(130 days)
WinSCAN is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC).
The WinSCAN System is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC). The operator selects the recommended parameters for nucleated cells, starts the scan and can walk away. The system locates and stores color images of the nucleated cells for each slide, the operator reviews the data on the WinSCAN monitor, and classifies the nucleated cells as nRBC or WBC based on stain and nuclear characteristics. Identification parameters include light adsorption, size and shape.
Automatic relocation, capture and archiving of the cell images are performed by the instrument based upon operator selection. The instrument also can be used in the manual mode to systematically scan a slide and examine each field.
The WinSCAN System consists of the following components:
- Intel-based PC with Windows operating system .
- Monitor, kevboard, mouse .
- Switchstick control unit t
- Color printer for images and texts .
- Microscope with brightfield capability and 10x, 20x, 40x and 50x . objectives
- Set of transmission and excitation filters .
- Camera and image acquisition board t
- Motorized devices: stage, filter wheels, focus drive 0
- Motor controller unit and manual controls .
The Intel-based PC operates the instrument through the software program that coordinates control of the automated microscope. The base operating system of the PC consists of MicroSoft Windows 3.1 and DOS version 6.0. When the WinSCAN icon is selected from the Program Manager, WinSCAN software is downloaded and operation of the system can begin. The WinSCAN program includes all facilities to scan, relocate, and acquire images from a microscope slide. The WinSCAN software also performs image acquisition and scan functions, as well as archiving and relocation of images and facilitates operator review of cells. Classification of cells is performed by the operator.
The Monitor. Kevboard, Mouse, and switchstick control unit provide user interfaces to the system. The monitor displays all information to the user, including menus for navigating through the WinSCAN operation, image acquisition results and allows for operator visual review for operator classification of the nucleated cells found on a slide. The keyboard allows for user entry of scan parameters, slide numbers, user identification, comments, and annotation of results. The mouse/keyboard is used to interact with the instrument and software, including selecting menus and functions of the WinSCAN software and instrument.
A color printer is included with the system that allows the user to print results of scanning or acquired images of cells.
Here's a breakdown of the acceptance criteria and study information for the WinSCAN Automated Imaging System, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Correlation coefficient with manual method | >/= 0.98 |
| Precision (Coefficient of Variation) | 24% - 26% CV (compared to 23% CV for manual microscopy) |
Study Details
2. Sample sizes for the test set and data provenance:
The document does not explicitly state the sample size used for the test set. It mentions "30 samples" for precision testing but doesn't specify if this is the full test set for correlation.
Data Provenance: Not specified.
3. Number of experts used to establish the ground truth for the test set and their qualifications:
The document does not provide information on the number of experts or their qualifications used to establish ground truth for the test set. The device itself relies on "operator interaction and decision making to identify nucleated red blood cells" and "operator review," implying human classification is integral, but it doesn't specify how the ground truth for testing was established.
4. Adjudication method for the test set:
The document does not specify an adjudication method. Given that cell classification is performed by an operator even in the device's intended use, it's possible that the "manual microscopy method" against which the device was compared served as a form of ground truth or reference, but no formal adjudication process is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size:
The document mentions a comparison to the "standard manual microscope method" and states that "Results of studies demonstrate that WinSCAN is an effective tool that aids the operator in locating and classifying nucleated cells." It does not explicitly describe an MRMC study designed to measure the improvement in human readers' performance with AI assistance. The study focuses on the correlation of the device's output (when used by an operator) with manual methods, and the precision of the system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No, a standalone study was not performed. The WinSCAN system is explicitly designed for "operator interaction and decision making" and "operator review" for cell classification. The device "aids the operator in locating" cells, but the final classification is human-driven.
7. The type of ground truth used:
The ground truth appears to be based on the "standard manual microscope method." This implies human expert classification from traditional microscopy.
8. The sample size for the training set:
The document does not specify a sample size for a training set. Given the date (1998) and the description of the device as an "automated imaging system" that aids an operator in classification rather than an autonomous decision-making AI, it's likely that the "AI" component is more geared towards image processing, cell location, and presentation, rather than complex machine learning that requires a distinct training phase as understood today.
9. How the ground truth for the training set was established:
As no training set is mentioned (or implied in a modern AI sense), the method for establishing its ground truth is not applicable/not provided. The device's "identification parameters include light adsorption, size and shape," suggesting rule-based or conventional image processing techniques, rather than supervised machine learning requiring labeled training data.
