(258 days)
Incstar Rubella IgG "fast" ELISA Kit
Not Found
No
The summary describes a standard ELISA test kit and an automated processor for running the test. There is no mention of AI or ML algorithms being used for data analysis, interpretation, or any other function. The performance studies focus on traditional metrics like sensitivity, specificity, and agreement with predicate devices and reference panels.
No.
The device is an in-vitro diagnostic test kit used to detect antibodies to rubella in human serum to assess immunological response or immune status. It does not treat or directly provide therapy to a patient.
Yes
The device is intended for the qualitative and quantitative determination of IgG antibodies to aid in the assessment of immunological response and in the diagnosis of recent or current infection with rubella.
No
The device is a test kit that includes reagents and is used in conjunction with a physical automated processor (MAGO® Plus Automated EIA Processor) or manually. It is not solely software.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states that the device is for the "qualitative and quantitative determination of IgG antibodies in human serum to aid in the assessment of the patient's immunological response to infection with rubella and in the determination of the immune status of individuals... The evaluation of acute and convalescent sera can aid in the diagnosis of recent or current infection with rubella." This clearly indicates that the device is intended to be used in vitro (outside the body) on human specimens (serum) to provide information for diagnostic purposes (aiding in the assessment of immune response and diagnosis of infection).
- Device Description: The description details an "enzyme-linked immunosorbent assay (ELISA) for the qualitative and quantitative detection of IgG to rubella in human serum." This describes a laboratory test performed on a human sample.
- Performance Studies: The performance studies involve testing human serum samples to evaluate the device's performance in detecting rubella IgG antibodies.
All of these points align with the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Diamedix Is-Rubella IgG Test Kit is an Enzyme Immunoassay (EIA) for the qualitative and quantitative determination of IgG antibodies in human serum to aid in the assessment of the patient's immunological response to infection with rubella and in the determination of the immune status of individuals, including females of child-bearing age. The evaluation of acute and convalescent sera can aid in the diagnosis of recent or current infection with rubella. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated EIA Processor.
Product codes (comma separated list FDA assigned to the subject device)
LFX
Device Description
The Is-Rubella IgG Test System is an enzyme-linked immunosorbent assay (ELISA) for the qualitative and quantitative detection of IgG to rubella in human serum
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
human serum
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
A total of three hundred and seventy-six prospective (fresh) sera were tested for the presence of rubella IgG antibodies using the Diamedix Is-Rubella IgG Test Kit and two other marketed tests at two independent sites (site #1, Miami, FL and site #2, Salt Lake City, Utah). Site #3 (Diamedix Corp. ( Miami, FL) tested 270 retrospective (frozen) sera both manually and using the MAGO Plus Automated EIA Processor. In addition, all sites tested a panel of 100 retrospective selected negative and low positive sera provided by Diamedix. Site #1 also tested the CDC Rubella Reference Panel.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Comparison Testing:
- Site #1: Tested 198 prospective samples.
- Relative Sensitivity: 97.1% (165/170)
- Relative Specificity: 100.0% (19/19)
- Overall Agreement: 97.4% (184/189)
- Discordant samples resolution: 4 samples negative by Is-Rubella IgG and positive by other EIA were negative by referee method; 1 equivocal.
- Site #2: Tested 178 prospective samples.
- Relative Sensitivity: 100.0% (127/127)
- Relative Specificity: 83.8% (31/37)
- Overall Agreement: 96.3% (158/164)
- Discordant samples resolution: 5 samples positive by Is-Rubella IgG and negative by other EIA were positive by referee method; 1 equivocal.
- Site #3 (Manual): Tested 270 retrospective (frozen) samples.
- Relative Sensitivity: 98.1% (204/208)
- Relative Specificity: 100.0% (49/49)
- Overall Agreement: 98.4% (253/257)
- Discordant samples resolution: Of 4 samples negative by Is-Rubella IgG but positive by other EIA, 2 were negative and 2 positive by referee EIA.
- Site #3 (MAGO Plus): Tested 263 retrospective (frozen) samples.
- Relative Sensitivity: 99.5% (209/210)
- Relative Specificity: 100.0% (43/43)
- Overall Agreement: 99.6% (252/253)
- Discordant samples resolution: 1 sample negative by Is-Rubella IgG but positive by other EIA was also negative by referee EIA.
- Retrospective Panel (All Sites): Tested a panel of approximately 50 negative and 50 low positive sera.
- Site #1: Relative Sensitivity 100%, Relative Specificity 100%
- Site #2: Relative Sensitivity 93.3%, Relative Specificity 100%
- Site #3: Relative Sensitivity 92.0%, Relative Specificity 100%
Correlation of Manual and MAGO Plus Results:
- 193 samples compared; highly reactive samples excluded.
- Correlation coefficient (r): 0.9550.
CDC Serum Panel Data:
- Panel: 82% positive, 18% negative samples.
- Outside testing site (Site #1) agreement: 100% for all samples.
- Diamedix agreement (manual and automated):
- Negative results: 100% (18 of 18)
- Positive specimens: 97.6% (80 of 82)
Linearity:
- Strongly positive serum samples serially diluted and assayed in duplicate.
- High degree of linearity demonstrated for both manual and MAGO Plus methods.
