(215 days)
The ENDO Absorbable Interference Screw is indicated for use in graft fixation in the repaid and reconstruction of the cruciate ligaments of the know, during open and arthroscoppic surgery.
The ENDO Surgical Concepts, Inc. Absorbable Interference Screws have technological characteristics that are the same as absorbable bone and interference screws currently in interstate commerce that have been registered with the Food and Drug Administration under a 510(k). The subject screws are made of absorbable polymers such as polylactic acid, polyglycolic acid, or other, or copolymers. The competitive screws are comparable in the range of sizes, screw pitch, and hex sizes. The interference screws are also comparable in terms of clinical function to metallic interference screws and ligament buttons that have been registered under a 510(k) and used for fixation of grafts, as described in the attached articles.
Here's a breakdown of the acceptance criteria and study information for the ENDO Absorbable Interference Screws, based on the provided text:
Acceptance Criteria and Device Performance
The document does not explicitly state numerical acceptance criteria in a table format. Instead, it relies on demonstrating substantial equivalence to legally marketed predicate devices and established clinical efficacy of the surgical procedure and similar devices. The core "acceptance criterion" is the demonstration that the ENDO Absorbable Interference Screws are safe and effective and function comparably to existing devices used for graft fixation in cruciate ligament repair.
Table 1: Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Inferred from Document) | Reported ENDO Absorbable Interference Screw Performance |
|---|---|
| Safety and Effectiveness for Intended Use: Bone to bone fixation, fixation of bone patella bone and semitendonosus or tendon grafts in the repair of cruciate ligaments during open or arthroscopic surgery. | Claimed Safe and Effective: "ENDO Surgical Concepts believes that the Absorbable Interference Screw System is safe and effective." |
| Technological Characteristics Equivalence: Comparable to absorbable bone and interference screws currently in interstate commerce (polylactic acid, polyglycolic acid, or other copolymers; comparable in range of sizes, screw pitch, hex sizes). | Claimed Equivalent: "The ENDO Surgical Concepts, Inc. Absorbable Interference Screws have technological characteristics that are the same as absorbable bone and interference screws currently in interstate commerce..." |
| Clinical Function Equivalence: Comparable to metallic interference screws and ligament buttons in clinical function. | Claimed Equivalent: "...interference screws are also comparable in terms of clinical function to metallic interference screws and ligament buttons." Specific studies cited show high percentages of good/excellent results for the surgical procedure using interference screws, and comparable clinical results between interference screws and ligament buttons. |
| Initial Pullout Strength: Comparable to metallic screws and well above forces expected in the immediate postoperative period. | Claimed Comparable: "In vitro testing has documented the initial pullout force for polylactic acid absorbable interference screw to be comparable to metallic screws with the force being well above the force expected to be applied to the construct in the immedicate postoperative period." |
| Minimization of known problems (e.g., graft advancement, screw damage to suture, tendon laceration): Designed to address "pitfalls" identified in literature. | Addressed in Design: Tapered tip, cannulated design with guide wire, radiused thread tips, associated cannula to protect sutures/tendons during insertion. |
| Minimization of known device malfunctions (e.g., screwdriver breakage, screw cracking): Designed to reduce issues seen in MDR reports for predicate devices. | Addressed in Design: Uses a stronger 3.5mm hex drive (vs. 2.5mm), one-piece driver shaft (no welds), designed with wall thickness and drive system to minimize cracking in the drive region. |
Study Information
The document relies primarily on a literature review and comparison to predicate devices, rather than a single, controlled clinical trial specifically for the ENDO Absorbable Interference Screws.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not applicable in the context of a dedicated "test set" for the ENDO device in a standalone study. The document refers to data from numerous previous studies, reviews, and clinical experiences.
- Data Provenance: The data referenced is from published orthopedic literature and the "reputation, clinical use, and clinical results" of the surgical procedure over "more than ten years of experience." Specific articles mentioned include authors like Lambert (1983), Kurosaka (1987), Fulkerson, Lemos, Matthews, Hulstyn, Cassin, Hecker, Shelbourne, O'Brien (1991), and Buss (1993). These likely represent data from various countries and institutions as is typical for broad orthopedic literature. The studies are predominantly retrospective analyses of surgical outcomes or in vitro mechanical testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of Experts: Not explicitly stated as a defined group for a specific "test set." The "ground truth" is established through the consensus and findings of the broader orthopedic community, as reflected in decades of published research, clinical practice, and orthopedic meetings.
- Qualifications of Experts: The experts are the authors and researchers in the orthopedic literature cited, presumably orthopedic surgeons, biomechanical engineers, and other medical professionals specializing in cruciate ligament repair and sports medicine. Their specific years of experience are not detailed in this summary.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not applicable. There was no single "test set" requiring adjudication by a panel for this submission. The "adjudication" is essentially the peer review process of clinical research and the general acceptance within the orthopedic field.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Comparative Effectiveness Study: No, this is not an AI-based device, and therefore, an MRMC study comparing human readers with and without AI assistance was not conducted. This document describes a medical device (surgical screw) for mechanical fixation.
