(117 days)
For in vitro diagnostic use with the CARESIDE Analyzer™ to measure prothrombin time from applied citrated whole blood or citrated plasma as a general screening procedure for the detection of possible clotting deficiencies in the extrinsic pathway or to monitor patients receiving anticoagulant therapy.
CARESIDE™ PT cartridges are used with the CARESIDE Analyzer™ to measure prothrombin time in citrated plasma with the sample applied either as citrated whole blood or citrated plasma. The CARESIDE™ PT cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma to a cartridge cuvette to initiate the measurement of prothrombin time. The cartridge (patent pending) contains all reagents necessary to measure prothrombin time.
Each CARESIDE™ PT cartridge consists of a cuvette with dried recombinant rabbit tissue factor with calcium ions mounted in a plastic cartridge with a hinged lid. The user opens the pouch, introduces the citrated whole blood or citrated plasma specimen into the cartridge Sample Well, closes the lid and inserts the cartridge into the CARESIDE Analyzer™.
Once loaded, the CARESIDE Analyzer™ scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the Sample Well into the cartridge channels and chambers. As the cartridge continues to spin. the blood cells are separated from the plasma and the cells accumulate in the separation well. Forty microliters of citrated plasma remain in the metering passage. Any excess sample flows into an overflow well. Once the analyzer completes any other tests, the sample is added to begin the coagulation test. The metered volume of sample is dispensed into the cuvette by a plunger that displaces a flexible seal that covers the Sample Well while a second plunger seals the cartridge vent. As the flexible seal is displaced, air is forced through the metering passage, forcing the sample out and into the cuvette to mix with the reagent. The cuvette is then positioned over an LED and the coagulation event is optically monitored. An onboard timer measures the coagulation time.
Photodiodes monitor the amount of scattered light transmitted by a 570 nanometer light emitting diode to optically monitor the progress of the clot formation. The time at which a clot is detected is reported as the prothrombin time. The results are also reported as the prothrombin ratio (% PT/mean normal PT reference interval) or as the INR (International Normalized Ratio).
The provided text describes the CARESIDE™ PT device and its performance in comparison to a predicate device, Innovin on the MLA Electra 900C. However, it does not explicitly state acceptance criteria or a detailed study plan with statistical endpoints in the format typically used for AI/ML device evaluations. The document is primarily a 510(k) summary, focusing on substantial equivalence.
Based on the provided information, here's an attempt to extract the requested details, with several points marked as "Not Stated" where the information is absent from the text.
Acceptance Criteria and Device Performance
The document does not explicitly state pre-defined acceptance criteria in terms of statistical metrics (e.g., minimum sensitivity, specificity, or agreement thresholds) for regulatory approval. Instead, it relies on demonstrating substantial equivalence to a predicate device through comparative performance characteristics.
The "Accuracy via Method comparison" section provides the most direct performance comparison, which suggests implicit acceptance of performance equivalent to or better than the predicate device.
