K Number
K973718

Validate with FDA (Live)

Date Cleared
1998-12-15

(442 days)

Product Code
Regulation Number
866.3120
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a qualitative in vitro test for the detection of C. trachomatis plasmid DNA in urine from males and females, in endocervical swab specimens, and in male urethral swab specimens as evidence of symptomatic or asymptomatic infection with C. trachomatis. C. trachomatis DNA is detected by Polymerase Chain Reaction (PCR) amplification of target DNA and by hybridization capture of amplified target using the COBAS AMPLICOR Analyzer.

Device Description

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a multiplex in vitro diagnostic test performed on the COBAS AMPLICOR Analyzer. The COBAS AMPLICOR Analyzer automates the annolification, the nucleic acid hybridization, and the colorimetric detection procedures of the Test. The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis also has an Internal Control that identifies. specimens that contain substances inhibitory to PCR.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the COBAS AMPLICOR™ CT/NG Test for Chlamydia trachomatis, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" in terms of specific sensitivity and specificity thresholds that needed to be met. Instead, it presents the device's performance in comparison to existing gold standards (culture and DFA). The reported performance across various patient groups and specimen types allows implicit evaluation against generally accepted standards for diagnostic tests.

MetricAcceptance Criteria (Implicit/General)Reported Device Performance (Range)
Analytical Sensitivity (Limit of Detection)Sufficient to detect C. trachomatis1 Inclusion Forming Unit (1 IFU) per test for all 15 Chlamydia serovars.
Analytical SpecificityNo cross-reactivity with common urogenital flora/pathogensNegative results for all 152 bacterial, 6 fungal, 1 protozoan, and 11 viral isolates at ≥ 10^4 copies of genomic DNA per test.
Precision (Reproducibility)Qualitatively correct results, low variability in absorbance100% qualitatively correct results across all specimen types, concentrations (0, 1.25, 3.75, 6.25 IFU/test for CTM; 0, 1, 3, 5 IFU/test for urine), and sites. Mean, Min, Max A660 values reported show consistency.
Clinical Sensitivity (Females - individual specimen)High clinical detection rateCTM: 94.9% (93.6-96.2%) Asymptomatic; 95.9% (94.6-97.2%) Symptomatic. Urine: 90.9% (84.5-97.3%) Asymptomatic; 90.3% (84.3-96.3%) Symptomatic.
Clinical Specificity (Females - individual specimen)High ability to correctly identify negative casesCTM: 98.3% (97.6-99.1%) Asymptomatic; 98.0% (97.1-98.8%) Symptomatic. Urine: 98.3% (97.4-99.1%) Asymptomatic; 96.9% (95.8-97.9%) Symptomatic.
Clinical Sensitivity (Males - individual specimen)High clinical detection rateCTM: 98.7% (97.1-100%) Asymptomatic; 96.8% (94.3-99.3%) Symptomatic. Urine: 89.9% (83.2-96.5%) Asymptomatic; 88.3% (83.8-92.8%) Symptomatic.
Clinical Specificity (Males - individual specimen)High ability to correctly identify negative casesCTM: 97.7% (96.6-98.9%) Asymptomatic; 94.6% (93.2-96.0%) Symptomatic. Urine: 96.3% (94.8-97.7%) Asymptomatic; 92.2% (90.5-93.8%) Symptomatic.
Clinical Sensitivity (Female Patient Status - all specimens)Higher detection with combined specimensCTM + URINE: 90.8% (84.7-96.9%) Asymptomatic; 95.2% (94.0-96.4%) Symptomatic. Total CTM: 89.4% (85.0-93.8%). Total Urine: 86.9% (82.0-91.8%). (Note: The document states "better concordance with culture/DFA positive patients when both swab and urine specimens are tested... resulted in fewer unverified positive test results and higher assay sensitivity")
Clinical Specificity (Female Patient Status - all specimens)High specificity with combined specimensCTM + URINE: 97.4% (96.4-98.4%) Asymptomatic; 96.4% (95.3-97.5%) Symptomatic. Total CTM: 98.2% (97.6-98.8%). Total Urine: 97.8% (97.1-98.4%).