6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done
- Standalone Performance Study: Not applicable. This is a physical medical device, not an algorithm. Its performance relies on its inherent mechanical properties and its interaction with the human body and surgical technique.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: The ground truth is primarily established through:
- Clinical Outcomes Data: High percentages of "good and excellent results" reported in orthopedic literature for the surgical procedure using interference screws.
- Expert Consensus: The long history of favorable results and the designation of the procedure as the "Gold Standard" in the orthopedic community.
- Pathology/Biomechanical Principles: In vitro mechanical testing demonstrating fixation strengths.
- Comparative Clinical Studies: Papers directly comparing interference screws with other fixation methods (e.g., ligament buttons) and showing comparable results (e.g., 93 out of 100 points on the HSS rating scale).
8. The sample size for the training set
- Training Set Sample Size: Not applicable. There is no "training set" in the context of an algorithm or AI. The design of the ENDO Absorbable Interference Screws was informed by a broad review of existing knowledge, clinical experience, and identified issues (MDR reports, literature on pitfalls), which could be considered analogous to "training data" for a product development process, but not in statistical machine learning terms.
9. How the ground truth for the training set was established
- Ground Truth for Training Set: As above, not a formal "training set." The information that guided the design of the ENDO screw came from:
- Literature reviews identifying "pitfalls in the use of interference screws" (e.g., Matthews' article).
- MDR reports detailing device malfunctions (e.g., screwdriver breakage, screw cracking) for predicate devices.
- Knowledge of biomechanical principles in surgical fixation.
- Existing established successful designs and techniques in the field.
The "ground truth" for design improvements was the identification of specific problems with previous devices or techniques that the ENDO screw aimed to mitigate.
{0}------------------------------------------------
8 1998 SEP
10.1 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the reguirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K980457
Applicant Information:
| Date Prepared: | 1/15/98 |
|---|---|
| Name: | ENDO Surgical Concepts, Inc. |
| Address: | 15 East Putnam Avenue #250 Greenwich, CT 06830 |
| Contact Person: | Alan Small |
| Phone Number: | 203-622-8423 |
| Fax Number | 203-622-1250 |
Device Information:
ENDO Absorbable Interference Screws Trade Name: Common Name: Bone or Interference Screw Classification Name: Bone or Interference Screw
Equivalent Device:
ENDO Surgical Concepts, Greenwich, CT, Interference Screws K935836 Linvatec, Clearwater, FL. Concept Absorbable Interference Screws or Bioscrew K933719, K952831
Acufex. Mansfield. MA Interference Screws and Ligament Button and Bioabsorbable Interference Screw K943548 Richards, Memphis, TN, Hewson Ligament Button
Intended Use:
Bone to bone fixation, fixation of bone patella bone and semitendonosus or tendon grafts in the repair of cruciate ligaments during open or arthroscopic surgery.
Summary:
The ENDO Surgical Concepts, Inc. Absorbable Interference Screws have technological characteristics that are the same as absorbable bone and interference screws currently in interstate commerce that have been registered with the Food and Drug Administration under a 510(k). The subject screws are made of absorbable polymers such as polylactic acid, polyglycolic acid, or other, or copolymers. The competitive screws are comparable in the range of sizes, screw pitch, and hex sizes. The interference screws are also comparable in terms of clinical function to metallic interference screws and ligament buttons
{1}------------------------------------------------
that have been registered under a 510(k) and used for fixation of grafts, as described in the attached articles.
The safety and effectiveness of interference screws in general is demonstrated in the reputation, clinical use, and clinical results that the surgical procedure of reconstruction of the anterior cruciate ligament with a bone patellar tendon bone graft fixed with interference screws has seen over the more than ten years of experience with its use. The high percentage of good and excellent results reported with this procedure at orthopedic meetings and in the orthopedic literature (some of which is attached to this application) has led it to be referred to as the "Gold Standard" to which comparison of new devices and procedures is made (Otero et al). In vitro mechanical testing has demonstrated the effectiveness of mechanical fixation of bone using interference screws.
Literature related to laboratory and clinical studies on interference screw fixation describe a long history of favorable results with the use of interference screws in graft fixation. The clinical article by Lambert in 1983 is the first documented use of interference screws and he described the use of the technique in 200 patients followed over 5 years. He used standard cancellous screws for fixation and found a high percentage of good to excellent results. The articles on in vitro testing document fixation strengths in the early postoperative period that are higher than other fixation methods, such as sutures in combination with a screw and washer . These forces are described as higher than the forces on the graft in the early rehabilitative period (Matthews et al, Hulstyn et al. Cassin et al). The study by Kurosaka in 1987 showed increased fixation strength with a larger diameter headless screw, compared to the standard cancellous screw, and this screw design has been used clinically since then with good results. Some effect on pullout strength was noticed by Fulkerson if the screw was inserted more than 30 degrees off axis, but a clinical paper by Lemos found the divergence to average below ten degrees. The comparison of arthroscopic versus conventional surgical technique was studied in vitro (Hecker et al. Cassin et al) and in vivo (Shelbourne et al) with similar results being attained. The clinical article by Shelbourne concluded that "similar early clinical results can be successfully achieved" with the techniques. Most of the articles have recommendations on surgical technique for optimum insertion. Other studies presented orally at local meetings have described the favorable use of interference fixation techniques.
