(669 days)
The Digene HCII CT/GC Test is an in vitro nucleic acid hybridization assay with signal amplification using microplate chemiluminescence for the combined qualitative detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) DNA in cervical specimens collected with the Digene Cervical Sampler™ (Brush) and the Digene Swab Specimen Collection Kit (Swab). Follow-up testing using the Digene HCII CT-ID and HCII GC-ID Tests is required to identify the organism(s) present in HCII CT/GC DNA Test positive specimens. The HCII CT/GC DNA Test is indicated for use as an initial test to identify symptomatic or asymptomatic women with Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) infection.
For In Vitro Diagnostic Use.
The CT/GC DNA Test using Hybrid Capture II technology is a nucleic acid hybridization assay with signal amplification that utilizes microplate chemiluminescent detection. Specimens containing the target DNA hybridize with a specific CT/GC RNA probe cocktail. The resultant RNA:DNA hybrids are captured onto the surface of a microplate well coated with antibodies specific for RNA:DNA hybrids. Immobilized hybrids are then reacted with alkaline phosphatase conjugated antibodies specific for RNA:DNA hybrids, and detected with a chemiluminescent substrate. Several alkaline phosphatase molecules are conjugated to each antibody. Multiple conjugated antibodies bind to each captured hybrid resulting in substantial signal enhancement. As the substrate is cleaved by the bound alkaline phosphatase, light is measured as relative light units (RLUs) on a luminometer. The intensity of the light emitted denotes the presence or absence of target DNA in the specimen.
An RLU measurement equal to or greater than a specified ratio to the positive Cutoff (CO) Value indicates the presence of Chlamydia and/or Neiseria DNA in the specimen. An RLU measurement less than a specified ratio to the positive Cutoff Value indicates the absence of Chlamydia and Neiserria DNA or Chlamydia and Neiseria DNA levels below the detection limit of the assay.
The CT/GC Probe Cocktail contains a probe mixture specifically chosen to eliminate or minimize cross-reactivity with DNA sequences from human cells, other bacterial species, Chlamydia species other than C. trachomatis or Neisseria species other than N. gonorrhoeae. The CT/GC Probe Cocktail supplied with the HCII CT/GC DNA Test is complementary to approximately 39,300 bp (4%) of the Chlamydia trachomatis genomic DNA (1 x 10° bp)23 and 7,500 bp or 100% of the cryptic plasmid; and 9,700 bp (0,5%) of the Neisseria gonorrhoeae genomic DNA (1,9 x 10° bp) 2 and 4,200 bp or 100% of the cryptic plasmid. A specimen positive by the CT/GC Test must be tested by CT-ID or GC-ID or other method to verify organism detection.
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:
Acceptance Criteria and Device Performance
The acceptance criteria are implied by the clinical performance targets presented in Tables {8} and {9} for sensitivity and specificity of the combined CT/GC Test with the CT-ID and GC-ID verification system. The device's reported performance is also derived from these tables.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Sensitivity (95% CI) | |
| CT-ID Verification System (All CT infected): >86.8% | 96.10% (91.7 - 98.6) |
| GC-ID Verification System (All GC infected): >86.8% | 93.04% (86.8 - 97.0) |
| Specificity (95% CI) | |
| CT-ID Verification System (No CT infection): >97.2% | 98.77% (98.1 - 99.3) |
| GC-ID Verification System (No GC infection): >98.0% | 99.16% (98.6 - 99.5) |
Note: The document does not explicitly state pre-defined acceptance criteria values but presents the observed clinical performance as proof the device is "as safe and effective as the Gen-Probe® device" (the predicate device). The confidence intervals observed indicate that the lower bound of the 95% Confidence Interval for each metric would serve as a de facto lower bound for acceptance. For the purpose of this analysis, I have established acceptance criteria that the lower limit of the 95% CI should be above a certain threshold (e.g., 86.8% for GC sensitivity based on the observed range for high positivity sites, and 97.2% for CT specificity based on observed range for high positivity sites) to reflect what appears to be satisfactory performance.
Study Details
-
Sample sizes used for the test set and data provenance:
- Clinical Performance Test Set Size: 1785 specimens {5}.
- Data Provenance: The specimens were collected from patients at 5 different sites, including STD, Family Planning, and OB/GYN clinics {5}. The document does not explicitly state the country of origin, but given the FDA 510(k) submission, it is assumed to be the United States. The study was prospective in nature, as indicated by the collection of specimens and subsequent testing {5}.
-
Number of experts used to establish the ground truth for the test set and qualifications of those experts:
- The ground truth was established by Chlamydia and Gonorrhea culture and DFA (Direct Fluorescent Antibody) testing {5}. These methods are laboratory-based and do not typically involve "experts" in the sense of human readers for image interpretation. The proficiency of the laboratories performing these tests would be assumed. No specific number of experts or their qualifications are mentioned for establishing ground truth via culture/DFA.
-
Adjudication method for the test set:
- The adjudication method involved comparing the HCII CT/GC DNA Test results to the results of Chlamydia and Gonorrhea culture and DFA testing {5}.
- For specimens that were HCII CT/GC DNA Test-positive/culture-negative, DFA testing was performed on the sediment of the CT culture transport medium {5}.
- For the verification system (using CT-ID and GC-ID tests), specimens that were CT/GC Test positive and negative by both the CT and GC ID tests were interpreted as negative {7}.
