K Number
K980190

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-03-06

(45 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the Chiron Diagnostics ACS 180: Automated Chemiluminescence Systems. The test is intended for use as an aid in monitoring patients previously treated for Stage III breast cancer. Serial testing for CA 27,29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression of disease in response to treatment.

Device Description

The Chiron Diagnostics ACS:180 BR assay is a fully automated, competitive, chemiluminescent assay. One reagent, designated Lite Reagent, is composed of a mouse monoclonal antibody specific for CA 27.29, labeled with acridinium ester. The antibody used in the assay, MAb B27.29, binds to a peptide epitope in the tandem repeat region of the MUC-1 gene product. The Solid Phase is composed of purified breast cancer antigen (CA 27.29) which is covalently coupled to paramagnetic particles (PMP). The patient serum sample is incubated with both reagents simultaneously for 7.5 minutes. The ACS: 180 system automatically performs the following steps: dispenses sample into a cuvette, dispenses Lite Reagent and Solid Phase and incubates for 7.5 minutes, separates, aspirates and washes the cuvettes, dispenses reagents which initiate the chemiluminescent reaction, reports results. An inverse relationship exists between the concentration of CA 27.29 in a sample and the relative light units (RLU) detected by the system.

AI/ML Overview

Chiron Diagnostics ACS:180 BR - Acceptance Criteria and Study Details

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state formal "acceptance criteria" in a numerical or target-based format for the clinical performance of the ACS:180 BR. Instead, it presents the assay's performance characteristics and then concludes that these were "substantially equivalent" to the predicate device. Therefore, the "acceptance criteria" can be inferred as achieving substantial equivalence to the predicate device's performance.

Performance MetricInferred Acceptance Criteria (Substantial Equivalence to Predicate Device)Reported Device Performance (ACS:180 BR)
For monitoring recurrence (1 Value > ULN):Similar sensitivity, specificity, PPV, NPV to predicate.Sensitivity: 60% (39, 79), Specificity: 93% (87, 96), PPV: 60% (39, 79), NPV: 93% (87, 96)
For monitoring recurrence (2 Consecutive Values > ULN):Similar sensitivity, specificity, PPV, NPV to predicate.Sensitivity: 48% (28, 69), Specificity: 98% (94, 100), PPV: 80% (52, 96), NPV: 91% (85, 95)
For managing metastatic disease (Sensitivity to Change):Similar sensitivity to predicate.Sensitivity to Change in Disease Status: 73% (58, 85)
For managing metastatic disease (Specificity to Change):Similar specificity to predicate.Specificity to Change in Disease Status: 71% (53, 85)
For managing metastatic disease (PPV to Change):Similar PPV to predicate.Positive Predictive Value to Change in Disease Status: 77% (61, 88)
For managing metastatic disease (NPV to Change):Similar NPV to predicate.Negative Predictive Value to Change in Disease Status: 67% (49, 81)
Correlation with Predicate Device (203 specimens):Strong correlation (e.g., r > 0.90, slope ~ 1, y-intercept ~ 0).r = 0.96, slope = 1.05, y-intercept = 6 U/mL
Correlation with Predicate Device (103 breast cancer specimens):Strong correlation (e.g., r > 0.90, slope ~ 1, y-intercept ~ 0).r = 0.96, slope = 1.04, y-intercept = 8 U/mL
Analytical Sensitivity<3.5 U/mL (Inferred from reported value)<3.5 U/mL
Assay Upper Limit of Normal (ULN)Established through testing healthy volunteers.38.6 U/mL (based on 314 healthy volunteers)

2. Sample Size and Data Provenance for Test Set

  • Sample Size for Recurrence Monitoring Study: 162 patients for the primary analysis of clinical performance as an aid in monitoring recurrence. These patients provided 942 specimens.
  • Sample Size for Metastatic Disease Management Study: 97 breast cancer patients with metastatic disease. Calculations were based on the 79 patients who had assay levels above the ULN.
  • Sample Size for Correlation Study: 203 specimens overall, with a subset of 103 women with histologically confirmed breast cancer.
  • Data Provenance: The studies were described as "multicenter-prospective studies" (for recurrence monitoring and metastatic disease management). The specific countries of origin are not explicitly stated, but the submission is to the U.S. FDA by a U.S. company, suggesting North American or international sites. The "healthy volunteer donors" for ULN determination were tested at "three laboratories."

