K Number
K990780
Date Cleared
1999-03-22

(13 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nuvolase 532 Laser System for Ophthalmology is intended for use in the treatment of vascular lesions of the skin, including capillary hemangioma (port wine stain), strawberry hemangioma, telangiectasia, and rosacea. It is also intended for pigmented lesions including freckles, age spots, café-au-lait, and lentigo.

Device Description

The Nuvolase 532 Cutaneous Laser System is a continuous-wave frequencydoubled diode-pumped Nd: Y AG laser system. Treatement beam power output for the system is 50 milliwatts to 1.5 watts at a wavelength of 532 nm. The maximum power level to tissue is 2.5 Watts CW. The aiming beam is provided by a red diode laser operating at 670 nm. Exposure durations for the Nuvolase 532 Laser System for Ophthalmology (in seconds) are 0.05, 0.1, 0.25, 0.5, 1.0, and continuous. Delivery of the beam occurs via fiber optic and handpiece.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Nuvolase 532 Cutaneous Laser System" and explicitly states it is a modification of a previously cleared device (Nuvolase 660 Laser System, K970667). The document does not contain any information regarding acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement.

The document is a regulatory submission for substantial equivalence based on technological characteristics and intended use, not a clinical study report. Therefore, I cannot provide the requested information.

Here's why each specific point cannot be addressed from the given text:

  1. A table of acceptance criteria and the reported device performance: No performance data or acceptance criteria are present. The submission focuses on demonstrating equivalence to a predicate device.
  2. Sample size used for the test set and the data provenance: No test set or study data is mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No ground truth establishment process is described.
  4. Adjudication method for the test set: Not applicable as there is no test set or study.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a laser system, not an AI-powered diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

Conclusion from the document:

The provided document details a 510(k) submission for a laser device, emphasizing its technological equivalence to a predicate device for the treatment of vascular and pigmented skin lesions. It is a regulatory submission asserting substantial equivalence, not a report of a study proving performance against acceptance criteria.

{0}------------------------------------------------

3/22/99

System for Opthalmology Nuvolase 532 L American Laser Medical, Inc. March 5, 1999

KA9078D

Summary of Safety and Effectiveness

Regulatory Authority:

Safe Medical Devices Act of 1990, 21 CFR 807.92

Company Name/Contact:

Daniel Hoefer American Laser Medical, Inc. 1832 South 3850 West Salt Lake City, UT 84104 (801) 972 1311, FAX (801) 972 5251

Name of Device:

Trade Name: Nuvolase 532 Cutaneous Laser System

Common Name: Dermatology Laser

Classification Name: Laser Instrument, Surgical, powered (per 21 CFR 878.4810)

Predicate Devices:

The Nuvolase 532 Laser System for Opthalmology has been modified from the American Laser Medical, Inc. Nuvolase 660 Laser System, K970667.

Description of Device:

The Nuvolase 532 Cutaneous Laser System is a continuous-wave frequencydoubled diode-pumped Nd: Y AG laser system. Treatement beam power output for the system is 50 milliwatts to 1.5 watts at a wavelength of 532 nm. The maximum power level to tissue is 2.5 Watts CW. The aiming beam is provided by a red diode laser operating at 670 nm. Exposure durations for the Nuvolase 532 Laser System for Ophthalmology (in seconds) are 0.05, 0.1, 0.25, 0.5, 1.0, and continuous. Delivery of the beam occurs via fiber optic and handpiece.

{1}------------------------------------------------

Intended Use:

The Nuvolase 532 Laser System for Ophthalmology is intended for use in the treatment of vascular lesions of the skin, including capillary hemangioma (port wine stain), strawberry hemangioma, telangiectasia, and rosacea. It is also intended for pigmented lesions including freckles, age spots, café-au-lait, and lentigo.

Technological Characteristics/Device Comparison:

The Nuvolase 532 Laser Sysem is a modification of the Nuvolase 660 Laser System, already in legal commercial distribution. Each of the systems optically pumps an Nd: Y AG crystal using 808 nm diodes to produce laser light at 1064 nm. This light passes through a second crystal which exhibits a non-linear optical response, re-emitting the laser energy at the first harmonic of the 1064 line, 532 nm greeen. The continuous wave beam is then shuttered electro-mechanically to produce the desired exposure durations. Each device is intended for treatement of vascular and pigmented lesions of the skin. The delivery system is equivalent in each case.

Conclusion:

The device modification does not affect the indications for use, materials, method of manufacture, or technology of the legally marketed device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three flowing lines representing the body and head. The figure is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" surrounding it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 2 1999

Mr. Daniel Hoefer Regulatory Affairs American Laser Medical, Inc. 1832 South 3850 West Salt Lake City, Utah 84104

Re: K990780 Trade Name: Nuvolase 532 Cutaneous Laser System Regulatory Class: II Product Code: GEX Dated: March 5, 1999 Received: March 9, 1999

Dear Mr. Hoefer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 – Mr. Daniel Hoefer

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name:_ NUVOLASE 532 CUTANEOUS LASER SYSTEM

Indications For Use:

...
..........................................................................................................................................................................

VASCULAR LESIONS:

    1. CAPILLARY HEMANGIOMA (PORT-WINE STAIN)
    1. STRAWBERRY HEMANGIOMA
    1. TELANGIECTASIA
    1. ROSACEA

PIGMENTED LESIONS:

    1. FRECKLES
    1. AGE SPOTS
    1. CAFE-AU-LAIT
    1. LENTIGO

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK990780
Prescription UseXOROver-The-Counter Use
(Per 21 CFR 801.109)(Optional Format 1-2-96)

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.