| Acceptance Criteria (Implied) | Reported Device Performance (CARESIDE™ PT) |
|---|---|
| Detection Limit: Similar to or better than predicate (7 sec) | 9 sec |
| Reportable Range: Similar to or better than predicate | 9 to 75 sec |
| Precision: Acceptable (No predicate target given) | Total CV, 10.2 sec, 4.1% |
| Accuracy (Correlation with Predicate): Strong correlation | CARESIDE™ = 0.98 (Innovin on Electra 900C ) + 0.60 sec,r = 0.98 (correlation coefficient) |
| Interference: No significant interference | No significant interference observed at tested concentrations of:Bilirubin 20 mg/dLHemoglobin 500 mg/dLTriglyceride 300 mg/dL |
| Safety and Effectiveness: As safe, effective, and performs as well as or better than the legally marketed predicate device. | "The nonclinical and clinical data provided demonstrate that the CARESIDE™ PT product is as safe, effective, and performs as well as or better than the legally marketed predicate device." |
Study Details
-
Sample size used for the test set and the data provenance:
- Sample Size: Not explicitly stated. The "Accuracy via Method comparison" implies a dataset was used for correlation, but the number of samples is not provided.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study appears to be clinical performance data for device comparison. It does not mention if the data was retrospective or prospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Stated. The ground truth in this context is the actual prothrombin time measured by a reference method (the predicate device). There are no "experts" involved in establishing this type of ground truth; it's a quantitative measurement.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. This type of adjudication is typically used for subjective assessments or when reconciling disagreements among human readers, especially in image-based diagnostic AI. For a quantitative measurement device like this, the ground truth is established by the reference method (predicate device), and the CARESIDE™ PT measurements are compared to it.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a comparison between two devices (CARESIDE™ PT vs. predicate device), not a study involving human readers with and without AI assistance for interpretation. The CARESIDE™ PT is a standalone diagnostic device.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone device comparison was done. The CARESIDE™ PT device, incorporating its own algorithm for clot detection, was compared against the predicate device (Innovin on Electra 900C) in a standalone manner to assess its measurement accuracy and other performance characteristics. The comparison is between the CARESIDE™ PT's measured coagulation time and the predicate device's measured coagulation time. The "Clot Detection Algorithm" for both devices is "Peak of second derivative of optical time course."
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for the "Accuracy via Method comparison" study effectively comes from the measurements provided by the predicate device (Innovin on Electra 900C). This is a form of reference standard measurement from an established, legally marketed device.
-
The sample size for the training set:
- Not applicable / Not stated. This document describes a medical device, specifically a test cartridge and analyzer, not a machine learning model that undergoes a distinct "training" phase with a separate "training set" in the common AI/ML sense. The device has an algorithm for clot detection, but its development process (including any data used to refine that algorithm) is not detailed here.
-
How the ground truth for the training set was established:
- Not applicable / Not stated. As above, the concept of a "training set" and its associated ground truth in the context of an AI/ML model is not directly addressed for this device.
{0}------------------------------------------------
CARESIDE, Inc. Page 10
CARESIDE™ PT Premarket March 30 1999
510(K) SUMMARY: CARESIDE™ PROTHROMBIN IV. TIME SAFETY AND EFFECTIVENESS
I. Applicant Information
- A. Applicant Name
- Applicant/Manufacturer Address B.
- Telephone Number C.
- D. Contact Person
- E. FAX Number
- F. e-Mail Address
- G. Date 510(k) Summary prepared
II. Device Information
- A. Device Name (Trade)
- Device Name (Classification) B.
- Device Classification C.
CARESIDE, Inc.
6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 AsarchK@CARESIDE.com March 30, 1999
CARESIDE™ PT
Prothrombin Time test system Hematology panel Prothrombin Time test system Regulation Number: 21 CFR 864.7750 Regulatory Class 2 Classification Number: 81GJS Performance Standards
- D. Special controls and performance standards
III. Substantial Equivalence Claim
- A. General equivalency claim
The ability to monitor coagulation events optically is widely recognized and accepted for use in coagulation tests.
Prothrombin time in vitro diagnostic products using optical and other detection technologies are already on the U.S. market.
B. Specific equivalency claim
This CARESIDE™ PT test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Innovin reagent for the quantitative measurement of prothrombin time on the MLA Electra 900C.
Dade Innovin on the MLA Electra 900C. Name of Predicate Device: K974343 (Innovin)/K884863-(Electra 900C) Predicate Device 510K number: 81GJS (Innovin)/81GKP (Electra 900C) Product Code:
{1}------------------------------------------------
IV. Device Description
CARESIDE™ PT cartridges are used with the CARESIDE Analyzer™ to measure prothrombin time in citrated plasma with the sample applied either as citrated whole blood or citrated plasma. The CARESIDE™ PT cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma to a cartridge cuvette to initiate the measurement of prothrombin time. The cartridge (patent pending) contains all reagents necessary to measure prothrombin time.