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the test set.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method like 2+1 or 3+1. For the recurrence monitoring study, "Breast cancer recurrence was established using clinical symptoms, and/or general or specific imaging techniques; e.g., x-ray, mammogram, magnetic resonance imaging, computerized tomography, ultrasound, bone or liver scan." This indicates a clinical diagnosis process, but no specific adjudication protocol is mentioned.

For the metastatic disease management study, disease progression or regression was used as an endpoint, but the adjudication of these changes is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was described. The device is an in vitro diagnostic test (an automated assay), not an image reading device that would typically involve multiple human readers. The comparison was the performance of the assay against clinical outcome and against a predicate device, rather than human reader performance with and without AI assistance.

6. Standalone (Algorithm Only) Performance

Yes, the studies presented are for the standalone performance of the ACS:180 BR assay. It's an automated chemiluminescence system measuring a biomarker, so its performance is inherently "algorithm only" in terms of measurement and result reporting. The clinical utility is then evaluated by how well these results correlate with actual disease status.

7. Type of Ground Truth Used

  • For Recurrence Monitoring: "Disease recurrence as defined by clinical outcome." This was established using "clinical symptoms, and/or general or specific imaging techniques; e.g., x-ray, mammogram, magnetic resonance imaging, computerized tomography, ultrasound, bone or liver scan." This indicates a combination of clinical follow-up and imaging.
  • For Metastatic Disease Management: "Changing disease status" (progression or regression) determined clinically, with a 20% increase in assay levels used as an indication of progressing disease.
  • For Specificity (Non-breast malignancies/Other conditions): Diagnosis of the specific malignancy or condition.
  • For Correlation with Predicate Device: The value obtained from the predicate device (Biomira TRUQUANT® BR™ RIA).
  • For ULN determination: Serum samples from "healthy volunteer donors."

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning. The ACS:180 BR is a quantitative immunoassay, not an AI/ML algorithm that is "trained" in the typical sense. The "training" in this context would be the assay development and optimization, which isn't described with a specific dataset size.

9. How Ground Truth for Training Set Was Established

As noted above, a distinct "training set" with established ground truth, as understood in machine learning, is not applicable or described for this type of in vitro diagnostic assay. Instead, the assay is developed based on chemical and biological principles to detect a specific antigen. The performance studies described (clinical performance, analytical sensitivity, specificity, etc.) serve as validation of the assay's accuracy and utility.

{0}------------------------------------------------

K '180140

MAR - 6 1998

Chiron Corporation 4560 Horton Street Emeryville, California 94608-2916 510.655.8730 Fax 510.923.3307

SUMMARY OF SAFETY AND EFFECTIVENESS - ACS: 180 BR

This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safe Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.

Date of Summary Preparation:

Company Name:

CHIRON DIAGNOSTICS

Company Contact:

Chiron Diagnostics Corporation 333 Coney Street East Walpole, MA 02032

January 16, 1998

Nancy Hornbaker Regulatory Affairs Chiron Diagnostics Chiron Corporation 4560 Horton Street Emeryville, CA 94608

510.923.2758 Telephone Number: 510.923.3344 Fax:

ACS:180 BR Automated Chemiluminescence System

Automated Tumor Associated Antigen

Class II device

Biomira TRUQUANT® BR™ RIA PMA P950011; 510(k) K965141

Device Name:

Common or Usual Name:

Classification:

Predicate Device:

{1}------------------------------------------------

Intended Use and Indications for Use:

Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the Chiron Diagnostics ACS 180: Automated Chemiluminescence Systems. The test is intended for use as an aid in monitoring patients previously treated for Stage III breast cancer. Serial testing for CA 27,29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression of disease in response to treatment.

Description of the Device:

The Chiron Diagnostics ACS:180 BR assay is a fully automated, competitive, chemiluminescent assay. One reagent, designated Lite Reagent, is composed of a mouse monoclonal antibody specific for CA 27.29, labeled with acridinium ester. The antibody used in the assay, MAb B27.29, binds to a peptide epitope in the tandem repeat region of the MUC-1 gene product. The Solid Phase is composed of purified breast cancer antigen (CA 27.29) which is covalently coupled to paramagnetic particles (PMP). The patient serum sample is incubated with both reagents simultaneously for 7.5 minutes.

The ACS: 180 system automatically performs the following steps:

  • . dispenses sample into a cuvette
  • dispenses Lite Reagent and Solid Phase and incubates for 7.5 minutes 행
  • separates, aspirates and washes the cuvettes 1
  • dispenses reagents which initiate the chemiluminescent reaction 월
  • reports results

An inverse relationship exists between the concentration of CA 27.29 in a sample and the relative light units (RLU) detected by the system.

