K Number
K982586
Date Cleared
1998-08-11

(18 days)

Product Code
Regulation Number
878.4460
Panel
HO
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free Surgeon's glove is a sterile powder free medical device intended to be worn I oncer free bargen ersonnel to protect a surgical wound from contamination.

Device Description

Class I Powder free Surgical Glove 79KGO that meets all the requirements of ASTM D3577.

AI/ML Overview

Here's a summary of the acceptance criteria and the study for the Shiva Medicare Limited Powder-Free Surgeon's Glove, as derived from the provided document:

This document describes the performance of a medical device, specifically a "Powder free Surgeon's Glove" manufactured by Shiva Medicare Limited, and its adherence to the ASTM D3577 standard. The primary goal of the submission appears to be demonstrating substantial equivalence to a legally marketed predicate device (Class I Surgeon's Glove (Powder free) 79KGO that meets all the requirements of ASTM D3577).

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicAcceptance Criteria (ASTM D3577 Requirement)Reported Device Performance (SML Value)Meets Criteria?
Dimensions
Length (various sizes)245 mm minimum (for 5.5) / 265 mm minimum270-272 mmYes
Width (various sizes)+/- 6 mm of specified nominal width68-112 mm (within tolerance for each size)Yes
Thickness (Cuff)0.1 mm minimum0.12 mmYes
Thickness (Palm)0.1 mm minimum0.16 mmYes
Thickness (Fingertip)0.1 mm minimum0.19 mmYes
Physical Properties (Before Ageing)
Tensile Strength24 mpa27 mpaYes (Exceeds)
Elongation at break %750%850%Yes (Exceeds)
Modulus at 500% elongation5.5 mpa (max)3 mpaYes (Below max)
Physical Properties (After Ageing)
Tensile Strength18 mpa min20 mpaYes (Exceeds)
Elongation at break %560% min750%Yes (Exceeds)
Performance Requirements
SterilityAs per USP*As per IP* (Indian Pharmacopea)Yes (Equivalent standard used)
Freedom from HolesS4 (AQL 1.5)S4 (AQL 1.5)Yes
DimensionS2 (AQL 4)S2 (AQL 4)Yes
Physical PropertyS2 (AQL 4)S2 (AQL 4)Yes
Powder ContentNot explicitly stated in D3577

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).