K Number
K981485
Manufacturer
Date Cleared
1999-11-29

(584 days)

Product Code
Regulation Number
866.3390
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digene HCII GC-ID Test is a nucleic acid hybridization assay with signal amplification using microplate chemiluminescence for the qualitative detection of N. gonorrhoeae DNA in cervical specimens collected using the Digene Cervical Sampler™ (Digene cervical brush and Digene Specimen Transport Medium) and in cervical specimens collected using the Digene Swab Specimen Collection Kit ([Swab SCK] Dacron® swab and Digene Specimen Transport Medium). The Digene HCII GC-ID Test is indicated for use as an aid in diagnosing infection with N. gonorrhoeae in symptomatic or asymptomatic females.

The HCII GC-ID Test may be used alone or as a supplemental test to the Digene HCII CT/GC Test to detect N. gonorrhoeae in specimens that are positive by the HCII CT/GC Test.

Device Description

The Digene GC-ID Test using Hybrid Capture II technology is a nucleic acid hybridization assay with signal amplification that utilizes microplate chemiluminescent detection. Specimens containing the target DNA hybridize with a specific GC RNA probe cocktail. The resultant RNA:DNA hybrids are captured onto the surface of a microplate well coated with antibodies specific for RNA:DNA hybrids. Immobilized hybrids are then reacted with alkaline phosphatase conjugated antibodies specific for RNA:DNA hybrids, and detected with a chemiluminescent substrate. Several alkaline phosphatase molecules are conjugated to each antibody. Multiple conjugated antibodies bind to each captured hybrid resulting in substantial signal amplification. As the substrate is cleaved by the bound alkaline phosphatase, light is emitted which is measured as relative light units (RLUs) on a luminometer. The intensity of the light emitted denotes the presence or absence of target DNA in the specimen.

An RLU measurement equal to or greater than the Cutoff Value indicates the presence of GC DNA in the specimen. An RLU measurement less than the Cutoff Value indicates the absence of GC DNA or GC DNA levels below the detection limit of the assay.

The GC Probe Cocktail contains a probe mixture specifically chosen to eliminate or minimize cross-reactivity with DNA sequences from human cells, other bacterial species, or Neisseria species other than gonorrhoeae. The GC Probe Cocktail supplied with the Digene GC-JD Test is complementary to approximately 9,700 bp or 0.5% of the Neisseria gonorrhoeae genomic DNA (1.9 x 108 bp) . One probe is complementary to 100% of the cryptic plasmid of 4,200 bp.

AI/ML Overview

The Digene HCII GC-ID Test is a nucleic acid hybridization assay for the qualitative detection of N. gonorrhoeae DNA. The acceptance criteria and the study proving the device meets these criteria are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria for the clinical performance (sensitivity, specificity, PPV, NPV). However, it presents the calculated performance characteristics, which serve as the "reported device performance." The non-clinical performance (precision and analytical sensitivity) demonstrates compliance with established laboratory testing standards.

Non-Clinical Performance:

