K Number
K973393
Device Name
CHEMMATE LCA
Date Cleared
1997-09-25

(86 days)

Product Code
Regulation Number
866.5550
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To qualitatively aid in the identification by light microscopy of human cells of lymphoid orgin, by recognizing Leukocyte Common Antigen (LCA) in normal and pathologic paraffin embedded tissues processed in neutral buffered formalin, B5, or Bouin's fixative.

Device Description

The ChemMate™ LCA (CD45) is comprised of a mixture of two mouse monoclonal antibodies, clones PD7/26/16 and 2B11; each is of the IgG1 class of immunoglobulins. Each clone reacts with a different epitope of a family of membrane glycoproteins present on most human leukocytes known as the leukocyte common antigen (LCA).

AI/ML Overview

This submission pertains to the ChemMate™ LCA (CD45) antibody, intended for the qualitative identification of human cells of lymphoid origin in paraffin-embedded tissues via light microscopy. The document provides detailed performance characteristics regarding the immunoreactivity and reproducibility of the device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria (e.g., "sensitivity must be >90%"). Instead, it describes the expected immunoreactivity profiles (reactive, variably reactive, nonreactive) for various tissue and cell types, both normal and pathological. The reported performance is then presented as the percentage of cases showing reactivity or non-reactivity within these categories, often accompanied by 95% confidence intervals where the sample size was sufficient.

Given the qualitative nature of the intended use (aid in identification by light microscopy), the acceptance criteria appear to be the demonstration of consistent and expected reactivity patterns as described in the literature and confirmed by the studies. The device “meets” acceptance criteria by demonstrating these expected immunoreactivity profiles.

CategoryTissue/Cell Type (Examples)Expected Reactivity (Implicit Acceptance Criteria based on literature/prior knowledge)Reported Device Performance (from Tables II, III, V; representative examples)
Normal Tissues (Table 1A & 1B broadly)Lymphocytes (Follicular center, mantle zones, interfollicular, thymic)Reactive100% reactive (lymphocytes in Table 1B from BioTek Solutions study)
Polymorphonuclear leukocytesVariably/Weakly ReactiveVariably reactive (Table 1A) / Unreactive (myeloid cells in Table III)
Epithelium (various types)Nonreactive100% nonreactive (e.g., adrenal, breast, renal epithelium)
B-Cell Non-Hodgkin's Lymphoma (Table II)Follicular center cell - small cleavedHighly Reactive15/15 (100%) reactive
Immunoblastic sarcomaHighly Reactive14/14 (100%) reactive
T-Cell Non-Hodgkin's Lymphoma (Table II)Immunoblastic sarcomaHighly Reactive10/11 (90.9%) reactive
LymphoblasticVariably Reactive / Lower Reactivity (stated in product limitations)2/4 (50%) reactive
Other Hematopoietic Neoplasms (Table III)Chronic Lymphocytic Leukemia (B-cell & T-cell types)Highly Reactive3/3 (100%) reactive for both types
Hairy Cell LeukemiaHighly Reactive5/5 (100%) reactive
Acute Myelogenous Leukemia (myeloid cells)Nonreactive / Very Weakly Reactive (stated in product limitations)0/7 (0%) reactive
Multiple Myeloma (plasma cells)Highly Reactive10/10 (100%) reactive
Non-Lymphoid Pathological Tissues (Table IV & V)Squamous cell carcinoma, Adenocarcinoma, Melanoma, SarcomaNonreactive100% nonreactive across many categories (e.g., 54/54 adenocarcinoma, 5/5 melanoma)

2. Sample Sizes Used for the Test Set and Data Provenance:

The document describes two main types of studies:

  • Reproducibility Study:

    • Sample size: 271 tissue specimens (both normal and tumor, as referenced in Tables IB and V).
    • Data Provenance: Not explicitly stated by country, but conducted by Ventana Medical Systems, Inc. (the manufacturer). Retrospective, as these are "serial sections" from existing tissue banks.
  • Immunoreactivity Studies (Various tables refer to different data sources):