- Manual Linearity: Slope 101.51, R-squared 0.9933
- MAGO Plus Linearity: Slope 168.47, R-squared 0.9969
Correlation to WHO Standard:
- Calibrated against WHO 1st International Standard for Anti-Rubella Immunoglobulin (RUBI-1-94).
- R^2: 0.9846, indicating strong positive correlation.
Quantitative Data:
- Serum pairs showed a 2.8-fold (mean 4.8-fold) increase in Is-Rubella IgG IU/ml values corresponded to a four-fold titer increase in rubella IgG antibody levels.
Cross-Reactivity Data:
- 49 sera positive for other viruses but negative for rubella IgG were tested.
- No cross-reactivity expected from VZV, HSV, CMV, Toxoplasma, EBV, Parvovirus B19, Measles IgG analytes.
Precision: (Intra-assay and Interassay CV% values are provided across different sites and methods for various serum samples and controls).
- Site #1 (Manual):
- Interassay CV% ranged from N/A (for very low values) to 23.14% (10 STD).
- Site #2 (Manual):
- Interassay CV% ranged from N/A to 20.12% (Serum E).
- Site #3 (Manual):
- Interassay CV% ranged from N/A to 23.82% (Serum C).
- Site #3 (MAGO Plus):
- Interassay CV% ranged from N/A to 18.43% (Serum C).
- Inter-Site Precision (Manual):
- Inter-site CV% ranged from N/A to 25.69% (Serum G).
- Lot-to-Lot Precision (Manual):
- Lot-to-lot CV% ranged from N/A to 11.78% (LPC).
- Lot-to-Lot Precision (MAGO Plus):
- Lot-to-lot CV% ranged from N/A to 20.57% (Serum H).
Expected Values:
- Incidence of rubella IgG antibodies varies among populations.
- Studies on 100 healthy South Florida donors (52 female, 48 male): 89% positive, 9% negative, 2% equivocal.
- 37 female donors of child-bearing age (18-45 years): 79% positive, 16% negative, 5% equivocal.
- 45 pregnant females (15 from each trimester): 92% negative, 4% negative, 4% equivocal for anti-rubella IgG (Note: discrepancy in data, likely typo, "92% negative, 2 (4%) were negative" could mean 92% positive given context of other studies).
- Total of 294 samples from females of childbearing age in clinical studies: 76% positive, 19% negative, 5% equivocal.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
- Relative Sensitivity: 97.1% (Site #1), 100.0% (Site #2), 98.1% (Site #3 Manual), 99.5% (Site #3 MAGO Plus)
- Relative Specificity: 100.0% (Site #1), 83.8% (Site #2), 100.0% (Site #3 Manual), 100.0% (Site #3 MAGO Plus)
- Overall Agreement: 97.4% (Site #1), 96.3% (Site #2), 98.4% (Site #3 Manual), 99.6% (Site #3 MAGO Plus)
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Incstar Rubella IgG "fast" ELISA Kit
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 866.3510 Rubella virus serological reagents.
(a)
Identification. Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).(b)
Classification. Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.
0
JAN 28 1999
Appendix E. 510(k) Summary of Safety and Effectiveness
The following section is included as required by the Safe Medical Device Act (SMDA) of 1990.
Name: Address:
i
Diamedix Corporation 2140 N. Miami Avenue Miami, FL 33127
Dr. Lynne Stirling Contact Person: Phone Number: 305-324-2354 Fax Number: 305-324-2585
1
510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
Applicant Information:
Date Prepared: | May 14, 1998 |
---|---|
Name: | Diamedix Corporation |
Address: | 2140 N. Miami Avenue |
Miami, FL 33127 |
Contact Person: | Dr. Lynne Stirling |
---|---|
Phone Number: | 305-324-2354 |
Fax Number: | 305-324-2585 |
Device Information:
Trade Name: Is-Rubella IgG Test System Common Name: Rubella IgG EIA Test Classification Name: Enzyme linked immunosorbent assay, rubella (866.3510)
Equivalent Device:
Incstar Rubella IgG "fast" ELISA Kit
Device Description: The Is-Rubella IgG Test System is an enzyme-linked immunosorbent assay (ELISA) for the qualitative and quantitative detection of IgG to rubella in human serum
Intended Use: The assay is intended for use in detecting IgG antibodies to rubella antigen in human serum. The results of the assay can be used as an aid in the assessment of the patient's immunological response to infection with rubella and in the determination of immune status of individuals, including females of child-bearing age. The evaluation of paired sera can aid in the diagnosis of current or recent infection.
Principle of the Procedure: The Is-Rubella IgG Test System is an enzyme-linked immunosorbent assay to detect IgG to rubella in human serum. Partially purified rubella antigen is attached to a solid phase microtiter well. Diluted test sera are added to each well. If antibodies which recognize the rubella antigen are present in the patient sample they will bind to the antigen on the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human immunoglobulin (conjugate) is added to each test well. If antibody is present the enzyme-linked antibody will bind to it. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is then added to each well. If enzyme is present from the prior step, the reaction is stopped and the color intensity is measured photometrically producing an indirect measure of the specific antibody present in the patient sample.