2. Sample Size and Data Provenance (Clinical Study)

  • Test Set Sample Size:
    • Total Patients: 4277
    • Total Specimens (initially collected): 8523 (both swab and urine from 4201 patients; urine only from 76 patients)
    • Specimens included in primary analysis (with IC): 8397
    • Specimens included in analysis (without IC): 8478 (after excluding initial equivocal and repeatedly inhibitory results)
  • Data Provenance: Retrospective, collected from six geographically diverse sites. Specific countries are not mentioned, but "US Highway 202 Somerville, New Jersey" for Roche suggests the study was likely conducted in the United States.

3. Number of Experts and Qualifications for Ground Truth

  • The document does not specify the "number of experts" or their "qualifications" involved in establishing the ground truth.
  • The ground truth method used was standard cell culture with immunofluorescent staining for C. trachomatis for positive samples.
  • For swab specimens that were culture negative by the COBAS test, DFA (Direct Fluorescent Antibody) testing was performed for the presence of C. trachomatis. These are established laboratory methods.

4. Adjudication Method for the Test Set

  • The document describes a hierarchical adjudication method for establishing the ground truth, rather than a traditional consensus by independent experts on the device's results.
  • Primary Ground Truth: C. trachomatis cell culture (with cyclohexamide treated McCoy cells stained with fluorescein-labeled monoclonal antibody).
  • Secondary Ground Truth (for culture-negative COBAS-positive swabs): DFA (Direct Fluorescent Antibody) for C. trachomatis.
  • Tertiary Check (for information only): Alternate PCR testing (MOMP gene target) was performed on COBAS AMPLICOR positive, culture/DFA negative specimens, but these results were explicitly not used to calculate the clinical performance characteristics. This indicates a strict reliance on the established culture/DFA as the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done in the context of human readers directly interacting with or improving with AI assistance. This document describes the performance of a standalone in vitro diagnostic (PCR-based) test. The "readers" here are the laboratory technicians conducting and interpreting the instrument's output.

6. Standalone Performance

  • Yes, a standalone study was done. The entire clinical performance section (Tables 3-7 and associated text) describes the performance of the COBAS AMPLICOR CT/NG Test as a diagnostic device without human-in-the-loop assistance in its interpretation beyond standard laboratory procedures (e.g., reading optical density and confirming results). It directly compares the device's output to the established gold standard.

7. Type of Ground Truth Used

  • The primary ground truth used was a combination of expert-performed laboratory methods:
    • Cell Culture (with immunofluorescent staining) for C. trachomatis.
    • Direct Fluorescent Antibody (DFA) for C. trachomatis (used for culture-negative COBAS-positive samples).
  • The MOMP PCR assay was an investigational check for false positives, but its results were not incorporated into the official performance metrics, highlighting adherence to established methods for ground truth.

8. Sample Size for the Training Set

  • The document does not provide details on a separate "training set" as would be typical for machine learning algorithms. This device is an in vitro diagnostic (IVD) test, and its development typically involves internal validation and optimization against characterized samples, followed by clinical validation. The data presented is primarily for the clinical validation.

9. How the Ground Truth for the Training Set Was Established

  • As there is no explicit "training set" discussed in the context of this 510(k) summary for an IVD, the method for establishing its ground truth is not detailed. However, for a device like this, the development process would have involved using characterized C. trachomatis strains and clinical samples with known infection status (determined by culture, etc.) to optimize the assay's primers, probes, and reaction conditions.

{0}------------------------------------------------

510(k) Summary

1973718

DEC 15 1998

COBAS AMPLICOR™ CT/NG Test for Chlamydia trachomatis Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876-1760 (908) 253-7200

Intended Use:

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a qualitative in vitro test for the detection of C. trachomatis plasmid DNA in urine from males and females, in endocervical swab specimens, and in male urethral swab specimens as evidence of symptomatic infection with C. truchomatis. C. trachomatis DNA is detected by Polymerase Chain Reaction (PCR) amplification of target DNA and by hybridization capture of amplified target using the COBAS AMPLICOR Analyzer.