Acceptance Criteria CategorySpecific MetricTarget/Range (Implicit)Reported Device Performance
PrecisionWithin Instrument Variability (Relative Light Units/Cutoff - RLU/CO %CV)Low variability %CV, suggesting consistent results within the same instrument. (Although not explicitly stated, typical precision studies aim for low %CV, often below 15-20% for quantitative assays, with potentially higher allowances for very low concentrations near the detection limit.)%CVs for RLU/CO: - Panel 1: 10.68%- Panel 2: 11.50%- Panel 3 (low concentration): 4.65%- Panel 4: 5.41%- Panel 5: 6.50%- Panel 6: 5.83%Qualitative results were 100% (54/54) in agreement with expected results.
Between Instrument Variability (RLU/CO %CV)Low variability %CV when using different instruments.%CVs for RLU/CO: - Panel 1: 1.73%- Panel 2: 1.56%- Panel 3 (low concentration): 1.10%- Panel 4: 1.37%- Panel 5: 0.69%- Panel 6: 1.91%
Within Run Variability (RLU/CO %CV)Low variability %CV for tests run in the same batch.%CVs for RLU/CO: - Panel 1: 28.26%- Panel 2: 34.94%- Panel 3 (low concentration): 10.89%- Panel 4: 8.85%- Panel 5: 13.48%- Panel 6: 13.12%
Total Precision (RLU/CO %CV)Overall low variability %CV over repeated tests under various conditions.%CVs for RLU/CO: - Panel 1: 28.20%- Panel 2: 35.42%- Panel 3 (low concentration): 11.96%- Panel 4: 9.08%- Panel 5: 13.26%- Panel 6: 12.90%
Analytical SensitivityLimit of Detection (LOD) (CFU/assay)Ability to detect N. gonorrhoeae at low concentrations across a range of isolates, auxotypes, and strains. (No specific numerical target is stated, but the aim is to establish the LOD for the test.)Varied from 25 to 50,000 CFU/assay for 114 N. gonorrhoeae isolates. Average detectable limit: 974 to 2887 CFU/assay.Overall average LOD: 1931 CFU/assay (3.8 x 104 CFU/ml).

Clinical Performance (Brush Specimens - All Patients Combined, using 1.0 RLU/CO cutoff):

Acceptance Criteria CategorySpecific MetricTarget/Range (Implicit)Reported Device Performance (All Sites Combined, 1.0 cutoff)95% Confidence Interval (All Sites Combined, 1.0 cutoff)
Clinical PerformanceSensitivityDemonstrates effective detection of N. gonorrhoeae compared to culture. (No specific numerical threshold is given, but typically clinical diagnostic tests aim for high sensitivity).92.55%85.3-97.0%
SpecificityDemonstrates accurate identification of true negatives compared to culture. (Typically, clinical diagnostic tests aim for high specificity).98.50%98.6-99.6%
PPVHigh probability that a positive test result correctly indicates infection.82.08%81.3-94.8%
NPVHigh probability that a negative test result correctly indicates no infection.99.44%98.9-99.8%

Note: The document provides results for both 1.0 and 2.5 RLU/CO cutoffs. The table above uses the 1.0 RLU/CO cutoff as it generally represents higher sensitivity, and the primary table uses this as the first reported value. Differences for the 2.5 cutoff are parenthetically provided in the original text.

2. Sample Sizes and Data Provenance

  • Test Set Sample Size (Clinical Study): 1825 clinical specimens.
    • Brush specimens: 1359 (Table 3)
    • Swab specimens: 466 (Table 4)
  • Data Provenance: Prospective. Specimens were collected from patients at 5 different clinical sites (STD, Family Planning, and OB/GYN clinics) for the multi-center clinical trial. Countries of origin are not specified but implied to be within the US given the FDA submission.

3. Number of Experts and Qualifications for Ground Truth

  • The ground truth for the clinical study was primarily established using Gonorrhea culture. This is a laboratory-based method, not typically requiring "experts" in the sense of physicians or radiologists making subjective calls, but rather trained microbiologists performing and interpreting the cultures.
  • For specimens that were Digene GC-ID Test-positive/culture-negative, PCR testing was performed. This also relies on laboratory expertise. The document states that "Digene GC-ID Test results were NOT resolved by PCR test results and therefore PCR had no impact on the calculations of the Digene GC-ID Test performance characteristics." This implies PCR was used for further investigation but not to alter the primary ground truth against which sensitivity/specificity were calculated.

4. Adjudication Method for the Test Set

The primary ground truth for the clinical study was Gonorrhea culture. For discordant results (GC-ID positive/Culture negative), PCR testing was performed for additional information, but the document explicitly states that the Digene GC-ID Test results were not resolved by PCR test results for the calculation of performance characteristics. Therefore, there was no expert adjudication process to determine a "final" ground truth from multiple conflicting results; culture was the single arbiter for the main performance calculations.

