K Number
K990915
Date Cleared
1999-04-21

(34 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for IgG Antinuclear Antibody (ANA) HEp2 is intended for the qualitative and semi-quantitative detection of IgG (Immunoglobulin G) antinuclear antibodies in human serum. Detection of ANA in humans can be used as an aid in the diagnosis of autoimmune connective tissue diseases.

Device Description

The ANA IgG IFA is an Indirect Fluorescence Assay (IFA) for the detection of IgG antibodies to ANA antigens in human serum.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Stellar Bio Systems' Indirect Fluorescence Assay for Anti-Nuclear Antibodies (ANA) IgG Antibody, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

The device's performance was evaluated relative to a commercially available ANA IFA kit. The acceptance criteria are implicitly defined by the reported performance metrics, which demonstrate the device's ability to provide comparable results to an existing device.

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Stellar ANA IFA Kit)
Relative Sensitivity~93.5% (referencing alternate IFA)93.5% (95% CI: 87.0% - 97.3%)
Relative Specificity~97.8% (referencing alternate IFA)97.8% (95% CI: 92.2% - 99.7%)
Relative Agreement~95.4% (referencing alternate IFA)95.4% (95% CI: 91.5% - 97.9%)
Titer Agreement (Identical)N/A (reported for 25 positive sera)12/25 sera had identical titers
Titer Agreement (± 1 dilution)N/A (reported for 25 positive sera)11/25 sera were within ± one, two-fold dilution
Titer Agreement (± 2 dilutions)N/A (reported for 25 positive sera)2/25 sera were within ± two, two-fold dilutions
ReproducibilityN/A (reported for 4 sera)35/36 endpoint titers identical, 1/36 within ± one dilution

Study Details

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: 199 sera (100 from patients diagnosed with SLE, 99 from normals).
  • Data Provenance: The samples were "frozen retrospective sera." The "99 sera were from normals with various ages, gender, and geographical areas." This suggests the data is retrospective and likely from mixed geographical origins, though specific countries are not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided in the summary. The study used a "commercially available ANA IFA" as the comparator (or "alternate IFA") to establish "relative" truth, but it does not specify how the initial diagnosis of SLE or the determination of "normals" was made, nor the involvement of experts in establishing the reference standard.

4. Adjudication method for the test set:

  • This information is not provided. The comparison was made against an "Alternate IFA," but the specifics of how discrepancies between the new device and the alternate were handled (if at all) are not detailed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted, and the device is not an AI-assisted device. It is a laboratory assay (Indirect Fluorescence Assay) for detecting antibodies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • This device is an in-vitro diagnostic assay, not an algorithm or software. Its performance is inherent to the chemical and biological reactions involved in the test. Therefore, the concept of "standalone (algorithm only)" is not applicable. The reported performance is the standalone performance of the assay itself.

7. The type of ground truth used:

  • The "ground truth" for this study was relative to another commercially available ANA IFA. The summary explicitly states: "Please be advised that 'relative' refers to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease."
  • For the SLE patients, their diagnosis (described as "patients diagnosed with SLE") served as an initial categorization, but the comparator IFA was the reference standard against which the new device was measured for "relative" performance.

8. The sample size for the training set:

  • This information is not applicable/provided. This device is a manual laboratory assay, not a machine learning model that requires a "training set" in the computational sense. The described evaluations are performance studies.

9. How the ground truth for the training set was established:

  • This information is not applicable as there is no training set in the context of this type of device.

{0}------------------------------------------------

4/21/99

K99 0915

510k Summary of Safety and Effectiveness

. his summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

Applicant Information:

Date Prepared:March 12, 1999
Name:Stellar Bio Systems, Inc.
Address:9075 Guilford RoadColumbia, MD 21046
Contact Person:John Brewer
PhoneNumber.410-381-8550
Fax Number.410-381-8984

Device Information:

Trade Name:Indirect Fluorescence Assay for Anti-Nuclear Antibodies (ANA) IgG Antibody
Common Name:ANA IgG IFA Test
Classification Name:ANA IgG Serological Reagent

quivalent Device: . H A

Device Description: The ANA IgG IFA is an Indirect Fluorescence Assay (IFA) for the detection of IgG antibodies to ANA antigens in human serum.

