(188 days)
G960236
Not Found
No
The summary describes a passive-fixation endocardial pacing lead and its performance characteristics. There is no mention of AI or ML in the intended use, device description, or performance studies. The focus is on the physical design, materials, and electrical performance of the lead.
Yes
The device is a pacing lead, designed for use with implantable pulse generators for therapeutic purposes in patients requiring cardiac pacing.
No
Explanation: The device, SYNOX leads, is described as a pacing lead designed for use with implantable pulse generators for long-term pacing and sensing. Its function is to deliver electrical impulses and sense cardiac activity, which is a therapeutic function rather than a diagnostic one. While it senses, this sensing is for the purpose of closed-loop pacing, not for independently diagnosing conditions.
No
The device description clearly describes a physical, implantable pacing lead with specific materials, shapes, and dimensions, not a software-only product.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the SYNOX leads are designed for use with implantable pulse generators for cardiac pacing and sensing. This is a therapeutic and diagnostic function performed in vivo (within the living body), not in vitro (outside the living body, typically on biological samples).
- Device Description: The description details an implantable medical device (pacing leads) designed to be placed within the heart.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), reagents, or laboratory procedures, which are hallmarks of IVD devices.
The device is a cardiac pacing lead, which is a type of implantable medical device used to deliver electrical impulses to the heart.
N/A
Intended Use / Indications for Use
SYNOX passive-fixation endocardial leads are safe and effective bipolar leads designed for use with implantable pulse generators which require bipolar 3.2 mm IS-1 compatible pacing leads. SYNOX bipolar passive-fixation endocardial leads are available in straight and "J"-shaped conformations, for pacing and sensing in the ventricle and atrium, respectively. The designation SX xx-JBP refers to SYNOX "J"-shaped leads, which are available in lengths of 45 and 53 cm; SX xx-BP refers to SYNOX straight leads, which are available in lengths of 53 or 60 cm.
The SX xx-JBP model has a permanent bend proximal to both lead electrodes, approximately 40 mm from the lead tip, resulting in the distal portion of the lead body having what is commonly referred to as a "U" or "J" shape. This lead shape facilitates placement in the right atrial appendage.
SYNOX leads are designed for use with implantable pulse generators which require pacing leads with a bipolar 3.2 mm IS-1 connector configuration; they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pulse generator therapy is medically indicated. This indication follows that recommended in the Class I definition of the ACC/AHA Task Force Report, entitled "Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmic Devices" (JACC, Vol. 18, No. 1, July 1991:1 - 13).
Product codes (comma separated list FDA assigned to the subject device)
DTB
Device Description
SYNOX passive-fixation endocardial leads are safe and effective bipolar leads designed for use with implantable pulse generators which require bipolar 3.2 mm IS-1 compatible pacing leads. SYNOX bipolar passive-fixation endocardial leads are available in straight and "J"-shaped conformations, for pacing and sensing in the ventricle and atrium, respectively. The designation SX xx-JBP refers to SYNOX "J"-shaped leads, which are available in lengths of 45 and 53 cm; SX xx-BP refers to SYNOX straight leads, which are available in lengths of 53 or 60 cm.
The SX xx-JBP model has a permanent bend proximal to both lead electrodes, approximately 40 mm from the lead tip, resulting in the distal portion of the lead body having what is commonly referred to as a "U" or "J" shape. This lead shape facilitates placement in the right atrial appendage.
These leads provide long-term safe and effective pacing through overall quality of design, manufacture and material biocompatibility. All SYNOX patient contact materials are commonly used in market-released leads. Acute and chronic biocompatibility tests have been performed, as well as long-term implantation studies. In addition, corrosion studies have been completed to address both long-term toxicity and durability of the surface iridium treatment. The testing conducted for biocompatibility as well as extensive clinical experience confirms that iridium is safe for use as an implantable material, and analyses supporting this view have been published within technical journals. Long-term corrosion testing results substantiate that iridium is a non-toxic and durable material for use in implantable devices.
