(11 days)
No
The document describes a new RF coil for an existing MRI system and mentions standard image processing techniques, but there is no mention of AI, ML, or related concepts like deep learning, training sets, or performance metrics typically associated with AI/ML devices.
No
The device is described as an "imaging device" and "Diagnostic Device" intended to provide "information that can be useful in diagnosis determination," indicating it is for diagnostic purposes, not therapeutic.
Yes
The "Intended Use / Indications for Use" section states: "When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination." Additionally, the "Device Description" section explicitly refers to the device as "Magnetic Resonance Diagnostic Device."
No
The device description explicitly states it is an "additional RF Coil (Shoulder Coil)" which is a hardware component for an MRI system. The software is mentioned as not needing revision to support the coil, indicating the coil itself is the primary enhancement.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The description clearly states that the device is an MR system and a Shoulder Coil used for imaging. It produces images of internal structures based on magnetic resonance properties.
- Intended Use: The intended use is to provide the physician with physiological and clinical information obtained non-invasively and without the use of ionizing radiation, through the creation of images.
- No Sample Analysis: There is no mention of analyzing samples taken from the body. The device directly interacts with the patient's body to generate images.
Therefore, the QD Shoulder Coil and the AIRIS II MRI system are imaging devices used for diagnostic purposes, but they do not fall under the definition of in vitro diagnostics.
N/A
Intended Use / Indications for Use
The QD Shoulder Coil provides imaging of the shoulder articular anatomy and other large ioints.
The MR system is an imaging device, and is intended to provide the physician with r ho "ift" of count" information, obtained non-invasively and without the use of ionizing prodiation. The MR system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Anatomical Region: Head, Body, Spine, Extremities
Nucleus excited: Proton
Diagnostic uses:
2D T1- / T2-weighted imaging
T1. T2. proton density measurements
MR Angiography
image processing
Imaging capabilities:
2D Spin Echo (SE); also with rephase
2D Gradient Field Echo (GE); also with rephase (2D GR)
2D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
2D Inversion Recovery (IR)
2D Fast Spin Echo (FSE); also with rephase
2D Fast Inversion Recovery (FIR)
2D Dual Slice acquisition (SEDS)
3D Spin Echo (SE)
3D Gradient Field Echo (GE)
3D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
MR Angiography (2D TOF, 3D TOF, half echo, high resolution/high definition, sloped slab profile, magnetization transfer contrast)
MR Fluoroscopy
RF Coil Uniformity
Adaptive Image post-processing
ACR/NEMA/DICOM 3 compliant
Product codes
90LNH, 90MOS
Device Description
FUNCTION: Identical to the AIRIS II 510(k) K974212.
The AIRIS II MRI Magnetic Resonance Diagnostic Device is being enhanced by one additional RF Coil (Shoulder Coil) to increase the clinical utility of the AIRIS II in the stationary configuration. The Shoulder Coil can operate either in the quadrature mode or in the phased array mode with the AIRIS II.
AIRIS II software did not need to be revised in order to support full functionality of these coils.
SCIENTIFIC CONCEPTS: Magnetic Resonance (MR) is based on the fact that certain atomic nuclei have electromagnetic properties which cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nucleus used in current imaging experiments in magnetic resonance. When placed in a magnetic field, there is a slight net orientation or alignment of these atomic nuclei with the magnetic field. The introduction of a short burst of radiofrequency (RF) excitation of wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a reorientation of the proton's magnetization vector. When the RF excitation is removed, the proton relaxes and returns to its original orientation. The rate of relaxation is exponential, and varies with the character of the proton and its adjacent molecular environment. This reorientation process is characterized by two exponential relaxation times called T1 and T2 which can be measured.
These relaxation events are accompanied by an RF emission or echo which can be measured and used to develop a representation of these emissions on a three dimensional matrix. Spatial localization is encoded into the echo by varying the RF excitation and by appropriately applying magnetic field gradients in x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of NMR characteristics of the nuclei under consideration can be constructed by using image processing techniques similar to those used in CT.
For magnetic fields up to 1.5T, the RF frequencies commonly used range up to 65MHz. The RF fields have pulse powers from several watts to greater than 10 kilowatts, and repeat at rates from once every few seconds to greater than fifty per second. The timevarying magnetic gradient fields have a typical duration of sub-millisecond to several milliseconds.
PHYSICAL AND PERFORMANCE CHARACTERISTICS: MR is currently of great interest because it is capable of producing high quality anatomical images without the associated risks of ionizing radiation. In addition, the biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In x-ray imaging, differences in x-ray attenuation, largely based on differences in electro density are responsible for the contrast observed in x-ray images. In MR imaging, differences in proton density, blood flow, and relaxation times T1 and T2 all may contribute to image contrast. In addition, by varying the duration and spacing of the RF pulses, images may be produced in which the contrast is primarily dependent on T1 relaxation, T2 relaxation, proton density, or a combination of all three.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Magnetic Resonance (MR)
Anatomical Site
Head, Body, Spine, Extremities, shoulder articular anatomy, other large joints
Indicated Patient Age Range
Not Found
Intended User / Care Setting
physician
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Hitachi AIRIS II with Shaped Shoulder Coil
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
DEC | | 1998
Attachment 1 510(k) Summary of Safety and Effectiveness
1
SUBMITTER INFORMATION: 1.0
- Hitachi Medical Systems America Submitter: 1.1 1959 Summit Commerce Park Twinsburg, OH 44087 PH: 330 425-1313 FX: 330 425-1410
- James Jochen Rogers 1.2 Contact:
- November 19, 1998 Date: 1.3
DEVICE NAME: 2.0
- Radiology 2.1 Classification Panel:
- 892.1000 Magnetic Resonance Diagnostic Device 2.2 Classification Number:
- System, Nuclear Magnetic Resonance Imaging 2.3 Product Nomenclature:
90LNH 2.4 Product Code(s):
- 90MOS (Magnetic Resonance Specialty Coil)
- AIRIS II Trade/Proprietary Name: 2.5
- 2.6 PREDICATE DEVICE(s):
Hitachi AIRIS II with Shaped Shoulder Coil
3.0 DEVICE DESCRIPTION:
- FUNCTION 3.1
Identical to the AIRIS II 510(k) K974212.
