K Number
K982696

Validate with FDA (Live)

Device Name
INTERPLANT
Date Cleared
1999-04-01

(241 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Burdette Medical Systems, Inc. Interplant™ system is used in planning and performing a prostate brachytherapy implant procedure to provide ultrasound probe alignment and radioactive seed implantation. The Interplant™ device will use standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds for the treatment of prostate cancer. The system will allow volume measurement and dose calculation, relevant prostate anatomy and dose visualization, and post-treatment seed localization.

Device Description

The Interplant™ system for brachytherapy was developed to address the need for a comprehensive system and procedure for prostate seed implants. The Interplant in system compresses the requirement for separate multiple procedures into one online setting, This product integrates all of the delivery system components (needles, template guide, probe stepper) with the dose planning system for 125-Iodine (125 I) and 103-Paladium (103 Pd) implants. The system is designed for real-time dose planning in the Operating Room (OR) and for performing post-implant seed localization assessment. Real time ultrasound image plane acquisition with automatic probe position determination makes it possible to complete the image acquisition needed for dose planning within about a minute. The interplant™ system provides the needed components for seed implant procedures except the seeds themselves and uses existing ultrasound imaging equipment. This will enable the physician to focus on the procedure, providing immediate feedback for improved quality assurance.

The Real-Time Brachytherapy Radiation Seed Implant Program and System includes all technology, computer software, and hardware, and methods associated with the implant of radioactive "seeds" or pellets in the prostate. System elements include the user interface, real-time image acquisition, image contouring, dose calculation and display software, dose volume histograms, isodose surface contours, post-implant seed localization, and the patient scheduling spreadsheet software.

The System consists of a two-dimensional and three-dimensional image visualization and brachytherapy dose system employing 2D ultrasound imaging for use in radioactive seed implants of the prostate. The software for the brachytherapy seed implant and dose calculation system was developed on a Pentium-based processor with supporting graphics and digitizing hardware.

AI/ML Overview

This 510(k) summary does not contain information on acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) submission for Burdette Medical Systems, Inc.'s Interplant™ system, which is a brachytherapy dose planning and implant system. It primarily focuses on demonstrating substantial equivalence to already marketed predicate devices.

Here's what can be extracted from the provided text, and what is missing:

Missing Information:

The provided text does not contain any of the following information:

  • A table of acceptance criteria and reported device performance.
  • Sample size used for the test set.
  • Data provenance (country of origin, retrospective/prospective).
  • Number of experts used to establish ground truth for the test set and their qualifications.
  • Adjudication method for the test set.
  • Whether a multi-reader, multi-case (MRMC) comparative effectiveness study was done.
  • Effect size of human reader improvement with AI vs. without AI assistance.
  • Whether a standalone (algorithm only) performance study was done.
  • Type of ground truth used (expert consensus, pathology, outcomes data).
  • Sample size for the training set.
  • How the ground truth for the training set was established.

Information Extracted from the Document:

While the core request about acceptance criteria and a study proving their fulfillment cannot be answered from this document, here's what the document does provide:

  • Device Name: Interplant™
  • Intended Use: To provide precision ultrasound probe alignment and radioactive seed implantation in the treatment of prostate cancer. The system allows volume and dose calculation, image visualization, and post-treatment seed localization. It uses standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds.
  • Technological Characteristics:
    • Integrates delivery system components (needles, template guide, probe stepper) with a dose planning system for 125-I and 103-Pd implants.
    • Designed for real-time dose planning in the Operating Room (OR) and post-implant seed localization assessment.
    • Real-time ultrasound image plane acquisition with automatic probe position determination allows image acquisition for dose planning within about a minute.
    • Includes a user interface, real-time image acquisition, image contouring, dose calculation and display software, dose volume histograms, isodose surface contours, post-implant seed localization, and patient scheduling spreadsheet software.
    • Employs 2D and 3D image visualization and brachytherapy dose system using 2D ultrasound imaging for prostate seed implants.
    • Software developed on a Pentium-based processor with supporting graphics and digitizing hardware.
  • Regulatory Classification: Class II, 21 CFR 892.5730/Procode: 90 MUJ (Medical charged-particle radiation therapy system).
  • Predicate Devices:
    1. Multimedia Medical Systems, Inc. B3DTUA TherapacPLUS dose planning
    2. SSGI Prowess 500 and Prowess 2000 dose planning K924240
    3. CIVCO Medical Instruments Ultra-Step stepper K875241
    4. B&K Medical Systems HP-100 stepper/stand
    5. Tayman Medical Accuseed stand/stabilizer K963302
    6. COTAN stand/stabilizer
    7. Northwest stand/stabilizer
    8. Maroon Bels, Inc. Brachystand and Brachystepper K972672
  • Submission Type: 510(k) Premarket Notification, seeking substantial equivalence.