Description of the Device:

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a multiplex in vitro diagnostic test performed on the COBAS AMPLICOR Analyzer. The COBAS AMPLICOR Analyzer automates the annolification, the nucleic acid hybridization, and the colorimetric detection procedures of the Test. The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis also has an Internal Control that identifies. specimens that contain substances inhibitory to PCR.

Similarities and Differences to Predicate Device:

The COBAS AMPLICOR CTNG Test for Chlamydia trachomatis is substantially equivalent to other commercially available in vitro diagnostic devices for the detection of Chiamydia trachomatis in urogenital swab and urine specimens. These methods include culture with immunofluorescent staining, ELISA, DFA, and nucleic acid hybridization. A commonality among all of these devices is that the unique biochemical properties of the target organism are all encoded in the DNA of the organism, essentially reducing each device to a test for genetic (i.e., phenotypic or genotypic) characteristics of the COBAS AMPLICOR CTNG Test for Chlamydia trachomatis detects DNA from the cryptic plasmid of the Chlamydia trachomatis organism while cell culture detects the complete viable inclusion forming unit. The clinical performance of the COBAS AMPLICOR Chlamydia trachomatis Test has been shown to be substantially equivalent to cell culture methods.

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is also similar to the AMPLICOR Chlamydia trachomatis Test and to the COBAS AMPLICOR Chlamydia trachomatis Test performed on the COBAS AMPLICOR Analyzer. Improvements in the COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis allow the automated multiplex amplification of Chlamydia and an Internal Control that is used to detect the presence of PCR inhibitors. All of these tests use the same oligonucleotide primers and probe for the detection of Chlamydia trachomatis and have similar detection reactions that are based on the absorbance measurement of a chromophore that is produced by the oxidation of 3,3',5,5'-tetramethybenzidine by hydrogen peroxide in the presence of horseradish peroxidase.

Non-Clinical Performance:

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis was shown to have an analytical sensitivity (limit of detection) of 1 Inclusion Forming Unit (1 IFU) per test for all 15 Chlamydia serovars (A, B, Ba, C, D, E, F, G, H, I, J, K, LGV1, LGV2, LGV3).

f:rey/regdocs/cing/cicobas/fidaresp12-14/final 510kaum/l

{1}------------------------------------------------

The analytical specificity of the COBAS AMPLICOR CT/NG Test for C. trachomatis was tested against 132 bacteria, 6 fungi, 1 protozoon and 11 virus isolates that may be isolated from the urogenital tract. The 1.52 backers, or c. g. C. trachomatis gave negative results for each isolate present in culture transport media and normal human urine at ≥ 104 copies of genomic DNA per test.

The precision of the COBAS AMPLICOR CT/NG Test for Chlamydia Irachomatis on the COBAS AMPLICOR Analyzer was determined for a panel of culture transport media specimens containing 0, 1.25, 3.75 and 6.25 Chlamydia trachomatis IFU/test and urine specimens containing 0, 1, 3 and 5 Chlamydia trachomatis IFU/test. Three independent operators at three different geographical sites tested the panel once a day for three days in duplicate. The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis gave 100% qualitatively correct results across all specimen types, concentrations and sites. The results of this study are presented in Tables 1 and 2.

C. trachomatis Spiked CTM (IFU/test)
01.253.756.25
Replicates72363636
% Correct Results100100100100
Mean A6600.0053.2503.1533.176
Minimum A6600.0002.7322.6862.681
Maximum A6600.0184.0003.6103.962

Table 1 COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis CTM Specimen Reproducibility

Table 2
COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis
Urine Specimen Reproducibility
C. trachomatis Spiked CTM (IFU/test)
0135
Replicates72363636
% Correct Qualitative100100100100
Mean A6600.0043.3933.2683.270
Minimum A6600.0002.6912.3812.264
Maximum A6600.0234.0004.0004.000

Clinical Performance:

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis was evaluated in a clinical study conducted at six geographically diverse sites. Swab (endocervical for females, urethral for males) and urine specimens were obtained from all patients entered into the study. Swab specimens were placed in culture transport media (CTM) used at each site. All swab specimens were tested by standard culture with cyclohexamide treated McCoy cells stained with fluorescein-labeled monoclonal antibody for C. trachomatis. Swab specimens that were culture negative by the COBAS Test were tested by DFA for the presence of C. trachomatis. The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis was performed on all endocervical swab and urine specimens obtained from female patients,

freu/reudocs/cing/crocobas/fdaresp12-14/final 510ksum/2

{2}------------------------------------------------

and all urethral swabs and urine specimens from male patients. COBAS testing was repeated for all specimens with initial results in the range of 0.2 to 0.8 Awa and when IC results were inhibited (negative).

A total of 8523 specimens collected from 4277 patients met the criteria for inclusion in the clinical study (patient was not on antibiotics, a valid culture result was obtained, specimen met storage requirements etc.). Both a swab and urine specimen was entered into the study for 4201 patients; a urine specimen only was entered into the study from 76 patients. Of the 8523 specimens included in the study, 45 specimens gave initial test results in the Equivocal Range and were excluded from the data analyses. Two specimens that were initially inhibitory gave results in the Equivocal Range upon repeat testing. These specimens are excluded from the analyses when the Internal Control was used but included in the analyses when the Internal Control was not used. In addition, 79 specimens were repeatedly inhibitory and were excluded from the data analyses which include the use of the Internal Control because the results were not interpretable. Therefore, 8397 specimens were included in the analyses when the Internal Control result was used and a total of 8478 specimens were included in the analyses when the Internal Control results were not used.

The clinical performance of the test was evaluated by comparing the results of the 8478 swab and urine specimens to the composite results of the comparative tests (culture, sub-culture and DFA). Alternate PCR testing using oliognucleotide primers targeted for a region of the C. trachomatis MOMP gene was performed on COBAS AMPLICOR positive, culture/DFA negative specimens. The MOMP test results were not used to calculate the clinical performance characteristics of the test and are reported for information purposes only. Of the 266 COBAS AMPLICOR positive, culture/DFA negative specimens that were classified as false positive results in this study, 185 were positive for C. trachomatis when that specimen or the matching urine or swab specimen from that patient was tested by the MOMP assay. These data suggest that many specimens considered as false positive in the Clinical Data Performance Tables did contain C. trachomatis DNA.

The results from the clinical study are shown in Tables 3 and 4. Table 3 shows the clinical performance of the COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis in comparison to the endocervical culture/DFA results for female patients and to the urethral culture/DFA results for male patients. In this Table, True Positive (TP) represents the number of concordant positive culture or DFA and COBAS Test results. True Negative (TN) represents the number of concordant negative culture and COBAS results. False Negative (FN) represents the number of culture positive results. False Positive (FP) represents the number of culture and DFA negative, COBAS positive results.

Table 4 shows the clinical performance of the COBAS AMPLICOR CT/NG Test for Chlamvaia trachomatis for testing both swab and urine specimens from female patients combined and separately, for each specimen type, in comparison to the patient infected status. Female patient infected status was determined by endocervical or urethral culture/DFA positive results. The data in Table 4 show that there is better concordance with culture/DFA positive patients when both swab and urine specimens are tested by the COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis. The testing of both swab and urine specimens by the COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis resulted in fewer unverified positive test results and higher assay sensitivity as compared to single specimen (swab or urine) testing only.

A summary of the test results obtained in the clinical study performed for the COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is contained in Tables 5 and 6. Table 5 summarizes the combinations of test results obtained for female patients; Table 6 summarizes the combinations of test results obtained for male patients. These tables show that patients with a positive result in both a urine and a swab specimen had a lower rate of unverified positives relative to culture and DFA) than single positive specimen results. Testing of both specinen types may be useful for increasing the confidence in a positive result using the COBAS AMPLICOR CT/NG Test for Chlumydia trachomatis, particularly for low prevalence populations.