    • Table IA (Reactive and Nonreactive Normal Tissues/Cells - Literature Review): A total of 45 tissues were examined.
      • Data Provenance: This table largely synthesizes data from published literature (references 1, 7, 17, 18, 21), implying multicentric and retrospective data collection.
    • Table IB (Reactive and Nonreactive Normal Tissues/Cells - BioTek Solutions Study): 126 normal tissues.
      • Data Provenance: Conducted by BioTek Solutions (a company, likely related to Ventana) using the ChemMate™ LCA-CD45 and TechMate™ Automated Staining System. Retrospective.
    • Table II (Reactive B- and T- Cell Non-Hodgkin's Lymphoma - Literature Review): 80 specimens.
      • Data Provenance: Predominantly literature-based, with some fixed in multiple fixatives. Retrospective.
    • Table III (Other Hematopoietic Neoplasms - Literature Review): Not a single sample size for the entire table, but individual counts for each disease type (e.g., 3/3 for CLL, 5/5 for Hairy Cell Leukemia). Total reactive/non-reactive cases shown in the table: (70 reactive + 115 non-reactive) = 185 unique cases across various tumor types. However, this is also a compilation of literature.
      • Data Provenance: Literature-based. Retrospective.
    • Table IV (Nonreactive Non-Lymphoid Pathological Tissues - Literature Review): A total of 162 tissues.
      • Data Provenance: Literature-based. Retrospective.
    • Table V (Reactive and Nonreactive Pathological Tissues/Cells - BioTek Solutions Study): 145 pathological tissues.
      • Data Provenance: Conducted by BioTek Solutions using the ChemMate™ LCA-CD45 and TechMate™ Automated Staining System. Retrospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

The document implicitly relies on consensus from pathologists and researchers who published the referenced literature, as well as pathologists involved in internal studies (BioTek Solutions).

  • For the literature-based data (Tables IA, II, III, IV): The ground truth for diagnosis and reactivity was established by the original authors/pathologists of the cited publications. While specific numbers and qualifications of these experts aren't given in this 510(k) summary, it's assumed they are qualified pathologists and researchers in the field of immunohistochemistry and hematopathology.
  • For the BioTek Solutions studies (Tables IB, V): The studies state "All tissues were stained using the TechMate™ Automated Staining System." and the intent is for interpretation "by a pathologist within the context of clinical data, gross and microscopic morphological data and multiple chemical and immunohistochemical stains." It is implied that qualified pathologists made the assessments, but the specific number of experts or their years of experience are not mentioned.

4. Adjudication Method for the Test Set:

No specific adjudication method (e.g., 2+1, 3+1) is described for the ground truth establishment in either the literature-based data or the BioTek Solutions studies. The reported findings are presented as observations or consensus from the respective studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC study comparing human readers with AI assistance versus without AI assistance was done. This device is an antibody reagent, not an AI-powered diagnostic tool. The performance is assessed in terms of its biological reactivity and reproducibility, not human interpretive improvement.

6. Standalone Performance Study:

Yes, the studies described are essentially "standalone" in nature for the device (the antibody). The performance refers to the antibody's ability to consistently stain specific cell populations. The outcomes (e.g., positive or negative staining) are then interpreted by human pathologists, but the study itself (immunoreactivity and reproducibility) focuses on the consistent behavior of the reagent. The "Performance Characteristics" section directly addresses this.

7. Type of Ground Truth Used:

The ground truth used is primarily expert consensus (pathology/histopathology). The tissue diagnoses (e.g., specific types of lymphomas, carcinomas, normal tissues) were established by qualified pathologists based on morphological, clinical, and often other immunohistochemical markers, as is standard practice in surgical pathology. The reactivity of the LCA (CD45) antibody was then assessed against these established diagnoses.

8. Sample Size for the Training Set:

This product is an antibody reagent, not a machine learning algorithm. Therefore, there is no "training set" in the computational sense. The product's development and optimization involved empirical testing to ensure optimal predilution and performance with specific detection kits and automated staining systems. The "training" in this context would be the iterative development and formulation of the antibody cocktail and its associated staining protocols.