A. Explanation of Device Function
Each CARESIDE™ PT cartridge consists of a cuvette with dried recombinant rabbit tissue factor with calcium ions mounted in a plastic cartridge with a hinged lid. The user opens the pouch, introduces the citrated whole blood or citrated plasma specimen into the cartridge Sample Well, closes the lid and inserts the cartridge into the CARESIDE Analyzer™.
Once loaded, the CARESIDE Analyzer™ scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the Sample Well into the cartridge channels and chambers. As the cartridge continues to spin. the blood cells are separated from the plasma and the cells accumulate in the separation well. Forty microliters of citrated plasma remain in the metering passage. Any excess sample flows into an overflow well. Once the analyzer completes any other tests, the sample is added to begin the coagulation test. The metered volume of sample is dispensed into the cuvette by a plunger that displaces a flexible seal that covers the Sample Well while a second plunger seals the cartridge vent. As the flexible seal is displaced, air is forced through the metering passage, forcing the sample out and into the cuvette to mix with the reagent. The cuvette is then positioned over an LED and the coagulation event is optically monitored. An onboard timer measures the coagulation time.
Test Cartridge Architecture: '' '''' '''' ''' ' ' ' ' '
Diluent > Reconstituted PT Reagent Dried PT Reagent -PT Reagent + Plasma --> Stable Clot
Photodiodes monitor the amount of scattered light transmitted by a 570 nanometer light emitting diode to optically monitor the progress of the clot formation. The time at which a clot is detected is reported as the prothrombin time. The results are also reported as the prothrombin ratio (% PT/mean normal PT reference interval) or as the INR (International Normalized Ratio).
- B. Test Summary
The prothrombin time is a screening test for the extrinsic and common coagulation pathway that was originally described by Quick in 1935. The prothrombin time test is sensitive to the coagulation factor abnormalities and to factor inhibitors affecting coagulation II, V, VII, and X, plus proteins C and S. Each component of the prothrombin reagents (calcium ions, tissue factor, and phospholipids) is required to initiate clotting. Differences among PT reagent preparations affect their sensitivity to reductions in coagulation factors. Accordingly, PT reagents are characterized by an international sensitivity index (ISI).
Various methods have been used historically to report and interpret prothrombin time results in order to account for differences in thromboplastin sensitivity. For example, prothrombin time results have been expressed as the ratio between a patient's prothrombin time and the mean normal prothrombin time reference interval.
{2}------------------------------------------------
Prothrombin time results are also reported as the International Normalized Ratio as recommended by the International Committee on Thrombosis and Hemostasis and the International Council for Standardization in Hematology. The INR is the equivalent PT ratio that would be obtained if the appropriate International Reference Preparation (ISI = 1.0. World Health Organization) were used as the source of thromboplastin in the performance of the prothrombin time.
The time required for the formation of a fibrin clot provides information regarding the presence and activity of coagulation factors I (fibrinogen), II, V, VII, and X. The factors II. VII. and X are dependent upon the co-factor vitamin K for activation by posttranslational carboxylation. The prothrombin time may be increased in patients with hepatic disease, vitamin K deficiency or in patients on vitamin K antagonist anticoagulant therapy such as warfarin. Likewise, significant alterations in the intake of vitamin K will affect the prothrombin time.
EXTRINSIC COAGULATION PATHWAY
Tissue Factor > Factor VIIa Factor VII
VIIa, Ca" Factor Xa Factor X -
Xa Va > Thrombin Prothrombin -
Thrombin > Fibrin monomers Fibrinogen
xilla_ca_2__________________________________________________________________________________________________________________________________________________________________ Fibrin
--
Ca*2 > Crosslinked fibrin (Stable clot) Fibrin polymers -
V. Intended Use
- A. Intended Use
The CARESIDE™ PT cartridge is intended for in vitro diagnostic use in conjunction with CARESIDE Analyzer™ to quantitatively measure prothrombin time in applied citrated whole blood or citrated plasma.
B. Indications for Use ~
To be used with the CARESIDE Analyzer™ to measure PT from applied citrated whole blood or citrated plasma specimens as a general screening procedure for the detection of possible clotting deficiencies in the extrinsic pathway or to monitor patients receiving anticoagulant therapy.