2
Performance Characteristics
A. Comparison Testing
A total of three hundred and seventy-six prospective (fresh) sera were tested for the presence of rubella IgG antibodies using the Diamedix Is-Rubella IgG Test Kit and two other marketed tests at two independent sites (site #1, Miami, FL and site #2, Salt Lake City, Utah). Site #3 (Diamedix Corp. ( Miami, FL) tested 270 retrospective (frozen) sera both manually and using the MAGO Plus Automated EIA Processor. In addition, all sites tested a panel of 100 retrospective selected negative and low positive sera provided by Diamedix. Site #1 also tested the CDC Rubella Reference Panel (See Section C. )
Site #1 tested 198 samples submitted for immune status screening. Samples were obtained from the S. Florida area. Table 1 compares the results obtained for the Is-Rubella IgG Test Kit and their currently used testing method.
Site #2 tested 178 samples submitted for ToRCH screening. Samples were obtained from the West region. Table 2 compares the results obtained for the Is-Rubella IgG Test Kit and their currently used testing method.
| TABLE 1
Is-Rubella IgG - Site #1 | | | TABLE 2
Is-Rubella IgG - Site #2 | | | | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-------------------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------|
| | Positive | Negative | Equivocal | | Positive | Negative | Equivocal |
| Comp.
EIA 1
Positive | 165 [93] | 5 [2] | 8 [7] | Comp.
EIA 2
Positive | 127 [50] | 0 | 0 |
| Negative | 0 | 19 [7] | 1 [1] | Negative | 6 [3] | 31 [12] | 3 |
| | | | | Equivocal | 7 [3] | 2 [1] | 2 [2] |
| | 95% CI*
Relative Sensitivity 165/170 = 97.1%
Relative Specificity 19/19 = 100.0%
Overall Agreement** 184/189 = 97.4% | | | | 95% CI*
Relative Sensitivity 127/127 = 100.0%
Relative Specificity 31/37 = 83.8%
Overall Agreement** 158/164 = 96.3% | | |
| | | 93.3-99.0
82.4-100.0
93.9-99.1 | | | | 97.1-100.0
68.0-93.8
92.2-98.7 | |
For Site #1, further resolution of the discordant samples was performed by testing such samples in a referee EIA method. Four of the samples negative by the Is-Rubella IgG Test Kit and positive by the other EIA were negative by the referee method; the remaining sample was equivocal.
For Site #2, further resolution of the discordant samples was performed in a similar manner. Five of the six samples positive by the Is-Rubella IgG Test Kit and negative by the other EIA were positive by the referee method; the remaining sample was equivocal.
Site #3 (Diamedix Corp.) tested 270 samples (all frozen) by the manual method and 263 of these samples (seven being ONS) by the MAGO Plus method. Samples were obtained from S. Florida blood donors. Tables 3 and 4 compare the results obtained for the Is-Rubella IgG Test Kit and another marketed EIA method.
| TABLE 3
Is-Rubella IgG - Site #3 : Manual | | | | | TABLE 4
Is-Rubella IgG - Site #3 : | | | |
|----------------------------------------------|----------------------|-----------------|-----------|---------------------|---------------------------------------|----------------------|------------------|----|
| | Positive | Negative | Equivocal | | | Positive | Negative | |
| Other
EIA | Positive | 204 [73] | 4 [2] | 6 [5] | | Positive | 209 | 1 |
| | Negative | 0 | 49 [30] | 0 | Other
EIA | Negative | 0 | 43 |
| | Equivocal | 3 [1] | 3 [2] | 1 | | Equivocal | 3 | 1 |
| | Relative Sensitivity | 204/208 = 98.1% | | 95% CI*
95.1-99. | | Relative Sensitivity | 209/210 = 99.5% | |
| | Relative Specificity | 49/49 = 100.0% | | 92.7-100.0 | | Relative Specificity | 43/43 = 100.0% | |
| | Overall Agreement** | 253/257 = 98.4% | | 96.1-99.6 | | Overall Agreement** | 252/253 = 99.6 % | |
TABLE 4 |
---|
Measles-Rubella IgG - Site #3 : MAGO Plus |
= 100.0%
Equivocal
3
0 3
95% CI*
97.4-100.0
91.8-100.0
97.8-100.0
] denotes samples from females of childbearing age (18-45 years) [
- 95% Confidence Intervals (CI) calculated by the Exact Method
** Equivocal results were excluded from calculations
000278
3
For Site #3 (manual testing), further resolution of the discordant sera revealed that of the Is-Rubella IgG Test Kit but positive in the other EIA two were negative and two were positive by a referee EIA method. For MAGO Plus testing, the sample that was negative in the Is-Rubella IgG Test Kit but positive in the other EIA was also negative by a referee EIA method.
In addition to the samples tabulated above, each site tested the same panel consisting of approximately 50 negative and 50 low positive sera provided by diamedix. Table 5 shows the results of this testing compared to the predicate method.
Other EIA | Pos | Neg | Equ | Relative Sensitivity | Relative Specificity | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Is-Rubella IgG | Pos | Pos | Pos | Neg | Neg | Neg | Equ | Equ | Equ | 95%CI | 95% CI | |||||||||
Site #1 | 52 | 0 | 1 | 0 | 43 | 2 | 0 | 1 | 1 | 100% | 93.2-100 | 100% | 91.8-100 | |||||||
Site #2 | 42 | 3 | 8 | 0 | 45 | 0 | 0 | 2 | 0 | 93.3% | 81.7-100 | 100% | 92.1-100 | |||||||
Site #3 | 46 | 4 | 3 | 0 | 45 | 0 | 0 | 2 | 0 | 92.0% | 80.0-97.8 | 100% | 92.1-100 |
TABLE 5 : All Sites - Retrospective Panel of Negative and Weakly Positive Sera
NOTE : Please be advised that relative refers to the comparison of the assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be comparison assay's accuracy to predict disease.