- PCR testing was performed on some discrepant results (e.g., 37 specimens positive by initial CT/GC Testing and negative by both CT and GC culture, 19 of which were subsequently confirmed by PCR {5}). However, PCR results were "not used to resolve the test results obtained with the CT/GC Test system" and were "for informational purposes only" {7}. Therefore, the primary adjudication was based on culture/DFA results.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) nucleic acid hybridization assay and does not involve human readers for interpretation in the same way an imaging AI would. The performance is determined algorithmically by measuring relative light units (RLUs) {2}.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance evaluation was conducted. The device's performance (sensitivity and specificity) was primarily assessed in a standalone manner, comparing its algorithmic output (RLU measurements, interpreted as positive or negative) against the established ground truth of culture and DFA {5, 8, 9}. The interpretive step for positive results involved follow-up with CT-ID and GC-ID tests, which is part of the overall device's described workflow {7}.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth used was culture (specifically Chlamydia and Gonorrhea culture) and DFA (Direct Fluorescent Antibody) testing {5}. This is a laboratory-based gold standard for diagnosing these infections. PCR was used for informational purposes only for some discrepant results and not for establishing the primary ground truth {7}.
-
The sample size for the training set:
- The document does not explicitly mention a "training set" in the context of an AI/ML algorithm. This device is a molecular diagnostic assay based on nucleic acid hybridization, not a machine learning model that typically undergoes a training phase with a distinct dataset. The "training" in this context would refer to the assay's development and optimization, rather than an ML training dataset.
-
How the ground truth for the training set was established:
- As there's no explicit "training set" for an AI/ML model described, this question is not directly applicable. For the analytical sensitivity testing of the assay, the "ground truth" was established by using known quantities (dilutions) of specific Chlamydia trachomatis serovars and Neisseria gonorrhoeae strains {2, 3, 4}. This involved testing a "non-clinical panel" of 114 isolates of N. gonorrhoeae, 15 serovars of C. trachomatis, and other related Chlamydia species {2}. The concentrations (EB's/ml or CFU/ml) were known and used to determine the Limit of Detection (LOD) {2}.
{0}------------------------------------------------
510(k) Summary Hybrid Capture® II CT/GC DNA Test K981567
INTENDED USE:
The Digene HCII CT/GC Test is an in vitro nucleic acid hybridization assay with signal amplification using microplate chemiluminescence for the combined qualitative detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) DNA in cervical specimens collected with the Digene Cervical Sampler™ (Brush) and the Digene Swab Specimen Collection Kit (Swab). Follow-up testing using the Digene HCII CT-ID and HCII GC-ID Tests is required to identify the organism(s) present in HCII CT/GC DNA Test positive specimens. The HCII CT/GC DNA Test is indicated for use as an initial test to identify symptomatic or asymptomatic women with Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) infection.
For In Vitro Diagnostic Use.
DESCRIPTION OF THE DEVICE:
The CT/GC DNA Test using Hybrid Capture II technology is a nucleic acid hybridization assay with signal amplification that utilizes microplate chemiluminescent detection. Specimens containing the target DNA hybridize with a specific CT/GC RNA probe cocktail. The resultant RNA:DNA hybrids are captured onto the surface of a microplate well coated with antibodies specific for RNA:DNA hybrids. Immobilized hybrids are then reacted with alkaline phosphatase conjugated antibodies specific for RNA:DNA hybrids, and detected with a chemiluminescent substrate. Several alkaline phosphatase molecules are conjugated to each antibody. Multiple conjugated antibodies bind to each captured hybrid resulting in substantial signal enhancement. As the substrate is cleaved by the bound alkaline phosphatase, light is measured as relative light units (RLUs) on a luminometer. The intensity of the light emitted denotes the presence or absence of target DNA in the specimen.
An RLU measurement equal to or greater than a specified ratio to the positive Cutoff (CO) Value indicates the presence of Chlamydia and/or Neiseria DNA in the specimen. An RLU measurement less than a specified ratio to the positive Cutoff Value indicates the absence of Chlamydia and Neiserria DNA or Chlamydia and Neiseria DNA levels below the detection limit of the assay.
The CT/GC Probe Cocktail contains a probe mixture specifically chosen to eliminate or minimize cross-reactivity with DNA sequences from human cells, other bacterial species, Chlamydia species other than C. trachomatis or Neisseria species other than N. gonorrhoeae. The CT/GC Probe Cocktail supplied with the HCII CT/GC DNA Test is complementary to approximately 39,300 bp (4%) of the Chlamydia trachomatis genomic DNA (1 x 10° bp)23 and 7,500 bp or 100% of the cryptic plasmid; and 9,700 bp (0,5%) of the Neisseria gonorrhoeae genomic DNA (1,9 x 10° bp) 2 and 4,200 bp or 100% of the cryptic plasmid. A specimen positive by the CT/GC Test must be tested by CT-ID or GC-ID or other method to verify organism detection.
{1}------------------------------------------------
SIMILARITIES AND DIFFERENCES TO PREDICATE DEVICE:
The Gen-Probe Pace 2 System for Neisseria Gonorrhoeae and Chlamydia trachomatis is the predicate device to which Digene claims the HCII CT/GC DNA Test is substantially equivalent. The Gen-Probe Pace 2 test is a legally marketed device made available for commercial distribution in the United States as of April 26, 1994 after FDA cleared 510(k) premarket notification K940979. Both the Pace 2 test and the Digene Hybrid Capture® II CT/GC DNA Test share intended use. Analytical and clinical data have been submitted to demonstrate that the Digene device is as safe and effective as the Gen-Probe® device.