9. How the Ground Truth for the Training Set Was Established:

As there is no "training set" in the AI sense, this question is not applicable. The development of the antibody involved standard laboratory and histopathological methods to characterize its specificity and sensitivity, likely using well-characterized tissue samples from disease banks, with ground truth established by expert pathologists as described above.

{0}------------------------------------------------

14343

SEP 2 5 1997

510(k) SUMMARY

September 22, 1997 Date:

VENTANA MEDICAL SYSTEMS, INC. 3865 North Business Center Drive Tucson, Arizona 85705

Telephone: (520) 887-2155 (520) 887-2558 Facsimile:

  • Stephen A. Tillson, Ph.D. Contact: Vice President Scientific Affairs/ Quality Assurance
    Registration # : 2028492

ChemMate™ LCA (CD45) Trade Name:

Class II

FOR IN VITRO DIAGNOSTIC USE. Intended Use:

ChemMate™ LCA (CD45; Clones PD7/26/16 and B11) is intended for laboratory use to qualitatively aid in the identification by light microscopy of human cells of lymphoid origin, by recognizing Leukocyte Common Antigen (LCA), in normal and pathological paraffin embedded tissues processed in interpreted by a pathologist within the context of clinical data, gross and microscopic morphological data and multiple chemical and immunohistochemical stains.

This mouse monoclonal antibody has been optimally prediluted for use with the ChemMate™ SDK605 and SDK305 Secondary Detection -Peroxidase/DAB kits. Additionally, the prediluted ChemMate™ LCA (CD45) antibody as well as Peroxdase/DAB kits have been optimized for use with the TechMate™ for automated immunohistochemical staining.

{1}------------------------------------------------

510K SUMMARY OF SAFETY AND EFFECTIVENESS

SUMMARY AND EXPLANATION

The ChemMate™ LCA (CD45) is comprised of a mixture of two mouse monoclonal antibodies, clones PD7/26/16 and 2B11; each is of the IgG1 class of immunoglobulins.1 Please refer to Warnke, et. al.1 for a description of the cultivation, characterization and selection of these two clones. Each clone reacts with a different epitope of a family of membrane glycoproteins present on most human leukocytes known as the leukocyte common antigen (LCA). Tissues analyzed were either frozen or were embedded in paraffin and fixed with one of the following fixatives: formol saline, neutral buffered formalin, zinc sulfate [1%] formalin, B5, Bouin's, or acetic acid [2%] formol saline. 13 The LCA family consists of five to eight glycoproteins (molecular weight 180 to 220 kD)4-6.19 The clone PD7/26/16, included in the Fourth International Workshop on Human Leukocyte Differentiation Antigens (Vienna, 1989), was confirmed to react with an epitope expressed on three of the glycoproteins in the LCA family (MW 190, 205, and 220 kD) and has been designated as CD45RB.8 The clone 2B11, included in the Third International Workshop on Human Leukocyte Differentiation Antigens (Oxford, 1986), was confirmed to react with epitopes on four of the LCA glycoproteins (MW 180, 190, 205, and 220 kD).4 For additional information of the CD45 isoforms, see Knapp, et al.6

A unique feature of the ChemMate™ LCA (CD45) is it's ability to recognize epitopes on lymphocytes that survive the rigors of processing and fixation. The resultant advantage in the use of this antibody is its applicability with formalin-fixed specimens.1 Excessive exposure of tissue to formalin beyond what is required for adequate fixation will result in increased cross linking and loss of many reactive epitopes. For example, in fresh (frozen) tissue, cross linking is eliminated hence antigenic reactivity is well preserved. It is in such frozen specimens that reactivity with the clones PD7/26/16 and 2B11 is not only demonstrated among lymphoid cell populations but may also be rarely demonstrated in some myeloid populations.3