B. Correlation of Manual and MAGO Plus Results
The Is-Rubella IgG Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence of the manual and MAGO Plus Procedures, the results of 193 samples tested assayed manually and by MAGO Plus were compared. Highly reactive samples that exceeded the reportable range and were excluded from this comparison. A scattergram and regression line of the results obtained with 95% confidence intervals is shown in Figure 1.
Image /page/3/Figure/7 description: This image is a scatter plot comparing MAGO PLUS IU/ml to MANUAL IU/ml. The data points are clustered around a regression line, indicating a positive correlation between the two variables. The equation of the regression line is Y = 1.6349 + 0.9947 X, and the correlation coefficient (r) is 0.9550, suggesting a strong positive linear relationship.
Image /page/3/Figure/8 description: The image shows the title of a figure. The title is "FIGURE 1 : Manual vs MAGO Plus Correlation". The title is written in a bold, sans-serif font.
C. CDC Serum Panel Data
)
The following information was obtained wth the Centers for Disease Control and Prevention (CDC) serum panel for rubella serology assays which was tested by the Is-Rubella IgG Test Kit both manually and using the MAGO Plus Automated Processor. For independent verification, this panel was also tested manually by outside site #1. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. Results were submitted to the CDC for their interpretation and evaluation. This does not imply an endorsement of the assay by the CDC.
The panel consists of 82% positive and 18% negative samples. The outside testing site correctly identified all samples (100% agreement). Of the results obtained by Diamedix, there was 100% (18 of 18) agreement with the negative results using both the manual and automated methods and 97.6% (80 of 82) agreement with the positive specimens using both the manual and automated methods.
4
C. Linearity
ﻤﺴﻠﺴﺔ
ﻟﻤﺴﺘﻘﻠﺔ
Several strongly positive serum samples were serially diluted and separate dilutions were assayed, in duplicate, in the Is-Rubella IgG Test Kit both manually and using the MAGO Plus Automated EIA Processor. Representative linear regression graphs and scattergrams of the mean results with 95% confidence intervals are presented in Figures 3 and 4 for one patient sample. The results demonstrate a high degree of linearity throughout the assay when samples are tested either manually or by MAGO Plus.
Image /page/4/Figure/2 description: The image contains two line graphs comparing manual linearity and MAGO Plus linearity. The graph on the left, labeled "FIGURE 2: Manual Linearity," shows a slope of 101.51, a Y-intercept of 0.54, and an R-squared value of 0.9933. The graph on the right, labeled "FIGURE 3: MAGO Plus Linearity," shows a slope of 168.47, a Y-intercept of -1.54, and an R-squared value of 0.9969. Both graphs plot dilution on the x-axis and IU/ml on the y-axis, with data points and trendlines indicating a positive correlation between dilution and IU/ml.
D. Correlation to WHO Standard
The Is-Rubella IgG Test Kit has been calibrated against the WHO 1st International Standard for Anti-Rubella Immunoglobulin (code RUBI-1-94). To demonstrate the accuracy of the quantitative procedure, several dilutions of the WHO Standard were prepared and assayed manually in triplicate versus the Is-Rubella IgG Test Kit standard curve. The linear regression graph and scattergram of the mean results with 95% Confidence Intervals is shown in Figure 4.
Image /page/4/Figure/5 description: The image is titled "FIGURE 4: WHO IU/ml vs Is-Rubella IgG IU/ml". The image contains the equation Y = 24080 + 07327 X. The number 80 is also present in the image.
Image /page/4/Figure/6 description: This image is a scatter plot that shows the relationship between Is-Rubella IgG IU/ml and WHO IU/ml. The x-axis represents WHO IU/ml, and the y-axis represents Is-Rubella IgG IU/ml. The plot includes a regression line with the equation Y = 2.4080 + 0.7327 X, and the coefficient of determination (R^2) is 0.9846, indicating a strong positive correlation between the two variables.
E. Quantitative Data
Serum pairs were obtained by preparing multiple two-fold dilutions of several strongly positive sera. Ratios for dilutions representing a four-fold difference in antibody level were evaluated as a serum pair both manually and using the MAGO Plus. Overall, it was estimated that a 2.8 told (mean 4.8-fold ) increase in Is-Rubella IgG IU/ml values corresponded to a four-fold titer increase in rubella IgG antibody levels.
5
F. Cross-Reactivity Data
Sera containing IgG antibodies to viruses potentially cross-reactive to rubella have been tested in the Is-Rubella IgG Test Kit. Forty-nine sera negative for IgG antibodies to rubella IgG Test Kit as well as in another marketed test but positive for one or more viruses were evaluated. The data in the following table suggest that no cross-reactivity should be expected with the Is-Rubella IgG Test Kit from these analytes.
| TABI | | 1 | | t
0 |
------ | -- | --- | -- | -------- |
---|
Analyte | Rubella IgG | VZV IgG | HSV IgG | CMV IgG | Toxoplasma IgG | EBV IgG | Parvovirus B19 IgG | Measles IgG |
---|---|---|---|---|---|---|---|---|
No.of Pos. | ||||||||
Samples | 0 | 42 | 48 | 43 | 5 | 48 | 12 | 47 |
G. Precision
)
Seven serum samples, spanning the reportable range, as well as the 10 IU/ml Standard and kit Low Positive and Negative Controls were tested quantitatively and values calculated from IU/ml results. Sites #1 and #2 tested samples in triplicate in three separate runs on three different days. Site #3 (Diamedix Corp.) tested samples in triplicate in two separate runs on three different days both manually and using the MAGO Plus Automated EIA Processor. Note that for MAGO Plus (Table 10) the 50 IU/ml Standard was replaced with an additional weakly positive sample. Tables 7-13 show the intra- and interassay precision for each site, the inter-site precision for manual and MAGO Plus.
