K Number
K974492
Date Cleared
1998-02-26

(90 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SC6000/SC6000P enhanced with ST Segment Analysis is intended to be used in the environment where patient care is provided by Healthcare Professionals, trained in the use of the device, i.e. physicians, nurses, and technicians, who will determine when use of ST Segment Analysis is indicated, based upon their professional assessment of the patient's medical condition.

ST Segment Analysis is intended for use in the adult population.

The SC6000/SC6000P is not for home use.

Device Description

The SC 6000 / SC 6000P Bedside Monitoring Series Enhanced with ST Segment Analysis is an updated software version of the SC 6000 / SC 6000P Bedside Monitoring Series. The modification adds software to determine the ST Segment of the ECG signal and to compute the deviation of this ST Segment from the isoelectric point (baseline). This is the same algorithm that is used in the Siemens 1481 (T) Digital Telemetry System with ST Segment Analysis Option (K951371). The hardware of the SC 6000 / SC 6000P (510(k) K944350) is unchanged.

The ST Segment Analysis is not active when the SC 6000/SC 6000P is in the neonatal mode.

The modified software (version VC0) is not compatible with all previously sold versions of the monitor. Therefore, a software upgrade will be offered to the owners of units with previous software revisions. No hardware changes are required for the upgrade.

AI/ML Overview

The provided text describes the Siemens SC6000/SC6000P Bedside Monitoring Series enhanced with ST Segment Analysis. The submission is a 510(k) for a modified device, meaning it seeks to prove substantial equivalence to a predicate device. Therefore, the "study" described is primarily focused on demonstrating this equivalence rather than a de novo clinical trial.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to the predicate device, especially regarding ST Segment deviation measurement accuracy.

Acceptance Criteria / Performance MetricPredicate Device (Siemens 1481T Digital Telemetry with ST Segment Analysis Option)Applicant Device (Siemens SC6000/SC6000P Series Enhanced with ST Segment Analysis)
ST Segment deviation measurement accuracy± 1mm / ± 0.1mVSame (± 1mm / ± 0.1mV)
Intended Use (for ST Segment Analysis)To detect ST segment deviation values.Same.
Intended PopulationAdultAdult
Intended EnvironmentWhere patient care is provided by healthcare professionals.Same.
ISO point adjustment rangeComplex start to fiducial pointSame
ISO point default30 msec before QRS onsetSame.
ST measurement point adjustment rangeFiducial point to complex endSame
ST complex length900 msecSame
Sample Rate100 samples per secondSame
Update interval20 Seconds15 Seconds
Leads processedAny two of I, II, III, VAny one of I, II, III, V, aVR, aVL, aVF

Note: The document explicitly states "The ST Segment Analysis of the SC6000/SC6000P series patient monitors is equivalent to the ST Segment Analysis of the predicate device." This statement, along with the "Same" annotations in the table, indicates that the acceptance criteria for the ST Segment analysis are met by matching the performance of the predicate device. The changes in "Leads processed" and "Update interval" are noted but not presented as deviations from an acceptance criterion for ST segment accuracy itself, but rather as functional differences that were presumably deemed acceptable.

2. Sample size used for the test set and the data provenance

The document does not specify a sample size for a test set or data provenance (e.g., country of origin, retrospective/prospective). As this is a 510(k) submission primarily relying on software equivalence to a predicate device, it appears to leverage existing data or validation processes from the predicate device rather than conducting a new, independent clinical study with a defined test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not provide information on the number or qualifications of experts used to establish ground truth for a test set. This type of detail is typically associated with de novo clinical studies involving human interpretation, which is not the primary focus of this 510(k) submission for a software modification to a monitoring device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

The document does not mention any adjudication method for a test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was reported. This device is an ST Segment monitor, and the focus is on the algorithm's performance in measuring ST deviation, not on human reader improvement with AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone assessment of the algorithm's performance is implicit in the comparison of the ST Segment Analysis algorithm to the predicate device. The document explicitly states: "The ST Segment Analysis of the SC6000/SC6000P series patient monitors is equivalent to the ST Segment Analysis of the predicate device." The comparison table directly addresses the accuracy ("± 1mm / ± 0.1mV") of the algorithm itself. This suggests that the algorithm's output (ST segment deviation values) is being evaluated directly against predefined accuracy standards inherent to the predicate device's performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The explicit type of ground truth used is not specified. However, given the context of a 510(k) submission for a monitoring device's ST segment analysis, the implicit ground truth would likely be a combination of:

  • Physiological standards: The accepted medical definition and measurement of ST segment deviation.
  • Comparison to predicate device: The performance of the predicate device (Siemens 1481T Digital Telemetry) itself serves as the benchmark against which the modified algorithm's "truth" or accuracy is compared.
  • Compliance with industry standards: The device complies with "Performance Measurements for Algorithms to Detect Transient Ischemic ST Segment Changes", IEEE 1992 and "ST Segment Monitor Preliminary Guidance", US Department of Health and Human Services, July 1994, which would imply certain performance expectations that serve as ground truth for algorithm design and testing.