Intended Use: The Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for IgG Antinuclear Antibody (ANA) HEp2 is intended for the qualitative and semi-quantitative detection of IgG (Immunoglobulin G) antinuclear antibodies in human serum. Detection of ANA in humans can be used as an aid in the diagnosis of autoimmune connective tissue diseases.

Principle of Procedure:

Stellar Bio Systems' fluorescent antibody assays use the indirect method of antibody detection and titer determination. Diluted patient serum samples are applied to fixed antigens provided on paint delineated wells on glass microscope slides. During a 30 minute incubation, antibody specific for ANA antigens forms an antigen/antibody complex with the fixed antigens. In a brief washing step, nonspecific antibody and other unreacted serum proteins are eliminated. Fluorescein-conjugated goat anti-human IgG is then applied to the wells of the glass slide. The anti-IgG conjugate combines with human IgG, if present, during a 30 minute incubation. After a brief wash to remove unreacted conjugate, the slides are viewed by fluorescence microscopy. A positive antibody reaction is denoted by bright green fluorescence at the antigen sites.

{1}------------------------------------------------

Performance Characteristics

  1. Relative Sensitivity and Specificity - The Stellar ANA IFA kit was evaluated relative to a commercially available ANA IFA. The samples were frozen retrospective sera. 100 sera were from patients diagnosed with SLE. Ninety-nine sera were from normals with various ages, gender, and geographical areas. The data in Table 1 summarizes the data.

Table 1 Relative Sensitivity and Specificity of the Stellar ANA IFA Kit Relative to Alternate IFA

PositiveNegativeTotal
AlternateIFAPositive1007107
Negative28890
Total10295197
95% Confidence Interval
Relative Sensitivity = 100/107 = 93.5%87.0% - 97.3%
Relative Specificity = 88/90 = 97.8%92.2% - 99.7%
Relative Agreement = 188/197 = 95.4%91.5% - 97.9%

Please be advised that 'relative' refers to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease.

    1. Titer Agreement: 25 positive sera were serially two-fold diluted and the endpoint titer was determined on the Stellar ANA IFA and a commercially available ANA IFA. The endpoint titer results are as follows.
  • Stellar ANA IFA

{2}------------------------------------------------

Identical Titer12/25
± one, two-fold dilution11/25
± two, two-fold dilutions2/25

.

  1. Reproducibility: Three positive with various titers (1:80, 1:640, 1:2560) and one negative sera were serially diluted and assayed three times each on three different assays. 35/36 of the end point titers were identical. 1/36 of the end point titers was within ± one, two-fold dilution.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 1 1999

Mr. John Brewer Stellar Bio Systems, Inc. 9075 Guilford Road Columbia, Maryland 21046

Re: K990915

Trade Name: Indirect Fluorescence Assay for Anti-Nuclear Antibodies (ANA) IgG Antibody Regulatory Class: II Product Code: DHN Dated: March 12, 1999 Received: March 18, 1999

Dear Mr. Brewer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Dutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

K 990915 510(k) Number: Not Known

Device Name: Indirect Fluorescence Assay for Anti-Nuclear Antibodies (ANA) IgG Antibody

Indications For Use: The Stellar Bio Systems' Indirect Fluorescence Assay (IFA) for IgG Antinuclear Antibody (ANA) HEp2 is intended for the qualitative and semi-quantitative detection of IgG (Immunoglobulin G) antinuclear antibodies in human serum. Detection of ANA in humans can be used as an aid in the diagnosis of autoimmune connective tissue diseases.

Cste E. Mafin

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K990915

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use V (Per 21 CFR 801.109)

OR

Over-The Counter Use (Optional Format 1-2-96)

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).