TABLE 7 : Site #1- Intra-Assay and Interassay Precision
SERUM | INTRA-ASSAY DAY 1 | INTRA-ASSAY DAY 2 | INTRA-ASSAY DAY 3 | INTERASSAY (n=9) | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
MEAN | ||||||||||||
IU/ml | SD | CV% | MEAN | |||||||||
IU/ml | SD | CV% | MEAN | |||||||||
IU/ml | SD | CV% | MEAN | |||||||||
IU/ml | SD | CV% | ||||||||||
A | 0.1 | 0.10 | N/A | 0.0 | 0.00 | N/A | 0.0 | 0.06 | N/A | 0.0 | 0.07 | N/A |
B | 0.3 | 0.06 | N/A | 0.1 | 0.06 | N/A | 0.3 | 0.00 | N/A | 0.3 | 0.10 | N/A |
C | 20.9 | 0.95 | 4.55 | 20.8 | 3.07 | 14.76 | 21.9 | 2.32 | 10.59 | 21.2 | 2.05 | 9.67 |
D | 25.8 | 1.76 | 6.82 | 18.3 | 0.51 | 2.79 | 21.2 | 1.56 | 7.36 | 21.8 | 3.48 | 15.96 |
E | 42.4 | 0.65 | 1.53 | 39.5 | 3.86 | 9.77 | 34.9 | 2.65 | 7.59 | 38.9 | 4.03 | 10.36 |
F | 42.7 | 4.12 | 9.65 | 41.9 | 4.43 | 10.57 | 37.2 | 2.23 | 5.99 | 40.6 | 4.11 | 10.12 |
G | 23.1 | 4.04 | 17.49 | 20.5 | 2.76 | 13.46 | 20.0 | 2.00 | 10.00 | 21.2 | 3.00 | 14.15 |
10 STD | 10.9 | 1.88 | 17.25 | 14.1 | 2.87 | 20.35 | 10.3 | 2.43 | 23.59 | 11.8 | 2.73 | 23.14 |
50 STD | 50.5 | 1.19 | 2.36 | 55.9 | 3.30 | 5.90 | 45.9 | 0.06 | 0.13 | 50.8 | 4.67 | 9.19 |
LPC | 24.5 | 0.06 | 0.24 | 26.5 | 0.91 | 3.43 | 21.0 | 1.11 | 5.29 | 24.0 | 2.51 | 10.46 |
NC | 2.7 | 0.06 | N/A | 2.8 | 0.06 | N/A | 2.2 | 0.06 | N/A | 2.6 | 0.25 | N/A |
TABLE 8 : Site #2 - Intra-Assay and Interassay Precision
| SERUM | INTRA-ASSAY
DAY 1 | | | INTRA-ASSAY
DAY 2 | | | INTRA-ASSAY
DAY 3 | | | INTERASSAY (n=9) | | |
|--------|----------------------|------|-------|----------------------|------|-------|----------------------|------|-------|------------------|------|-------|
| | MEAN
IU/ml | SD | CV% | MEAN
IU/ml | SD | CV% | MEAN
IU/ml | SD | CV% | MEAN
IU/ml | SD | CV% |
| A | 0.3 | 0.06 | N/A | 0.3 | 0.12 | N/A | 0.5 | 0.00 | N/A | 0.4 | 0.12 | N/A |
| B | 0.5 | 0.06 | N/A | 0.4 | 0.06 | N/A | 0.4 | 0.12 | N/A | 0.4 | 0.11 | N/A |
| C | 17.2 | 1.25 | 7.27 | 17.1 | 2.19 | 12.81 | 16.0 | 2.10 | 13.13 | 16.8 | 1.75 | 10.42 |
| D | 14.3 | 0.98 | 6.85 | 14.0 | 1.57 | 11.21 | 12.8 | 0.25 | 1.95 | 13.7 | 1.16 | 8.47 |
| E | 50.6 | 3.16 | 6.25 | 31.8 | 1.85 | 5.82 | 43.2 | 0.59 | 1.37 | 41.9 | 8.43 | 20.12 |
| F | 28.0 | 1.45 | 5.18 | 26.3 | 2.24 | 8.52 | 25.2 | 2.50 | 9.92 | 26.5 | 2.20 | 8.30 |
| G | 14.2 | 1.82 | 12.82 | 12.5 | 1.27 | 10.16 | 10.6 | 1.93 | 18.21 | 12.5 | 2.13 | 17.04 |
| 10 STD | 9.7 | 0.06 | 0.62 | 10.6 | 0.85 | 8.02 | 9.7 | 0.25 | 2.58 | 10.0 | 0.63 | 6.30 |
| 50 STD | 45.2 | 0.66 | 1.46 | 46.2 | 0.64 | 1.39 | 45.8 | 3.53 | 7.71 | 45.7 | 1.87 | 4.09 |