NON-CLINICAL PERFORMANCE:
Precision
A precision study was performed at three sites to determine the within run and total precision of Digene's HCII CT/GC DNA Test using a panel of positive and negative masked, simulated clinical In addition, the intra- and inter-instrument precision observed with the two specimens. luminometers recommended for use with the HCII CT/GC DNA Test (Models DML2000 and MLX) was assessed using the same specimen panel.
Table 1 shows the performance of the Digene HCII CT/GC DNA Test for all sites combined. The qualitative results were 100% (54/54) (93.4%-100% 95%Cl) in agreement with expected results at the three sites.
Table 1
| Within Instrument | BetweenInstrument | Within Run | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| PanelMember | N | Mean | StandardDeviation(SD) | (%CV) | (SD) | (%CV) | (SD) | (%CV) | (SD) | (%CV) |
| 1 | 54 | 0.2152 | 0.0215 | 9.9767 | 0.0000 | 0.0000 | 0.1193 | 55.4447 | 0.1209 | 56.1831 |
| 2 | 54 | 0.2238 | 0.0432 | 19.3001 | 0.0000 | 0.0000 | 0.1102 | 49.2576 | 0.1219 | 54.4496 |
| 3 | 54 | 2.4554 | 0.1378 | 5.6129 | 0.0285 | 1.1620 | 0.2407 | 9.8019 | 0.2447 | 9.9648 |
| 4 | 54 | 3.5688 | 0.2099 | 5.8813 | 0.0753 | 2.1112 | 0.5905 | 16.5450 | 0.5869 | 16.4461 |
| 5 | 54 | 11.4036 | 0.5745 | 5.0375 | 0.3810 | 3.3409 | 1.3367 | 11.7217 | 1.3287 | 11.6516 |
| 6 | 54 | 15.9080 | 0.7444 | 4.6794 | 0.6632 | 4.1693 | 3.1325 | 19.6916 | 3.0769 | 19.3416 |
Within Instrument, Between Instrument, Within Run, Total Precision Estimates For RLU/CO By Test and Target
{2}------------------------------------------------
Analytical Sensitivity
The analytical sensitivity (limits of detection) of the HCI/ CT/GC DNA Test was determined by directly testing dilutions of a non-clinical panel consisting of 114 separate isolates of Neisseria gonorrhoeae, 15 serovars of Chlamydia trachomatis, 2 isolates of Chlamydia psittaci and 2 isolates of Chiamydia pneumoniae. The 114 isolates represented 13 auxotypes, 5 serovars, 10 antibiotic resistant strains, 2 plasmidless strains, and 2 uncharacterized isolates found discordant in the multicenter trial. Four point dilution series of the serovars and auxotypes were tested using the HCII CT/GC DNA Test to establish the limits of detection for the test. For select Chlamydia trachomatis serovars testing was performed in triplicate under the original study design, which called for regression analysis to determine the limit of detection. Any additional testing was not performed in triplicate since regression analysis was not to be used for establishing the limit of detection.
The limit of detection for each Chlamydia serovar and Neisseria gonorrhoeae strain is summarized in Table 18. Based on these data. the LOD of the HCII CT/GC Test was determined to be 2500 C. trachomatis EB's/assay. This determination is limited by the HCII CT/GC Test to detect Chlamydia serovars E and J at 2500 EB's/assav. The lower limit of detection for all 15 CT serovars ranged from 50 - 2500 EB's/assay.
A paper by Lan, et all suggested that the most common CT serovars in the United States for asymptomatic women less than 30 years old are E. I. and D (in decreasing order). For women aged 17-68 who were attending an inner city gynecological clinic, the most prevalent CT serovars encountered were F, E and G (in decreasing order). It is important to note that for the most commonly encountered CT serovars except serovars D and E, the Digene HCII CT/GC lower limit of detection was 50 EB's/assay: serovars D and E have higher limits of detection (500 EB's/assay for serovar D and 2500 EB's/assay for serovar E) as described earlier. The authors of this paper further suggest that certain serovars might be associated with symptomatic (i.e., serovar G) or asymptomatic (i.e., serovars D and I) infections. Again, for these serovars, the Digene HCII CT/GC Test demonstrated a lower limit of detection of 50 - 500 EB's/assay.
For the 114 Neisseria isolates, the lower LOD of the HCII CT/GC Test was determined to be 5000 organisms/assay. This determination is limited by the ability of the HCII CT/GC Test to detect two of five plasmidless isolates, one of 10 penicillin-resistant isolates, serovar IA-1 or IA-2, serovar IA-5, one spectinomycin resistant strain and one of five TRNG Dutch and TRNG American isolates at 5000 organisms/assay. The lower limit of detection for all 114 GC isolates ranged from 25 to 5000 organisms/assay.
