(133 days)
The ID Biomedical Velogene™ Rapid MRSA Identification Assay is intended as a qualitative assay for the definitive identification of methicillin resistance in presumptively identified cultures of Staphylococcus aureus by detecting the presence of the mecA gene. The presence of the mecA gene confers resistance to methicillin.
The Velogene™ Rapid MRSA Identification Assay is a DNA probe based diagnostic device that is based on Cycling Probe™ Technology (CPT) to generate spectrophotometric or visual results. The assay consists of two reagent kits, MRSA Lysis/Cycle Kit and MRSA Microwell Detection Kit.
Here's a breakdown of the acceptance criteria and study details for the Velogene™ Rapid MRSA Identification Assay, based on the provided text:
Acceptance Criteria and Device Performance
The primary acceptance criterion is substantial equivalence to the predicate device, the Oxacillin Screen Agar (NCCLS approved test for MRSA detection). This substantial equivalence is demonstrated through high agreement in identifying methicillin resistance in S. aureus samples.
| Acceptance Criteria Category | Acceptance Criteria (Implicit) | Reported Device Performance (Velogene™ Rapid MRSA Identification Assay) |
|---|---|---|
| Overall Agreement with Predicate Device | High agreement with Oxacillin Screen Agar in clinically relevant S. aureus samples. | 99.3% agreement (420/423) in routine clinical samples. |
| Reproduction of MRSA Results | High reproducibility for MRSA results. | 99.8% (449/450) reproducible for MRSA (OD650 ≤ 0.18, Visual: Clear). |
| Reproduction of MSSA Results | High reproducibility for MSSA results. | 99.8% (449/450) reproducible for MSSA (OD650 > 0.18), 100% (450/450) visual reproducibility. |
| Analytical Sensitivity (MRSA) | Detect MRSA at clinically relevant concentrations. | Detects 93.75 x 10 to 375 x 10 MRSA CFU/reaction (OD650 < 0.18 or visual clear). |
| Analytical Specificity (S. aureus) | Specific for S. aureus mecA gene detection, limiting false positives from other species. | Variations observed with coagulase-negative staphylococci (Staphylococcus epidermidis, S. warneri, S. sciuri), indicating the assay should only be used for coagulase-positive Staphylococcus. |
| Challenge Panel Accuracy | High accuracy in identifying known MRSA/MSSA/BORSA isolates. | 95.7% accurate across 4 clinical sites for a challenge panel. |
| BORSA Panel Identification | Correctly identify BORSA samples. | 100% correct identification (32/32) visually and spectrophotometrically for BORSA samples. |
Study Details
2. Sample size used for the test set and the data provenance:
-
Clinical Test Set:
- Sample Size: 423 coagulase-positive S. aureus samples.
- Data Provenance: Routinely submitted samples for microbiological identification at 4 geographically distributed U.S. sites. This indicates a prospective collection of real-world clinical samples.
-
Reproducibility Panel:
- Sample Size: 800 samples (10 isolates in a panel, 2 replicates per panel member, 5 days, 2 lots, 4 clinical sites; 10 * 2 * 5 * 2 * 4 = 800).
- Data Provenance: Tested at 4 clinical sites.
-
Challenge Panel:
- Sample Size: 75 isolates (49 homoresistant/heteroresistant MRSA, 25 MSSA, 1 BORSA).
- Data Provenance: Tested by each of the 4 clinical sites.
-
BORSA Panel:
- Sample Size: 32 samples (8 samples per site across 4 clinical sites).
- Data Provenance: Tested across 4 clinical sites.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The text does not explicitly state the number or qualifications of experts used to establish the initial ground truth for S. aureus classification prior to testing with the predicate device. However, the ground truth for discrepant isolates was established by nuc/mecA PCR, which is a molecular, objective gold standard.
4. Adjudication method for the test set:
For the initial comparison, the "ground truth" was largely established by the predicate device (Oxacillin Screen Agar). For the 3 discrepant isolates between the Velogene™ assay and the Oxacillin Screen Agar, the adjudication method was nuc/mecA PCR. This serves as a definitive molecular gold standard to resolve discrepancies.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, this is not an MRMC comparative effectiveness study in the context of human readers and AI assistance. This study evaluates a diagnostic assay's performance against a predicate device. The "readers" are the methods of interpretation (spectrophotometer vs. visual) of the Velogene™ assay, and its comparison is against a growth-based culture method. There's no AI involved.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The Velogene™ Rapid MRSA Identification Assay can be interpreted visually or using a spectrophotometer. While the spectrophotometer provides an objective reading (OD650 ≤ 0.18 or > 0.18) similar to an "algorithm only" interpretation, visual interpretation is also an option, implying a human-in-the-loop component for some results. However, the performance data presents both spectrophotometric and visual results as highly concordant, suggesting objective interpretation is a primary mode. The PCR results validating discrepancies are also a standalone, objective measure.
