K Number
K990093
Date Cleared
1999-05-11

(120 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • Assessment of symptoms that may be related to rhythm disturbances of the heart: Patients with . palpitations; The evaluation of arrhythmia's in patients from pediatric to adult age.
  • . Assessment of risk in patients with or without symptoms of arrhythmia.
  • . Assessment of efficacy of Antiarrhythmic therapy.
  • . Assessment of Pacemaker Function.
  • . Assessment of real time ST segment analysis
  • . Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease and arrhythmia analysis during exercise testing.
  • . Assessment of transferred SpO2 level date from an MDE Escort Bedside monitor to EasiView for display and alarm.
  • . Assessment of transferred Invasive and Non-Invasive blood pressure, both systolic and diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm.
  • Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor to ● EasiView for display and alarm.
  • Assessment of transferred respiration data from an MDE Escort Bedside monitor to Easi View for display.
  • Assessment is for single-hospital environment.
Device Description

A Zymed Telemetry monitoring system consists of a series of interface devices to include ECG transmitters, multi-parameter transmitters, a central Telemetry monitoring computerized unit with a strip chart recorder, Easi 5 lead and laser printer. The Zymed central monitor supports up to eight patients for real time cardiac monitoring. The system displays each patient's ECG continuously on the screen while performing real time ECG waveform analysis for all eight patients. This analysis permits immediate detection and classification of abnormal beats, cardiac rhythm disturbances and variations. The system will also display each patient's Oxygen Saturation, Respiration, Non-Invasive Blood Pressure, Invasive Blood Pressure and Temperature.

Each ECG transmitter's frequency can be programmed to operate at any frequency within the entire VHF band. For US domestic sites, the transmitters will comply with FCC band allocations (174-216 Mhz). In addition to ECG data, the transmitters also detect and transmit cardiac pacemaker information. Other information including transmitter status and individual lead impedance is also transmitted to the Zymed system for overall system safety and efficacy.

The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for the diagnosis of patients with various heart conditions. The system also provides tools to review a patient's cardiac performance. On-line review mechanisms as well as detailed analysis screens have been designed into the system to facilitate and to enhance the patient's diagnosis and treatment. Features such as individual ECG printouts, multi-channel automatic ST analysis, trend data analysis, and Full Disclosure data further enhance the system's qualities as a valuable and practical clinical tool.

The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for multi-parameter monitoring of the patient. The Zymed system provides a means to monitor multiple patients from a central monitoring station for ECG, Oxygen Saturation, Non-Invasive Blood Pressure, Invasive Blood Pressure, Temperature, and Respiration.

The system has the following options available:

Choice of 4, 6, or 8 bed central monitor Full disclosure screen and printout (full resolution programmable from 0 to 168 hours) Choice of 6 lead sets, based on transmitter capability Full arrhythmia analysis to include multi-channel automatic ST Analysis Laser Printer, print server options Strip Chart Networking

AI/ML Overview

The provided text describes the Zymed Telemetry System: Model EasiView Telemetry Central Station Monitor and its performance evaluation for 510(k) clearance. Here's an analysis of the acceptance criteria and study as requested:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a quantitative manner for all metrics. However, it implicitly uses the performance of industry-accepted databases and comparison to predicate devices as a benchmark for equivalence. The reported device performance is presented as being "typical" and "nearly identical data" to the comparator.

Metric (Implicit Acceptance Criteria)Reported Device Performance
Arrhythmia Analysis
QRS Sensitivity (SE)"Typical for the real time monitoring environment"
QRS Positive Predictivity (+P)"Typical for the real time monitoring environment"
QRS False Positive Rate (FPR)"Typical for the real time monitoring environment"
Ventricular Sensitivity (SE)"Typical for the real time monitoring environment"
Ventricular Positive Predictivity (+P)"Typical for the real time monitoring environment"
Ventricular False Positive Rate (FPR)"Typical for the real time monitoring environment"
Couplets Sensitivity (SE)"Typical for the real time monitoring environment"
Couplets Positive Predictivity (+P)"Typical for the real time monitoring environment"
Couplets False Positive Rate (FPR)"Typical for the real time monitoring environment"
Short runs Sensitivity (SE)"Typical for the real time monitoring environment"
Short runs Positive Predictivity (+P)"Typical for the real time monitoring environment"
Short runs False Positive Rate (FPR)"Typical for the real time monitoring environment"
Long runs Sensitivity (SE)"Typical for the real time monitoring environment"
Long runs Positive Predictivity (+P)"Typical for the real time monitoring environment"
Long runs False Positive Rate (FPR)"Typical for the real time monitoring environment"
ST Analysis
Episode Sensitivity (ESE)"Typical for the real time monitoring environment"
Episode Positive Predictivity (E+P)"Typical for the real time monitoring environment"
Duration Sensitivity (DSE)"Typical for the real time monitoring environment"
Duration Positive Predictivity (D+P)"Typical for the real time monitoring environment"
High Heart Rates"Within recommended guidelines in excess of 300 bpm"
Noise Performance"Equal to or better than the old system when looking at baseline, electrode or muscle as the cause of noise."