In paraffin-embedded, fixed specimens, the strongest labeling is seen on lymphoid cells, where it is most intense on the cell surface and less marked within the cytoplasm. Weaker membranous and/ or cytoplasmic staining is seen on macrophages and histiocytes, while other myeloid cells variably demonstrate even weaker staining or are frequently unreactive.17 ChemMate™ LCA (CD45) may be of value in assisting in the recognition of certain neoplasms, since neoplastic cells may, in a proportion of cases, fail to display the usual morphologic features of normal leukocytes in H & E sections, or may display atypical morphologic features. 810 Studies have also demonstrated LCA reactivity in many lymphomas and other hematopoietic neoplasms of Ivmphoid

{2}------------------------------------------------

type while neoplasms of epithelial, mesenchymal, or neural derivation remain unreactive. 137,11 The antibody is reactive with most, but not all neoplastic B and T cells in non-Hodgkin's lymphomas and leukemias of B and T cell type, malignant lymphoma of the thyroid, and hairy cells. 72-16 In a study of non-Hodgkin's lymphomas that compared genotypic analysis with immunophenotyping in paraffin-embedded tissue sections (fixed in B5 or 10% buffered formalin), LCA was positive in 42 of 44 cases (95%).24 Rare Reed-Sternberg cells and their variants have been reported to reveal membrane staining for LCA. although most Reed-Sternberg cells appeared negative or indeterminant due to the presence of adjacent strongly LCA positive reactive Ivmphoid cells. Among neoplastic diseases of non-lymphocvtic leukocvtic origin. (monocytic, myeloid and histiocytic neoplasms) staining is less frequent. less sensitive and more variable involving both a membranous and cytoplasmic pattern. A summary of immunoreactivity found in paraffinembedded cells/tissues and neoplasms (from references 1 and 7) may be found in Tables I-V of Performance Characteristics in this insert.

The antibody has been nonreactive with most carcinomas and tissues of epithelial, mesenchymal, and neural origin, as well as malignant melanomas, rhabdomyosarcomas and Ewing's sarcoma 17. These observations are also borne out in the immunoreactivity profiles of Table IV and V in this insert. It should however be noted that the interpretation of any positive LCA staining or its absence should be complemented by morphological studies and proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified individual.

Product Specific Limitations:

    1. In poorly fixed tissue specimens, nonspecific staining of non-lymphoid tissues may be observed. particularly epithelium and smooth muscle.
    1. Association between lymphocytes and secretory ductal epithelium in the production of immunoglobulin is well known. LCA staining may be exhibited in the epithelium of both fibroadenoma and fibrocystic disease of the breast.20 Although LCA reactivity has not been extensively studied in maliqnancies of the breast such as secretory carcinoma, signet ring cell lobular or ductal carcinoma, or for cystosarcoma phylloides, it is possible that LCA staining may also occur in some of these conditions. LCA reactions in breast tissues should not present differential difficulties in that histologic patterns in normal and abnormal breast tissues should be recognizable. Additionally, LCA is generally employed in a panel typically including cytokeratins and other markers (e.g., vimentin, S100).
    1. It has been reported that the sensitivity of clones PD7/26/16 and 2B11 is 90- . 93% for non-Hodgkin's lymphomas. False negative results have been seen

{3}------------------------------------------------

in acute lymphoblastic leukemia.722,24 It is recommended that a panel of antibodies be used to rule out lymphoid tumors.23

    1. Among neoplastic diseases of non-lymphocytic leukocytic origin, (monocytic, myeloid and histiocytic neoplasms) staining is less frequent, less sensitive and more variable involving both a membranous and cytoplasmic pattern.
    1. LCA positive lymphocytes closely apposed to carcinoma cells give the appearance of membrane reactivity to the latter cells'.
    1. To avoid false positive interpretations, cytoplasmic staining unaccompanied bv membrane staining should not be considered a positive result.

Performance Characteristics:

Reproducibility: ChemMate™ LCA (CD45), and Negative Control Reagent have been tested on serial sections of 271 tissue specimens (both normal and tumor specimens were included in the study; see Tables IB and V). Runs were performed a total of three times, with each run being performed on a different day. Consistent staining results were obtained.