In the section "Retesting of HCII CT/GC Test Positive Specimens using the GC-ID Test," combined GC Culture and PCR was used as a reference standard, and an equivocal zone was defined for the GC-ID test, where specimens falling in this zone could be retested.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not mentioned or conducted. This device is an in vitro diagnostic test that directly detects nucleic acids, not an imaging device requiring human interpretation, so human reader studies are not applicable.

6. Standalone (Algorithm Only) Performance

Yes, the study primarily evaluated the standalone performance of the Digene HCII GC-ID Test. The reported sensitivity, specificity, PPV, and NPV are measures of the algorithm's (test's) performance independent of human interpretation or assistance, against a gold standard (culture).

7. Type of Ground Truth Used

  • Clinical Study: The primary ground truth for the clinical performance evaluation was Gonorrhea culture.
  • Discordant Resolution (for investigational purposes, not main metrics): PCR testing was used to further investigate Digene GC-ID Test-positive/culture-negative specimens.
  • Retesting HCII CT/GC Positive Specimens: Combined GC Culture and PCR.

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" for the Digene HCII GC-ID Test in the context of an algorithm or machine learning model. This device is a biochemical assay, not a software algorithm that undergoes machine learning training. The data presented is for validation and performance characteristics.

9. How the Ground Truth for the Training Set Was Established

As noted above, there is no mention of a "training set" in the machine learning sense. The device's design and probe cocktail selection (complementary to N. gonorrhoeae genomic DNA, minimizing cross-reactivity) would be based on scientific knowledge and laboratory development, not iterative training on a labeled dataset.

{0}------------------------------------------------

510(k) Summary Digene GC-ID Test Hvbrid Capture® II

INTENDED USE:

The Digene HCII GC-ID Test is a nucleic acid hybridization assay with signal amplification using microplate chemiluminescence for the qualitative detection of N. gonorrhoeae DNA in cervical specimens collected using the Digene Cervical Sampler™ (Digene cervical brush and Digene Specimen Transport Medium) and in cervical specimens collected using the Digene Swab Specimen Collection Kit ([Swab SCK] Dacron® swab and Digene Specimen Transport Medium). The Digene HCII GC-ID Test is indicated for use as an aid in diagnosing infection with N. gonorrhoeae in symptomatic or asymptomatic females.

The HCII GC-ID Test may be used alone or as a supplemental test to the Digene HCII CT/GC Test to detect N. gonorrhoeae in specimens that are positive by the HCII CT/GC Test.

DECRIPTION OF THE DEVICE:

The Digene GC-ID Test using Hybrid Capture II technology is a nucleic acid hybridization assay with signal amplification that utilizes microplate chemiluminescent detection. Specimens containing the target DNA hybridize with a specific GC RNA probe cocktail. The resultant RNA:DNA hybrids are captured onto the surface of a microplate well coated with antibodies specific for RNA:DNA hybrids. Immobilized hybrids are then reacted with alkaline phosphatase conjugated antibodies specific for RNA:DNA hybrids, and detected with a chemiluminescent substrate. Several alkaline phosphatase molecules are conjugated to each antibody. Multiple conjugated antibodies bind to each captured hybrid resulting in substantial signal amplification. As the substrate is cleaved by the bound alkaline phosphatase, light is emitted which is measured as relative light units (RLUs) on a luminometer. The intensity of the light emitted denotes the presence or absence of target DNA in the specimen.

An RLU measurement equal to or greater than the Cutoff Value indicates the presence of GC DNA in the specimen. An RLU measurement less than the Cutoff Value indicates the absence of GC DNA or GC DNA levels below the detection limit of the assay.

The GC Probe Cocktail contains a probe mixture specifically chosen to eliminate or minimize cross-reactivity with DNA sequences from human cells, other bacterial species, or Neisseria species other than gonorrhoeae. The GC Probe Cocktail supplied with the Digene GC-JD Test is complementary to approximately 9,700 bp or 0.5% of the Neisseria gonorrhoeae genomic DNA (1.9 x 108 bp) . One probe is complementary to 100% of the cryptic plasmid of 4,200 bp.