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: The document does not explicitly state a specific "sample size" in terms of number of patients or recordings. It refers to the AHA (American Heart Association), MIT (Massachusetts Institute of Technology), and European ST-T (EST) databases. These are public databases of ECG recordings used for algorithm testing, implying a collection of pre-recorded data.
  • Data Provenance:
    • Country of Origin: Not specified in the document, but the names (AHA, MIT, European ST-T) suggest origins in the US and Europe.
    • Retrospective or Prospective: The use of pre-existing, established databases (AHA, MIT, EST) indicates a retrospective study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not provide information on the number or qualifications of experts used to establish the ground truth within the AHA, MIT, and EST databases. The ground truth for these databases is typically established by expert cardiologists or technicians during the creation of the databases themselves, often through manual annotation and consensus. This information is inherent to the databases, not detailed in this specific submission.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test set. Given the use of established databases, the ground truth labels within those databases would have been pre-adjudicated during their creation.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The study focuses on the standalone performance of the device's algorithms against established databases and a comparison of the device's technical characteristics and overall performance (e.g., noise handling, high heart rate) to a predicate device ("EasiView (Old) telemetry system" and "MDE Escort Monitor System (Old)"). There is no mention of human readers or AI assistance in the context of improving human reader performance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone (algorithm only) performance assessment was done. The performance evaluation was conducted against the AHA, MIT, and European ST-T databases, where the device's algorithms processed these pre-recorded ECGs and generated performance metrics (SE, +P, FPR, ESE, E+P, DSE, D+P) for arrhythmia and ST analysis.

7. The Type of Ground Truth Used

The ground truth used was based on expert consensus annotations within the established AHA, MIT, and European ST-T databases. These databases are recognized in the field for providing annotated ECG waveforms that serve as the "true" events against which algorithms are measured.

8. The Sample Size for the Training Set

The document does not specify the sample size for any training set. It only discusses the evaluation of the device's performance against the AHA, MIT, and EST databases, which are typically used for testing, not training, of algorithms after their development.

9. How the Ground Truth for the Training Set Was Established

Since the document does not mention the use of a distinct training set or its associated ground truth establishment, this information is not available in the provided text. The submission focuses on the evaluation of the already developed device.

{0}------------------------------------------------

MAY 1 1 1999

7 vmed

K990093

1

Zymed Inc. 20 North Aviador Street Camarillo, California 93010-8348 800.235.5941 - 805.987.9611 Fax 805.987.9532

510(k) Summary

Submitter:

Dudley Harris, Director of Regulatory Affairs/QA Zymed Medical Instrumentation 20 North Aviador Street Camarillo, CA 93010 805/987-9532 Fax: 800/235-5941 (401) Phone: Date of Summary: 1-8-99 D. Harris - see above Contact:

Trade Name: Common Name: Classification Name: (per 21 CFR 870.1025)

Zymed Telemetry System: Model EasiView Telemetry Central Station Monitor Detection and Alarm, Arrhythmia

Legally marketed device to which S.E. is claimed. Zymed Telemetry System: Model EasiView - 510(k) K 980186 Medical Data Electronics: Escort 100/300 Series B Patient Monitor System - 510(k) K930004/A

Description: A Zymed Telemetry monitoring system consists of a series of interface devices to include ECG transmitters, multi-parameter transmitters, a central Telemetry monitoring computerized unit with a strip chart recorder, Easi 5 lead and laser printer. The Zymed central monitor supports up to eight patients for real time cardiac monitoring. The system displays each patient's ECG continuously on the screen while performing real time ECG waveform analysis for all eight patients. This analysis permits immediate detection and classification of abnormal beats, cardiac rhythm disturbances and variations. The system will also display each patient's Oxygen Saturation, Respiration, Non-Invasive Blood Pressure, Invasive Blood Pressure and Temperature.

Each ECG transmitter's frequency can be programmed to operate at any frequency within the entire VHF band. For US domestic sites, the transmitters will comply with FCC band allocations (174-216 Mhz). In addition to ECG data, the transmitters also detect and transmit cardiac pacemaker information. Other information including transmitter status and individual lead impedance is also transmitted to the Zymed system for overall system safety and efficacy.

The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for the diagnosis of patients with various heart conditions. The system also provides tools to review a patient's

{1}------------------------------------------------

cardiac performance. On-line review mechanisms as well as detailed analysis screens have been designed into the system to facilitate and to enhance the patient's diagnosis and treatment. Features such as individual ECG printouts, multi-channel automatic ST analysis, trend data analysis, and Full Disclosure data further enhance the system's qualities as a valuable and practical clinical tool.

The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for multi-parameter monitoring of the patient. The Zymed system provides a means to monitor multiple patients from a central monitoring station for ECG, Oxygen Saturation, Non-Invasive Blood Pressure, Invasive Blood Pressure, Temperature, and Respiration.

