K Number
K983395
Device Name
FETALGARD 3000 FETAL MONITOR
Manufacturer
Date Cleared
1999-08-12

(321 days)

Product Code
Regulation Number
884.2740
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements. This data is intended to aid in assessing the well-being of the fetus during pregnancy (antepartum), and labor and delivery (intrapartum). The modifications described in this submission do not implement any changes in intended use.
Device Description
The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements. Uterine contractions are monitored using an external tocotonometer or an intrauterine pressure transducer. Fetal heart rate is measured using an external pulsed Doppler ultrasound transducer or directly with a spiral scalp electrode. Maternal heart rate and respiration are measured using standard ECG electrodes. Heart rate, respiration rate, and uterine activity are presented graphically on an LCD display or chart recorder and digitally on LED displays.
More Information

Not Found

No
The description focuses on standard physiological monitoring and graphical display, with no mention of AI/ML terms or capabilities.

No
The device is described as a monitoring system that displays data to aid in assessing fetal well-being, not to treat or cure a condition.

Yes
The device monitors and displays physiological data (maternal labor, fetal heart rate, uterine contractions, maternal heart rate, respiration) with the stated intention to "aid in assessing the well-being of the fetus," which is an assessment of a medical condition.

No

The device description explicitly mentions hardware components such as transducers, electrodes, LCD displays, and chart recorders, indicating it is not solely software.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. These tests are performed outside of the body (in vitro).
  • Fetalgard 3000 Function: The Fetalgard 3000 Fetal Monitor measures physiological signals directly from the patient's body (maternal labor, fetal heart rate, maternal heart rate, respiration). It uses external transducers and electrodes placed on the body, or an intrauterine pressure transducer and a spiral scalp electrode.
  • No Sample Analysis: The device does not analyze samples taken from the body. It is a monitoring system that captures and displays real-time physiological data.

Therefore, based on the provided information, the Fetalgard 3000 Fetal Monitor is a monitoring device that operates in vivo (within the living body), not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements. This data is intended to aid in assessing the well-being of the fetus during pregnancy (antepartum), and labor and delivery (intrapartum). The modifications described in this submission do not implement any changes in intended use.

Product codes (comma separated list FDA assigned to the subject device)

85 HGM

Device Description

The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements. Uterine contractions are monitored using an external tocotonometer or an intrauterine pressure transducer. Fetal heart rate is measured using an external pulsed Doppler ultrasound transducer or directly with a spiral scalp electrode. Maternal heart rate and respiration are measured using standard ECG electrodes. Heart rate, respiration rate, and uterine activity are presented graphically on an LCD display or chart recorder and digitally on LED displays.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Pulsed Doppler ultrasound

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The modifications to the device have undergone a subset of the original verification and validation test scenarios to show that all functions that may have been affected by the modifications are still in compliance with the original user requirements, and that safety and performance have not been adversely affected. This involved the use of both simulated input signals and tape recordings of actual physiological waveforms.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K950420

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 884.2740 Perinatal monitoring system and accessories.

(a)
Identification. A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.(b)
Classification. Class II (performance standards).

0

AUG 12 1999

510(k) Summary for Analogic Corporation Modifications to Fetalgard 3000 Fetal Monitor

DATE THIS SUMMARY WAS PREPARED: September 11, 1998

SUBMITTER'S NAME AND ADDRESS:

Analogic Corporation 8 Centennial Drive Peabody, MA 01960

CONTACT PERSON:

Steven Clarke, Regulatory Specialist
Telephone(978)977-3000 extension 2388
Facsimile(781)245-1274

DEVICE NAME:

Proprietary Name:Fetalgard 3000 Fetal Monitor (Modification to
Common Name:Perinatal Monitoring System
Classification Name:Perinatal Monitoring System and Accessories

PREDICATE DEVICE:

The legally marketed device to which equivalence is being claimed is:

Fetalgard 3000 Fetal Monitor

DEVICE DESCRIPTION:

.

The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements.

K98395

1

Uterine contractions are monitored using an external tocotonometer or an intrauterine pressure transducer.

Fetal heart rate is measured using an external pulsed Doppler ultrasound transducer or directly with a spiral scalp electrode.

Maternal heart rate and respiration are measured using standard ECG electrodes.

Heart rate, respiration rate, and uterine activity are presented graphically on an LCD display or chart recorder and digitally on LED displays.

INTENDED USE:

The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements. This data is intended to aid in assessing the well-being of the fetus during pregnancy (antepartum), and labor and delivery (intrapartum). The modifications described in this submission do not implement any changes in intended use.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:

The modified Fetalgard 3000 Fetal Monitor is nearly identical to the predicate Fetalgard 3000 Fetal Monitor which was cleared for marketing under Premarket Notification K950420.

The tocotonometer has been redesigned using the same principle of operation. with changes in the design details to enhance reliability and improve ease of manufacture.

The top housing of the ultrasound transducer will be manufactured out of a different material, which is equivalent in terms of mechanical properties and biocompatibility.

The system software has undergone several evolutionary revisions to improve the detection and display of fetal heart rate under adverse conditions.

2

NONCLINICAL TESTS USED IN DETERMINATION OF SUBSTANTIAL EQUIVALENCE:

The modifications to the device have undergone a subset of the original verification and validation test scenarios to show that all functions that may have been affected by the modifications are still in compliance with the original user requirements, and that safety and performance have not been adversely affected. This involved the use of both simulated input signals and tape recordings of actual physiological waveforms.

CONCLUSIONS FROM NONCLINICAL TESTING:

The testing of the modified Fetalgard 3000 Fetal Monitor demonstrates the performance is substantially equivalent to the predicate device.

3

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 12 1999

Mr. Steven A. Clarke, RAC Regulatory Affairs Manager Analogic Corporation 8 Centennial Drive Peabody, MA 01960

Re: K983395 Fetalguard 3000 Fetal Monitor Dated: July 29, 1999 Received: July 30, 1999 Regulatory Class: II 21 CFR §884.2740/Procode: 85 HGM

Dear Mr. Clarke:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), piease contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

[signature]

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

510(k) Number (if known): 1

Device Name: _ Fetalgard 3000 Fetal Monitor

Indications For Use:

The Fetalgard 3000 Fetal Monitor is a Perinatal Monitoring System for showing graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating display of uterine contraction and fetal heart rate measurements. This data is intended to aid in assessing the well-being of the fetus during pregnancy, labor, and delivery.

David C. Seppam
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109)

Analogic Corporation Fetalgard 3000 510(k)