(24 days)
HEALTH + AID® PREMIUM Pre-powdered Latex Examination Glove is a single use device intended for medical purposes that is worn on the hand of health care and similar personnel to prevent contamination between the health care personnel and the patient.
100mcg/gm or less total water extractable protein per gram
HEALTH + AID® PREMIUM Pre-powdered Latex Examination Gloves meet the current specifications listed under the ASTM Specification D 3578 - 95, Standard Specification for Rubber Examination Gloves. They are
This document describes the 510(k) summary for HEALTH + AID® PREMIUM Pre-powdered Latex Examination Gloves.
Here's an analysis of the provided text with respect to your specific questions:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Meets ASTM D 3578-95, Standard Specification for Rubber Examination Gloves | Gloved meet all the current specifications listed under the ASTM Specification D 3578-95, Standard Specification for Rubber Examination Gloves. |
| Meets 21 CFR 800.20 | Gloves meet all the current specifications listed under 21 CFR 800.20 |
| Primary skin irritation testing in the rabbit | No irritation |
| Delayed contact sensitization testing in the guinea pig | No sensitization |
| Total Water Extractable Protein | 100mcg/gm or less total water extractable protein per gram (this is an indication for use, not specifically an acceptance criteria for performance against the predicate, but it is a claim being made for the device) |
2. Sample size(s) used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes used for testing. It refers to "Testing is performed as per ASTM D 3578-95 and 21 CFR 800.20." These standards, particularly ASTM D 3578-95, would define the sampling plans and testing methodologies.
The provenance of the data is not specified beyond being generated by the device manufacturer, M.R.G. Industries Sdn. Bhd., located in Malaysia. It is implied to be prospective testing conducted by the manufacturer for regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as the device is a medical glove, not a diagnostic or AI-driven decision support system requiring expert-established ground truth. The acceptance is based on physical and biological testing against established standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for the type of device and testing described. Adjudication methods are typically relevant for subjective assessments, particularly in clinical trials or AI performance evaluations, not for objective material property testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical glove, not an AI or imaging device that would involve human readers or MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a medical glove, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device is based on objective measurements against established performance standards and biocompatibility tests. For example:
- Physical properties (e.g., tensile strength, elongation, dimensions) are measured against criteria specified in ASTM D 3578-95.
- Freedom from holes is tested per ASTM D 3578-95.
- Biocompatibility (skin irritation and sensitization) is assessed through animal studies (rabbit and guinea pig).
- Water extractable protein is measured objectively.
8. The sample size for the training set
Not applicable. The device is a physical product (glove) and does not involve AI or machine learning algorithms that require a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
3
510(k) SUMMARY
HEALTH + AID® PREMIUM PRE-POWDERED FRÉE LATEX EXAMINATION GLOVES
| Submitter's Name : | M.R.G.Industries Sdn. Bhd. |
|---|---|
| Submitter's Address : | PT 4004, Kamunting Industrial Estate34600 Taiping PerakMalaysia |
| Submitter's Phone Number | 605 891 5555 |
| Submitter 's Fax Number : | 605 891 2688 |
| Name of Contact Person : | Goh, See Khek |
| Date of Preparation : | May 23, 1998 |
| Name of Device : | |
| Trade Name : | HEALTH + AID® PREMIUM PRE-POWDEREDLATEX EXAMINATION GLOVES |
| Common Name : | Latex examination gloves |
| Classification Name : | Patient Examination Gloves |
| Legally Marketed Device to WhichEquivalency is Being Claimed : | HEALTH + AID® PREMIUM Pre-powderedLatex Examination Gloves as described in the510(k) notification are substantially equivalent tothe Class 1 patient examination glove 80LYY. Itmeets all the current specifications listed underthe ASTM Specification D 3578 - 95, StandardSpecification for Rubber Examination Gloves. |
| Description of the Device : | HEALTH + AID® PREMIUM Pre-powderedLatex Examination Gloves meet the currentspecifications listed under the ASTMSpecification D 3578 - 95, Standard Specificationfor Rubber Examination Gloves. They are |
| Intended Use of the Device: | HEALTH + AID® PREMIUM Pre-powderedLatex Examination Gloves are intended for singleuse for medical purposes and are worn on thehand of health care and similar personnel toprevent contamination between the health carepersonnel and the patients. |
| Summary of TechnologicalCharacteristics Compared to the PredicateDevice : | There are no different technologicalcharacteristics. Gloves are made from naturalrubber compound and the initial products arepowdered natural latex examination gloves. |
| Brief Discussion of Nonclinical Tests : | Testing is performed as per ASTM D 3578-95and 21 CFR 800.20. Gloves meet all the currentspecifications listed under the ASTMSpecification D 3578 - 95, Standard Specificationfor Rubber Examination Gloves.Primary skin irritation testing in the rabbit anddelayed contact sensitization testing in the guineapig indicate no irritation or sensitization. |
| Brief Discussion of Clinical Tests : | No new clinical tests were conducted under this510(k). |
| Conclusions Drawn for the Nonclinicaland Clinical Tests : | Nonclinical laboratory and animal data indicatethat the powder free product meets allperformance and biocompatability requirements. |
| Other Information Deemed Necessary byFDA : | Not applicable |
{1}------------------------------------------------
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 22 1998
Mr. See Khek Goh ·M.R.G. Industries Sdn. Bhd. PT 4004, Kamunting Industrial Estate P.O. Box 9 34600 Taiping Perak MALAYSIA
Re : K981893 Health + Aid® Pre-Powdered Latex Trade Name: Examination Gloves with 100 mcgm or Less Total Water Extractable Protein Regulatory Class: I Product Code: LYY Dated: May 29, 1998 Received: May 29, 1998
Dear Mr. See Khek Goh:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{3}------------------------------------------------
Page 2 - Mr. See Khek Goh
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA described in your sie in promalence of your device to a legally marketed predicate device results in a classification for your marketed predicate actros your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fday.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{4}------------------------------------------------
Attachment 2
INDICATIONS FOR USE
| Applicant | : | M.R.G. INDUSTRIES SDN. BHD.PT 4004, Kamunting Industrial EstateP.O. Box 934600 Taiping PerakMALAYSIA |
|---|---|---|
| 510(k) Number(if known) | : | K981893 * |
| Device Name | : | HEALTH + AID® PREMIUM PRE-POWDERED LATEXEXAMINATION GLOVE (PROTEIN LABEL CLAIM) |
| Indications For Use | : | 100mcg/gm or less total water extractableprotein per gram |
| HEALTH + AID® PREMIUM Pre-powdered Latex Examination Glove is a single use | ||
| device intended for medical purposes that is worn on the hand of health care and similar |
personnel to prevent contamination between the health care personnel and the patient.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Chris S.
(Division Sign-Off)
( Division of Dental, Infection Control,
and General Hospital, Infection Control, and General Hospital, Infection -510(k) Number
Over-The-Counter OR
Image /page/4/Picture/10 description: The image shows a simple line drawing. There is a horizontal line across the bottom of the image. Above the line is an "X" shape. The lines of the "X" intersect near the middle of the image.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.