Immunoreactivity: The following immunoreactivities have been demonstrated in paraffin-embedded tissues. The secondary staining systems were either ABC peroxidase or Indirect secondary antibody labeled peroxidase. The lists provided below (see Tables IA, II - IV) are not exhaustive but characterize the types of immunoreactivity reported in the literature for clones contained in the monoclonal antibody cocktail of ChemMate™ LCA (CD45). Tables IIB and V are summaries of the immunoreactivity studies performed by BioTek Solutions using ChemMate™ LCA (CD45) and the ChemMate™ Secondary Detection -Peroxidase/DAB Kit.

{4}------------------------------------------------

TABLE IA TABLE TA

REACTIVE AND NONREACTIVE NORMAL TISSUES/CELLS*,7,17,18,21

REACTIVE CELLSVARIABLY REACTIVECELLSNONREACTIVECELLS
Follicular centerlymphoid cellsInterfollicular transformedlymphoid cells immunoblasts)Polymorphonuclear leukocytes,rare myeloid precursors withweak staining
Follicular mantle zonesEpithelioid histiocytesErythroid cells
Interfollicular smalllymphocytesSinus histiocytesMegakaryocytes
Perisinusoidal cellsPlasma cellsTingible body macrophages withoccasional small cytoplasmicgranules
Splenic white pulpMonocytesMacrophagesInterdigitating reticulum cellswith rare, weak membrane• staining
Splenic red pulplymphoid cellsSplenic macrophages and sinuslining cells
Thymic lymphocytesThymic epithelial cells
Mast cellsEpithelium
OsteoclastLangerhans' cells, skinConnective tissue

Table IA Note: A total of 45 tissues were fixed in either 10% neutral buffered formalin, B5, or Zenker's acetic acid solution. Three specimens were fixed in multiple fixatives (B5, Zenker's, formalin, and Bouin's). Tissues were embedded in paraffin, and the secondary staining systems were either ABC peroxidase or Indirect secondary antibody labeled peroxidase. The tissues evaluated included:

bone marrow [6]dermatopathic lymphadenitis [4]
lymph nodes (nonspecific reactive hyperplasia) [10]normal skin [2]
prostate (severe chronic prostatitis) [1]
salivary gland (benign lymphoepithelial lesion) [1]spleen (hyperplastic changes) [5]
sarcoidosis [5]thymus [3]
talc granuloma [1]toxoplasmic lymphadenitis [4]
thyroid (Hashimoto's thyroiditis) [3]

{5}------------------------------------------------

TABLE IB REACTIVE AND NONREACTIVE NORMAL TISSUES/CELLS

REACTIVEVARIABLY REACTIVENONREACTIVE
CELLSCELLSCELLS
lymphocytesadrenal glandular epithelium
breast epithelium & stroma
neurons
glia
squamous epithelium
heart muscle
colonic epithelium
small intestinal epithelium
renal epithelium
hepatocytes / liver bile ducts
alveolar epithelium
striated muscle
ovarian stroma & epithelium
pancreatic epithelium
prostate epithelium
gastric epithelium
testicular & interstitial epithelium
thyroid epithelium
uterine epithelium, muscle & stroma

Table IB Note: In studies performed by BioTek Solutions, a total of 126 normal tissues were fixed in 10% neutral buffered formalin, embedded in paraffin, immunostained with ChemMate™LCA-CD45 and the ChemMate™ Secondary Detection - Peroxidase/DAB Kit. All tissues were stained using the TechMate™ Automated Staining System. The tissues evaluated included:

adrenal [3]skin [2]
brain [7]small intestine [2]
breast [50]spleen [3]
heart [3]squamous epithelial [2]
kidney [5]stomach [1]
large intestine [2]striated muscle [2]
liver [5]testis [5]
lung [5]thyroid [6]
ovary [5]tonsil [3]
pancreas [5]uterus [5]
prostate[5]

