K Number
K980751

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-02-24

(363 days)

Product Code
Regulation Number
880.5725
Age Range
12 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Provides controlled delivery of fluids, blood and blood products and can be used in enteral, epidural, subcutaneous, arterial and intravenous applications.

Device Description

Linear Peristaltic Infusion Pump

AI/ML Overview

Here's an analysis of the provided text regarding the Sabratek 4040 Infusion Pump, focusing on acceptance criteria and the supporting study:

1. Table of Acceptance Criteria and Reported Device Performance

Based on the provided information, the key performance characteristic highlighted and compared is the Flow Accuracy.

Acceptance Criteria (Predicate Devices - 6060 & 3030)Reported Device Performance (Sabratek 4040 Pump)Meets Acceptance Criteria?
+/-6% (Sabratek 6060 Pump)+/-5%Yes
+/-5% (Sabratek 3030 Pump)+/-5%Yes

It's important to note that the document primarily focuses on demonstrating substantial equivalence to predicate devices, rather than setting explicit acceptance criteria for a new device against a predefined benchmark. The "acceptance criteria" here are derived from the performance of the predicate devices. The Sabratek 4040 meets or exceeds the flow accuracy of both predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "Non-clinical tests included accuracy testing and comparison with Sabratek's stated accuracy claim and recording of data."

  • Sample Size: Not specified. The document does not provide any details regarding the number of units tested, the number of measurements taken, or specific test protocols that would allow for the determination of sample size.
  • Data Provenance: The document does not explicitly state the country of origin. Given the submission is to the U.S. FDA, it is highly likely that the testing was conducted domestically (US) or by a manufacturer associated with US regulatory requirements. There is no information to indicate if the data was retrospective or prospective, but for device approval, it would typically be prospective testing specifically for the submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not applicable. This device is an infusion pump, and its "ground truth" for performance is established through objective engineering and metrological measurements (e.g., flow rate calibration, volume dispensed) rather than subjective expert interpretation (like a radiologist reading an image). There would be no need for medical experts to establish a "ground truth" for the device's accuracy performance.

4. Adjudication Method for the Test Set

  • Not applicable. As the "ground truth" is based on objective measurements, there would be no subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is not relevant for an infusion pump. This type of study is typically used for diagnostic or screening devices where human readers interpret medical images or data, and the AI's impact on their performance is evaluated.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, implicitly. The "accuracy testing" mentioned is a standalone performance test of the device itself. The infusion pump operates without human intervention in its core pumping and accuracy functions. The stated performance of "+/-5%" for Flow Accuracy is a standalone performance metric of the device.

7. The Type of Ground Truth Used

  • Engineering/Metrological Measurements. The ground truth for infusion pump accuracy is objective and quantifiable. It involves precisely measuring the volume of fluid dispensed over a given time and comparing it to the programmed flow rate. This would involve calibrated equipment and measurement standards.

8. The Sample Size for the Training Set

  • Not applicable. This device is a mechanical/electronic infusion pump, not an AI/ML-based device. Therefore, there is no "training set" in the context of machine learning. The device's performance is based on its engineering design, calibration, and manufacturing specifications.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As stated above, there is no training set for this device.

{0}------------------------------------------------

2/04/99

510(k) Submission, Sabratek 4040 Infusion Pump

K980751

Sabratek, Niles, IL 60714

The following is a Summary of the Sabratek 4040 Infusion Pump substantial equivalence and safety and efficacy.

Classification Name:Infusion Pump
Common/Usual Name:Infusion Pump
Proprietary Name:Sabratek 4040 Infusion Pump
Classification:Class II Medical Device
Performance Standards:No applicable performance standardsHave been issued under Section 514 ofThe Food, Drug and Cosmetic Act.
Predicate Device:Sabratek 6060 HOMERUN® Volumetric Infusion Pump

K941984 and Sabratek 3030 Infusion Pump, K914581.