7. The type of ground truth used:
- Primary Ground Truth for Clinical Comparison: The Oxacillin Screen Agar (a NCCLS approved test for MRSA detection) served as the primary comparative ground truth.
- Definitive Ground Truth for Discrepancy Resolution: nuc/mecA PCR was used as the definitive ground truth for the 3 discrepant isolates, confirming the presence or absence of the mecA gene. This is a molecular/pathology-based ground truth.
- Analytical Sensitivity/Specificity/Reproducibility: Used defined control strains (S. aureus ATCC 29213 - mecA negative, S. aureus ATCC 33592 - mecA positive) and specific isolates for analytical specificity testing. These are laboratory-defined ground truths.
8. The sample size for the training set:
The document describes performance evaluation studies and does not explicitly mention a dedicated training set or its sample size. This is common for traditional diagnostic assays where performance is validated against established methods and/or molecular gold standards, rather than "trained" like machine learning models. The study focuses on validation of the assay.
9. How the ground truth for the training set was established:
As no explicit training set is described, this question is not applicable. The assay's principles are based on known molecular biology (detection of the mecA gene).
{0}------------------------------------------------
510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Trade Name: | Velogene™ Rapid MRSA Identification Assay |
|---|---|
| Common Name: | DNA probe test to identify the presence of the mecA gene in S. aureus .The presence of mecA confers resistance to methicillin. |
| Classification Name: | Manual antimicrobial susceptibility test system |
Comparative Device
The ID Biomedical Velogene™ Rapid MRSA Identification Assay was compared to the Oxacillin Screen Agar (Mueller Hinton Agar with 4% NaCl and Oxacillin (6 µg/mL)), a NCCLS approved test for detection of MRSA.
Device Description
The Velogene™ Rapid MRSA Identification Assay is a DNA probe based diagnostic device that is based on Cycling Probe™ Technology (CPT) to generate spectrophotometric or visual results.
The assay consists of two reagent kits, MRSA Lysis/Cycle Kit and MRSA Microwell Detection Kit.
| MRSA Lysis/Cycle Kit | |
|---|---|
| MRSA Lysis Reconstitution Buffer (1 x 3 mL) | Buffered Solution |
| MRSA Lysis Reagent, Lyophilized (2 x 0.5 mL) | Lyophilized Buffer and Enzymes |
| MRSA Cycle Reconstitution Buffer (1 x 6 mL) | Buffered Solution |
| MRSA Cycle Reagent, Lyophilized (48 x 0.2 mL) | Lyophilized Buffer, Enzyme and DNA/RNA Probe |
| Tubes, 1.5 mL (50) | N/A |
ID Biomedical Corporation
8855 Northbrook Court, Burnaby, BC V5J 5J1 Canada Telephone: 604-431-9314 Fax: 604-431-9378
Component Name
Reagent Description
{1}------------------------------------------------
Component Name
Reagent Description
MRSA Microwell Detection Kit
| MRSA Cycle Stop Reagent (1 x 6 mL) | Buffered Solution containing Antibody |
|---|---|
| Streptavidin Coated Microwells (48) | Streptavidin Coated Microtiter Well |
| Wash Buffer (1 x 50 mL) | Buffered Solution |
| Detection Substrate Reagent (1 x 12 mL) | Tetramethylbenzidine and Hydrogen PeroxideSolution |
| Detection Stop Reagent (1 x 5.5 mL) | Buffered Solution |
| Transfer Pipettes (50) | N/A |
| Microwell Frame (1) | N/A |
Principles of the Assay:
The Velogene™ Rapid MRSA Identification Assay is a DNA probe assay that uses Cycling Probe™ Technology (CPT).