| LPC | 21.1 | 0.44 | 2.09 | 24.2 | 2.36 | 9.75 | 22.0 | 3.12 | 14.18 | 22.4 | 2.39 | 10.67 |
| NC | 2.7 | 0.06 | N/A | 2.3 | 0.85 | N/A | 2.9 | 0.28 | N/A | 2.6 | 0.44 | N/A |
6
TABLE 9 : Site #3-Intra-Assay and Interassay Precision (Manual)
SERUM | INTRA-ASSAY DAY 1 | INTRA-ASSAY DAY 2 | INTRA-ASSAY DAY 3 | INTERASSAY (n=18) | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
MEAN IU/ml | SD | CV% | MEAN IU/ml | SD | CV% | MEAN IU/ml | SD | CV% | MEAN IU/ml | SD | CV% | |
A | 0.0 | 0.05 | N/A | 0.1 | 0.08 | N/A | 0.2 | 0.16 | N/A | 0.1 | 0.13 | N/A |
B | 0.2 | 0.04 | N/A | 0.2 | 0.04 | N/A | 0.2 | 0.20 | N/A | 0.2 | 0.11 | N/A |
C | 14.2 | 0.99 | 6.97 | 14.6 | 1.23 | 8.42 | 14.3 | 6.12 | 42.80 | 14.4 | 3.43 | 23.82 |
D | 15.1 | 1.13 | 7.48 | 14.3 | 1.38 | 9.65 | 15.8 | 3.94 | 24.94 | 15.1 | 2.44 | 16.16 |
E | 37.4 | 1.93 | 5.16 | 35.8 | 2.63 | 7.35 | 35.3 | 7.52 | 21.30 | 36.1 | 4.54 | 12.58 |
F | 33.9 | 2.18 | 6.43 | 34.6 | 2.33 | 6.73 | 28.4 | 4.18 | 14.72 | 32.3 | 4.04 | 12.51 |
G | 20.3 | 2.32 | 11.43 | 19.8 | 1.49 | 7.53 | 15.9 | 2.44 | 15.35 | 18.6 | 2.85 | 15.32 |
10 STD | 9.2 | 0.48 | 5.22 | 9.4 | 0.26 | 2.77 | 12.5 | 1.83 | 14.64 | 10.4 | 1.88 | 18.08 |
50 STD | 53.3 | 3.35 | 6.29 | 51.1 | 2.12 | 4.15 | 53.9 | 9.17 | 17.01 | 52.8 | 5.55 | 10.51 |
LPC | 24.7 | 3.62 | 14.66 | 24.6 | 1.11 | 4.51 | 30.4 | 4.90 | 16.12 | 26.6 | 4.38 | 16.47 |
NC | 2.8 | 0.12 | N/A | 2.8 | 0.12 | N/A | 3.1 | 0.33 | N/A | 2.9 | 0.24 | N/A |
TABLE 10 : Site #3- Intra-assay and Interassay Precision (MAGO Plus)
SERUM | INTRA-ASSAY DAY 1 | INTRA-ASSAY DAY 2 | INTRA-ASSAY DAY 3 | INTERASSAY (n=18) | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
MEAN IU/ml | SD | CV% | MEAN IU/ml | SD | CV% | MEAN IU/ml | SD | CV% | MEAN IU/ml | SD | CV% | |
A | 0.2 | 0.08 | N/A | 0.2 | 0.10 | N/A | 0.2 | 0.10 | N/A | 0.2 | 0.09 | N/A |
B | 0.4 | 0.06 | N/A | 0.2 | 0.10 | N/A | 0.3 | 0.05 | N/A | 0.3 | 0.12 | N/A |
C | 16.8 | 2.46 | 14.64 | 15.4 | 2.93 | 19.03 | 19.6 | 2.87 | 16.67 | 17.2 | 3.17 | 18.43 |
D | 15.3 | 1.65 | 10.78 | 12.7 | 1.45 | 11.42 | 16.0 | 2.68 | 14.64 | 14.6 | 2.38 | 16.30 |
E | 39.6 | 2.47 | 6.24 | 35.6 | 2.51 | 7.05 | 39.8 | 3.00 | 7.54 | 38.3 | 3.18 | 8.30 |
F | 37.7 | 1.98 | 7.69 | 38.1 | 3.23 | 8.48 | 34.4 | 3.37 | 9.80 | 36.7 | 3.43 | 9.35 |
G | 17.1 | 2.90 | 8.36 | 16.6 | 1.80 | 10.84 | 18.1 | 1.26 | 6.96 | 17.3 | 1.55 | 8.96 |
H | 14.2 | 1.43 | 13.94 | 15.4 | 3.21 | 20.84 | 17.5 | 1.23 | 7.03 | 15.6 | 2.59 | 16.60 |
10 STD | 17.0 | 2.19 | 12.88 | 14.1 | 0.89 | 6.31 | 18.6 | 1.23 | 6.61 | 16.5 | 2.39 | 14.48 |