8. The sample size for the training set

The document does not specify a sample size for the training set.

9. How the ground truth for the training set was established

The document does not provide information on how the ground truth for a training set was established. Given the age of the submission (1997) and the nature of the device (ECG monitoring with ST segment analysis), it's highly probable that traditional signal processing and rule-based algorithms were in use, rather than machine learning models requiring explicit training sets with labeled ground truth in the modern sense. The "ground truth" for developing such algorithms would typically be based on established physiological principles, expert knowledge of ECG interpretation, and potentially large databases of ECG recordings analyzed by cardiologists.

{0}------------------------------------------------

K974492

FEB 2 6 1998

510(k) SUMMARY as required per 807.92(c)

    1. Submitters Name, Address:
      Siemens Medical Systems, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (978) 750-7500 Fax: (978) 777-3398 Official Correspondent: David Simard, Director Quality Assurance & Regulatory Affairs Contact person for this submission: Jacqueline Emery Date submission was prepared: November 14, 1997
    1. Trade Name. Common Name and Classification Name:

A. Trade Name:

Siemens SC6000/ SC6000P Bedside Monitoring Series enhanced with ST Segment Analysis

Common NameClassificationNumberClassRegulation Number
Cardiac Monitor74DRTII21 CFR 870.2300
Pulse Rate Monitor74BWSII21 CFR 870.2300
Pulse Oximeter74DQAII21 CFR 870.2700
Breathing Frequency Monitor73BZQII21 CFR 868.2375
Clinical Electronic Thermometer80BWXII21 CFR 880.2910
Indwelling Blood Pressure Monitor74CAAII21 CFR 870.1110
Noninvasive Blood Pressure Monitor74DXNII21 CFR 870.1130
Heart Rate Monitor, Neonatal74FLOII21 CFR 870.2300
Ventilatory Effort Monitor (Apnea Detector)73FLSII21 CFR 868.2375
Monitor Blood Pressure, Neonatal, Invasive74FLPII21 CFR 870.1110
ST Segment Monitor with Alarm74 MLDIII21 CFR 870.1025
Arrhythmia Detector & Alarm System74DSIIII21 CFR 870.1025
  • B. Common Name, Classification Name, Class and Regulation Number:
    1. Predicate Device Identification:
      The Siemens 1481 (T) Digital Telemetry System with ST Segment Analysis Option (K951371)

1

COMPANY CONFIDENTIAL

Siemens Medical Systems, Inc. Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923 USA

Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)

{1}------------------------------------------------

Siemens SC6000/SC6000P Portable Bedside Monitoring Series with ST Segment Analysis

3. Device Description:

The SC 6000 / SC 6000P Bedside Monitoring Series Enhanced with ST Segment Analysis is an updated software version of the SC 6000 / SC 6000P Bedside Monitoring Series. The modification adds software to determine the ST Segment of the ECG signal and to compute the deviation of this ST Segment from the isoelectric point (baseline). This is the same algorithm that is used in the Siemens 1481 (T) Digital Telemetry System with ST Segment Analysis Option (K951371). The hardware of the SC 6000 / SC 6000P (510(k) K944350) is unchanged.

The ST Segment Analysis is not active when the SC 6000/SC 6000P is in the neonatal mode.

The modified software (version VC0) is not compatible with all previously sold versions of the monitor. Therefore, a software upgrade will be offered to the owners of units with previous software revisions. No hardware changes are required for the upgrade.

    1. Intended Use:
      The intended use of the SC 6000/SC 6000P Bedside Monitoring Series is to measure heart rate, respiration rate, invasive pressure, non-invasive pressure, arrhythmia (adult) temperature, arterial oxygen saturation, pulse rate, central apnea, and ST Segment Analysis (adult). The device will produce visual and audible alarms if any of the above parameters vary beyond preset limits and produce timed or alarm recordings.