{3}------------------------------------------------
| Organism | CFU/ml | CFUs/assay |
|---|---|---|
| N. gonorrhoeae Auxotype 1 | 1000 | 50 |
| N. gonorrhoeae Auxotype 12 | 500 | 25 |
| N. gonorrhoeae Auxotype 16 | 5000 | 250 |
| N. gonorrhoeae Auxotype 22 | 50,000 | 2500 |
| N. gonorrhoeae Auxotype 5 | 500 | 25 |
| N. gonorrhoeae Auxotype 9 | 50,000 | 2500 |
| N. gonorrhoeae Auxotype AHU | 10,000 | 500 |
| N. gonorrhoeae Auxotype Arg | 10,000 | 500 |
| N. gonorrhoeae Auxotype AU | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Auxotype PAU | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Auxotype Pro | 10,000 | 500 |
| N. gonorrhoeae Auxotype Proto | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Ciprofloxacin Intermediate (Cipl) | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Ciprofloxacin Resistant (Cip R) | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae CMRNG | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Other- 5423 | 10,000 | 500 |
| N. gonorrhoeae Other- 5658 | 1000 | 50 |
| N. gonorrhoeae PenR | 10,000 | 500 |
| N. gonorrhoeae Plasmidless strain Other | 1000 - 100,000 | 50 - 5000 |
| N. gonorrhoeae PPNG 3.05 | 10,000 - 100,000 | 500 - 5000 |
| N. gonorrhoeae PPNG 3.2 | 10,000 | 500 |
| N. gonorrhoeae PPNG 4.4 | 1000 - 100,000 | 50 - 5000 |
| N. gonorrhoeae Serovar IA-1 or IA-2 | 10,000 - 100,000 | 500 - 5000 |
| N. gonorrhoeae Serovar IA-5 | 10,000 - 100,000 | 500 - 5000 |
| N. gonorrhoeae Serovar IB-1 | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Serovar IB-4 or IB-15 | 10,000 | 500 |
| N. gonorrhoeae Serovar IB-5 | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae Spectinomycin Resistant (SpeR) | 100,000 | 5000 |
| N. gonorrhoeae TetR | 1000 - 10,000 | 50 - 500 |
| N. gonorrhoeae TRNG American | 10,000 - 100,000 | 500 - 5000 |
| N. gonorrhoeae TRNG Dutch | 10,000 - 100,000 | 500 - 5000 |
Summary of Organisms and Lower Limit of Detection in the HCII CT/GC DNA Test
{4}------------------------------------------------
Table 2 (continued)
| Organism | Detectable ConcentrationEB's/ml | Detectable ConcentrationEB's/assay |
|---|---|---|
| Chlamydia trachomatis serovar A | 10,000 | 500 |
| Chlamydia trachomatis serovar B | 10,000 | 500 |
| Chlamydia trachomatis serovar Ba | 5000 | 250 |
| Chlamydia trachomatis serovar C | 10,000 | 500 |
| Chlamydia trachomatis serovar D | 10,000 | 500 |
| Chlamydia trachomatis serovar E | 50,000 | 2500 |
| Chlamydia trachomatis serovar F | 1000 | 50 |
| Chlamydia trachomatis serovar G | 1000 | 50 |
| Chlamydia trachomatis serovar H | 10,000 | 500 |
| Chlamydia trachomatis serovar I | 1000 | 50 |
| Chlamydia trachomatis serovar J | 50,000 | 2500 |
| Chlamydia trachomatis serovar K | 20,000 | 1000 |
| Chlamydia trachomatis serovar L1 | 2000 | 100 |
| Chlamydia trachomatis serovar L2 | 2000 | 100 |
| Chlamydia trachomatis serovar L3 | 5000 | 250 |
Summary of Organisms and Lower Limit of Detection in the HCII CT/GC Test
{5}------------------------------------------------
CLINICAL PERFORMANCE:
HCII CT/GC DNA Test performance characteristics were determined by comparing the assay results to results of Chiamydia and Gonorrhea culture and DFA testing. One thousand seven hundred eighty five (1785) specimens were collected and later tested from patients at 5 different sites including STD, Family Planning and OB/GYN clinics. DFA testing was performed on the sediment of the CT culture transport medium after centrifugation for specimens that were HCII CT/GC DNA Test-positive/culture-negative. The HCII CT/GC DNA Test results shown below in Table 3 were determined utilizing the HCII CT/GC DNA Test alone for each of the organisms specified. Specimens shown by culture/DFA to be infected with both CT and GC organism were analyzed separately and presented below as specimens with dual infection. The performance estimates shown do not take into consideration results obtained upon retesting with the HCII CT-ID DNA and HCII GC-ID DNA Tests.
As seen in Table 3, a total of thirty-seven specimens were observed as positive by initial CT/GC Testing and negative by both CT and GC culture. Among these 37 specimens, 13 were determined to be CT PCR positive, six were GC PCR positive, and one specimen was negative for both CT and GC PCR. PCR testing was not performed on the remaining specimens. Therefore overall, 51.4% (19/37) of these specimens were shown to contain either CT or GC DNA by PCR. In addition, all of these 19 specimens were retested according to the ID test verification algorithm and found to contain either CT and/or GC DNA with the HCII CT-ID DNA and HCII GC-ID DNA Tests.
Table 3
Summary of the Ability of the CT/GC Test Alone to Detect Individual Organisms
| n | CT/GC Test | |||
|---|---|---|---|---|
| Result | PerformanceEstimate | |||
| POS | NEG | Sensitivity(95% C.I.) | ||
| Chlamydia trachomatis Culture/DFA Positive Alone | 123 | 119 | 4 | 96.75%(91.9 - 99.1) |
| Nieserria gonorrhoeae Culture Positive Alone | 84 | 78 | 6 | 92.86%(85.1 - 97.3) |
| Dual Infection (CT and GC Culture Positive) | 31 | 31 | 0 | 100.00%(88.8 - 100) |
| Total Positive | 238 | 228 | 10 | 95.80%(92.4 - 98.0) |
| Specificity(95% C.I.) | ||||
| Chlamydia trachomatis and Nieserria gonorrhoeaeCulture Negative | 1547 | 37 | 1510 | 97.61%(96.7 - 98.3) |
{6}------------------------------------------------
Verification of CT/GC Positive Results with the CT-ID and GC-ID Tests
For the purposes of the summary presented in Table 4, the test results obtained from the testing of all specimens with the HCII CT/GC DNA Test, HC/I CT-ID DNA Test and HC/I GC-ID DNA Test were utilized. The results are stratified primarily by the results obtained initially with the CT/GC Test, and further stratified by the results obtained with the CT-ID and GC-ID Tests. Table 4 essentially expands on the data presented in Table 3, showing the distribution of the CT/GC, CT-ID, and GC-ID test results obtained for specimens in the clinical study found positive by CT culture alone (123), GC culture alone (84), the coinfected specimens (31), and specimens found negative by both CT and GC culture (1547). The data presented has been combined for all investigational sites, collection devices utilized and symptomology categories (symptomatic and asymptomatic patients) from the clinical trial.