{6}------------------------------------------------

NO. REACTIVE/NO. TESTED%95% CONFIDENCEINTERVAL
A. TISSUE: B-CELL
LymphoplasmacyticFollicular center cell4/4100%39.8-100%
-small cleaved15/15100%78.2-100%
-large cleaved4/666.7%22-95%
-mixed small and large5/5100%48-100%
-small non-cleaved5/5100%48-100%
-large non-cleaved11/1291.7%62.5-99.7%
Immunoblastic sarcoma14/14100%76.8-100%
B. TISSUE: T-CELL
Mycosis fungoides1/1100%--
Immunoblastic sarcoma10/1190.9%58-99.5%
lymphoblastic2/450%9.0-91%
Unclassified3/3100%29.2-100%

TABLE II REACTIVE B- AND T- CELL NON-HODGKIN'S LYMPHOMA'

Table II Note: Confidence intervals were not calculated for those cases with a sample size of one. The reactivity however was included for anecdotal reference information.

A total of 80 specimens were fixed in either formalin (25 specimens), B5 (48 specimens), or Zenker's solution (2). Five specimens were fixed in multiple fixatives (B5, formalin, Zenker's, and Bouin's). The specimens were embedded in paraffin, and the secondary staining systems were either ABC peroxidase or indirect secondary antibody labeled peroxidase.

{7}------------------------------------------------

TABLE III OTHER HEMATOPOIETIC NEOPLASMS '

TISSUE /RELEVANT CELL TYPENO. REACTIVE/NO. TESTED%95% CONFIDENCEINTERVAL
Chronic Lymphocytic Leukemia
-B-cell Type: lymphoid cells3/3100%29.2-100%
-T-cell Type: lymphoid cells3/3100%29.2-100%
Acute Lymphoblastic Leukemia:
-lymphoblasts3/837.5%8.5-75.5%
Prolymphocytic (B-cell) Leukemia:
-prolymphocytes1/1100%--
Hairy Cell Leukemia:
-Hairy Cells5/5100%48-100%
Acute Myelogenous Leukemia:
- myeloid cells0/70%0-41%
Chronic Myelogenous Leukemia:
- myeloblasts0/20%0-84.7%
Erythroleukemia:
-erythroblasts0/20%0-84.7%
Monocytic Leukemia:
-monoblasts1/1100%--
Polycythemia Vera:
-erythroid cells0/20%0-84.7%
Extramedullary Hematopoiesis:
-myeloid cells0/30%0-70.8%
-megakaryocytes0/30%0-70.8%
Multiple Myeloma:
-plasma cells10/10100%68.5-100%
Plasma Cell Leukemia:
-plasma cells0/10%--
Systemic Mast Cell Disease:
-mast cells3/3100%29.2-100%
TISSUENO. NONREACTIVE/NO. TESTED%95% CONFIDENCEINTERVAL
Squamous cell carcinoma4/4100%39.8-100%
Nasopharyngeal squamous cell carcinoma1/1100%--
Adenocarcinoma54/54100%93-100%
Ovarian serous cystadenocarcinoma1/1100%--
Transitional cell carcinoma2/2100%15.8-100%
Bladder-invasive transitional cell carcinoma1/1100%--
Metastatic carcinoma5/5100%48-100%
Basal-cell carcinoma1/1100%--
Breast-infiltrating lobular carcinoma2/2100%15.8-100%
Small cell anaplastic lung carcinoma10/10100%68.5-100%
Breast-infiltrating ductal carcinoma2/2100%15.8-100%
Pulmonary large-cell undiff. carcinoma1/1100%--
Pulmonary small-cell undiff. carcinoma2/2100%15.8-100%
Nasopharyngeal undiff. carcinoma7/7100%59-100%
Laryngeal undiff. carcinoma1/1100%--
Medullary carcinoma, thyroid1/1100%--
Thymoma (epithelial)4/4100%39.8-100%
Amelanocytic malignant melanoma5/5100%48-100%
Malignant melanoma3/3100%29.2-100%
Alveolar rhabdomyosarcoma3/3100%29.2-100%
Ewing's sarcoma9/9100%66-100%
Neuroblastoma2/2100%15.8-100%
Soft-tissue fibrosarcoma1/1100%--
Soft-tissue malignant fibrous histiocytoma4/4100%39.8-100%
Eosinophilic granuloma1/1100%--
Granular cell tumor1/1100%--
Adrenal Pheochromocytoma2/2100%15.8-100%
Adrenal neuroblastoma3/3100%29.2-100%
Glioblastoma1/1100%--
Neurofibroma/neuroma6/6100%54.1-100%
Oligodendroglioma1/1100%--
Cerebellar astrocytoma1/1100%--
Meningioma1/1100%--
Chordoma1/1100%--
Skeletal giant cell tumor1/1100%--
Carcinoid, lung1/1100%--
Ileal carcinoid tumor2/2100%15.8-100%
Colonic carcinoid carcinoma1/1100%--
Wilm's tumor1/1100%--
Germ cell tumor7/7100%59-100%
Meduloblastoma1/1100%--
Ependymoma1/1100%--
Testicular embryonal carcinoma1/1100%--
Testicular spermatocytic seminoma1/1100%--
Testicular seminoma1/1100%--