{2}------------------------------------------------

Specific Performance Characteristics:

Analytical Sensitivity

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Comments of the montage of the management of the management of the management with a fire a many of the commended to be a li control and the commended to be a li controlled t
<3.5 U/mL
Analytical Sensitivity

Assay Upper Limit of Normal

A total of 314 healthy volunteer donors were tested at three laboratories to determine the normal range of the assay and to determine the upper limit of normal (ULN). Following good laboratory practice, each laboratory should determine its own normal range and ULN.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------And Separate de March****
1Normalunder. 00010. 14001L
Marcel Comments of Children and Children and Children Children Children Children Children Children Children Children Children Children Children Children Andress of Children C

Assay Specificity

The specificity of the assay was determined by testing serum samples from patients with active malignancies other than breast and serum from patients with other diseases and conditions. Results from the clinical study are presented below.

Patients with non breast malignancies frequently have raised levels of the CA 27.29 antigen as indicated by their specificities. CA 27.29 is normally expressed by most epithelial tissues and over-expressed by many epithelial cancers. It is not a breast specific antigen.

Specificity of the ACS180:BR in Patients with Malignancies other than Breast

MalignancyNSpecificity
Colon4374.4%
Liver2045.0%
Lung4757.5%
Ovary5044.0%
Pancreas4553.3%
Prostate3482.4%
Stomach2993.1%
Uterus3086.7%

{3}------------------------------------------------

ConditionNSpecificity
Benign Breast Disease
Breast Adenoma6195.08%
Fibrocystic Breasts6898.53%
Cirrhosis2576.00%
Endometriosis2491.67%
Lactating Woman3789.19%
Mild Chronic Hepatitis2885.71%
Ovarian Cyst5094.00%
Pregnancy4997.96%
Renal Impairment2085.00%
Severe Chronic Hepatitis2095.00%

Specificity of the ACS:180 BR in Patients with Other Diseases And Conditions

Potentially Interfering Substances

There are no known cross-reactants for CA 27.29 as measured by the ACS:180 BR assay.

The potential interference of chemotherapeutic agents, therapeutic drugs, and tumor marker antigens was tested by adding these substances to five serum pools containing CA 27.29 ranging from 20.0 to 445.8 U/mL. The level of CA 27.29 in each of these pools was then determined using the ACS:180 BR assay and normalized to the level without the respective drugs or antigens.

SubstanceMean %RecoverySubstanceMean %Recovery
Acetaminophen (10X)96.3Granisetron HCl (10X)99.9
Cimetidine (40X)100.9Lorazepam (10X)98.0
Ciprofloxacin (10X)100.1Megestrol acetate (5X)105.2
Codeine (10X)96.9Methotrexate (10X)105.8
Cyclophosphamide (1X)96.2Morphine (10X)95.6
Dexamethasone (10X)100.2Ondansetron (10X)93.3
Diphenhydramine HCl (10X)101.8Paclitaxel (2X)101.7
Doxorubicin (10X)95.9Prochlorperazine (10X)96.4
Etopside (2X)99.5Tamoxifen (10X)96.6
5-Fluorouracil (10X)104.2Vinorelbine tartrate (10X)104.9

{4}------------------------------------------------

AntigenMean % Recovery
Carcinoembryonic antigen(CEA)103.8
Ovarian Cancer antigen(CA125)105.9
GI Cancer antigen(CA 19.9)103.6

Correlation with Predicate Device

The performance of the ACS:180 BR assay was compared with that of the predicate device in a study of 203 specimens. These specimens had CA 27.29 levels that spanned the range from 7 to 994 U/mL. The results of the linear regression analysis indicated that the two methods were correlated. The correlation coefficient (r) was 0.96; the slope was 1.05 and y-intercept was 6 U/mL.

The CA 27.29 results from a subset of this population, 103 women with histologically confirmed breast cancer, were also highly correlated (r = 0.96). In this analysis, the slope was 1.04 and the y-intercept was 8 U/mL.

