K Number
K980639
Manufacturer
Date Cleared
1998-06-15

(116 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • For in vitro diagnostic use only.
  • For the qualitative and semi-quantitative detection of IgG antibodies to SSB antigen in human serum by enzyme immunoassay.
  • For use as an aid in the diagnosis of systemic rheumatic disease, particularly Sjogren's Syndrome.
  • For manual use, or for use with the HyPrep System Plus semi-automated fluid handler.
Device Description

The SeraQuest Anti-SSB test is a solid-phase enzyme immunoassay (EIA), which is performed in microwells, at room temperature, in three thirty minute incubations. It has been developed to detect IgG antibodies which are directed against SSB nuclear antigen (Anti-SSB), in human serum. The Calibrators in the SeraQuest Anti-SSB test set have been assigned Index values based on an in-house standard. Test results are reported as Index values.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study used to demonstrate the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicitly Derived from Predicate Performance)Reported Device Performance (SeraQuest Anti-SSB)
Relative sensitivity comparable to predicate device (Shield Diastat Anti-SSB)96.4% (95% CI: 89.6% to 100%)
Relative specificity comparable to predicate device (Shield Diastat Anti-SSB)97.2% (95% CI: 95.3% to 99.3%)
Overall agreement comparable to predicate device (Shield Diastat Anti-SSB)99.6% (95% CI: 95.3% to 99.1%)

Note: The document doesn't explicitly state quantitative acceptance criteria. Instead, it presents a comparison to a predicate device (Shield Diastat Anti-SSB) and implies that performance "comparable" to the predicate, as demonstrated by the study results, is acceptable for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): 288 specimens
  • Data Provenance: Not explicitly stated (e.g., country of origin). The study involved "human serum" specimens. The study design appears to be retrospective, as existing specimens were tested with both the new device and the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The study uses the Shield Diastat Anti-SSB test as the comparator or "reference" for evaluating the SeraQuest Anti-SSB test, rather than an independent expert-established ground truth.

4. Adjudication Method for the Test Set

This information is not provided. Given that the predicate device's results were used as the comparator, it's unlikely a separate adjudication process was involved beyond comparing the two test results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study compares the performance of two diagnostic tests (SeraQuest Anti-SSB vs. Shield Diastat Anti-SSB), not the performance of human readers with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The SeraQuest Anti-SSB test is a lab-based enzyme immunoassay, which is an "algorithm only" device in the context of its operation and interpretation. The results were compared directly against another lab-based assay (the predicate).

7. The Type of Ground Truth Used

The "ground truth" for this study was the results from the predicate device, the Shield Diastat Anti-SSB test. The study evaluates the substantial equivalence of the new device's performance against an existing, legally marketed device.

8. The Sample Size for the Training Set

This information is not applicable/provided. The SeraQuest Anti-SSB test is a biochemical assay (Enzyme Immunoassay) and does not involve AI or machine learning algorithms that would require a "training set" in the conventional sense. Any internal optimization or development of the assay would have used internal validation data, but this is not a training set as understood in AI/ML contexts.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided for the reasons stated in point 8.

{0}------------------------------------------------

APPENDIX 3 (revised 4/18/98).

Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. 17, g

JUN 1 5 1998

510(k) SUMMARY

K980639

Applicant:Quest International, Inc.1938 N.E. 148th TerraceNorth Miami, FL 33181
Registration No.1061839
Contact Person:Robert A. Cort, V.P., Quality Assurance
Telephone:(305) 948-8788
Telefax:(305) 948-4876
Manufacturing Site:Same as above
Device:SeraQuest® Anti-SSB
Device Name:Antinuclear antibody immunological test system (21CFR § 866.5100)
Device Classification:Class II (performance standards)

escription:

The SeraQuest Anti-SSB test is a solid-phase enzyme immunoassay (EIA), which is performed in microwells, at room temperature, in three thirty minute incubations. It has been developed to detect IgG antibodies which are directed against SSB nuclear antigen (Anti-SSB), in human serum.

The Calibrators in the SeraQuest Anti-SSB test set have been assigned Index values based on an in-house standard. Test results are reported as Index values.