-
- 10 - 10 8 1. - . .
{3}------------------------------------------------
VI. Technological Characteristics
- Similarities A.
| CARESIDE™ PT | Innovin on Electra 900C | |
|---|---|---|
| Intended Use | A general screen for the detection of clotting deficiencies in the extrinsic pathway or to monitor patients receiving anticoagulant therapy. | For use in prothrombin time determinations and prothrombin time-based assays. |
| Indications | For in vitro diagnostic use.For professional laboratory use. | Same |
| Measurement | Quantitative | Same |
| Method Principle | Classical one-stage prothrombin time (optical clot detection in the presence of recombinant rabbit tissue factor and synthetic phospholipids.) | Same, except type recombinant tissue factor is human in lieu of rabbit |
| Clot Detection Algorithm | Peak of second derivative of optical time course | Peak of second derivative of optical time course |
| Specimen dilution | Not required | Same |
| Materials | Dried recombinant rabbit brain tissue factor, synthetic phospholipids, and calcium. | Dried recombinant human tissue factor and phospholipids (thromboplastin) with calcium |
| Detection principle | Photometric | Same |
| Test time | Approx. 4 minute warm-up (on-board) plus clot time. | Approx. 1 minute warm-up (on-board) plus clot time |
| Sample Type | Applied citrated whole blood or citrated plasma | Citrated plasma |
| Specimen volume | 40 µl test volume(300 ± 50 µl applied volume whole blood or plasma) | 100 µl (plasma) |
| Quality Control | 2 levels | Same |
| Reporting Units | Sec, PT Ratio, INR | Sec and (PT Ratio or INR) |
| Reaction Temp. | 37 °C | 37.6 °C |
B. Differences
| CARESIDE™ PT | Innovin on Electra 900C | |
|---|---|---|
| Direct bloodspecimen | Yes, whole blood | No, requires separation ofwhole blood prior to sampleapplication |
| Accuratepipetting | Not required | Required |
| Reagent pre-warming | On-board | On-board |
·
.
{4}------------------------------------------------
i
C. Comparative Performance Characteristics
| CARESIDE™ PT | Innovin on Electra 900C | |
|---|---|---|
| Detection limit | 9 sec | 7 sec |
| Reportable range | 9 to 75 sec | 7 to 106 sec |
| Precision | Total CV, 10.2 sec, 4.1% | Not available |
| Accuracy viaMethodcomparison | CARESIDE™ = 0.98 (Innovin on Electra 900C ) + 0.60 sec,r = 0.98 | |
| Interference | No significant interference observed at tested concentration of interferent:Bilirubin 20 mg/dLHemoglobin 500 mg/dLTriglyceride 300 mg/dL | No data provided. |
D. Conclusion
The nonclinical and clinical data provided demonstrate that the CARESIDE™ PT product is as safe, effective, and performs as well as or better than the legally marketed predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle with its wings spread, which is a common symbol of the United States. The eagle is depicted in a simple, graphic style, with bold lines and a minimalist design.
DEPARTMENT OF HEALTH & HUMAN SERVICES
JUL 26 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Kenneth B. Asarch, Pharm. D., Ph.D. Vice President, Quality Systems and Regulatory Affairs Careside Inc. 6100 Bristol Parkway Culver City, California 90230
Re: K991065 Trade Name: CARESIDE™ PT Regulatory Class: II Product Code: GJS Dated: June 17, 1999 Received: June 21, 1999
Dear Dr. Asarch:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
VI. INDICATIONS FOR USE
510(k) Number:
Device Name:
CARESIDE™ PT
Indications for use:
For in vitro diagnostic use with the CARESIDE Analyzer™ to measure prothrombin time from applied citrated whole blood or citrated plasma as a general screening procedure for the detection of possible clotting deficiencies in the extrinsic pathway or to monitor patients receiving anticoagulant therapy.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lita E. Madson
Prescription Use M (Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
§ 864.7750 Prothrombin time test.
(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).