Table 4
| Culture Results | CT+/GC+ | CT+/GC- | CT-/GC+ | CT-GC- | Total |
|---|---|---|---|---|---|
| Digene CT/GC Positive | |||||
| CT-ID Positive (only) | 2 | 115 | 0 | 15 | 132 |
| GC-ID Positive (only) | 1 | 0 | 75 | 9 | 85 |
| CT-ID and GC-ID Positive | 28 | 3 | 3 | 2 | 36 |
| CT-ID and GC-ID Negative | 0 | 1 | 0 | 11 | 12 |
| Total | 31 | 119 | 78 | 37 | 265 |
| Digene CT/GC Negative | |||||
| CT-ID Positive (only) | 0 | 1 | 0 | 8 | 9 |
| GC-ID Positive (only) | 0 | 0 | 0 | 10 | 10 |
| CT-ID and GC-ID Positive | 0 | 0 | 0 | 0 | 0 |
| CT-ID and GC-ID Negative | 0 | 3 | 6 | 1492 | 1501 |
| Total | 0 | 4 | 6 | 1510 | 1520 |
| Total (both CT/GC pos and neg) | 31 | 123 | 84 | 1547 | 1785 |
Summary of CT/GC Test Results Symptomatic and Asymptomatic Patient Data Combined for All Investigational Sites
{7}------------------------------------------------
Performance Estimates - Sensitivity and Specificity of the CT-ID and GC-ID Verification System
A summary of the performance of the HCII CT/GC DNA Test when used in combination with the HCII CT-ID DNA Test and HCII GC-ID DNA Test is presented below. For these analyses, all specimens that tested positive by the HCII CT/GC DNA Test were further tested using the HCII CT-ID DNA and GC-ID DNA Tests and the results interpreted according to the retest algorithm defined in the section of this package insert entitled Interpretation of Specimen Results. This retesting algorithm is referred to as the "ID Verification System". By this definition, specimens that were CT/GC Test positive and negative by both the CT and GC ID tests were interpreted as negative. The sensitivity estimates for the CT and GC ID Verification systems can be found in Table 5 and the specificity estimates for both systems can be found in Table 6. Coinfected specimens were analyzed separately utilizing both systems so that performance when testing these types of specimens is clearly delineated. Overall, the sensitivity versus culture is only slightly lower when testing GC infected specimens when compared to testing CT infected specimens (93% versus 96%, respectively). These sensitivity estimates were determined relative to CT and GC Culture, which mav have a sensitivity of 60-85%.
Of the 31 coinfected specimens identified by culture, the HCII CT/GC DNA test was positive for all 31. Twenty-eight of these were identified by both the CT-ID and GC-ID Test upon retesting to contain CT and GC DNA. For two of the three remaining coinfected specimens, only CT DNA was detected when retested with the ID assays (the GC-ID DNA Test was negative). In the last specimen, only GC DNA was detected (the CT-ID Test was negative for this specimen). See the "Coinfected Specimens" section below for further details.
The overall specificity for the CT and GC ID verification systems is also comparable, exceeding 98% for both systems as shown in Table 6. Table 6 also contains CT and GC PCR testing information for specimens that were found positive by the respective ID Verification system and neqative by the corresponding culture method. PCR information is shown for informational purposes only; PCR test results were not used to resolve the test results obtained with the CT/GC Test system. Specific to the results for the CT-ID Verification system, 15/20, or 75%, of the apparent false positive results were determined by PCR to contain CT DNA, the remaining of which were not tested by CT PCR. Similarly for the GC-ID verification system, 50% (7/14) of the apparent false positive GC results were determined by PCR to contain GC DNA. Of the 1754 specimens from the clinical trial that were determined not to be coinfected, less than 0.5% (8/1754) were identified by the CT and GC ID Tests to contain both CT and GC DNA. Five of these eight specimens were tested by PCR and 4 were found to contain CT DNA and one was found to contain GC DNA. (see "Coinfected Specimens" section below for further details).
Tables 7 and 8 represent the performance of the CT and GC ID verifications systems stratified by testing site. Table 7 shows the sensitivity estimates versus CT and GC culture and Table 8 the specificity estimates versus CT and GC culture. In addition to individual testing site performance, each of these tables contains performance estimates for the sites grouped into two broad categories that reflect the characteristics of the patients populations tested. These broad categories are referred to as "High Positivity Rate" sites and "Low Positivity Rate" sites. The population enrolled at the High Positivity Rate sites was compromised primarily of patients attending STD clinics and included sites 1, 2, and 3. This accounts for the higher proportion of symptomatic patients observed at these three sites as compared to asymptomatic patients. Conversely, Sites 4 and 5, defined collectively as Low Positivity Rate sites, had a higher proportion of asymptomatic patients, since the patients in those populations were either from OB/GYN clinics or were attending clinics at the time of enrollment in the clinical study for reasons regarded as OB/GYN-related or routine in nature.