The AIRIS II MRI Magnetic Resonance Diagnostic Device is being enhanced by one additional RF Coil (Shoulder Coil) to increase the clinical utility of the AIRIS II in the stationary configuration. The Shoulder Coil can operate either in the quadrature mode or in the phased array mode with the AIRIS II.
AIRIS II software did not need to be revised in order to support full functionality of these coils.
3.2 SCIENTIFIC CONCEPTS
Magnetic Resonance (MR) is based on the fact that certain atomic nuclei have electromagnetic properties which cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nucleus used in current imaging experiments in magnetic resonance. When placed in a magnetic field, there is a slight net orientation or alignment of these atomic nuclei with the magnetic field. The introduction of a short burst of radiofrequency (RF) excitation of wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a reorientation of the proton's magnetization vector. When the RF excitation is removed, the proton relaxes and returns to its original orientation. The rate of relaxation is exponential, and varies with the character of the proton and its adjacent molecular environment. This reorientation process is characterized by two exponential relaxation times called T1 and T2 which can be measured.
2
These relaxation events are accompanied by an RF emission or echo which can be measured and used to develop a representation of these emissions on a three dimensional matrix. Spatial localization is encoded into the echo by varying the RF excitation and by appropriately applying magnetic field gradients in x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of NMR characteristics of the nuclei under consideration can be constructed by using image processing techniques similar to those used in CT.
For magnetic fields up to 1.5T, the RF frequencies commonly used range up to 65MHz. The RF fields have pulse powers from several watts to greater than 10 kilowatts, and repeat at rates from once every few seconds to greater than fifty per second. The timevarying magnetic gradient fields have a typical duration of sub-millisecond to several milliseconds.
PHYSICAL AND PERFORMANCE CHARACTERISTICS 3.3
MR is currently of great interest because it is capable of producing high quality anatomical images without the associated risks of ionizing radiation. In addition, the biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In x-ray imaging, differences in x-ray attenuation, largely based on differences in electro density are responsible for the contrast observed in x-ray images. In MR imaging, differences in proton density, blood flow, and relaxation times T1 and T2 all may contribute to image contrast. In addition, by varying the duration and spacing of the RF pulses, images may be produced in which the contrast is primarily dependent on T1 relaxation, T2 relaxation, proton density, or a combination of all three.
DEVICE INTENDED USE: 4.0
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician. these images provide information that can be useful in diagnosis determination.
- Anatomical Region: Head, Body, Spine, Extremities ●
- Nucleus excited: Proton
- . Diagnostic uses: 2D T1- / T2-weighted imaging
- T1. T2. proton density measurements
MR Angiography
image processing
Imaging capabilities:
- 2D Spin Echo (SE); also with rephase
- 2D Gradient Field Echo (GE); also with rephase (2D GR)
- 2D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
- 2D Inversion Recovery (IR)
- 2D Fast Spin Echo (FSE): also with rephase
- 2D Fast Inversion Recovery (FIR)
- 2D Dual Slice acquisition (SEDS)
- 3D Spin Echo (SE)
3
3D Gradient Field Echo (GE) 3D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing MR Angiography (2D TOF, 3D TOF, half echo, high resolution/high definition, sloped slab profile, magnetization transfer contrast) MR Fluoroscopy RF Coil Uniformity Adaptive Image post-processing ACR/NEMA/DICOM 3 compliant
DEVICE TECHNOLOGICAL CHARACTERISTICS: 5.0
Identical to the Predicate Device.
4
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 11 1998
James Jochen Rogers Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park Twinsburg, Ohio 44087
K984278 Re:
QD Multiple Array Shoulder Coil Part #MR-MSC-51Q/M Dated: November 3, 1998 Received: November 30, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS
Dear Mr. Rogers:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitto diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.ida.gov/cdrh/dsmaldsmam.html".
Sincerely yours,
Lillian Yiz, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
510(k) Number (if known):
Device Name: Additional RF Coils for AIRIS II (Shoulder Coil)
Indications for Use:
The QD Shoulder Coil provides imaging of the shoulder articular anatomy and other large ioints.
The MR system is an imaging device, and is intended to provide the physician with r ho "ift" of count" information, obtained non-invasively and without the use of ionizing prodiation. The MR system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
- Anatomical Region: Head, Body, Spine, Extremities
- Proton Nucleus excited:
- Diagnostic uses:
2D T1- / T2-weighted imaging T1. T2. proton density measurements MR Angiography image processing
· Imaging capabilities:
2D Spin Echo (SE); also with rephase
2D Gradient Field Echo (GE); also with rephase (2D GR)
-
2D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
2D Inversion Recovery (IR) -
2D Fast Spin Echo (FSE); also with rephase
-
2D Fast Inversion Recovery (FIR)
-
2D Dual Slice acquisition (SEDS)
-
3D Spin Echo (SE)
-
3D Gradient Field Echo (GE)
-
3D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
-
MR Angiography (2D TOF, 3D TOF, half echo, high resolution/high definition, sloped slab profile, magnetization transfer contrast)
-
MR Fluoroscopy
RF Coil Uniformity
Adaptive Image post-processing
ACR/NEMA/DICOM 3 compliant
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801-109
he Counter Use
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K984278
1970 THORTOON THORTIN MAGS: 20 EC BC 77 950