SIMILARITIES AND DIFFERENCES TO PREDICATE DEVICE:

The Gen-Probe® Pace® 2 System for Neisseria Gonorrhoeae is the predicate device to which Digene claims the HCII GC-ID Test is substantially equivalent. The Gen-Probe Pace 2 test is a legally marketed device made available for commercial distribution in the United States as of April 26, 1994 after FDA cleared 510(k) premarket notification K920301. Both the Pace 2 test and the Digene Hybrid Capture® II GC-ID Test share the same intended use. Analytical and clinical data have been submitted to demonstrate that the Digene device is as safe and effective as the Gen-Probe® device.

{1}------------------------------------------------

NON-CLINICAL PERFORMANCE:

Precision

A precision study was performed at three sites to determine the within run and fotal precision of Digene's HCII GC-ID Test using a panel of positive and negative masked, simulated clinical specimens. In addition, the intra- and inter-instrument precision observed with the two luminometers recommended for use with the HCII GC-ID Test (Models DML2000 and MLX) was assessed using the same specimen panel. One of the three sites experienced difficulties with other Hybrid Capture II assays being performed as part of this study that were attributable to assay technique most likely caused by improper or inadequate training. Although, the HCII GC-ID Test precision testing results were unaffected, the technologist performing the testing was retrained in the proper assay technique.

Table 1 shows the performance of the Digene GC-ID Test for all sites combined (including the site that experienced technical problems prior to retraining the technologist in the proper assay technique). The assay demonstrated equivalent precision after technologist retraining, however, for panel member 3 (which contained low concentrations of GC organism), the RLU/CO values observed were within or near the assay's equivocal zone of 1.0-2.5. For the purposes of these data analyses, all of those RLU/CO values that fell within the equivocal zone or exceeded 2.5 were interpreted as positive. Although not evident from this table, the qualitative results were 100% (54/54) (93.4%-100% 95%CI) in acreement with expected results at the three sites.

PanelMemberNMeanWithin InstrumentBetweenInstrumentWithin RunTotal
(SD)(%CV)(SD)(%CV)(SD)(%CV)(SD)(%CV)
1540.09740.010410.68180.00171.73280.027528.25560.027528.1978
2540.09670.011111.50310.00151.56180.033834.93620.034235.4230
3543.23350.15024.64620.03561.09970.352010.88690.386611.9551
4543.84070.20785.40920.05251.36710.34018.85410.34879.0802
55416.16761.05076.49860.11220.69402.178813.47662.143713.2589
65418.07041.05395.83210.34561.91242.370113.11582.331612.9027

Table 1 Within Instrument, Between Instrument, Within Run, Total Precision Estimates For RLU/CO By Test and Target

For panel member 3, which has a low concentrations of GC organism, the RLU/CO values observed were within or near the assay's equivocal zone of 1.0-2.5. For the purposes of these data analyses, all of those RLU/CO values that fell within the equivocal zone or exceeded 2.5 were interpreted as positive.

Analytical Sensitivity

The analytical sensitivity (limits of detection) of the Digene GC-ID Test was determined by directly comparing a dilution series of a specimen panel consisting of 114 separate isolates of Neisseria gonorrhoeae. The 114 isolates represented 13 auxotypes, 5 serovars. 10 antibiotic resistant strains, 6 plasmidless strain isolates, and 2 uncharacterized isolates found discordant in the multicenter trial. Four-point dilution series of each of the isolates were tested once using the Digene GC-ID Test to establish the limits of detection for the test. The limit of detection for each Neiserria auxotype is summarized in Table 2. The detectable limit range stated was the dilution

{2}------------------------------------------------

of each auxotype that was detected within or very near to the assay's equivocal zone of 1.0-2.5 RLU/CO.