Principle:

Diluted samples are incubated in wells coated with SSB antigen. Antibodies against SSB (if present) are immobilized in the wells. Residual sample is eliminated by washing and draining, and coniugate (enzyme-labeled antibodies to human IgG) is added and incubated, If IgG antibodies to SSB are present, the conjugate will be immobilized in the wells. Residual conjugate is eliminated by washing and draining, and the enzyme substrate is added and incubated. In the presence of the enzyme, the substrate is converted to a vellow end-product which is read photometrically at 405 nm.

Intended Use:

For the qualitative and semi-quantitative detection of human antibodies to SSB nuclear antigen in uman serum by enzyme immunoassay. For manual use, or for use with the HyPrep System Plus. ാ In Vitro Diagnostic Use Only.

{1}------------------------------------------------

redicate Device:

:

The SeraQuest Anti-SSB test is substantially equivalent in intended use and performance, to the Shield Diastat Anti-SSB test, Shield Diagnostics, Dundee, DD2 1SW.

Summary of technological characteristics:

CharacteristicSeraQuest Anti-SSBShield Diastat Anti-SSB
Description:Enzyme ImmunoassayEnzyme Immunoassay
Intended Use:The detection ofantibodies against SSBin human serum.The detection ofantibodies against SSBin human serum.
Solid Phase:Plastic MicrowellPlastic Microwell
Antigen Source:Calf ThymusCalf Thymus
Number of Incubation Periods:ThreeThree
ample Dilution:1:501:500
Sample IncubationDuration:30 minutes60 minutes
Incubation Temperature:Room temperatureRoom temperature
Ezyme-labeled Conjugate:
AntibodyGoat anti-human IgGGoat anti-human IgG
EnzymeAlkaline phosphataseAlkaline phosphatase
Conjugate Volume:100 μl100 μl
Conjugate IncubationDuration:30 minutes30 minutes
Substrate:p-Nitrophenylphosphatephenolthaleinmonophosphate
Subtrate Volume:Substrate IncubationDuration:100 μl30 minutes100 μl30 minutes

{2}------------------------------------------------

top Reagent:0.5 M TrisodiumphosphateSodiumHydroxide
Stop Reagent Volume:100 μl100 μl
Readout:Spectrophotometric405 nmSpectrophotometric540-565 nm

Summary of Clinical Testing:

Of the 288 specimens tested, 27 were positive, and 252 were negative in both the SeraQuest and Shield tests (please see Table 1 below). Seven specimens which were negative by the Shield test, were positive in the SeraQuest test. One specimen which was negative in the Shield test, was equivocal in the SeraQuest test. One specimen which was positive in the Shield test, was negative in the SeraQuest test.

TABLE 1.

RESULTS OF SeraQuest Anti-SSB ASSAYS AND SHIELD Anti-SSB ASSAYS OF 288 SERUM SPECIMENS.

SHIELDAnti-SSBASSAYSeraQuest Anti-SSB
PositiveEquivocalNegative%95 % Cl
Positive2701Relative sensitivity 96.489.6 to100*
Negative71252Relative specificityOverall agreement97.299.695.3 to 99.3*95.3 to 99.1*

ア Excluding equivocal results.

★ Calculated by the normal method.

Reference: Gardner, M.J. and Altman, D.G., Confidence Intervals Rather Than Hypothesis Testing. Brit. Med. J., 292: 746-750, 1986.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing another, representing care and support.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 1 5 1998

Mr. Robert A. Cort Vice President, Quality Assurance Ouest International, Inc. 1938 N.E. 148th Terrace North Miami, Florida 33181

Re : K980639/S1 Trade Name: SeraQuest® Anti-SSB Test System Requlatory Class: II Product Code: LLL Dated: April 20, 1998 Received: April 24, 1998

Dear Mr. Cort:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. 173

APPENDIX 8 (REVISED 4/17/98)

Page 1

510(k) Number (if known): _K980639

Device Name: SeraQuest Anti-SSB

Indications For Use:

  • For in vitro diagnostic use only. 1.
    1. For the qualitative and semi-quantitative detection of IgG antibodies to SSB antigen in human serum by enzyme immunoassay.
    1. For use as an aid in the diagnosis of systemic rheumatic disease, particularly Sjogren's Syndrome.
    1. For manual use, or for use with the HyPrep System Plus semi-automated fluid handler.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rita G. Makin

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use - -

(Optional Format 1-2-96)

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).