f:rey/regdocs/cony/cicobas/fdaresp12-14/final 510ksum/3

{3}------------------------------------------------

The clinical sensitivity and specificity of the COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis has not been reliably determined for detecting those patients with clinically active infection that can be transmitted to partners or cause Chlamydia-related sequelae. In the clinical study described here, 24.4% of COBAS positive results were from patients with negative cultures and DFA tests. The significance of those results that were COBAS positive, but culture and DFA negative is unknown. A proportion of these COBAS positive specimens (63.8%) were also positive by an alternate target PCR assay; however, the performance of this alternate target assay has not been established

{4}------------------------------------------------

SexSpecimenSymptomTPTNFPFNNo. Inhib.% Repeatedly InhibitoryTotalSensitivity (95% CI)Specificity (95% CI)MOMP+/FP
FemaleCTMAsymptomatic75(75)1013(1017)17(16)4(4)20.20%1111(1112)94.9%(93.6-96.2)(94.9%)(93.6-96.2)98.3%(97.6-99.1)(98.5%)(97.7-99.2)9/17(10/16)
Symptomatic93(93)1025(1031)21(21)4(4)60.58%1149(1149)95.9%(94.6-97.2)(95.9%)(94.6-97.2)98.0%(97.1-98.8)(98.0%)(97.2-98.8)11/21(11/21)
FemaleURINEAsymptomatic70(70)1011(1023)18(18)7(8)131.26%1119(1119)90.9%(84.5-97.3)(89.7%)(83.0-96.5)98.3%(97.4-99.1)(98.3%)(97.5-99.1)13/18(13/18)
Symptomatic84(84)1018(1030)33(33)9(9)121.15%1156(1156)90.3%(84.3-96.3)(90.3%)(84.3-96.3)96.9%(95.8-97.9)(96.9%)(95.9-97.9)14/33(14/33)
Total for Females322(322)4067(4101)89(88)24(25)330.73%4535(4536)93.1%(90.4-95.7)(92.8%)(90.1-95.5)97.9%(97.4-98.3)(97.9%)(97.5-98.3)47/89(48/88)
MaleCTMAsymptomatic76(76)608(612)14(14)1(1)40.65%703(703)98.7%(97.1-100)(98.7%)(97.1-100)97.7%(96.6-98.9)(97.8%)(96.6-98.9)5/14(5/14)
Symptomatic183(183)977(994)56(54)6(6)141.40%1236(1237)96.8%(94.3-99.3)(96.8%)(94.3-99.3)94.6%(93.2-96.0)(94.8%)(93.5-96.2)32/56(32/54)
MaleURINEAsymptomatic71(71)616(617)24(24)8(8)10.16%720(720)89.9%(83.2-96.5)(89.9%)(83.2-96.5)96.3%(94.8-97.7)(96.3%)(94.8-97.7)14/24(14/24)
Symptomatic173(168)976(1004)83(80)23(30)272.63%1282(1282)88.3%(83.8-92.8)(84.8%)(79.9-89.8)92.2%(90.5-93.8)(92.6%)(91.1-94.2)62/83(60/80)
Total for Males503(498)3177(3227)177(172)38(45)461.41%3941(3942)93.0%(90.8-95.1)(91.7%)(89.4-94.0)94.7%(94.0-95.5)(94.9%)(94.2-95.7)113/177(111/172)

Table 4 Clinical Performance Of COBAS AMPLICOR CT/NG Test for Chlamydia trachomutis Including and Excluding the Internal Controll

' Test results without the Internal Control shown in parentheses.

True Positive (TP) represents the number of concordant positive culture or DFA and COBAS Test results. True Negative (TN) represents the number of concordant negative culture and COBAS results. False Negative (FN) represents the number of culture positive, COBAS negative results. False Positive (FP) represents the number of culture and DFA negative, COBAS positive results.