Comparative papers from the same institution (The Hospital for Special Surgery in NY, NY) published in The Journal of Bone and Joint Surgery by O'Brien in 1991 and Buss in 1993 compare directly the use of interference screws with ligament buttons for fixation of a patellar tendon graft in anterior cruciate reconstruction surgery. In both 93 of a possible 100 points were attained on the rating scale of the Hospital for Special Surgery. The patients were followed for a minimum of two years. Sutures are used to hold the bone
{2}------------------------------------------------
plug to a ligament button, whereas with interference screws direct bone to bone apposition is attained. This clinical study documents a direct comparison of the use of interference screws with another well known method, ligament buttons, in a total of 150 patients followed for a minimum of two years.
In vitro testing has documented the initial pullout force for polylactic acid absorbable interference screw to be comparable to metallic screws with the force being well above the force expected to be applied to the construct in the immedicate postoperative period. Clinical studies have shown the polylactic acid absorbable interference screw to be equivalent to metallic interference screws
ENDO Surqical Concepts, Inc. has made a reasonable search of available information about interference screws and their use in an effort to have a comprehensive background on the types of problems to which the devices are susceptible. While it is meant to be thorough such a search is necessarily limited by available resources. This search has brought to light the description in the literature in an article by L. Matthews, called "Pitfalls in the Use of Interference Screws for Anterior Cruciate Ligament Reconstruction" (article attached) of some potential problems with the use of interference screws. These include: inadvertent graft advancement, screw damage to the passing suture, and tendon laceration by screw threads. ENDO Surgical has used the product design to minimize these potential problems. The author of the article states that "an interference screw with a more tapered and less blunt tip may engage the bone pluq more readily and thereby reduce risk of miaration before graft engagement". The ENDO screw has a tip that is tapered to aid in the insertion process. A cannulated screw with quide wire helps to maintain alignment of the screw with the hole and the ENDO screw is used with a quide wire. In the area of screw damage to the passing suture and tendon laceration ENDO has designed the screw with thread tips that are radiused slightly broader than the competitive products to make the thread less sharp and prone to cutting. ENDO is also providing the associated cannula to cover the screw and protect the sutures and tendons from the screw during insertion. The author of the paper L. Matthews also describes some modifications of surgical techniques that can reduce the risk of these potential problems. This demonstrates that the use of the screws is effected by surgical technique.
The search also identified some MDR reports regarding device malfunctions on interference screw products manufactured by Acufex Microsurgical and Linvatech (formerly Concept, Inc.) in a search of the DIOGENES database, These reports related to the breakage or bending of the screwdrivers used to insert the interference screw and the cracking of absorbable interference screws on insertion in the hex region. The ENDO system can use a 3.5 mm hex drive, which is substantially stronger than a 2.5 mm driver due to its larger tip size. The driver shaft is also made of one piece of material, with no welds that can act
{3}------------------------------------------------
as potential sites for failure or crack initiation. The fewer than twenty MDR reports located for 1991, 1992 and 1993 compare to the annual volume use of interference screws of approximately 200,000 units in each of those years. The percentage of MDR reports for the number of devices used is approximately .003%, based on these numbers. The ENDO absorbable screw has been designed with a wall thickness and drive system that should minimize the potential problem of cracking in the drive region.
The use of interference screws in graft fixation is believed to be safe and effective based on the over ten years of clinical use with good results and mechanical testing studies supporting the biomechanical principles of interference fixation. The available MDR reports relate mostly to the insertion instruments and drive features not to the interference screws fixation thus the rate of complications of the interference screws appears to be very low. Even with this low rate ENDO has improved on these aspects of the device to try to reduce the complication rate further and learn from the experience of others. ENDO Surgical Concepts believes that the Absorbable Interference Screw System is safe and effective.
Alan D. Small
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
8 1998 SEb
Mr. Alan A. Small President ENDO Surgical Concepts, Inc. 15 East Putnam Avenue #250 Greenwich, Connecticut 06830
K980457 Re: ENDO Absorbable Interference Screw Trade Name: Regulatory Class: II Product Codes: HWC and MRY Dated: June 30, 1998 Received: July 6, 1998
Dear Mr. Small:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ಗ್ಗ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Alan A. Small
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamaim.nhtml".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K980457
Device Name: ENDO Absorbable Interference Screw
Indications For Use:
The ENDO Absorbable Interference Screw is indicated for use in graft fixation in the repaid and reconstruction of the cruciate ligaments of the know, during open and arthroscoppic surgery.
(PLRASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
code
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K980457
cription Use 21 CFR 801.109)
OR
Over-The-Counter Use
N/A