K Number
K974308

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-01-27

(71 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Peripheral Vascular Array is a phased array device which operates in the receiveonly mode. The device is intended to receive signal from hydrogen protons during MR imaging. The Peripheral Vascular Array, as with any surface coil utilized in MR imaging, is intended to provide improved image quality as a result of the high Signal to Noise ratio (SNR) produced when placing a surface coil close to the anatomy of interest. Higher SNR allows the user to prescribe thinner slices and higher matrices necessary for the high spatial resolution needed in making diagnosis.

The Peripheral Vascular Array is primarily designed for imaging of the vascular structures of the abdomen, pelvis, and lower limbs. Soft tissue imaging of these areas may also be evaluated using this device. The design of this coil allows the user to easily image the large peripheral vascular system associated with the abdomen, pelvis and lower limbs without moving the patient of the coil.

Patient fitness and suitability to undergo an MRI exam utilizing the Peripheral Vascular Array must be determined by the individual physician trained in the field of Diagnostic Magnetic Resonance Imaging.

Device Description

The Peripheral Vascular Array is primarily designed for imaging of the vascular structures of the abdomen, pelvis, and lower limbs. Soft tissue imaging of these areas may also be evaluated using this device. The design of this coil allows the user to easily image the large peripheral vascular system associated with the abdomen, pelvis and lower limbs without moving the patient or the coil.

The Peripheral Vascular coil is designed as a durable item with an expected life of at least 2500 procedures, equivalent to at least five years of service. The specific application of the Peripheral Vascular coil involves MR Imaging at 1.5 Tesla using the GE 1.5T Signa systems with the phased array option.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the MEDRAD 1.5T Peripheral Vascular Array:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative "acceptance criteria" in a pass/fail threshold manner. Instead, it describes performance metrics that were evaluated and compared to a predicate device. The implied acceptance criterion is "equivalent or increased SNR" and "improved image quality and morphological detail" compared to the predicate.

Performance MetricAcceptance Criteria (Implied)Reported Performance
Signal-to-Noise Ratio (SNR)Equivalent or increased SNR compared to predicate device (GE 1.5T Body Coil)"The results were compared to verify the equivalent or increased SNR of the proposed coil.""...substantiate improved SNR and morphological detail with the proposed coil."
Image UniformityCharacterized using NEMA MS-6-91"Contours of the images obtained with the coil were constructed for the axial image, sagittal image, and the coronal image." (Implies satisfactory characterization)
Image Clarity / Morphological DetailImproved compared to predicate device"...substantiate improved SNR and morphological detail with the proposed coil.""Image clarity, morphological detail... demonstrate that the Medrad 1.5T Peripheral Vascular Array will produce the required detailed resolution..."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The document mentions "clinical testing" where "Images were obtained for the proposed Medrad 1.5T Peripheral Vascular Array and the predicate." However, the number of patients or images in this test set is not provided.
  • Data Provenance: Not explicitly stated. Based on the context of a 510(k) submission to the FDA, it is highly likely that this was prospective data collected specifically for the regulatory submission to demonstrate the device's performance. The country of origin is not mentioned, but given "Medrad, Inc." is an American company and the submission is to the FDA, it's reasonable to infer the data was likely collected in the USA.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The document mentions that patient fitness and suitability for an MRI exam "must be determined by the individual physician trained in the field of Diagnostic Magnetic Resonance Imaging." However, this refers to the physicians performing the clinical use, not necessarily the expert readers interpreting the test set images for ground truth establishment.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs Without AI Assistance

  • MRMC Study: No, an MRMC study was not conducted. This device is a hardware component (MRI surface coil), not an AI algorithm.
  • AI Improvement: This question is not applicable as the device is not an AI software/algorithm. The comparison is between two hardware coils (the proposed coil vs. the predicate body coil).

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • This question is not applicable as the device is a hardware component (MRI surface coil), not a standalone algorithm.

7. The Type of Ground Truth Used

The concept of "ground truth" in the typical sense (e.g., pathology, outcomes) for diagnostic algorithm performance is not directly applicable here. For the phantom testing, the "ground truth" was established by NEMA Standard No. 6, Characterization of Special Purpose Coils for Diagnostic Magnetic Resonance Images, which defines the methodology for measuring SNR and uniformity. For the clinical testing, the "ground truth" was essentially the visual assessment of "improved SNR and morphological detail" by presumably experienced radiologists comparing images from the two coils. The document does not specify a separate, independent ground truth method like pathology for the clinical images.

8. The Sample Size for the Training Set

This question is not applicable as the device is a hardware component and does not involve training an algorithm.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as the device is a hardware component and does not involve a training set or ground truth establishment in that context.

{0}------------------------------------------------

897 4308

Image /page/0/Picture/1 description: The image shows the word "MEDRAD" in a bold, sans-serif font. The letters are all capitalized and slightly slanted to the right. A small registered trademark symbol is located to the right of the letter "D". The word appears to be a logo or brand name.

JAN 27 1998

QA-97-1222

510(k) SUMMARY

MEDRAD 1.5T PERIPHERAL VASCULAR ARRAY

OFFICIAL CONTACT:

CLASSIFICATION NAME:

COMMON/USUAL NAME:

Medrad, Inc.