Additional qualification testing results validate the safety and effectiveness of the lead design and materials used. In all cases the lead or lead component met or exceeded test specifications.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
ventricle, atrium, right atrial appendage
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
NONCLINICAL STUDIES:
Qualification testing was performed to evaluate the final device as well as various manufacturing processes. All applicable national and international standards and/or criteria were evaluated. In all cases, test specifications were met. Tests were performed in the following categories:
- mechanical, electrical and environmental testing
- IS-1 BP connector testing
- Si tube abrasion testing
- transport and sterility testing
- introducer and stylet testing
Further, in order to address the issue of constant flexion experienced by endocardial leads, the SYNOX lead body was subjected to accelerated flex testing to evaluate its performance relative to other market-released BIOTRONIK leads. SYNOX conductor coils exhibited superior fatigue resistance as compared to the market-released control lead: therefore, fatigue resistance of the SYNOX lead is expected to be at least as good as the control lead.
An additional test was developed to evaluate the rigidity of the lead in its distal portion, as suggested in recent FDA guidance. Test results demonstrate that SYNOX lead tip rigidity is less than that of other marketed endocardial leads.
An in vivo endocardial lead clinical study was sponsored by BIOTRONIK GmbH & Co in Hungary. Ten dogs were chronically implanted with three different configurations of endocardial leads with the objective to assess the electrical performance, biocompatibility, and biostability, including corrosion resistance, of endocardial leads with fractal iridium surface structured electrodes. The fractal surface lead results were compared to results from a control lead which contained porous, sintered electrode surfaces. The implant duration was three months for half of the study group and six months for the rest of the study population. The SYNOX tip electrode with a surface area of 1.3 mm2 was one of the lead configurations tested.
Electrochemical and corrosion resistance analysis results from the in vivo Hungarian dog investigation indicate that fractal iridium structured electrode surfaces are stable against any measurable corrosion attack when electrically loaded and chronically implanted. Results from a histological evaluation of the tissue in the vicinity of the electrode tips showed no observable iridium, titanium or platinum in either the three month implant samples or the six month samples. Based upon these findings and other performance data collected, the investigators concluded that the endocardial leads tested have excellent long term electrophysiological properties.
CLINICAL STUDIES:
A prospective, non-randomized clinical study was initiated on January 27, 1997, under IDE G960236 to evaluate the safety and effectiveness of the SYNOX pacing lead. This study is currently ongoing; the control lead used in this clinical investigation is the BIOTRONIK POLYROX lead, which is the predicate device to the SYNOX lead.
Primary objective: determine whether the SYNOX pacing leads are safe and effective endocardial pacing leads.
Key results:
- There were no reports of any unanticipated adverse device effects occurring during the clinical investigation.
- There have been no reports of safety issues related to the chronic use of the pacing lead.
- One lead that was explanted due to loss of capture met all electrical lead specifications upon analysis.
- The other two SYNOX lead explants were not due to lead performance but rather because the patient required an ICD implant.
- The four patient deaths during the study have not been related to the pacing system.
- All of the anticipated adverse events reported, except for the one lead explant described above, were resolved by the investigator through reprogramming, drug therapy, or lead repositioning.
- The combined rates of anticipated and unanticipated adverse events for the SYNOX leads are not statistically different from the rates for the control leads.
- As expected, due to the decreased lead tip area (1.3 mm2 for SYNOX vs 3.5 mm² for POLYROX) the mean measured atrial and ventricular lead pacing impedance for SYNOX leads was more than 150 ohms over the mean measured atrial and ventricular lead pacing impedance for the control leads at all of the follow-up intervals.
- The mean capture and sensing thresholds of the SYNOX leads were equal to or better than the control leads for all of the follow-up intervals.
- The atrial sensing and pacing performance of the SYNOX leads were comparable to the control leads.
- The ventricular sensing and pacing performance of the SYNOX leads were comparable to the control leads.
In addition to the SYNOX IDE lead study, another SYNOX lead study was conducted in France to evaluate the performance of the ventricular SYNOX pacing leads. Study results were consistent with the findings of the SYNOX IDE clinical study.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Not Found
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.