Conclusion:

This document is a regulatory submission for premarket notification (510(k)), not a study report. The primary goal of a 510(k) is to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than to present detailed clinical study results with acceptance criteria. Therefore, the specific information requested about acceptance criteria and a study proving performance against them is not typically found in this type of summary. Such information would be expected in a more comprehensive clinical or validation study report, which is usually referenced or summarized within a 510(k) but not fully detailed.

{0}------------------------------------------------

4/1/99

Burdette Medical Systems, Inc. 206 N. Randolph Street, Suite 301 Champaign, IL 61820 Tel: (217) 239-0900 Fax: (217) 239-0905

K982694

510(k) Summary

Submitter: Burdette Medical Systems, Inc.

Address: 206 N. Randolph Suite 301 Champagne IL 61820

Phone number: (217) 239-0900

Fax number: (217) 239-0905

Contact person: Everette C. Burdette, Ph.D.

Date prepared: July 27, 1998

Trade name: Interplant™

Common name: Brachytherapy dose planning and implant system

Classification name: 892.5050 Medical charged-particle radiation therapy system Class II

Substantial equivalence claimed to:

    1. Multimedia Medical Systems, Inc. B3DTUA TherapacPLUS dose planning
    1. SSGI Prowess 500 and Prowess 2000 dose planning K924240
    1. CIVCO Medical Instruments Ultra-Step stepper K875241
    1. B&K Medical Systems HP-100 stepper/stand
    1. Tayman Medical Accuseed stand/stabilizer K963302
    1. COTAN stand/stabilizer
    1. Northwest stand/stabilizer
    1. Maroon Bels, Inc. Brachystand and Brachystepper K972672

Description:

The Interplant™ system for brachytherapy was developed to address the need for a comprehensive system and procedure for prostate seed implants. The Interplant in system compresses the requirement for separate multiple procedures into one online setting,

This product integrates all of the delivery system components (needles, template guide, probe stepper) with the dose planning system for 125-Iodine (125 I) and 103-Paladium (103 Pd) implants. The system is designed for real-time dose planning in the Operating Room

{1}------------------------------------------------

(OR) and for performing post-implant seed localization assessment. Real time ultrasound image plane acquisition with automatic probe position determination makes it possible to complete the image acquisition needed for dose planning within about a minute. The interplant™ system provides the needed components for seed implant procedures except the seeds themselves and uses existing ultrasound imaging equipment. This will enable the physician to focus on the procedure, providing immediate feedback for improved quality assurance.

Intended use:

The Burdette Medical Systems, Inc. interplant™ system is used to provide precision ultrasound probe alignment and radioactive seed implantation in the treatment of prostate cancer. The system will allow volume and dose calculation, image visualization, and posttreatment seed localization.

Summary of technological characteristics:

The Real-Time Brachytherapy Radiation Seed Implant Program and System includes all technology, computer software, and hardware, and methods associated with the implant of radioactive "seeds" or pellets in the prostate. System elements include the user interface, real-time image acquisition, image contouring, dose calculation and display software, dose volume histograms, isodose surface contours, post-implant seed localization, and the patient scheduling spreadsheet software.

The System consists of a two-dimensional and three-dimensional image visualization and brachytherapy dose system employing 2D ultrasound imaging for use in radioactive seed implants of the prostate. The software for the brachytherapy seed implant and dose calculation system was developed on a Pentium-based processor with supporting graphics and digitizing hardware.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 1999

Everette C. Burdette, Ph.D. President Burdette Medical Systems 206 N. Randolph, Suite 301 Champaign, IL 61820

Re:

K982696 Interplant™ (Brachytherapy Dose Planning and Implant System) Dated: February 14, 1999 Received: February 24, 1999 Regulatory class: II 21 CFR 892.5730/Procode: 90 MUJ

Dear Dr. Burdette:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requilations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal register , Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal. Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

Device Name: Interplant™

Indications for Use:

The Burdette Medical Systems, Inc. Interplant™ system is used in planning and performing a prostate brachytherapy implant procedure to provide ultrasound probe alignment and radioactive seed implantation. The Interplant™ device will use standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds for the treatment of prostate cancer. The system will allow volume measurement and dose calculation, relevant prostate anatomy and dose visualization, and post-treatment seed localization.

n982696

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use __

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982696

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.