ELEMENT OFCOMPARISONSABRTATEK 4040 PUMPSABRATEK 6060 PUMPSABRATEK 3030 PUMP
510(k)#K941984
Pump typeLinear PeristalticRotary PeristalticLinear Peristaltic
Intended use(s)Intravenous, Intra-arterial,subcutaneous & intracavityinfusionsIntravenous, Intra-arterial,subcutaneous & intracavityinfusionsLinear Peristaltic
Administration Sets-Captive Standard SetYesYesNo
-Captive 1.2 Micron FilterYesYesNo
-Captive .22 Micron FilterYesYesNo
-Standard SabrasetYesNoYes
-Approved Competitive SetsYesNoYes
Free Flow Clamp-Anti Free Flow Clamp (Set)YesYesNo
-Free Flow Clamp (Pump)YesNoYes
Specific Drug, biologic use:-Blood/Blood ProductsBlood/Blood ProductsNoNo
Components-Pumping MechanismYesYesYes
-BatteryYesYesYes
-Pressure SensorYesYesYes
-Ultrasound Air In LineYesYesYes
Pumping mechanismLinear PeristalticRotary PeristalticLinear Peristaltic
Power RequirementsSwitchable90-132/138-264Vac 50/60hz12 - 18 Volts DC, 50 mamaximum110V AC, 10 watts, 50/60hz
Battery Life6-6.5 hrs at 125 ml/hr25 hrs at 125 ml/hr7-8 hrs at 125 ml/hr
Recharge Time24 hrsN/A24 hrs
Fluid Path Material:
-TubingPVCPVCPVC
-SpikeABSABSABS
-Check ValveN/AN/AN/A
-Silicone Rubber TubingSilicone RubberSilicone RubberSilicone Rubber
-1.2 or .22 Micron FilterModified acrylic/suporModified acrylic/suporModified acrylic/supor
-MLL (Male Luer Lock)ABSABSABS
-Connecting BarbsAcrylicAcrylicAcrylic
-Y-SiteAcrylicAcrylicAcrylic
-Fluid BagN/AEVAN/A
-Solvent bonding SolutionCyclohexanoneCyclohexanoneCyclohexanone
Infusion Parameters:
-Flow Rate Range0.1 - 999 ml/hr0.1 400 ml/hr1- 999 ml/hr
-Flow Accuracy, %+/-5%+/-6%+/-5%
-KO rate0.1 - 10 ml/hr0.1- 10 ml/hr1 - 9 ml/hr
-Infusion volume limit0.1 - 999.9 ml1.1 - 9999 ml1 - 999 ml
-Prime/purge modeNoYesNo
Profiles
-ContinuousYesYesYes
-Auto-RampYesYesYes
-IntermittentYesYesYes
-25 PeriodYesYesYes
-PCANoYesNo
Alarm Type:
-Air-In-LineAudible, LCDAudible, LCDAudible, LCD
-Cassette Not InstalledN/AAudible, LCDN/A
-Check Internal 9V BatteriesN/AAudible, LCDN/A
-Complete XX ml (or mg)Audible, LCDAudible, LCDAudible, LCD
-Door OpenAudible, LCDAudible, LCDAudible, LCD
-Down OcclusionAudible, LCDAudible, LCDAudible, LCD
-Empty Bag or Up OcclusionAudible, LCDAudible, LCDAudible, LCD
-Empty BatteryAudible, LCDAudible, LCDAudible, LCD
-HoldAudible, LCDAudible, LCDAudible, LCD
-Low BagNoneAudible, LCDNone
-Low BatteryAudible, LCDAudible, LCDAudible, LCD
-MalfunctionAudible, LCDAudible, LCDAudible, LCD
-Move TubingAudible, LCDNoneAudible, LCD
-Nicad DepletedAudible, LCDAudible, LCDAudible, LCD
-Rate Exceeds Microset LimitsN/AAudible, LCDN/A
-Release/Remove PCA CordN/AAudible, LCDN/A
-ReprogramAudible, LCDAudible, LCDAudible, LCD
-Stuck KeyAudible, LCDAudible, LCDAudible, LCD
-High PressureAudible, LCDAudible, LCDAudible, LCD
-Low PressureAudible, LCDAudible, LCDAudible, LCD
External CommunicationCapabilityYesYesYes

{1}------------------------------------------------

510(k) Submission, Sabratek 4040 Infusion Pump

Sabratek, Niles, IL 60714

:

{2}------------------------------------------------

IndicationsProvides controlled delivery of fluids, blood and bloodproducts and can be used in enteral, epidural, subcutaneous,arterial and intravenous applications.
ContraindicationsThe Sabratek 4040 Infusion Pump is contraindicated for usewith a blood set, other than those specifically designated assuch and manufactured by Sabratek Corporation.
Non-Clinical TestsNon-clinical tests included accuracy testing and comparisonwith Sabratek's stated accuracy claim and recording of data.
ConclusionsThe Sabratek 4040 Infusion Pump is equivalent in safety andefficacy to its predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 4 1999

Mr. Charles P . Gill Manager, Regulatory Affairs Sabratek Corporation 8111 North St.Louis Avenue Skokie, Illinois 60076

K980751 Re : Sabratek 4040 Infusion Pump Trade Name: Requlatory Class: II Product Code: FRN Dated: November 13, 1998 December 3, 1998 Received:

Dear Mr. Gill

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal You may, therefore, Food, Drug, and Cosmetic Act (Act). market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಕ್ಕೆ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of

{4}------------------------------------------------

Page 2 - Mr. Gill

the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdag.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Directbr Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

ﻴﺔ

510(k) Number (if Known)

Device Name: Sabratek 4040 Infusion Pump

Provide controlled delivery of fluids, blood and blood products and can be Indications for use: used in enteral, epidural, subcutancous, arterial and intravenous applications.

Prescription Device. Federal Law (US) restricts this device to sale by or on the order of a physician.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use -----

Labacia Cucente (Optional Format 1-2-96)
(Division)

(Division Sign-Off) Division of Denial. Infection Control, and General Hospital Devices

510(k) Number K980751

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).