The analytical sensitivity of the GC-ID Test varied from 25 to 50,000 CFU/assay for the 114 Neisseria gonorrhoeae isolates tested, including auxotypes, serovars, plasmidless, and antibioticresistant strains. Only one of the six plasmidless strains and one of five of the Neisseria gonorrhoeae IA-5 serovars tested were detected at 50,000 CFU/assay; none of the other 112 isolates were detected at concentrations in excess of 5000 CFU/assay. The average detectable limit for all 114 isolates ranged from 974 to 2887 CFU/assay when taking into consideration isolate dilutions that fell both within the assay's equivocal zone and above 2.5 RLU/CO. The overall average limit of detection was 1931 CFU/assay (3.8 x 104 CFU/ml). Clinical specimens that contain organism at or near the limit of detection may need to be retested by an alternate test procedure or on a new specimen from the patient as defined in the Interpretation of Results section of this package insert."

AuxotypeDetectable Concentration
CFU/mlCFUs/assay
N. gonorrhoeae Auxotype 1100050
N. gonorrhoeae Auxotype 12500-500025-250
N. gonorrhoeae Auxotype 16103-10450-500
N. gonorrhoeae Auxotype 22104-105500-5000
N. gonorrhoeae Auxotype 5500-500025-250
N. gonorrhoeae Auxotype 95 x 1042500
N. gonorrhoeae Auxotype AHU (5 isolates)104-105500-5000
N. gonorrhoeae Auxotype Arg (5 isolates)104-105500-5000
N. gonorrhoeae Auxotype AU (5 isolates)103-10450-500
N. gonorrhoeae Auxotype PAU (5 isolates)103-10550-5000
N. gonorrhoeae Auxotype Pro (5 isolates)104-105500-5000
N. gonorrhoeae Auxotype Proto (5 isolates)103-10450-500
N. gonorrhoeae Ciprofloxacin Intermediate (Cipl) (5 isolates)103-10550-5000
N. gonorrhoeae Ciprofloxacin Resistant (Cip R) (4 isolates)103-10450-500
N. gonorrhoeae CMRNG (5 isolates)104-105500-5000
N. gonorrhoeae Other- 5423104-105500-5000
N. gonorrhoeae Other- 5658103-10450-500
N. gonorrhoeae PenR (5 isolates)104-105500-5000
N. gonorrhoeae Plasmidless strains (6 isolates)103-10650-50,000
N. gonorrhoeae PPNG 3.05 (5 isolates)104-105500-5000
N. gonorrhoeae PPNG 3.2104-105500-5000
N. gonorrhoeae PPNG 4.4 (4 isolates)103-10550-5000
N. gonorrhoeae Serovar IA-1 or IA-2 (5 isolates)104-105500-5000
N. gonorrhoeae Serovar IA-5 (4 isolates)104-106500-50,000
N. gonorrhoeae Serovar IB-1 (5 isolates)103-10450-500
N. gonorrhoeae Serovar IB-4 or IB-15 (5 isolates)103-10550-5000
N. gonorrhoeae Serovar IB-5 (5 isolates)103-10550-5000
N. gonorrhoeae Spectinomycin Resistant (SpeR)1055000
N. gonorrhoeae TetR (5 isolates)103-10550-5000
N. gonorrhoeae TRNG American (5 isolates)104-105500-5000
N. gonorrhoeae TRNG Dutch (5 isolates)104-105500-5000
N. gonorrhoeae Type Strain500-500025-250

Table 2: Summary of Detectable Limits of Sensitivity for GC Auxotypes, Serovars, Plasmidless, and Antibiotic-resistant Strains

{3}------------------------------------------------

CLINICAL PERFORMANCE:

Digene GC-ID Test performance characteristics were determined by comparing the assay results to results of Gonormea culture. 1825 specimens were tested from patients at 5 different sites including STD, Family Planning and OB/GYN clinics. PCR testing was performed for specimens that were Digene GC-ID Test-positive/culture-negative. Digene GC-ID Test results were NOT resolved by PCR test results and therefore PCR had no impact on the calculations of the Digene GC-ID Test performance characteristics. Results from the clinical trial for specimens collected with the Digene Cervical Sampler (cervical brush) are shown in Table 3 and specimens collected with the Digene Swab SCK in Table 4.