LPC | 33.3 | 1.20 | 3.60 | 27.7 | 3.13 | 11.30 | 36.3 | 1.57 | 4.33 | 32.4 | 4.19 | 12.93 |
NC | 4.2 | 0.28 | N/A | 4.1 | 0.24 | N/A | 4.9 | 0.12 | N/A | 4.4 | 0.41 | N/A |
TABLE 11 : Inter-Site Precision (manual)
SERUM | Site #1 | Site #2 | Site #3 | INTERSITE | ||
---|---|---|---|---|---|---|
MEAN | ||||||
IU/ml | MEAN | |||||
IU/ml | MEAN | |||||
IU/ml | MEAN | |||||
IU/ml | SD | CV% | ||||
A | 0.0 | 0.4 | 0.1 | 0.2 | 0.21 | N/A |
B | 0.3 | 0.4 | 0.2 | 0.3 | 0.10 | N/A |
C | 21.2 | 16.8 | 14.4 | 17.5 | 3.45 | 19.71 |
D | 21.8 | 13.7 | 15.1 | 16.9 | 4.33 | 25.62 |
E | 38.9 | 41.9 | 36.1 | 39.4 | 3.63 | 9.21 |
F | 40.6 | 26.5 | 32.3 | 33.1 | 7.09 | 21.42 |
G | 21.2 | 12.5 | 18.6 | 17.4 | 4.47 | 25.69 |
10 STD | 11.8 | 10.0 | 10.4 | 10.7 | 0.95 | 8.88 |
50 STD | 50.8 | 45.7 | 52.8 | 49.8 | 3.66 | 7.35 |
LPC | 24.0 | 22.4 | 26.6 | 24.3 | 2.12 | 8.72 |
NC | 2.6 | 2.6 | 2.9 | 2.7 | 0.17 | N/A |
TABLE 12 : Lot-to-Lot Precision (manual)
| SERUM | Lot 31107
INTERASSAY | | | Lot 41307
INTERASSAY | | | Lot 50707
INTERASSAY | | | LOT-TO-LOT
INTERASSAY | | |
|--------|-------------------------|------|-------|-------------------------|------|-------|-------------------------|------|-------|--------------------------|------|-------|
| | MEAN | SD | CV% | MEAN | SD | CV% | MEAN | SD | CV% | MEAN | SD | CV% |
| | IU/ml | | | IU/ml | | | IU/ml | | | IU/ml | | |
| A | 0.1 | 0.05 | N/A | 0.1 | 0.10 | N/A | 0.1 | 0.13 | N/A | 0.1 | 0.00 | N/A |
| B | 0.2 | 0.07 | N/A | 0.2 | 0.22 | N/A | 0.2 | 0.11 | N/A | 0.2 | 0.00 | N/A |
| C | 13.9 | 1.54 | 11.08 | 13.0 | 1.45 | 11.15 | 14.4 | 3.43 | 23.82 | 13.8 | 0.71 | 5.14 |
| D | 13.4 | 1.86 | 13.88 | 13.6 | 2.12 | 15.59 | 15.1 | 2.44 | 16.16 | 14.0 | 0.93 | 6.64 |
| E | 35.2 | 3.36 | 9.55 | 33.2 | 4.21 | 12.68 | 36.1 | 4.54 | 12.58 | 34.8 | 1.48 | 4.25 |
| F | 33.3 | 1.91 | 5.74 | 33.5 | 1.81 | 5.40 | 32.3 | 4.04 | 12.51 | 33.0 | 0.64 | 1.94 |
| G | 16.9 | 1.82 | 10.77 | 20.4 | 2.24 | 10.98 | 18.6 | 2.85 | 15.32 | 18.6 | 1.75 | 9.41 |
| 10 STD | 9.1 | 0.50 | 5.49 | 9.0 | 0.80 | 8.89 | 10.4 | 1.88 | 18.08 | 9.5 | 0.78 | 8.21 |
| 50 STD | 50.2 | 3.79 | 7.55 | 49.7 | 3.29 | 6.62 | 52.8 | 5.55 | 10.51 | 50.9 | 1.66 | 3.26 |
| LPC | 21.0 | 1.75 | 8.33 | 24.6 | 3.14 | 12.76 | 26.6 | 4.38 | 16.47 | 24.1 | 2.84 | 11.78 |
| NC | 2.8 | 0.21 | N/A | 3.0 | 0.34 | N/A | 2.9 | 0.24 | N/A | 2.9 | 0.10 | N/A |
Image /page/6/Picture/8 description: The image shows the number 000282 in a clear, sans-serif font. The numbers are evenly spaced and of uniform size, creating a symmetrical appearance. The image is simple and focuses solely on the numerical sequence.