One Mearaa Drive clanoia, PA 15051-0730 (412) 767-2400

Mary Ann Greenawalt, J.D. Sr. Regulatory Affairs Associate Medrad, Inc. One Medrad Drive Indianola, PA 15051

Magnetic Resonance Diagnostic Accessory [21 CFR 892.1000]

MRI Surface Coil

PROPRIETARY NAME: Medrad 1.5T Peripheral Vascular Array

PREDICATE DEVICES:

GE Signa 1.5T Body Coil

DEVICE DESCRIPTION:

The Peripheral Vascular Array is primarily designed for imaging of the vascular structures of the abdomen, pelvis, and lower limbs. Soft tissue imaging of these areas may also be evaluated using this device. The design of this coil allows the user to easily image the large peripheral vascular system associated with the abdomen, pelvis and lower limbs without moving the patient or the coil.

The Peripheral Vascular coil is designed as a durable item with an expected life of at least 2500 procedures, equivalent to at least five years of service. The specific application of the Peripheral Vascular coil involves MR Imaging at 1.5 Tesla using the GE 1.5T Signa systems with the phased array option.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "MEDRAD" in a bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance. There is a small symbol or mark to the upper right of the letter "D", possibly a trademark symbol.

QA-97-1222

INTENDED USE:

The Peripheral Vascular Array is a phased array device which operates in the receive-only mode. The device is intended to receive signal from hydrogen protons during MR imaging. The Perioheral Vascular Array, as with any surface coil utilized in MR imaging, is intended to provide improved image quality as a result of the high Signal to Noise ratio (SNR) produced when placing a surface coil close to the anatomy of interest. Higher SNR allows the user to prescribe thinner slices and higher matrices necessary for the high spatial resolution needed in making diagnosis.

The Peripheral Vascular Array is primarily designed for imaging of the vascular structures of the abdomen. pelvis, and lower limbs. Soft tissue imaging of these areas may also be evaluated using this device. The design of this coil allows the user to easily image the large peripheral vascular system associated with the abdomen, pelvis and lower limbs without moving the patient or the coil.

Patient fitness and suitability to undergo an MRI exam utilizing the Peripheral Vascular Array must be determined by the individual physic ian trained in the field of Diagnostic Magnetic Resonance Imaging.

PERFORMANCE TEST DATA:

Signal to Noise Ratio (SNR)

The Medrad 1.5T Peripheral Vascular Array was evaluated using National Electric Manufacturer's Association (NEMA) Standard No. 6, Characterization of Special Purpose Coils for Diagnostic Magnetic Resonance Images.

Both the Medrad 1.5T Peripheral Vascular Array and the predicate GE 1.5T Body Coil were evaluated with a loaded phantom to determine the SNR for both coils. The coils were evaluated using the same scan parameters to produce identical images. The results were compared to verify the equivalent or increased SNR of the proposed coil.

Image Uniformity Testing:

The Medrad 1.5T Peripheral Vascular Array was evaluated using NEMA Standard MS-6-91 to characterize the uniformity of the proposed coil. Contours of the images obtained with the coil were constructed for the axial image, sagittal image, and the coronal image.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "MEDRAD" in a stylized, bold, sans-serif font. The letters are all capitalized and appear to be slightly slanted to the right. A registered trademark symbol is present to the right of the letter "D".

QA-97-1222

Clinical Testing:

Images were obtained for the proposed Medrad 1.5T Peripheral Vascular Array and the predicate. The GE 1.5T Body Coil results were compared to substantiate improved SNR and morphological detail with the proposed coil.

SAR: Not applicable.

Extensive safety, verification, durability, and clinical CONCLUSION: testing was conducted with the Medrad 1.5T Peripheral Vascular Array to substantiate the claims of the proposed device and to verify that the proposed device is substantially equivalent to the predicate devices.

Image clarity, morphological detail and increased SNR demonstrate that the Medrad 1.5T Peripheral Vascular Array will produce the required detailed resolution in surface coil imaging.

end

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 27 1998

Mary Ann Greenwalt, J.D. Sr. Regulatory Affairs Associate Medrad, Inc. One Medrad Drive Indianola. PA 15051

Re:

K974308 Medrad 1.5T Peripheral Vascular Array Dated: November 14, 1997 Received: November 17, 1997 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS

Dear Ms. Greenwalt:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/edrh/dsmamain.html" -------

Sincerely yours,

Sincerely yours,

Lillian Yin, Ph.D.

Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Page of İ

510 (k) NUMBER (IF KNOWN) : K974308

DEVICE NAME : MEDRAD PERIPHERAL VASCULAR ARRAY

INDICATIONS FOR USE:

INDICATIONS FOR USE

The Peripheral Vascular Array is a phased array device which operates in the receiveonly mode. The device is intended to receive signal from hydrogen protons during MR imaging. The Peripheral Vascular Array, as with any surface coil utilized in MR imaging, is intended to provide improved image quality as a result of the high Signal to Noise ratio (SNR) produced when placing a surface coil close to the anatomy of interest. Higher SNR allows the user to prescribe thinner slices and higher matrices necessary for the high spatial resolution needed in making diagnosis.

The Peripheral Vascular Array is primarily designed for imaging of the vascular structures of the abdomen, pelvis, and lower limbs. Soft tissue imaging of these areas may also be evaluated using this device. The design of this coil allows the user to easily image the large peripheral vascular system associated with the abdomen, pelvis and lower limbs without moving the patient of the coil.

Patient fitness and suitability to undergo an MRI exam utilizing the Peripheral Vascular Array must be determined by the individual physician trained in the field of Diagnostic Magnetic Resonance Imaging.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter-Use (Optional Format 1-2-96)

Daniel A. Seggern
(Division Sign-Off)

Division of Reproductive, Abdomina and Radiological Devices 510(k) Number

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.