(220 days)
This product is intended for use with the Huntleigh DVT Flowtron pump to aid in the prevention of deep vein thrombosis in surgical, trauma and certain medical patients at risk for DVT formation. It is not intended for use on patients with known DVT or with any other intermittent compression device. This device is restricted to sale by or on the order of a physician.
The RTM DVT PS100 ProSoft Cuff is a cotton garment designed to be wrapped around the lower extremity (calf), secured with looped tape and connected to Huntleigh DVT Flowtron pump. The outer shell of the garment is a blend of cotton and linen and encases a plastic inflatable bladder that is connected to the pump via poly vinyl tubing.
This submission describes a medical device, the RTM DVT PS100 ProSoft Cuff, seeking substantial equivalence to a predicate device, the Huntleigh Flowtron DVT System L501 Garment. The "study" conducted for this submission focuses on demonstrating technological equivalence rather than clinical effectiveness against specific performance criteria related to DVT prevention.
Here's an analysis of the provided information based on your questions:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a submission demonstrating substantial equivalence to a predicate device, the acceptance criteria are based on directly comparing the technological characteristics of the new device to the predicate device. There aren't explicitly stated numerical performance thresholds in the way one might see for a diagnostic device's sensitivity or specificity. Instead, the acceptance criterion is a qualitative assessment of "almost identical" or "virtually no differences detected."
| Acceptance Criteria (for Technological Equivalence) | Reported Device Performance |
|---|---|
| Air Flow/Volume Test: Inflation/deflation times for the RTM DVT PS100 ProSoft Cuff should be nearly identical to the Huntleigh L501 garment. | Inflation/deflation times were "almost identical with less than one second differences detected." |
| Cycle Test (with counter pressure): Inflation/deflation times for the RTM DVT PS100 ProSoft Cuff (attached to a person's calf) should be virtually identical to the Huntleigh L501 garment (attached to a person's calf). | "Virtually no difference in the times detected." |
2. Sample Size for the Test Set and Data Provenance
The documentation does not explicitly state a numerical sample size for the "Air Flow/Volume Test" or the "Cycle Test." It implies a comparison of one RTM DVT PS100 ProSoft Cuff bladder against one Huntleigh L501 bladder. For the "Cycle Test," it implies testing one of each cuff on "a person's calf."
- Sample Size: Not explicitly stated as a numerical value for multiple instances; implied n=1 for each device type in each test scenario (unattached bladder and attached to a calf).
- Data Provenance: The tests were conducted by RT Medical. The location of these tests is not specified, but it can be assumed to be within the United States, given that the company address is in Ohio. The data is prospective as it was generated specifically for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
There were no external experts used to establish ground truth in the traditional sense for these technological tests. The "ground truth" was established by direct measurement of physical properties (inflation/deflation times) of the devices themselves. The comparison was against the predicate device, not an expert-annotated dataset.
4. Adjudication Method for the Test Set
No adjudication method was mentioned or necessary, as the "ground truth" was based on direct physical measurements and comparison between the two devices.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The study described is a technological comparison of the new device to a predicate device, not a study of clinical effectiveness or human reader performance. Therefore, there is no mention of effect size for human readers with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable to the RTM DVT PS100 ProSoft Cuff. This device is a mechanical compression cuff, not a software algorithm or AI-powered device. Therefore, the concept of "standalone (algorithm only)" performance does not apply.
7. The Type of Ground Truth Used
The "ground truth" used was the measured physical performance of the predicate device (Huntleigh L501 garment). The RTM DVT PS100 ProSoft Cuff's performance was compared directly to these measurements.
8. The Sample Size for the Training Set
There is no training set for this type of device submission. This is a physical medical device, not a machine learning model.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable as there is no training set for a physical medical device like the RTM DVT PS100 ProSoft Cuff.
{0}------------------------------------------------
DEC 1 4 1998
Image /page/0/Picture/1 description: The image shows a logo for "RT International Services, Inc." The logo features the letters "RT" in a stylized font inside of a circle. The word "International" is written in a cursive font, and the words "Services, Inc." are written in a sans-serif font below the cursive word.
P.O. BOX 1179 • CUYAHOGA FALLS, OH 44223-0179 • 1 (800) 837-0529 • FAX (216) 686-0597
November 4, 1998
Center for Devices and Radiological Health Food and Drug Administration 9200 Corporate Blvd. Rockville, MD 20850
RE: 510(k) Summary, K981643
Contact Person: Karen L. Diehl, RN P.O. Box 1179 Cuyahoga Falls, OH 44223 1-800-837-0529
RTM DVT PS100 ProSoft Cuff Device Name:
Compression Cuff Common Name:
Classification Name: Powered, inflatable tube massager, calf cuff CFR Section: 890.5650
-
- Claiming Substantial Equivalence to Huntleigh Flowtron DVT System L501 Garment (K881632).
Description: The RTM DVT PS100 ProSoft Cuff is a cotton garment designed to be wrapped around the lower extremity (calf), secured with looped tape and connected to Huntleigh DVT Flowtron pump. The outer shell of the garment is a blend of cotton and linen and encases a plastic inflatable bladder that is connected to the pump via poly vinyl tubing.
- Claiming Substantial Equivalence to Huntleigh Flowtron DVT System L501 Garment (K881632).
-
Intended Use: This product is intended for use with the Huntleigh DVT Flowtron pump to aid in the prevention of deep vein thrombosis in surgical, trauma and certain medical patients at risk for DVT formation. It is not intended for use on patients with known DVT or with any other intermittent compression device. This device is restricted to sale by or on the order of a physician.
{1}------------------------------------------------
Technological Characteristics:
The RTM DVT PS100 ProSoft Cuff and the Huntleigh L501 garment operate essentially the same technologically with virtually no differences detected during testing. They are both plastic bladders connected to the Huntleigh DVT Flowtron pump, inflated intermittently and applied to the lower extremities for DVT prevention.
To determine that the two devices are technologically equivalent RT Medical conducted an "Air Flow/Volume Test" and a "Cycle Test." The Air Flow/ Volume Test consisted of inflating the RTM bladder and the Huntleigh L501 bladder and timing the inflation and deflation times. They were found to be almost identical with less than one second differences detected in the inflation/ deflation times. The Cycle Test was accomplished in the same manner, however, the garment was attached to a person's calf to provide counter pressure and mimic a clinical environment. Inflation/deflation times for both cuffs were obtained and recorded with, again, virtually no difference in the times detected. The conclusion was drawn that the two devices are technologically and substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus-like symbol with three wavy lines representing the human form. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol.
DEC 1 4 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Karen L. Diehl Director, Clinical Support Services R.T. Medical Services, Inc. 730 W. Portage Trail Extension Akron, OH 44313
K981643 Re : RTM DVT ProSoft Cuff Model Number PS100 Requlatory Class: II (Two) Product Code: JOW November 5, 1998 Dated: Received: November 12, 1998
Dear Ms. Diehl:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of The general controls provisions of the Act include the Act. requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Paqe 2 - Ms. Karen L. Diehl
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callatian, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
| Page | of |
|---|---|
| ------ | ---- |
510(k) Number (if known): K981643
Device Name: RTM DVT ProSoft Cuff
Indications For Use:
This product is intended for use with the Huntleigh DVT Flowtron pump to aid in the prevention of deep vein thrombosis in surgical, trauma and certain medical patients at risk for the formation of DVT. It is not intended for use on patients with known DVT or with any other intermittent compression device. Federal law restricts this device to sale by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bera R. Bymperly
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number____ K 981643
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
§ 870.5800 Compressible limb sleeve.
(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).