(238 days)
Meniett 20 is indicated for symptomatic treatment of Menière's disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops.
Meniett 20 is an electronically controlled membrane pump. Meniett 20 generates dynamic pressure and consists of a pump house, electronic hardware, software and encapsulation. In connection with the pump house there is a tube to which an ear cuff is connected and through which air pressure pulses from Meniett 20 are transferred to the patient's ear canal. The construction is encapsulated in a plastic cover designed specially for Meniett 20. There is an opening for replacing the battery. With the help of a pulse shape selector, concealed in the battery compartment, there is a possibility to switch between three different pulse shapes. To start and stop the treatment there is an on/off button on the front panel. On the front panel are four indicators for communication with the patient.
The provided document is a 510(k) summary for the Pascal Medical AB Meniett 20 device, which is a medical device for symptomatic treatment of Meniere's Disease. It does not contain information about acceptance criteria or a detailed study proving the device meets those criteria in the format requested. However, it does describe non-clinical and clinical test results.
Based on the available text, here's what can be extracted and inferred regarding the closest equivalent to acceptance criteria and study details:
Closest Equivalent to Acceptance Criteria and Device Performance:
The document summarizes the device's technical specifications and clinical effect, which serve as the de facto "performance" even if not explicitly stated as "acceptance criteria."
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Product life meets specifications | "The results of these tests confirmed that the proposed device meets the specifications for these parameters." (Performed in non-clinical in vitro testing) |
| Presentation of data meets specifications | "The results of these tests confirmed that the proposed device meets the specifications for these parameters." (Performed in non-clinical in vitro testing) |
| Environmental parameters meet specifications | "The results of these tests confirmed that the proposed device meets the specifications for these parameters." (Performed in non-clinical in vitro testing) |
| Power supply and battery life meet specifications | "The results of these tests confirmed that the proposed device meets the specifications for these parameters." (Performed in non-clinical in vitro testing) |
| Sound levels meet specifications | "The results of these tests confirmed that the proposed device meets the specifications for these parameters." (Performed in non-clinical in vitro testing) |
| Environmental operating temperatures meet specifications | "The results of these tests confirmed that the proposed device meets the specifications for these parameters." (Performed in non-clinical in vitro testing) |
| Device achieves intended clinical effect (relief of vertigo attacks, improved balance and functionality) | "The results showed that the use of Meniett 20 gave the intended clinical effect in all the patients except two." (Clinical test on 20 patients) |
| No worsening of patient condition | "None of the patients became worse while using Meniett 20." (Clinical test on 20 patients) |
| No other adverse effects observed | "No other adverse effects of treatment have been observed in connection with the use of Meniett 20." (Clinical test on 20 patients) |
| Performs well under normal daily clinical use conditions | "It was demonstrated that the Meniett 20 performed well under the normal conditions of daily clinical use." (Concluded from clinical test) |
Study Details:
-
Sample size used for the test set and the data provenance:
- Sample Size (Clinical): 20 patients.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study involved "patients with a definite history and diagnosis of Meniere's disease and active vestibular symptoms." It is an observational study described in a regulatory submission, so it is retrospective in nature from the perspective of the submission, but the patient data would have been collected prospectively during the treatment period.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The diagnosis of Meniere's disease would typically be made by medical professionals, but the number and qualifications of those involved in establishing the "ground truth" (i.e., diagnosis and symptom assessment in the context of the study) are not detailed.
-
Adjudication method for the test set:
- This information is not provided in the document. The text states "The results showed that the use of Meniett 20 gave the intended clinical effect in all the patients except two," implying an assessment of clinical effect, but the method for dispute resolution or consensus among multiple assessors (if any) is not discussed.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This was not an MRMC study, nor does it involve AI assistance for human readers. This device is a treatment device, not an diagnostic imaging or AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This device is a physical medical device (membrane pump) for treatment, not an algorithm. Therefore, a standalone algorithm-only performance assessment does not apply. The device operates under patient use following medical instruction and surgical intervention (myringotomy).
-
The type of ground truth used:
- The "ground truth" for the clinical study was based on clinical diagnosis of Meniere's disease and patient-reported symptoms (e.g., attacks of vertigo, balance, functionality), presumably corroborated by medical professionals. The statement "The aim of treatment was to relieve patients from the attacks of vertigo and to improve balance and functionality" indicates subjective and objective clinical assessments.
-
The sample size for the training set:
- This device is not an AI/ML algorithm that requires a "training set" in the typical sense. The "Clinical Test Results" section describes a clinical study on 20 patients. There is no mention of a separate training set.
-
How the ground truth for the training set was established:
- As there is no "training set" for an AI/ML algorithm, this question does not apply. If interpreted as how the patients were selected for the clinical study (the "test set"), they were identified as "20 patients with a definite history and diagnosis of Meniere's disease and active vestibular symptoms." This diagnosis would be established by medical experts based on clinical criteria for Meniere's disease.
{0}------------------------------------------------
PASCAL MEDICAL
DEC 28 1999
Kg91562
Confidential
Filename Meniett 20 510(k) Supplement
FDA No. K991562
September 15th, 1999
SECTION XVII 510(K) SUMMARY
510(K) Summary
| Submitter: | Pascal Medical ABTeknoCenterS-302 50 Halmstad, Sweden | |
|---|---|---|
| Telephone:FAX: | 46-35-17 47 4046-35-12 69 21 | |
| Date prepared: | March 31, 1999 | |
| Device name: | Pascal Medical AB Meniett 20 | |
| Classification name: | Tympanometer | |
| Predicate Devices: | American Electromedics Corporation ScreeningTympanometer (K822475) & Hyper/Hypo baric Chambers |
This medical device is being submitted in this 510(K) Notification by virtue of the fact that it is substantially equivalent to hypo / hyperbaric pressure chambers which have been used previously for the treatment and temporary relief of symptoms associated with Meniere's Disease. The Meniett 20 is a device which functions in a range of performance specifications which are similar to a conventional tympanometer and is used in a manner similar to a hypo / hyper baric pressure chamber to create localized overpressure in the middle ear to relieve the symptoms associated with Meniere's Disease.