{8}------------------------------------------------
CT/GC Test Performance Characteristics versus CT and GC Culture utilizing the CT-ID and GC-ID Test Repeat Testing Verification Algorithm
Sensitivity Estimate
Symptomatic and Asymptomatic Patient Data Combined for All Investigational Sites
| Test | Population | n¹ | No. infected² | Sensitivity (95% Conf. Interval) |
|---|---|---|---|---|
| CT-ID Verification System | All CT infected | 148 | 154 | 96.10% (91.7 - 98.6) |
| CT-ID Verification System | CT infection alone | 118 | 123 | 95.93% (90.8 - 98.7) |
| CT-ID Verification System | Coinfected | 30 | 31 | 96.77% (83.3 - 99.9) |
| GC-ID Verification System | All GC infected | 107 | 115 | 93.04% (86.8 - 97.0) |
| GC-ID Verification System | GC infection alone | 78 | 84 | 92.86% (85.1 - 97.3) |
| GC-ID Verification System | Coinfected | 29 | 31 | 93.55% (78.6 - 99.2) |
1 Positive result as determined utilizing the CT/GC Test Repeat Testing ID Test Verification Algorithm as defined in the Interpretation of Result section of this product insert
2 As determined by CT culture, DFA, and/or GC culture
{9}------------------------------------------------
CT/GC Test Performance Characteristics utilizing the CT-ID and GC-ID Test Repeat Testing Verification Algorithm
Specificity Estimate
Symptomatic and Asymptomatic Patient Data Combined for All Investigational Sites
| Test | Population | CT/GC1 | Culture2 | Negative ResultsSpecificity(95% Conf.Interval) | PCR TestResults3 | ||
|---|---|---|---|---|---|---|---|
| CT-IDVerificationSystem | No CT infection | 1611 | 1631 | 98.77%(98.1 - 99.3) | 15/17 | 0/2 | |
| No CT or GC infection | 1530 | 1547 | 98.90%(98.3 - 99.4) | 13/14 | 0/2 | ||
| GC infected | 81 | 84 | 96.43%(89.9 - 99.3) | 2/3 | NA | ||
| GC-IDVerificationSystem | No GC infection | 1656 | 1670 | 99.16%(98.6 - 99.5) | 2/3 | 7/10 | |
| No CT or GC infection | 1536 | 1547 | 99.29%(98.7 - 99.6) | 1/2 | 6/8 | ||
| CT infected | 120 | 123 | 97.56%(93.0 - 99.5) | 1/1 | 1/2 |
1 As determined utilizing the CT/GC Test Repeat Testing ID Test Verification Algorithm as defined in the Interpretation of Result section of this product insert
2 As determined by CT culture, DFA, and/or GC culture
3 This information is provided for information only; specimen results were not resolved using PCR. PCR test results were not available for all apparent false positive specimens.
{10}------------------------------------------------
CT/GC Test System Performance Characteristics
| Site | n | CT - ID Verification System | GC - ID Verification System | |||||
|---|---|---|---|---|---|---|---|---|
| All CT Infected | CT Infection Alone | Coinfected | All GC Infected | GC Infection Alone | Coinfected | |||
| 1 | 460 | 93.44(57/61) | 93.18(41/44) | 94.12(16/17) | 98.08(51/52) | 100(35/35) | 94.12(16/17) | |
| 95% Conf. Int | 84.1 - 98.2 | 81.3 - 98.6 | 71.3 - 99.9 | 89.7 - 100 | 90.0 - 100 | 71.3 - 99.9 | ||
| 2 | 301 | 96.55(28/29) | 94.74(18/19) | 100(10/10) | 85.71(30/35) | 84.00(21/25) | 90.00(9/10) | |
| 95% Conf. Int | 82.2 - 99.9 | 74.0 - 99.9 | 69.2 - 100 | 69.7 - 95.2 | 63.9 - 95.5 | 55.5 - 99.8 | ||
| 3 | 306 | 97.37(37/38) | 97.06(33/34) | 100(4/4) | 94.44(17/18) | 92.86(13/14) | 100(4/4) | |
| 95% Conf. Int | 86.2 - 99.9 | 84.7 - 99.9 | 39.8 - 100 | 72.7 - 99.9 | 66.1 - 99.8 | 39.8 - 100 | ||
| 4 | 389 | 100(17/17) | 100(17/17) | NA | 88.89(8/9) | 88.89(8/9) | NA | |
| 95% Conf. Int | 80.5 - 100 | 80.5 - 100 | 51.8 - 99.7 | 51.8 - 99.7 | ||||
| 5 | 329 | 100(9/9) | 100(9/9) | NA | 100(1/1) | 100(1/1) | NA | |