(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
0
1.1 Safety and Effectiveness Summary
SYNOX passive-fixation endocardial leads are safe and effective bipolar leads designed for use with implantable pulse generators which require bipolar 3.2 mm IS-1 compatible pacing leads. SYNOX bipolar passive-fixation endocardial leads are available in straight and "J"-shaped conformations, for pacing and sensing in the ventricle and atrium, respectively. The designation SX xx-JBP refers to SYNOX "J"-shaped leads, which are available in lengths of 45 and 53 cm; SX xx-BP refers to SYNOX straight leads, which are available in lengths of 53 or 60 cm.
The SX xx-JBP model has a permanent bend proximal to both lead electrodes, approximately 40 mm from the lead tip, resulting in the distal portion of the lead body having what is commonly referred to as a "U" or "J" shape. This lead shape facilitates placement in the right atrial appendage.
These leads provide long-term safe and effective pacing through overall quality of design, manufacture and material biocompatibility. All SYNOX patient contact materials are commonly used in market-released leads. Acute and chronic biocompatibility tests have been performed, as well as long-term implantation studies. In addition, corrosion studies have been completed to address both long-term toxicity and durability of the surface iridium treatment. The testing conducted for biocompatibility as well as extensive clinical experience confirms that iridium is safe for use as an implantable material, and analyses supporting this view have been published within technical journals. Long-term corrosion testing results substantiate that iridium is a non-toxic and durable material for use in implantable devices.
Additional qualification testing results validate the safety and effectiveness of the lead design and materials used. In all cases the lead or lead component met or exceeded test specifications.
Field clinical experience as well as the in vitro and qualification testing performed on the SYNOX lead show that the risk to the patient in using these leads is no greater than that of comparable endocardial leads. Potential adverse effects associated with the implantation of endocardial leads include, but are not limited to: fibrotic tissue formation, thrombosis, embolism, elevated thresholds, lead dislodgment, fibrillation, body rejection phenomena, cardiac tamponade, pneumothorax, muscle/nerve stimulation, infection, skin erosion, valve damage and ventricular ectopy.
1
Table 2 below summarizes some of the potential symptoms indicating a complication and possible corrective actions:
Table 2 |
---|
Lead Complications |
SYMPTOM | POTENTIAL COMPLICATION | POTENTIAL CORRECTIVE ACTION |
---|---|---|
Loss of pacing or | ||
sensing | • Electrode dislodgement | |
• Lead fracture | ||
• Setscrew penetration of | ||
lead insulation | ||
• Improper lead to pacemaker | ||
connection | • Reposition lead | |
• Replace lead | ||
• Replace lead | ||
• Reconnect lead to pacemaker | ||
Increase or decrease in | ||
threshold | • Fibrotic tissue formation | • Adjust pulse generator output; |
• Reposition lead |
1.2 Summary of Studies
1.2.1 NONCLINICAL STUDIES
Qualification testing was performed to evaluate the final device as well as various manufacturing processes. All applicable national and international standards and/or criteria were evaluated. In all cases, test specifications were met. Tests were performed in the following categories:
- mechanical, electrical and environmental testing ●
- IS-1 BP connector testing .
- Si tube abrasion testing .
- transport and sterility testing .
- introducer and stylet testing .
Further, in order to address the issue of constant flexion experienced by endocardial leads, the SYNOX lead body was subjected to accelerated flex testing to evaluate its performance relative to other market-released BIOTRONIK leads. SYNOX conductor coils exhibited superior fatigue resistance as compared to the market-released control lead: therefore, fatigue resistance of the SYNOX lead is expected to be at least as good as the control lead.
An additional test was developed to evaluate the rigidity of the lead in its distal portion, as suggested in recent FDA guidance'. Test results demonstrate that SYNOX lead tip rigidity is less than that of other marketed endocardial leads.
1 Draft Guidance for the Submission of Research and marketing for Permanent Pacemaker Leads, Version 2.1, March 24, 1997. U.S. Department of Health and Human Services, Food and Drug Administration.