CPT utilizes a fluorescein labeled, biotinylated DNA-RNA-DNA chimeric probe providing an RNase H sensitive cleavable linkage when bound to the complementary sequence of the mecA gene. The RNA portion of the chimeric probe is cleaved by RNase H when hybridized to the target DNA. The uncleaved probe (mecA negative) is detected by binding of the probe to a solid surface and attachment of an anti-fluorescein antibody conjugated with horseradish peroxidase, which converts a substrate to a colored end product. Cleavage of the probe (mecA positive) prevents binding of the probe-anti-fluorescein antibody complex, thus preventing formation of the colored end product.
Interpretation of Results:
Results may be read visually or using a spectrophotometer.
A methicillin resistant isolate (i.e. mecA gene is present) produces a colorless result or OD650mm of ≤0.18.
A methicillin sensitive isolate (i.e. mecA gene is absent) produces a distinctly blue color or OD650nm of > 0.18.
A result can be generated 90 minutes after primary isolation.
{2}------------------------------------------------
Intended Use of the Velogene™ Rapid MRSA Identification Assay
The ID Biomedical Velogene™ Rapid MRSA Identification Assay is intended as a qualitative assay for the definitive identification of methicillin resistance in presumptively identified cultures of Staphylococcus aureus by detecting the presence of the mecA gene. The presence of the mecA gene confers resistance to methicillin.
Clinical Significance
S. aureus may be resistant to ß-lactam antibiotics by one of three mechanisms:
-
- Classic resistance due to the possession of the chromosomal mecA gene which codes for the supplemental penicillin binding protein PBP2a. PBP2a protein functions as a transpeptidase involved in the synthesis of the cell wall and has a decreased affinity for ß-lactam antibiotics.
This category is further divided into the heteroresistant methicillin S. aureus (MRSA), in which only a small percentage of the population of mecA containing cells express the gene, and the homoresistant MRSA, in which all cells express the mecA gene.
- Classic resistance due to the possession of the chromosomal mecA gene which codes for the supplemental penicillin binding protein PBP2a. PBP2a protein functions as a transpeptidase involved in the synthesis of the cell wall and has a decreased affinity for ß-lactam antibiotics.
-
- Possession of modified penicillin binding proteins 1, 2 and 4 (MOD-SA). These proteins also have a decreased affinity for ß-lactam antibiotics.
-
- Hyper-B-lactamase producers (HBLP) continually produce high levels of B-lactamase.
The heteroresistant MRSA, MOD-SA and HBLP are usually referred to as the borderline resistant S. aureus (BORSA) because they can have an oxacillin MIC of 2 to 8 µg/mL.
Homoresistant and heteroresistant strains of S. aureus are considered to be resistant to all {}lactam antibiotics and are usually resistant to erythromycin, clindamycin, chloramphenicol, tetracycline, trimethoprim-sulfamethoxazole, and quinolone or aminoglycosides. The drug of choice for treating these strains of S. aureus is vancomycin.
The MOD-SA strains are considered to be resistant to all ß-lactam antibiotics and ß-lactam-B-lactamase inhibitor combinations. Infections caused by these isolates can be treated with vancomycin or some other non-B-lactam antibiotic.
{3}------------------------------------------------
HBLP, although identified as methicillin resistant by conventional methods due to a MC of 2 to 8 ug/mL, can in fact be treated with B-lactam-B-lactamase inhibitor combinations.
Device Comparison
ID Biomedical Corporation believes that the Velogene™ Rapid MRSA Identification Assay is substantially equivalent to oxacillin screen agar for detecting methicillin resistance (i.e. mecA) in presumptively identified colonies of coagulase positive Staphylococcus aureus.
Oxacillin Screen Agar:
Oxacillin screen agar is a growth based test which identifies methicillin resistance through the expression of the mecA gene that codes for the PBP2a, hyper B-lactamase production and possession of modified penicillin binding proteins 1, 2 and 4.
The oxacillin screen agar is inoculated with a suspension of an overnight culture of S. aureus (turbidity equivalent to a 0.5 McFarland standard) and incubated for 24 hours at 35°C. Plates are examined for evidence of growth. Growth in the presence of oxacillin indicates resistance to methicillin.
Results for methicillin resistance using the oxacillin screen agar require at least 24 hours after primary culture isolation.
Velogene™ Rapid MRSA Identification Assay:
The Velogene™ Rapid MRSA Identification Assay identifies methicillin resistance by detecting the nucleotide sequence specific for the mecA gene.