The performance characteristics of the GC-ID Test were calculated applying both a 1,0 and a 2,5 cutoff without consideration of the presumptive positive specimens falling in the equivocal zone described in the Interpretation of Results section of this package insert. Therefore, the performance of the test may vary in your laboratory depending on the distribution of values that fall within the equivocal zone and the repeat results obtained when retesting presumptive (equivocal zone) specimens is performed. As a point of reference, less than 0.9% of the specimens (17/1825) tested during the Multicenter clinical study used to establish this test's performance fell into this range.

Sufficient data have not been generated to accurately determine whether the sensitivity and positive predictive value of the Hybrid Capture® II GC-ID test using the Digene Swab SCK (Dacron® swab) is equivalent to the sensitivity and positive predictive value observed with specimens collected using the Digene Cervical Sampler™. Since the use of the Digene Cervical Sampler™ is contraindicated in the collection of cervical specimens from pregnant women, the ability of the test to detect the presence of GC DNA may be reduced in this population of patients or whenever a Dacron® swab is used for specimen collection".

Performance estimates for the assay are based on specimens stored at 2-8℃ or frozen and tested within 1-2 weeks of collection.

{4}------------------------------------------------

Table 3 Digene GC-ID Test versus GC Culture Results for Brush Specimens

Performance Characteristics were calculated utilizing RLU/CO cutoff values of 1.0 and 2.5 ane presented below.
Values stated parenthetically represent differences in the poin