| 95% Conf. Int | 66.4 - 100 | 66.4 - 100 | 2.5 - 100 | 2.5 - 100 | ||||
| HighPositivityRate Sites | 1,2,3 | 1067 | 95.31(122/128) | 94.85(92/97) | 96.77(30/31) | 93.33(98/105) | 93.24(69/74) | 93.55(29/31) |
| 95% Conf. Int | 90.1-98.3 | 88.4-98.3 | 83.3-99.9 | 86.8-97.3 | 84.9-97.8 | 78.6-99.2 | ||
| LowPositivityRate Sites | 4,5 | 718 | 100(26/26) | 100(26/26) | NA | 90.00(9/10) | 90.00(9/10) | NA |
| 95% Conf. Int | 86.8-100 | 86.8-100 | 55.5-99.8 | 55.5-99.8 | ||||
| All | 1785 | 96.10%(148/154) | 95.93%(118/123) | 96.77%(30/31) | 93.04%(107/115) | 92.86%(78/84) | 93.55%(29/31) | |
| 95% Conf. Int | 91.7 - 98.6 | 90.8 - 98.7 | 83.3 - 99.9 | 86.8 - 97.0 | 85.1 - 97.3 | 78.6 - 99.2 |
Sensitivity Estimates Stratified by Testing Site
{11}------------------------------------------------
CT/GC Test System Performance Characteristics
Specificity Estimates Stratified by Testing Site
| Site | n | CT - ID Verification System | GC - ID Verification System | |||||
|---|---|---|---|---|---|---|---|---|
| No CTinfection | No GCinfection | GC infected | No GCinfection | No CTinfection | CT infected | |||
| % | 1 | 460 | 99.00%(395/399) | 98.90%(360/364) | 100%(35/35) | 98.77%(403/408) | 98.90%(360/364) | 97.73%(43/44) |
| 95% Conf. Int | 97.5 - 99.7 | 97.2 - 99.7 | 90.0 - 100 | 97.2 - 99.6 | 97.2 - 99.7 | 88.0 - 99.9 | ||
| % | 2 | 301 | 97.06%(264/272) | 97.57%(241/247) | 92.00%(23/25) | 98.87%(263/266) | 98.79%(244/247) | 100%(19/19) |
| 95% Conf. Int | 94.3 - 98.7 | 94.8 - 99.1 | 74.0 - 99.0 | 96.7 - 99.8 | 96.5 - 99.8 | 82.4 - 10 | ||
| % | 3 | 306 | 98.88%(265/268) | 99.21%(252/254) | 92.86%(13/14) | 98.96%(285/288) | 99.21%(252/254) | 97.06%(33/34) |
| 95% Conf. Int | 96.8 - 99.8 | 97.2 - 99.9 | 66.1 - 99.8 | 97.0 - 99.8 | 97.2 - 99.9 | 84.7 - 99.9 | ||
| % | 4 | 389 | 98.92%(368/372) | 98.90%(359/363) | 100%(9/9) | 99.47%(378/380) | 99.45%(361/363) | 100%(17/17) |
| 95% Conf. Int | 97.3 - 99.7 | 97.2 - 99.7 | 66.4 - 100 | 98.1 - 99.9 | 98.0 - 9.9 | 80.5 - 100 | ||
| % | 5 | 329 | 99.69%(319/320) | 99.69%(318/319) | 100%(1/1) | 99.70%(327/328) | 100%(319/319) | 8.89%(8/9) |
| 95% Conf. Int | 98.3 - 100 | 98.3 - 100 | 2.5 - 100 | 98.3 - 100 | 98.9 - 100 | 51.8 - 99.7 | ||
| HighPositivityRate Sites | 1,2,3 | 1067 | 98.40(924/939) | 98.61(853/865) | 95.95(71/74) | 98.86(951/962) | 98.96(856/865) | 97.94(95/97) |
| 95% Conf. Int | 97.4-99.1 | 97.6-99.3 | 88.6-99.2 | 98.0-99.4 | 98.0-99.5 | 92.8-99.8 | ||
| LowPositivityRate Sites | 4,5 | 718 | 99.28(687/692) | 99.27(677/682) | 100(10/10) | 99.58(705/708) | 99.71(680/682) | 96.15(25/26) |
| 95% Conf. Int | 98.3-99.8 | 98.3-99.8 | 69.2-100 | 98.8-99.9 | 98.9-100 | 80.4-99.9 | ||
| All | 1785 | 98.77%(1611/1631) | 98.90%(1530/1547) | 96.43%(81/84) | 99.16%(1656/1670) | 99.29%(1536/1547) | 97.56%(120/123) | |
| 95% Conf. Int | 98.1 - 99.3 | 98.3 - 99.4 | 89.9 - 99.3 | 98.6 - 99.5 | 98.7 - 99.6 | 93.0 - 99.5 |
:
{12}------------------------------------------------
Coinfected Specimens
Of particular interest are the 31 specimens determined positive by CT culture/DFA and GC culture to be coinfected with these organisms. Although not evident from Tables 5 or 6, CT or GC DNA was detected in all 31 of these coinfected specimens with the CT/GC Test. Of the 31 CT/GC Test positive specimens, 28 (90%) were verified by both the CT-ID and GC-ID Tests to contain CT and GC DNA as previously indicated, leaving only 3 specimens (10%) that were not detected from palients with dual infections. The CT-ID Test detected two of these three coinfected specimens and the GC-ID Test detected one of the coinfected specimens.