{5}------------------------------------------------

SpecimenSymptomTotal%InhibitoryNo.Inhib.SensitivitySpecificity(95% CI)MOMP+/FF-
CTM + URINEAsymp1126(1227)0.00%090.8%(84.7-96.9)90.8%(84.7-96.9)97.4%(96.4-98.4)(97.7%)(96.9-98.6)18/27(18/26)
Symptomatic1169(1169)0.10%195.2%(94.0-96.4)95.2%(94.0-96.4)96.4%(95.3-97.5)(96.4%)(95.3-97.5)15/38(15/38)
CTMAsymp1111(1112)0.20%287.2%(80.2-94.3)87.2%(80.2-94.3)98.3%(97.6-99.1)(98.4%)(97.7-99.2)11/17(12/16)
Symptomatic1149(1149)0.58%691.3%(85.8-96.7)91.3%(85.8-96.7)98.1%(97.2-98.9)(98.1%)(97.3-98.9)16/20(12/20)
Totals - CTM2260(2261)0.39%8(89.4%)(85.0-93.8)(89.4%)(85.0-93.8)98.2%(97.6-98.8)(98.3%)(97.7-98.8))27/37(24/36)
URINEAsymp1119(1119)1.26%1385.7%(78.2-93.2)(84.7%)(77.1-92.4)98.4%(97.7-99.2)(98.5%)(97.7-99.2)13/16(13/16)
Symptomatic1156(1156)1.15%1287.9%(81.4-94.3)(87.9%)(81.4-94.3)97.1%(96.1-98.1)(97.2%)(96.2-98.2)14/30(14/30)
Totals - Urine2275(2275)1.212586.9%(82.0-91.8)(86.4%)(81.5-91.4)97.8%(97.1-98.4)(97.8%)(97.2-98.4))27/36(27/36)

Table S Performance of COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis vs Patient Status' Female Patients Including and Excluding the Internal Control²

4Culture and DFA results in this table include endocervical and urchral results 4 Test results without the Internal Control shown in parentheses.

Page 6 of 7

{6}------------------------------------------------

No.PatientsCultureStatusEndocervical And Urethral CultureResultsDFAResultsCOBASAMPLICORResults bySpecimen Type
EndocervicalOnlyUrethralOnlyBothPositiveSwabUrine
146+85259N/A++
11+803N/A+-
7+241N/A-+
11+182N/A--
6-+++
2-++-
1-+--
18--++
18--+-
26---+
1965-N/A--

Table 6 COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis Test Result Summary - Female Patients'

1 Results from 85 patients without matched CTM and urine results are excluded from the table

Table 7 COBAS AMPLICOR CT/NG Test for Chlamydia truchomatis Test Result Summary - Male Patients

No.PatientsUrethralCultureStatusDFAResultsCOBAS AMPLICOR ResultsBy Specimen Type
SwabUrine
215+N/A++
20+N/A+-
4+N/A--
16-+++
3-++-
48--++
17--+-
51---+
2--InhibInhib
1503-N/A--

· Results from 140 patients without matched CTM and urine results are excluded from the table


{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of a human figure, with three stylized profiles facing to the right. Below the profiles are two wavy lines.

DEC 15 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Alex Wesolowski Sr. Director, Regulatory and Clinical Affairs Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, NJ 08876-3771

Re: K973718

Trade Name: Roche COBAS Amplicor CT/NG Test for Chlamydia trachomatis Regulatory Class: I Product Code: MKZ Dated: September 18, 1998 Received: September 22, 1998

Dear Mr. Wesolowski:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{8}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Page_ 1 _of_1 _

510(k) Number (if known): K973718

Device Name: COBAS Amplicor CT/NG Test for C. trachomatis

Indications For Use:

The COBAS AMPLICOR CT/NG Test for Chlamydia trachomatis is a qualitative in vitro test for the detection of C. trachomatis plasmid DNA in urine from males and females, in endocervical swab specimens, and in male urethral swab specimens as evidence of symptomatic or asymptomatic infection with C. trachomatis. C. trachomatis DNA is detected by Polymerase Chain Reaction (PCR) amplification of target DNA and by hybridization capture of amplified target using the COBAS AMPLICOR Analyzer.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois
Division Sign Off

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K973718

Prescription Use X (Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 866.3120 Chlamydia serological reagents.

(a)
Identification. Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusChlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).(b)
Classification. Class I (general controls).