GC-ID:POSPOSNEGNEG
Culture:POSNEGPOSNEG
Siten=SensitivityPPVSpecificityNPVGC-ID+/CultTested + PCR
Symptomatic135139(38)7(3)1(2)304(308)97.50(95.00)84.78(92.68)97.75(99.04)99.67(99.35)5/7(2/3)
95% CI83.1-99.980.1-98.597.2-99.898.2-100
2188132416976.4786.6798.8397.691/2
95% CI50.1-93.259.5-98.395.8-99.994.2-99.4
3233146(3)1212(215)93.3370.00(82.35)97.25(98.62)99.540#/6
95% CI68.1-99.856.6-96.296.0-99.797.4-100
4163400159100.00100.00100.00100.00N/A
95% CI39.8-10039.8-10097.7-10097.7-100
All93570(69)15(8)6(7)844(851)92.11(90.79)82.35(89.61)98.25(99.07)99.29(99.18)6#/15
95% CI83.6-97.180.1-95.498.2-99.698.5-99.7
Asymptomatic110110(9)20(1)89100.00(90.00)83.33(81.82)97.80100.00(98.89)2/2
95% CI69.2-10051.6-97.992.3-99.795.9-100
21220010100.00100.00100.00100.00N/A
95% CI15.8-10015.8-10069.2-10069.2-100
3841(0)00(1)83100.00(0.00)100.00100.00100.00(98.81)N/A
95% CI(2.5-100)(2.5-100)(95.7-100)(95.7-100)
422642(0)1219(221)80.0066.67(100.00)99.10(100.00)99.551/2(N/A)
95% CI28.4-99.539.8-10098.3-10097.5-100
510001NANA100.00100.00NA
95% CI2.5-1002.5-100
All42417(15)4(2)1(3)402(404)94.44(83.33)80.95(88.24)99.01(99.51)99.75(99.26)3/4(2/2)
95% CI72.7-99.963.6-98.598.2-99.998.6-100
ALL145249(47)9(5)1(3)393(397)98.00(94.00)84.48(90.38)97.76(98.76)99.75(99.25)7/9(4/5)
95% CI89.4-10079.0-96.897.1-99.698.6-100
2200152417978.9588.2498.9097.811/2
95% CI54.4-94.063.6-98.596.1-99.994.5-99.4
331715(14)6(3)1(2)295(298)93.75(87.50)71.43(82.35)98.01(99.00)99.66(99.33)0#/6
95% CI69.8-99.856.6-96.297.1-99.898.1-100
438982(0)1378(380)88.8980.00(100.00)99.47(100.00)99.741/2(N/A)
95% CI51.8-99.763.1-10099.0-10098.5-100
510001NANA100.00100.00NA
95% CI2.5-1002.5-100
All135987(84)19(10)7(10)1246(1255)92.55(89.36)82.08(89.36)98.50(99.21)99.44(99.21)9#/19
95% CI85.3-97.081.3-94.898.6-99.698.9-99.8
GC-ID:Culture:POSPOSPOSNEGNEGPOSNEGNEGSensitivityPPVSpecificityNPVGC-ID+/Cult-Tested+PCR
Symptomatic172005100.00100.00100.00100.00N/A
95% CI15.8-10015.8-10047.8-10047.8-100
292132(0)176(78)92.8686.67(100.00)97.44(100.00)98.70(98.73)0/2
95% CI66.1-99.875.3-10095.4-10093.2-100
352003100.00100.00100.00100.00N/A
95% CI15.8-10015.8-10029.2-10029.2-100
516203(1)0159(161)NA0.0098.15(99.38)100.001#/3
95% CI2.5-10096.6-10097.7-100
All266175(1)1243(247)94.4477.27(94.44)97.99(99.60)99.59(99.60)1#/5
95% CI72.7-99.954.6-99.995.4-10097.8-100
Asymptomatic110001N/ANA100.002.5-100100.002.5-100N/A
2102008100.00100.00100.00100.00N/A
95% CI15.8-10015.8-10063.1-10063.1-100
320002NANA100.00100.00N/A
95% CI15.8-10015.8-100
410001N/ANA100.00100.00N/A
95% CI2.5-1002.5-100
5186100185100.00100.00100.00100.00N/A
95% CI2.5-1002.5-10098.0-10098.0-100
All200300197100.00100.00100.00100.00N/A
95% CI29.2-10029.2-10098.1-10098.1-100
All182006100.00100.00100.00100.00N/A
95% CI15.8-10015.8-10054.1-10054.1-100
2102152(0)184(86)93.7588.24(100.00)97.67(100.00)98.82(98.85)0/2
95% CI69.8-99.863.6-10091.9-10093.6-100
372005100.00100.00100.00100.00N/A
95% CI15.8-10015.8-10047.8-10047.8-100
410001N/ANA100.00100.00N/A
95% CI2.5-1002.5-100
534813(1)0344(346)100.0025.00(50.00)99.14(99.71)100.001#/3
95% CI2.5-1001.3-98.798.4-10098.9-100
All466205(1)1440(444)95.2480.00(96.24)98.88(99.78)99.771#/5
95% CI76.2-99.959.3-99.997.4-10098.0-100

1 This information is provided for information only, specimen results were not resolved using PCR. 2 Site number 4 did not have any swab specimens from symptomatic patients.

# in two cases PCR was not done. NA = Not Applicable

{5}------------------------------------------------

Table 4 Digene GC-ID Test versus GC Culture Results for Swab Specimens

Performance Characteristics were calculated utilizing RLU/CO cutoff values of 1.0 and 2.5 are presented below. Values stated parenthetically represent differences in the point estimates when performance was calculated considering the 2.5 RLU/CO Cutoff.

1 This information is provided for information only, specimen results were not resolved using

ced for in

PCR. Site number 4 did not have any swab specimens from symptomatic patients.

in two cases PCR was not done. #

NA = Not Applicable

.

2

{6}------------------------------------------------

The clinical sensitivity and specificity of GC-ID for detecting those patients with clinically active infection that can be transmitted to partners or cause GC-related sequelae has not been determined in comparison to all commercially-available NAA methods for detection of GC DNA. In clinical studies, testing by a modified commercial NAA assay showed positivity in some GC-ID positive specimens obtained from culture negative patients. Estimated sensitivity is based on the number of GC-ID positive results found in patients who were culture positive for N. gonorrhoeae. Therefore, the GC-ID sensitivity can only be deduced relative to culture positivity which may have a sensitivity of 60-85%.