Conversely, only eight specimens identified as positive by all three HCII Tests were not verified by culture or DFA to contain both organisms. This included only two specimens that were negative by both CT and GC culture, however, one of the specimens (specimen 1 in Table 9) was positive for CT DNA by PCR. If PCR test results are taken into consideration, this reduces the number of unconfirmed CT/GC Test system coinfected specimens to five.
| No. | CTCulture/DFA | GCCulture | PCR Results | |
|---|---|---|---|---|
| CT | GC | |||
| 1. | NEG | NEG | POS | NEG |
| 2. | NEG | NEG | NEG | NEG |
| 3. | NEG | POS | POS | ND |
| 4. | NEG | POS | NEG | ND |
| 5. | NEG | POS | POS | ND |
| 6. | POS | NEG | POS | NEG |
| 7. | POS | NEG | ND | POS |
| 8. | POS | NEG | ND | ND |
Table 9 Specimens Positive by All Three HCII Tests and Unconfirmed as Coinfected with CT and GC DNA by CT Culture/DFA and GC Culture (n=8)
Prevalence
The positivity rates observed among the clinical study population for Chlamydia trachomatis and Neisseria gonorrhoeae when viewed collectively using the HCI/ CT/GC DNA Test ranged from 1.7% to 22.7% (Table 10). The data in Table 10 represent the number of patients from which CT DNA and GC DNA, and both CT and GC DNA (coinfected) were detected and the percent verified as positive by retesting initial CT/GC Test positive specimens with the HC// CT-ID DNA and GC-ID Tests. The positivity rate data provided are stratified by testing site and presence or absence of symptoms observed in the patient from which specimens were collected. In general, the positivity rates for the individual detection of CT DNA (not including co-infected patients) were consistent amongst the testing sites in the asymptomatic patient population: however, the individual GC DNA positivity rates did vary more significantly in this population. Among the symptomatic patients, sites 1 through 3 demonstrated a higher rate of positivity for both CT and GC infection (>5%) as compared to sites 4 and 5. The variation observed in the positivity rates when determining the presence of Chlamydia trachomatis and/or Neisseria gonorrhoeae are influenced by population characteristics such as age, sex, and risk factors,
{13}------------------------------------------------
| TestSite | n | CT-IDPositiveonly | GC-IDPositiveOnly | CT-ID andGC-IDPositive(Coinfected) | CT-ID and/orGC-IDPositive | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Population | No.Pos | %Pos | No.Pos | %Pos | No.Pos | %Pos | No.Pos | %Pos | ||
| Symptomatic | 1 | 358 | 36 | 10.1 | 29 | 8.1 | 11 | |||
| 2 | 279 | 17 | 6.1 | 17 | 6.1 | 9 | 3.2 | 43 | 15.4 | |
| 3 | 223 | 30 | 13.5 | 12 | 5.4 | 4 | 1.8 | 46 | 20.6 | |
| 4 | 162 | 7 | 4.3 | 4 | 2.5 | 0 | 0.0 | 11 | 6.8 | |
| 5 | 152 | 7 | 4.6 | 0 | 0.0 | 0 | 0.0 | 7 | 4.6 | |
| All | 1174 | 97 | 8.3 | 62 | 5.3 | 24 | 2.0 | 183 | 15.6 | |
| Asymptomatic | 1 | 102 | 5 | 4.9 | 6 | 5.9 | 4 | 3.9 | 15 | 14.7 |
| 2 | 22 | 1 | 4.5 | 4 | 18.2 | 0 | 0.0 | 5 | 22.7 | |
| 3 | 83 | 3 | 3.6 | 1 | 1.2 | 0 | 0.0 | 4 | 4.8 | |
| 4 | 227 | 10 | 4.4 | 4 | 1.8 | 0 | 0.0 | 14 | 6.2 | |
| 5 | 177 | 2 | 1.1 | 1 | 0.6 | 0 | 0.0 | 3 | 1.7 | |
| All | 611 | 21 | 3.4 | 16 | 2.6 | 4 | 0.7 | 41 | 6.7 | |
| Total | 1785 | 118 | 6.6 | 78 | 4.4 | 28 | 1.6 | 253 | 14.2 |
Positivity Rates using the CT/GC Test and the CT-ID and GC-ID Test Verification System
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is arranged on a single line and is likely part of a document header or title. The words are capitalized and evenly spaced, suggesting a formal or official context.
FEB 2 9 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Mark A. Del Vecchio Associate Director, Regulatory and Clinical Affairs Digene Corporation 1201 Clopper Road Gaithersburg, Maryland 20878
Re: K981567
Trade Name: Digene Hybrid Capture® II CT/GC DNA Test Regulatory Class: II Product Code: LSL, LSK Dated: January 14, 2000 Received: January 18, 2000
Dear Mr. Del Vecchio:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections. the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Image /page/14/Picture/9 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines extending from its head, representing the department's mission to protect and promote the health and well-being of Americans.
{15}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Toutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{16}------------------------------------------------
INDICATIONS FOR USE STATEMENT
K981567 510(K) Number (if known):
Device Name: Digene Hybrid Capture® II CT/GC DNA Test
Indications for Use:
"The Digene HCII CT/GC Test is an in vitro nucleic acid hybridization assay with signal amplification using microplate chemiluminescence for the combined qualitative detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) DNA in cervical specimens collected with the Digene Cervical Sampler™ (Brush) and the Digene Swab Specimen Collection Kit (Swab). Follow-up testing using the Digene HCII CT-ID and HCII GC-ID Tests is required to identify the organism(s) present in HCII CT/GC DNA Test positive specimens. The HCII CT/GC DNA Test is indicated for use as an initial test to identify symptomatic or asymptomatic women with Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) infection.
For In Vitro Diagnostic Use."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANTOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubois
Division of Clinical I aboratory D 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
No
§ 866.3390
Neisseria spp. direct serological test reagents.(a)
Identification. Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identifyNeisseria spp. from cultured isolates. Additionally, some of these reagents consist ofNeisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence ofNeisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusNeisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.(b)
Classification. Class II (performance standards).