{0}------------------------------------------------
Summary of Safety and Effectiveness
Submitter
Applied Imaging 2380 Walsh Avenue, Building B Santa Clara, California 95051 Phone: 408-562-0250 Fax: 408-562-0264
Date summary was prepared
January 27, 1998
Name of the device
WinSCAN Automated Imaging System
Identification of predicate device
Micro 21 System (also known as IMS-200): K925670
Description of the device
The WinSCAN System is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC). The operator selects the recommended parameters for nucleated cells, starts the scan and can walk away. The system locates and stores color images of the nucleated cells for each slide, the operator reviews the data on the WinSCAN monitor, and classifies the nucleated cells as nRBC or WBC based on stain and nuclear characteristics. Identification parameters include light adsorption, size and shape.
{1}------------------------------------------------
Automatic relocation, capture and archiving of the cell images are performed by the instrument based upon operator selection. The instrument also can be used in the manual mode to systematically scan a slide and examine each field.
The WinSCAN System consists of the following components:
- Intel-based PC with Windows operating system .
- Monitor, kevboard, mouse .
- Switchstick control unit t
- Color printer for images and texts .
- Microscope with brightfield capability and 10x, 20x, 40x and 50x . objectives
- Set of transmission and excitation filters .
- Camera and image acquisition board t
- Motorized devices: stage, filter wheels, focus drive 0
- Motor controller unit and manual controls .
The Intel-based PC operates the instrument through the software program that coordinates control of the automated microscope. The base operating system of the PC consists of MicroSoft Windows 3.1 and DOS version 6.0. When the WinSCAN icon is selected from the Program Manager, WinSCAN software is downloaded and operation of the system can begin. The WinSCAN program includes all facilities to scan, relocate, and acquire images from a microscope slide. The WinSCAN software also performs image acquisition and scan functions, as well as archiving and relocation of images and facilitates operator review of cells. Classification of cells is performed by the operator.
The Monitor. Kevboard, Mouse, and switchstick control unit provide user interfaces to the system. The monitor displays all information to the user, including menus for navigating through the WinSCAN operation, image acquisition results and allows for operator visual review for operator classification of the nucleated cells found on a slide. The keyboard allows for user entry of scan parameters, slide numbers, user identification, comments, and annotation of results. The mouse/keyboard is used to interact with the instrument and software, including selecting menus and functions of the WinSCAN software and instrument.
A color printer is included with the system that allows the user to print results of scanning or acquired images of cells.
{2}------------------------------------------------
Intended Use
The WinSCAN System is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC) and enumerate them as #nRBC/100 white blood cells (WBC).
Comparison of device characteristics to predicate
The WinSCAN instrument is substantially equivalent to the Micro 21 system. Both instruments can automatically scan a slide, both instruments allow for operator review, and both instruments allow for viewing images on a computer monitor. Unlike the Micro 21 system, the WinSCAN system includes a manual microscope that allows the user to view slides manually for verification. Both systems can relocate a cell or object on a slide, even upon its removal and replacement.
Testing
Results of studies demonstrate that WinSCAN is an effective tool that aids the operator in locating and classifying nucleated cells. The test results demonstrate that the WinSCAN is able to scan slides prepared from peripheral blood and aid the operator in locating white blood cells and nucleated red blood cells. When compared to the standard manual microscope method, the correlation coefficient is >/= 0.98. Additionally, precision testing using replicate counts of 30 samples demonstrates that WinSCAN performs in a repeatable manner, with a mean coefficient of variation of 24% - 26% over multiple samples, instruments and operators, as compared to 23% CV% precision demonstrated by manual microscopy.
Conclusion
The WinSCAN instrument is substantially equivalent to the Micro 21 system. Both instruments can automatically scan a slide for nucleated cells, both instruments allow for operator review, and both instruments allow for viewing images on a computer monitor. Unlike the Micro 21, WinSCAN includes a manual microscope that the user can use to view nucleated cells on a slide manually for verification.
Both systems can relocate a cell or object on a slide, even upon its removal and replacement. Both systems require operator interaction for cell classification and for the enumeration of nRBCs/100 WBCs.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. The logo is black and white and has a simple, clean design. The eagle is facing to the right.
Public Health Service
JUN - 9 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Patricia A. Milbank, JD, RAC Vice President, Requlatory Affairs and Quality Assurance Applied Imaqinq 2380 Walsh Avenue, Bldg. B Santa Clara, California 95051
K980375/S1 Re : WinSCAN Automated Imaging System Trade Name: Requlatory Class: II Product Code: JOY Dated: May 6, 1998 Received: May 7, 1998
Dear Ms. Milbank:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure ............
{5}------------------------------------------------
Page of of ___
| 510(k) Number (if known): | K980375 |
|---|---|
| Device Name: | WinSCAN Automated Imaging System |
Indications For Use:
Intended Use
WinSCAN is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Deborah Morie
for Dr. Peter Massim
(Division Sign-Ofm
Division of Clinical Laboratory Devices
510(k) Numbor.
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
§ 864.5260 Automated cell-locating device.
(a)
Identification. An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)(b)
Classification. Class II (performance standards).