7
TABLE 13 : Lot-to-Lot Precision (Mago Plus)
| SERUM | Lot 31107
INTERASSAY | | | Lot 41307
INTERASSAY | | | Lot 50707
INTERASSAY | | | LOT-TO-LOT
INTERASSAY | | |
|--------|-------------------------|------|-------|-------------------------|------|-------|-------------------------|------|-------|--------------------------|------|-------|
| | MEAN
IU/ml | SD | CV% | MEAN
IU/ml | SD | CV% | MEAN
IU/ml | SD | CV% | MEAN
IU/ml | SD | CV% |
| A | 0.3 | 0.16 | N/A | 0.2 | 0.10 | N/A | 0.2 | 0.09 | N/A | 0.2 | 0.06 | N/A |
| B | 0.3 | 0.14 | N/A | 0.4 | 0.60 | N/A | 0.3 | 0.12 | N/A | 0.3 | 0.06 | N/A |
| C | 19.7 | 1.96 | 9.95 | 17.0 | 1.76 | 10.40 | 17.2 | 3.17 | 18.43 | 18.0 | 1.50 | 8.30 |
| D | 20.3 | 2.45 | 12.07 | 16.3 | 2.79 | 17.12 | 14.6 | 2.38 | 16.30 | 17.1 | 2.93 | 17.10 |
| E | 42.3 | 3.67 | 8.68 | 36.7 | 2.60 | 7.08 | 38.3 | 3.18 | 8.30 | 39.1 | 2.88 | 7.40 |
| F | 34.0 | 6.64 | 19.53 | 34.1 | 1.45 | 4.25 | 36.7 | 3.43 | 9.35 | 34.9 | 1.53 | 4.38 |
| G | 16.6 | 1.48 | 8.92 | 17.1 | 1.78 | 10.41 | 17.3 | 1.55 | 8.96 | 17.0 | 0.36 | 2.12 |
| H | 18.9 | 2.60 | 13.76 | 13.3 | 1.65 | 12.41 | 15.6 | 2.59 | 16.60 | 15.9 | 3.27 | 20.57 |
| 10 STD | 18.7 | 1.14 | 6.10 | 15.9 | 2.08 | 13.08 | 16.5 | 2.39 | 14.48 | 17.0 | 1.47 | 8.65 |
| LPC | 33.5 | 1.90 | 5.67 | 29.0 | 3.99 | 13.76 | 32.4 | 4.19 | 12.93 | 31.6 | 2.35 | 7.44 |
| NC | 4.4 | 0.43 | N/A | 3.7 | 0.36 | N/A | 4.4 | 0.41 | N/A | 4.2 | 0.40 | N/A |
Expected Values
)
The incidence of rubella IgG antibodies varies among populations depending on vaccination practices. In a recent national survey of military recruits the seronegativity rate was 17.4% for males and 12.8% for females.
In the present studies sera from 100 healthy South Florida donors (52 female and 48 male) were evaluated in the Is-Rubella IgG Test Kit. Of the 100 samples, 89 (89%) were found to be positive, 9 (9%) were negative and 2 (2%) were equivocal. Age distribution, geographic location and prevalence is provided in Table 14. Histograms demonstrating the distribution of IU/ml values are shown in Figures 5 and 6.
Thirty-seven of the female donors were of child-bearing age (18-45 years). Of the sera from these donors, 29 (79%) were positive, 6 (16%) were negative and 2 (5%) were equivocal. A total of 45 sera from pregnant females (15 from each trimester) were also tested in the Is-Rubella IgG Test Kit. Forty one (92%) were negative, 2 (4%) were negative and 2 (4%) were equivocal for anti-rubella IgG. In addition, a total of 294 samples from females of childbearing age were identified in the outside and in-house clinical studies (these included the 45 sera from pregnant females already referenced). Of these samples, 223 (76%) were positive, 56 (19%) were negative and 15 (5%) were equivocal for antirubella IgG when evaluated in the Is-Rubella IgG Test Kit.
Number of Donors | Prevalence | |
---|---|---|
Total Number | 100 | 89.0% |
Geographic Location: | ||
South-Eastern US | 100 | 89.0% |
Age | ||
10-19 | 13 | 92.3% |
20-29 | 23 | 73.9% |
30-39 | 40 | 92.5% |
40-49 | 13 | 92.3% |
50-59 | 5 | 100.0% |
60-69 | 6 | 100.0% |
| 6 | | | ﺎ
1 |
|---|---|--|--------|
| | 1 | | |
Image /page/7/Figure/8 description: The image contains two bar charts, labeled as Figure 5 and Figure 6. Figure 5 shows the Is-Rubella IgG Positive Population, with the x-axis representing IU/mL levels in ranges (0-10, 11-20, 21-30, 31-40, 41-50, and >50) and the y-axis representing frequency, with the >50 range having the highest frequency. Figure 6 shows the Is-Rubella IgG Negative Population, with the x-axis representing IU/mL levels in ranges (1-2, 3-4, 5-6, and 7-10) and the y-axis representing frequency, with the 1-2 range having the highest frequency. Both charts provide a visual comparison of the distribution of Is-Rubella IgG levels in positive and negative populations.
8
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 28 1999
Lynne Stirling, Ph.D. Vice President, Regulatory Affairs Diamedix Corporation 2140 North Miami Ave. Miami, FL 33127
Re: K981729 Trade Name: Is-Rubella IgG Test System Regulatory Class: III Product Code: LFX Dated: November 23, 1998 Received: November 24, 1998
Dear Dr. Stirling:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
9
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
10
Appendix G. Indications for Use Statement
INDICATIONS FOR USE STATEMENT
510(K) NUMBER : ______________________________________________________________________________________________________________________________________________________________
DEVICE NAME : Is-Rubella IgG Test System
Indications for Use : The Diamedix Is-Rubella IgG Test Kit is an Enzyme Immunoassay (EIA) for the qualitative and quantitative determination of IgG antibodies in human serum to aid in the assessment of the patient's immunological response to infection with rubellaand in the determination of the immune status of individuals, including females of child-bearing age. The evaluation of acute and convalescent sera can aid in the diagnosis of recent or current infection with rubella. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated EIA Processor.
Woody Dubois
(Division Sign Off)
Division of Clinical Laboratory Devices
510(k) Number K981729
PRESCRIPTION USE X