In the Velogene™ Rapid MRSA Identification Assay, overnight cultures of S. aureus are lysed to free the DNA from the cells. The DNA present in the lysed cell suspensions is processed using the method of CPT assay to detect the presence of the mecA gene. The presence of the mecA gene confers resistance to methicillin.
Results of the Velogene™ Rapid MRSA Identification Assay can be obtained in 90 minutes after primary culture isolation.
{4}------------------------------------------------
Summary of Similarities:
- · Require isolates to be presumptively identified as coagulase positive S. aureus;
- · Both tests identify methicillin resistance in S. aureus is due to the mecA gene;
- · Oxacillin screen agar and the Velogene™ Rapid MRSA Identification Assay identify methicillin resistant S. aureus.
Summary of Differences:
- · Time to result after primary culture: Velogene™ Rapid MRSA Identification Assay requires 90 minutes; Oxacillin screen agar requires 24 hours.
- · Method of detection:
Velogene™ Rapid MRSA Identification Assay is a DNA probe assay detecting the specific nucleotide sequence for the mecA gene;
Oxacillin screen agar is a growth based test which identifies methicillin resistance through the expression of the mecA gene that codes for PBP2a, hyper-production of {}-lactamase, or modification of other penicillin binding proteins.
- Clinical Importance:
Velogene™ Rapid MRSA Identification Assay can detect mecA mediated methicillin resistance in 90 minutes after primary culture isolation. Presence of the mecA gene indicates the isolate is heteroresistant or homoresistant MRSA. Absence of the mecA gene indicates the isolate is MSSA, hyper-ß-lactamase producer, or MOD-SA. Oxacillin screen agar requires 24 hours to identify methicillin resistance after primary culture isolation. Growth on the oxacillin screen agar indicates the isolate is heteroresistant or homoresistant, hyper-B-lactamase producer, or MOD-SA. No growth indicates the isolate is a MSSA, a heteroresistant MRSA, hyper-B-lactamase producer, or MOD-SA.
{5}------------------------------------------------
Performance Data
The Velogene™ Rapid MRSA Identification Assay was evaluated and compared to the predicate device, oxacillin screen agar, on 423 coagulase-positive S. aureus samples at 4 geographically distributed U.S. sites using routinely submitted samples for microbiological identification. There was 99.3% agreement between the two methods (420/423).
-
- One isolate was resistant in the Velogene™ Rapid MRSA Identification Assay but appeared sensitive by oxacillin screen agar and had an oxacillin MIC of 1.0 µg/mL. However, nuc/mecA PCR showed it to contain the mecA gene, and it is considered to be a heteroresistant MRSA. This type of isolate should be considered as resistant and the patient can only be treated effectively with vancomycin.
- One isolate, identified as MSSA, was sensitive in the Velogene™ Rapid MRSA 2. Identification Assay and appeared resistant (hazy growth) by oxacillin screen agar. This isolate was negative for the mecA gene in the nuc/mecA PCR assay, and therefore would be considered to be methicillin sensitive.
- One MOD-SA isolate was sensitive in the Velogene™ Rapid MRSA Identification Assay 3. but resistant by oxacillin screen agar. This isolate was negative for the mecA gene in the nuc/mecA PCR assay.
Thus the PCR results of the 3 discrepant isolates confirmed the results of the Velogene™ Rapid MRSA Identification Assay.
| Sample Type | Velogene™ Rapid MRSAIdentification Assay (OD650) | OxacillinScreen Agar | Agreement |
|---|---|---|---|
| MRSA | ≤0.18 | Growth | 208/208 (100%) |
| MSSA | >0.18 | No growth | 204/205 (99.5%) |
| Heteroresistant MRSA | ≤0.18 | Growth | 0/1 (0%) |
| HBLP | >0.18 | No growth | 5/5 (100%) |
| MOD-SA | >0.18 | No growth | 3/4 (75%) |
| TOTAL | 420/423 (99.3%) |
{6}------------------------------------------------
Analytical Sensitivity:
The analytical sensitivity of the Velogene™ Rapid MRSA Identification Assay was determined using S. aureus ATCC 29213 (mecA negative control) and S. aureus ATCC 33592 (mecA positive control).
Using a spectrophotometer, 93.75 (10) to 375 (10) MRSA CFU/reaction were detected (OD650mm of < 0.18). Visually, 93.75 (10) to 375 (10) MRSA CFU/reaction were detected (visual interpretation of clear).