Clinical Performance of the ACS:180 BR

Aid in Monitoring Patients Previously Treated for Stage II or Stage III Breast Cancer

One clinical study of the ACS:180 BR comprised 942 specimen from 162 patients who were enrolled in a multicenter-prospective study. These patients were previously treated for Stage II or Stage III breast cancer and had no evidence of disease (NED) at the time of study entry (at least one year post-surgery). Patients were monitored over the length of the study (2.4 -24 months) by determining the CA 27.29 levels in serial blood specimens drawn at each visit with their clinician. All patients included in the ACS:180 BR clinical study were lymph node positive at time of surgery and had at least three serial bleeds after study enrollment. Two laboratories tested all specimens using the ACS:180 BR assay and the predicate device.

Sensitivity, specificity, and predictive values (PPV and NPV) of the ACS:180 BR as an aid in monitoring the recurrence of breast cancer were calculated against disease recurrence as defined by clinical outcome . Of the 162 patients included in the

Image /page/4/Picture/9 description: The image shows a circular object that is partially shaded. The left side of the circle is lighter, while the right side is dark, creating a crescent shape. The shading suggests a light source illuminating the object from the left.

1 Breast cancer recurrence was established using clinical symptoms, and/or general or specific imaging techniques; e.g., x-ray, mammogram, magnetic resonance imaging, computerized tomography, ultrasound, bone or liver scan.

{5}------------------------------------------------

study, 25 showed clinical evidence of disease recurrence. The utility of the ACS:180 BR assay as an aid for the detection of breast cancer recurrence was analyzed under two scenarios:

    1. Using any single ACS:180 BR result greater than the ULN (38.6 U/mL) prior to disease recurrence
    1. Using two consecutive ACS:180 BR results greater than the ULN (38.6 U/mL) prior to disease recurrence.

Assay results above the ULN must have occurred prior at or before the clinical diagnosis of recurrence in order to be considered predictive of disease recurrence.

Assay ParameterOne value > ULNTwo Consecutive Values > ULN
Sensitivity60%(39, 79)*)48%(28,69)
Specificity93%(87, 96)98%(94, 100)
Positive Predictive Value60%(39, 79)80%(52, 96)
Negative Predictive Value93%(87, 96)91%(85, 95)
  • Numbers in parentheses represent the 95% confidence intervals.

By comparison, the sensitivity of the predicate device, using one value > ULN, ranged from 58% to 62% at the two sites, specificity ranged from 53% to 97%, PPV ranged from 23% to 73%, and NPV ranged from 84% to 94%. Using two consecutive values > ULN, the sensitivity of the predicate device at the two sites ranged from 38% to 42%, specificity ranged from 86% to 100%, PPV ranged from 42% to 100%, and NPV ranged from 86% to 92%.

Correlation with Predicate Device

The sensitivity, specificity and predictive values determined in this study were substantially equivalent to those of the predicate device.

{6}------------------------------------------------

Aid in Management of Breast Cancer Patients with Metastatic Disease

This clinical study was a multicenter prospective study of 97 breast cancer patients with metastatic disease. All patients received treatment during the study. Patients were enrolled at six sites and all ACS:180 Br testing was performed at three laboratories. The clinical sensitivity and specificity of the ACS:180 Br assay in relation to changing disease status in metastatic breast cancer patients were determined using a 20% increase in assay levels as an indication of progressing disease. The positive predictive value was calculated against disease progression. Calculations were based on the 79 patients who had assay levels above the ULN. Patients were considered to have completed the trial at the time of diagnosis of progressing or regressing disease. Patients with stable disease were followed from two to seven months. The following table summarizes the study results:

MeasureEstimate
Sensitivity to Change in Disease Status73%(58, 85)*
Specificity to Change in Disease Status71%(53, 85)
Positive Predictive Value to Change in Disease Status77%(61, 88)
Negative Predictive Value to Change in Disease Status67%(49, 81)
  • Numbers in parentheses represent the 95% confidence interval.

Results from this study demonstrated that changes in CA 27.29 levels, as measured by the ACS:180 Br assay, can be a predictor of disease status and response to therapy in patients with metastatic breast cancer.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

MAR - 6 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Nancy Hornbaker Regulatory Affairs Chiron Diagnostics Corporation 4560 Horton Street Emeryville, California 94608

Re : K980190 ACS:180 BR (CA 27.29) Automated Chemiluminescence Trade Name: System Regulatory Class: II Product Code: MOI Dated: January 16, 1998 Received: January 20, 1998

Dear Ms. Hornbaker:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth. in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{8}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Page ( of )

510(k) Number (if known):K,980190
Device Name:ACS: 180.

Indications For Use:

ADDENDUM

The Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the Chiron Diagnostics ACS:180° Automated Chemiluminescence Systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices K980190
510(k) Number

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.