating features are identical. The principal differences are that the control unit and cuff are an integral unit in the subject device but separate, requiring that the cuff be connected by the operator in the predicate device, and that operation is controlled by fuzzy logic in the subject device but by conventional automatic methods in the predicate device. It is concluded that there are no technologic differences between the subject and predicate devices that raise new questions concerning either safety or effectiveness.
    1. PERFORMANCE DATA. The measurement performance of the Jawon system has been evaluated in clinical studies conducted in accordance with ANSI/AAMI Standard SP10-1992 and found to comply fully with the accuracy criteria established in the standard. Safety testing including electrical characteristics of the system and components, life testing over 10,005 operational cycles, intra-device variability, environmental integrity under various operating and storage conditions including high and low altitude extremes, and resistance to vibration and shock has been conducted with satisfactory results. Similarly, electromagnetic compatibility and FCC Part 15 compliance studies have been conducted by ONETECH Testing & Evaluation Laboratories, and the device was found to comply with all relevant standards. Software verification and validation have been performed. It is concluded that the subject device complies with all applicable safety and performance standards.

2

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract image of an eagle or bird-like figure, with three stylized wing or feather shapes.

JUN 1 1991 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Jawon Medical Co., Ltd. c/o Ms. Carole Stamp Responsible Third Party Official 510(k) Program Manager TÜV Product Service Inc. 1775 Old Highway 8 NW, Suite 104 New Brighton, MN 55112-1891

Re: K991688 Wrist Type Digital Blood Pressure Monitor, Model HD-303S Regulatory Class: II (Two) Product Code: DXN Dated: May 14, 1999 Received: May 17, 1999

Dear Ms. Stamp:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

3

Page 2 - Ms. Carole Stamp

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Noninvasive Blood Pressure Measurement System Model HD-303S Wrist Type Digital Blood Pressure Monitor

Indications For Use:

Noninvasive measurement of systolic and diastolic blood pressure and heart rate in adult patients, i.e., age 18 and above, in a home care environment.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Otto K. Langurci

and Neurologica 510(k) Numbe

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use L

00023