Device Description:
Meniett 20 is indicated for symptomatic treatment of Menière's Disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Pascal Medical. The logo consists of a nautilus shell on the left, followed by the words "PASCAL MEDICAL" in all capital letters. The text is in a simple, sans-serif font.
Confidential
Filename Meniett 20 510(k) Supplement
FDA No. Date September 15th, 1999 K991562
Patient treatment
The treatment consists of two phases:
-
- Myrongotomiy
A surgical intervention whereby a conventional ventilation tube is surgically placed in the eardrum of the ear to be treated. This intervention is performed in exactly the same way and using the same materials as for other indications, e.g. otitis media.
- Myrongotomiy
-
- Application of pressure pulses to the middle ear
The therapy starts after the ventilation tube has been placed in the eardrum. The patient is instructed in how to place the ear-cuff in the external ear canal to minimize leakage to the outer environment. The patient is instructed to treat him/herself three times daily, morning, midday and evening and for three minutes each time. The total daily exposure of pressure pulses in the range of 10-16 mbar with this dosage is 64.8 seconds. Alteration of the standard dosage may be done in after consultation with the treating doctor.
- Application of pressure pulses to the middle ear
The treatment is continued for as long as the patient finds him/herself in a period of attacks of vertigo. During periods of remission there is no need for treatment
Meniett 20 is an electronically controlled membrane pump. Meniett 20 generates dynamic pressure and consists of a pump house, electronic hardware, software and encapsulation. In connection with the pump house there is a tube to which an ear cuff is connected and through which air pressure pulses from Meniett 20 are transferred to the patient's ear canal. The construction is encapsulated in a plastic cover designed specially for Meniett 20. There is an opening for replacing the battery. With the help of a pulse shape selector, concealed in the battery compartment, there is a possibility to switch between three different pulse shapes. To start and stop the treatment there is an on/off button on the front panel. On the front panel are four indicators for communication with the patient.
Intended Use:
Meniett 20 is indicated for symptomatic treatment of Menière's disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops.
Technological Characteristics:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Pascal Medical. The logo consists of a nautilus shell on the left and the words "PASCAL MEDICAL" on the right. The text is in a sans-serif font and is all caps.
Confidential
Filename Meniett 20 510(k) Supplement
Date FDA No. September 15th, 1999 K991562
Comparing the proposed device to the predicate devices, both devices utilize the same range of chemical compositions, packaging and formulations. There are no significant differences.
Summary of Non-Clinical Tests:
In vitro testing was performed to evaluate / assess the product life, presentation of data, environmental parameters, power supply and battery life, sound levels, and environmental operating temperatures. The results of these tests confirmed that the proposed device meets the specifications for these parameters.
Clinical Test Results:
20 patients with a definite history and diagnosis of Meniere's disease and active vestibular symptoms have been treated with middle ear pressure pulses using Meniett daily for a period of four weeks. Thereafter, the majority of patients have continued treatment up to 4 months. The dosage was standardized in all patients. The aim of treatment was to relieve patients from the attacks of vertigo and to improve balance and functionality. The results showed that the use of Meniett 20 gave the intended clinical effect in all the patients except two. None of the patients became worse while using Meniett 20. No other adverse effects of treatment have been observed in connection with the use of Meniett 20. Temporary relapses of symptoms have been observed in patients who were not able to continue the use Meniett 20. The symptoms were relieved again in these patients by a renewal of treatment with the Meniett 20. It was demonstrated that the Meniett 20 performed well under the normal conditions of daily clinical use.
Conclusions:
Testing performed on the Pascal Medical AB Meniett 20 indicates that it is safe, effective, and performs as intended, when used in accordance with the instructions for use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 8 1999
Jeffrey R. Sideman Pascal Medical AB 7307 Glouchester Drive Minneapolis, Minnesota 55435
Re: K991562 Trade Name: Meniett 20 Regulatory Class: Class II Product Code: 77ETY/874.1090 Dated: September 24, 1999 Received: September 29, 1999
Dear Mr. Sideman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Jeffrey R. Sideman
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page / of /
510(k) Number (if known): x 991562
Device Name: Pas call Midical AB Meniett 20
Indications For Use:
Image /page/5/Picture/4 description: The image shows the logo for Pascal Medical. The logo consists of a nautilus shell on the left and the words "PASCAL MEDICAL" on the right. The text is in a serif font and is all uppercase.
Pre-Market Notification
Pascal Medical AB Meniett 20 INDICATIONS FOR USE STATEMENT SECTION XII
Meniett 20 is indicated for symptomatic treatment of Menière's disease. The therapeutic effect of Meniett 20 is achieved by applying low frequency, low amplitude pressure pulses to the middle ear whereby inner ear endolymphatic fluids are assumed to be evacuated from the cochlea and thus relieve the patient of the symptoms associated with endolymphatic hydrops.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ellsworth L. Sharpe
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K991562
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
§ 874.1090 Auditory impedance tester.
(a)
Identification. An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.(b)
Classification. Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.