The system has the following options available:

Choice of 4, 6, or 8 bed central monitor Full disclosure screen and printout (full resolution programmable from 0 to 168 hours) Choice of 6 lead sets, based on transmitter capability Full arrhythmia analysis to include multi-channel automatic ST Analysis Laser Printer, print server options Strip Chart Networking

Intended use:

  • Assessment of symptoms that may be related to rhythm disturbances of the heart: Patients with . palpitations; The evaluation of arrhythmia's in patients from pediatric to adult age.
  • . Assessment of risk in patients with or without symptoms of arrhythmia.
  • . Assessment of efficacy of Antiarrhythmic therapy.
  • . Assessment of Pacemaker Function.
  • . Assessment of real time ST segment analysis
  • . Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease and arrhythmia analysis during exercise testing.
  • . Assessment of transferred SpO2 level date from an MDE Escort Bedside monitor to EasiView for display and alarm.
  • . Assessment of transferred Invasive and Non-Invasive blood pressure, both systolic and diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm.
  • Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor to ● EasiView for display and alarm.
  • Assessment of transferred respiration data from an MDE Escort Bedside monitor to Easi View for display.
  • Assessment is for single-hospital environment.

A review of the technological characteristics compared to the predicate devices are:

{2}------------------------------------------------

PlatformEasiView (New)telemetry systemEasiView (Old)telemetry system
TypeIBM PC AT CompatibleSame
CPU200 Mhz Pentium ProSame
RAM64 M BytesSame
Hard Disk1.6 G BytesSame
DisplaySVGASame
Transmitters:
3 channelyesyes
Tunable frequencies174-216 MhzSame
Software:
Number of patients4, 6, or 8Same
Operating SystemWindows NTDOS CompatibleSame
Data Storage24 hrs of EKG/ChannelSame
Number of leads for Analysis:3Same
Automatic ST AnalysisyesSame
Manual ST Analysisnono
PlatformEasiView (New)telemetry systemMDE Escort Monitor System (Old)monitoring system
Temperature:
Range20-50 degrees CSame
Alarms0-150 F (0-65 C)0.1-44.0 C
Non-Invasive BP
Measurement MethodOscillometricSame
Alarms
Systolic0-300 mmHg25-240 mmHg
MeanNA15-200 mmHg
Diastolic0-200 mmHg10-180 mmHg
Invasive BP
Number of Channels1 or 2Same
Alarms
Systolic0-300 mmHg0-240 mmHg
MeanNA0-200 mmHg
Diastolic0-200 mmHg0-180 mmHg
SpO2
Range20- 100%Same
Alarms0-100% by 1%0-100% by 1 %70-100% high limit50-99% low limit
Respiration
Range4-200Same

Performance was measured against industry accepted AHA (AHA), MIT (MIT) and European ST-T (EST) databases. Results were typical for the real time monitoring environment for the Easi View as targeted. Separate sensitivities (SE), positive predictivity (+P), and false positive rate (FPR) were examined for each database and measured for QRS, Ventricular, Couplets, Short runs and Long runs. Separate Episode

{3}------------------------------------------------

Sensitivities (ESE), Episode Positive Predictivity (E+P), Duration Sensitivity (DSE) and Durative Predictivity (D+P) were examined for the European ST-T(EST) database and measured for ST analysis. High heart rates to include pediatric patients were demonstrated to be within recommended guidelines in excess of 300 bpm, and performance in the presence of noise indicates the new system is equal to or better than the old system when looking at baseline, electrode or muscle as the cause of noise.

In summary, performance data between the two systems shows nearly identical data, and therefore, supports a claim of Substantial Equivalence.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 1 1999

Mr. Dudley Harris Director of RA/QA Zymed, Inc. 20 North Aviador Street Camarillo, CA 93010-8348

Re: K990093 Easiview Telemetry Central Station Monitor Regulatory Class: III (three) Product Code: DSI Dated: April 20, 1999 Received: April 21, 1999

Dear Mr. Harris:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food Drug and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this

{5}------------------------------------------------

response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2Misbranding by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J.Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number: K990093

Device Name: EasiView Telemetry Monitoring System

Indications for Use:

  • . Assessment of symptoms that may be related to rhythm disturbances of the heart; Patients with palpitations: The evaluations of arrhythmia's in patients from pediatric to adult age.
  • Assessment of risk in patients with or without symptoms of arrhythmia. .
  • Assessment of efficacy of Antiarrhythmia therapy. .
  • Assessment of Pacemaker Function.
  • . Assessment of real time ST segment analysis.
  • Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease . and arrhythmias analysis during exercise testing.
  • Assessment of transferred SpO2 level data from an MDE Escort Bedside monitor to ● EasiView for display and alarm.
  • Assessment of transferred invasive and Non-invasive blood pressure, both systolic and . diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm.
  • Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor . to EasiView for display and alarm.
  • Assessment of transferred respiration data from an MDE Escort Bedside monitor to EasiView . for display.
  • Assessment is indicated for single-hospital environment.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

F.K. Jones for the

Prescription Use
(CFR21 CFR 801.109)

Over-The-Counter Use

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.