Retesting of HCII CTIGC Test Positive Specimens using the GC-ID Test

A summary of the performance of the GC-ID Test when used for the retesting of specimens initially positive with the Digene Hybrid Capture® II CT/GC Test is presented below. The results have been stratified by the collection device used to collect the speciment the Digene Cervical Sampler (designated in Table 6 as "Brush") and the Digene Specimen Collection Kit (designated in Table 6 as "Swab"). A total of nine specimens (all using the Brush) that were positive by culture and PCR combined tested negative with the CT/GC test (0.6%, 9/1560). Twelve of the 144 CT/GC positive specimens that tested negative with the GC-ID Test also tested negative with the Digene CT-ID Test. Only two of those 144 specimens were GC culture/PCR positive and 118 were determined to be CT culture positive. No additional NAA testing was performed on these specimens.

As indicated in Table 6, five of the nine CT/GC positive specimens which fell into the equivocal zone upon retesting with the GC-ID Test confirmed positive by GC culture and PCR combined. As suggested by these five specimens, the usefulness of the GC-ID to confirm the presence of GC DNA in specimens that test positive by the CT/GC is not compromised by interpreting GC-ID Test equivocal specimens as presumptive positive, as instructed in the "Interpretation of Results" section of this package insert.

Table 5Summary of HCII GC-ID Test Results Obtained for Specimens Tested with the HC
CT/GC Test
n=1825
HCII Test ResultsCombined GC Culture and PCR
CT/GCGC-IDnBrushSwab
POSNEGPOSNEG
POSPOS112883210
EQUIV95202
NEG1441116126
TOTAL265941212228
NEGPOS31200
EQUIV72302
NEG1550611300414
TOTAL1560911350416

Of particular interest are the 32 specimens positive by HCII CT/GC and determined by GC culture and CT culture/DFA to be coinfected with these organisms; all but two (6.3%) of these coinfected specimens confirmed positive by the GC-ID Test. Based on these data, the GC-ID test is effective for confirming the presence of GC DNA in specimens that yield an initial positive result with the HCII CT/GC Test.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three lines that resemble a person with their arms raised.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV 2 9 1999

Mr. Mark A. Del Vecchio Associate Director, Regulatory and Clinical Affairs Digene 9000 Virginia Manor Road Beltsville, Maryland 20705

Re: K981485 Trade Name: Digene Hybrid Capture® II GC-ID Test Regulatory Class: II Product Code: LSL Dated: September 10, 1999 Received: September 13, 1999

Dear Mr. Del Vecchio:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{8}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours.

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory. Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(K) Number (if known): K981485

Device Name: Digene Hybrid Capture® II GC-ID Test

Indications for Use:

The Digene HCII GC-ID Test is a nucleic acid hybridization assay with signal amplification using microplate chemiluminescence for the qualitative detection of N. gonorrhoeae DNA in cervical specimens collected using the Digene Cervical Sampler™ (Digene cervical brush and Digene Specimen Transport Medium) and in cervical specimens collected using the Digene Swab Specimen Collection Kit ([Swab SCK] Dacron® swab and Digene Specimen Transport Medium). The Digene HCII GC-ID Test is indicated for use as an aid in diagnosing infection with N. gonorrhoeae in symptomatic females.

The HCII GC-ID Test may be used alone or as a supplemental test to the Digene HCII CT/GC Test to detect N. gonorrhoeae in specimens that are positive by the HCII CT/GC Test.

For In Vitro Diagnostic Use.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANTOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois

510(k) Numb

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use_

§ 866.3390

Neisseria spp. direct serological test reagents.(a)
Identification. Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identifyNeisseria spp. from cultured isolates. Additionally, some of these reagents consist ofNeisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence ofNeisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusNeisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.(b)
Classification. Class II (performance standards).