Analytical Specificity:
The analytical specificity of the Velogene™ Rapid MRSA Identification Assay was tested using twelve species of coagulase negative staphylococci containing 3 strains of mecA positive methicillin resistant Staphylococcus epidermidis (MRSE) and 4 strains of methicillin sensitive S. epidermidis (MSSE).
Variations were observed in the results of the MSSE, S. warneri and S. sciuri. For the MSSE, the ODssonm range was 0.179 to 0.387, with one isolate below the cut off OD650m value of 0.18. For the S. warneri, 1 of the 6 visual observations was clear rather than distinctly blue. For S. sciuri, one isolate was below the cut-off ODssomm value of 0.18 with a reading of 0.175.
Based on the results of the above testing, the Velogene™ Rapid MRSA Identification Assay should only be used for identifying methicillin resistance in coagulase positive staphylococci.
Reproducibility
Reproducibility of the Velogene" Rapid MRSA Identification Assay was determined inhouse using three lots of reagents, three operators and ten strains of Staphylococcus aureus (five MSSA and five MRSA) with ten replicates in a single run.
| OD650nm | Visual Score | |
|---|---|---|
| MRSA | 99.8% (449/450)OD650 of ≤ 0.18 | 99.8% (449/450)Clear |
| MSSA | 99.8% (449/450)OD650 of > 0.18 | 100% (450/450)Distinctly blue |
{7}------------------------------------------------
In addition, a reproducibility panel containing a positive and negative control. 2 MRSA, 3 MSSA and 3 borderline oxacillin resistant S. aureus (BORSA) isolates (10 total isolates) was tested at each of the 4 clinical sites. Each site ran 2 replicates of each panel member on 5 different days with 2 different lots. Blinding was achieved by making 10 panels of randomly ordered samples. There was 100% agreement with all 800 samples tested over the 4 sites. One hundred percent (240/240) of the 3 MRSA isolates gave a clear result with an OD<< of ≤0.18, and 100% (560/560) of the MSSA and BORSA isolates were distinctly blue with an OD650 of >0.18.
Challenge Panel
A challenge panel of 75 isolates, consisting of 49 homoresistant and heteroresistant MRSA (mecA positive), 25 MSSA (mecA negative), and 1 BORSA (mecA positive, oxacillin screen resistant, MIC 8μg/mL) was tested by each site.
For the 4 clinical sites, the Velogene™ Rapid MRSA Identification Assay was 95.7% accurate.
BORSA Panel
A total of 32 samples of BORSA was tested across the 4 clinical sites (8 samples per site).
The Velogene " Rapid MRSA Identification Assay correctly identified all 32 samples (100%) both visually and spectrophotometrically.
Conclusion of Performance Data Comparison
With a performance agreement of 99.3% in the clinical data between the Velogene™ Rapid MRSA Identification Assay and Oxacillin Screen Agar, the Velogene™ Rapid MRSA Identification Assay was found to be substantially equivalent to Oxacillin Screen Agar for identifying methicillin resistance in presumptively identified cultures of coagulase positive S. aureus.
{8}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/8/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of three human profiles facing right, arranged in a cascading manner. The profiles are black against a white background. Encircling the graphic is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
JUL - 9 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Robert N. Bryan, Ph.D. Vice President, Research and Development ID Biomedical Corporation 19204 North Creek Parkway Suite 100 Bothell, WA 98011
Re: K990640
Trade Name: Velogene™ Rapid MRSA Identification Assay Regulatory Class: II Product Code: MYI Dated: June 21, 1999 Received: June 22, 1999
Dear Dr. Bryan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{9}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
INDICATIONS FOR USE
510(k) Number:
Device Name:
Velogene™ Rapid MRSA Identification Assay
Indications for Use:
The ID Biomedical Velogene™ Rapid MRSA Identification Assay is a qualitative DNA probe test which utilizes Cycling Probe™ Technology (CPT) to detect the mecA gene in isolated colonies of presumptively identified Staphylococcus aureus.
Woody Duboise
Laboratory Devices 510(k) Number
PRESCRIPTION For USE_X
ID Biomedical Corporation 8855 Northbrook Court, Burnaby, BC V5J 5J1 Canada Telephone: 604-431-9314 Fax: 604-431-9378
Doc Ref.: 109-99-2
§ 866.1640 Antimicrobial susceptibility test powder.
(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).