2
An in vivo endocardial lead clinical study was sponsored by BIOTRONIK GmbH & Co in Hungary. Ten dogs were chronically implanted with three different configurations of endocardial leads with the objective to assess the electrical performance, biocompatibility, and biostability, including corrosion resistance, of endocardial leads with fractal iridium surface structured electrodes. The fractal surface lead results were compared to results from a control lead which contained porous, sintered electrode surfaces. The implant duration was three months for half of the study group and six months for the rest of the study population. The SYNOX tip electrode with a surface area of 1.3 mm2 was one of the lead configurations tested.
Electrochemical and corrosion resistance analysis results from the in vivo Hungarian dog investigation indicate that fractal iridium structured electrode surfaces are stable against any measurable corrosion attack when electrically loaded and chronically implanted. Results from a histological evaluation of the tissue in the vicinity of the electrode tips showed no observable iridium, titanium or platinum in either the three month implant samples or the six month samples. Based upon these findings and other performance data collected, the investigators concluded that the endocardial leads tested have excellent long term electrophysiological properties.
1.2.2 CLINICAL STUDIES
A prospective, non-randomized clinical study was initiated on January 27, 1997, under IDE G960236 to evaluate the safety and effectiveness of the SYNOX pacing lead. This study is currently ongoing; the control lead used in this clinical investigation is the BIOTRONIK POLYROX lead, which is the predicate device to the SYNOX lead.
The primary objective of the investigation is to determine whether the SYNOX pacing leads are safe and effective endocardial pacing leads. Results from the clinical investigation provide valid scientific evidence and reasonable assurance that SYNOX pacing leads are safe and effective when used in accordance with their labeling.
To date, the main objectives of the clinical investigation have been achieved:
- There were no reports of any unanticipated adverse device effects occurring during . the clinical investigation.
- There have been no reports of safety issues related to the chronic use of the pacing . One lead that was explanted due to loss of capture met all electrical lead. specifications upon analysis. The other two SYNOX lead explants were not due to lead performance but rather because the patient required an ICD implant. The four patient deaths during the study have not been related to the pacing system.
- . All of the anticipated adverse events reported, except for the one lead explant described above, were resolved by the investigator through reprogramming, drug therapy, or lead repositioning.
3
-
The combined rates of anticipated and unanticipated adverse events for the SYNOX . leads are not statistically different from the rates for the control leads.
All lead performance endpoints of the clinical investigation have been achieved: -
As expected, due to the decreased lead tip area (1.3 mm2 for SYNOX vs 3.5 mm² for . POLYROX) the mean measured atrial and ventricular lead pacing impedance for SYNOX leads was more than 150 ohms over the mean measured atrial and ventricular lead pacing impedance for the control leads at all of the follow-up intervals.
-
The mean capture and sensing thresholds of the SYNOX leads were equal to or better . than the control leads for all of the follow-up intervals.
-
The atrial sensing and pacing performance of the SYNOX leads were comparable to the . control leads.
-
The ventricular sensing and pacing performance of the SYNOX leads were comparable . to the control leads.
In addition to the SYNOX IDE lead study, another SYNOX lead study was conducted in France to evaluate the performance of the ventricular SYNOX pacing leads. Study results were consistent with the findings of the SYNOX IDE clinical study.
4
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with healthcare, consisting of a staff with two snakes coiled around it. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 0 1998
Mr. David Makanani Biotronik, Inc. 6024 Jean Road Lake Oswego, OR 97035
Re: K980869 SYNOX SX 53-BP, SX 60-BP, SX 45-JBF, SX 53-JBP, Trade Name: Models 12 III Regulatory Class: Product Code: DTB June 4, 1998 Dated: Received: June 5, 1998
Dear Mr. Makanani:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbrandinq and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for
5
Page 2 - Mr. David Makanani
devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
On August 16, 1993 the Final Rule for Device Tracking was published in the Federal Reqister, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirements of the regulation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please not the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
6
1.2 Indications for Use
SYNOX leads are designed for use with implantable pulse generators which require pacing leads with a bipolar 3.2 mm IS-1 connector configuration; they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pulse generator therapy is medically indicated. This indication follows that recommended in the Class I definition of the ACC/AHA Task Force Report, entitled "Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmic Devices" (JACC, Vol. 18, No. 1, July 1991:1 - 13).
Judy Asmussen/s Doyle Gantt
and Neurological Devices
510(k) Number K980869