2

COMPANY CONFIDENTIAL

Siemens Medical Systems, Inc. Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923 USA

Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)

{2}------------------------------------------------

SIEMENS MED SYS

Siemens SC6000/SC6000P Portable Bedside Monitoring Series with ST Segment Analysis

Substantial EquivalentDeviceSiemens Medical Systems1481T Digital Telemetrywith ST Segment AnalysisOptionApplicantSiemens Medical SystemsSC 6000/SC 6000P SeriesEnhanced with ST SegmentAnalysisExplanation of themodified version
Intended UseThe intended use of this deviceis to detect a patient's EKGsignals and to transmit thisdata via radiofrequency to acentral monitoring station. Atthe central monitoring station,cardiac arrhythmias, STsegment deviation values andheart rates are determined.The patient's pulse rate andarterial oxygen saturationvalues, heart rate values, STsegment deviation values aredisplayed and visual and auralalarms and recordings areinitiated if these parametersvary beyond preset limits.The intended use of this device is tomeasure heart rate, respiration rate,invasive pressure, non-invasivepressure, arrhythmia (adult only), STSegment Analysis (adult), temperature,arterial oxygen saturation, pulse rate,and (central) apnea. This device willproduce visual and audible alarms ifany of these parameters vary beyondpreset limits and produce timed oralarm recordings. This device willconnect to the Siemens SIRENET orInfinity(Olympus) networkThe intended use forthe ST SegmentAnalysis is the samefor both the SC6000and the predicate,1481T. The devicesthemselves havedifferent intended uses(patient monitor vs.telemetry system)
IntendedPopulationAdultAdult
IntendedEnvironmentWhere patient care is providedby healthcare professionalsSame
ST Segmentdeviationmeasurementaccuracy$ \pm $ 1mm / $ \pm $ 0.1mVSame
Leads processedAny two of I, II, III, VAny one of I, II, III, V, aVR, aVL,aVFThe SC6000 seriesmeasures only one leadand supportsaugmented leads.
ISO pointadjustment rangeComplex start to fiducial pointSame
ISO point default30 msec before QRS onsetSame .
ST measurementpoint adjustmentrangeFiducial point to complex endSame
ST complexlength900 msecSame
Sample Rate100 samples per secondSame
Update interval20 Seconds15 Seconds

3

COMPANY CONFIDENTIAL

Siemens Medical Systems, Inc. Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923 USA

Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)

{3}------------------------------------------------

Siemens SC6000/SC6000P Portable Bedside Monitoring Series with ST Segment Analysis

    1. Assessment of non-clinical performance data for equivalence: Currently there are no FDA standards for this device
    1. Assessment of clinical performance data for equivalence: The ST Segment Analysis of the SC6000/SC6000P series patient monitors is equivalent to the ST Segment Analysis of the predicate device.
    1. Biocompatability: Not applicable (Same as original submission)
    1. Sterilization: Not applicable (Same as original submission)
    1. Standards and Guidance:

Currently there are no FDA standards for this device. The Siemens Series SC 6000/SC 6000P Bedside Monitoring Series enhanced with ST Segment Analysis complies with: "Performance Measurements for Algorithms to Detect Transient Ischemic ST Segment Changes", IEEE 1992

"ST Segment Monitor Preliminary Guidance", US Department of Health and Human Services, July 1994.

COMPANY CONFIDENTIAL

Siemens Medical Systems, Inc. Electromedical Systems Group, PCS

16 Electronics Avenue Danvers, MA 01923 USA

Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, a symbol often associated with medicine and healthcare.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 6 1998

Mr. David Simard Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923

K974492 Re : Siemens SC6000/SC6000P Bedside Monitoring Series Enhanced with ST Segment Analysis III (three) Requlatory Class: Product Code: 74 MLD Dated: November 25, 1997 Received: November 28, 1997

Dear Mr. Simard:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Mr. David Simard

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K974492

Device Name: Siemens SC6000 / SC6000P Bedside Monitoring Series enhanced with ST Segment Analysis

Indications for Use:

The SC6000/SC6000P enhanced with ST Segment Analysis is intended to be used in the environment where patient care is provided by Healthcare Professionals, trained in the use of the device, i.e. physicians, nurses, and technicians, who will determine when use of ST Segment Analysis is indicated, based upon their professional assessment of the patient's medical condition.

ST Segment Analysis is intended for use in the adult population.

The SC6000/SC6000P is not for home use.

MRI Compatibility Statement:

The Siemens SC6000 / SC6000P Bedside Monitoring System is not compatible for use in a MRI magnetic field.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use_

(Optional Format 1-2-96)

C. Carey

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.