Table III Note: Confidence intervals were not calculated for those cases with a sample size of one. The reactivity however was included for anecdotal reference information.

A total of 51 specimens were fixed in either formalin (2 spleens, 1 lymph node, and 1 skin specimen), B5 (9 spleens, 5 lymph nodes, and 1 soft tissue mass), or Zenker's acetic acid solution (32 bone marrow specimens). The specimens were embedded in paraffin, and the secondary staining systems were either ABC peroxidase or indirect secondary antibody labeled peroxidase.

{8}------------------------------------------------

TABLE IV TABEE TV

NONREACTIVE NON-LYMPHOID PATHOLOGICAL TISSUES 17

Table IV Note: Confidence intervals were not calculated for those cases with a sample size of one. The reactivity however was included for anecdotal reference information.

{9}------------------------------------------------

A total of 162 tissues were fixed in either formol saline, neutral buffered formalin, 10% neutral buffered formalin. B5. Zenker's acetic acid solution, acetic acid (2%) formol saline, or Bouin's. Six specimens were fixed in multiple fixatives (B5, Zenker's, formalin, and Bouin's). Tissues were embedded in paraffin, and the secondary staining systems were either ABC peroxidase or indirect secondary antibody labeled peroxidase.

TABLE V REACTIVE AND NONREACTIVE PATHOLOGICAL TISSUES/CELLS

REACTIVEVARIABLY REACTIVENONREACTIVE
CELLSCELLSCELLS
lymphoma cellsbreast epithelium & stroma
carcinoid epithelium
colonic carcinoma
gastric carcinoma
hepatocellular carcinoma
lung carcinoma
ovarian carcinoma
prostatic carcinoma
thyroid carcinoma
endodermal sinus carcinoma
melanocytes
placental trophoblasts
Ewing's sarcoma
fibrosarcoma
seminoma seminiferous epithelium
teratoma epithelia & mesenchyme

Table V Note: In studies performed by BioTek Solutions, a total of 145 pathological tissues were fixed in 10% neutral buffered formalin, embedded in paraffin, immunostained with ChemMate™ LCA-CD45 and the ChemMate™ Secondary Detection - Peroxidase/DAB Kit. All tissues were stained using the TechMate™ Automated Staining System. The tissues evaluated included:

breast carcinoma [65] lymphoma [45] carcinoid [4] colonic carcinoma [3] gastric carcinoma [2] hepatocellular carcinoma [3] lung carcinoma [2] ovarian carcinoma [1] prostatic carcinoma [6]

ﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬﻬ

thyroid carcinoma [1] endodermal sinus carcinoma [1] melanoma [6] placental tumor [1] sarcoma, Ewing's [1] sarcoma, fibro [1] seminoma [2] teratoma [1]

{10}------------------------------------------------

References:

    1. Warnke RA, et al. Diagnosis of human lymphoma with monoclonal antileukocyte antibodies. N Enq J Med 1983; 309:1275
    1. Pulido R, et al. Comparative biochemical and tissue distribution study of four distinct CD45 antigen specificities. J Immun 1988; 140:3851
    1. Warnke RA, Rouse RV. Limitations encountered in the application of tissue section immunodiagnosis to the study of lymphomas and related disorders. Human Path 1985; 16:326
    1. Cobbold S, et al. Non-lineage, LFA-1 family and leuycocyte common antigens: new and previously defined clusters. In: McMichael AJ, et al., Eds. Leukocyte Typing III. White cell differentiation antigens. Oxford University Press, New York; 1987; 788
    1. Knapp W, et al., eds. Leukocyte Typing IV. White cell differentiation antigens. Oxford University Press., 1989
    1. Streuli M, Morimoto C, Schrieber M, Schlossman SF, and Saito H. Characterization of CD45 and CD45R monoclonal antibodies using transfected mouse cell lines that express individual human leukocyte common antigens. J Immunology 1988; 141:3910-3914
    1. Kurtin PJ, Pinkus GS. Leukocyte common antigen-a diagnosis discriminant between hematopoietic and nonhematopoietic neoplasms in paraffin sections using monoclonal antibodies. Human Path 1985;16:353
    1. Taylor CR. Kledzik G. Immunohistologic techniques in surgical pathologya spectrum of "new" special stains. Hum Pathol 1981; 12:590
    1. Rosai J. Borderline epithelial lesions of the breast. Am J Surg Pathol. 1991; 15:209-221
    1. Van Ginneken AM. Van der Lei J. Understanding differential diagnostic disagreement in pathology. Proc Annu Symp Comput Appl Med Care. 1991: 99-103
    1. Gatter KC, et al. The differential diagnosis of routinely processed anaplastic tumors using monoclonal antibodies. Am J Clin Path 1984;82-83
    1. Meis JM. et al. A comparative marker study of large cell lymphoma. Hodqkin's disease, and true histiocytic lymphoma in paraffin-embedded tissue. Am J Clin Pathol 1986:86-591
    1. Hall PA, et al. Paraffin section immunohistochemistry. I. Non-Hodgkin's Ivmphoma. Histopathology 1988:13:149
    1. Faure P, et al. Diagnostic features of primary malignant lymphomas of the thyroid with monoclonal antibodies. Cancer 1988; 61:1852
    1. Shvero J, et al. Anaplastic thyroid carcinoma.. A clinical, histologic, and immunohistochemical study. Cancer 1988; 62:319
    1. Michie SA, et al. A panel approach to the evaluation of the sensitivity and specificity of antibodies for the diagnosis of routinely processed histologically undifferentiated human neoplasms. Am J Clin Pathol 1987: 88:457

{11}------------------------------------------------

    1. Hall PA, et al. Demonstration of lymphoid antigens in decalcified bone marrow trephines. J Clin Pathol 1987; 40:870
    1. Athanasou NA. et al. Leucocyte common antigen is present on osteoclasts. J. Pathol 1987; 153:121
    1. Kiernan JA. Histological and histochemical Methods: Theory and Practice. New York: Pergamon Press 1981
    1. Herman. GE. Elfont E. Aberrant CD45 (Leukocyte Common Antigen) Staining of Non-Malignant Breast Lesions in Zinc Formalin Fixed Tissue. J Histotechnology 1993; 16(2):151
    1. Athanasou NA, Puddle B, Quinn J, Woods CG. Use of Monoclonal Antibodies to Recognise Osteoclasts in Routinely Processed Bone Biopsy Specimens. J Clin Pathol 1991; 44:664
    1. Wieczorek R. Buck D. Bindl. J and Knowles. DM. Monoclonal Antibody Leu-22 (L60) Permits the Demonstration of Some Neoplastic T Cells in Routinely Fixed and Paraffin-Embedded Tissue Sections. Hum Pathol 1988; 19:1434
    1. Cartun, RW, Coles FB, and Pastuszak WT. Utilization of Monoclonal Antibody L26 in the Identification and Confirmation of B-Cell Lymphomas. Am J Pathol 1987; 129:415
    1. Elghetany MT, Kurec AS, Schuehler K, Forbes BA, Duggan DB, and Davey FR. Immunophenotyping of Non-Hodgkin's Lymphomas in Paraffin-Embedded Tissue Sections. A Comparison with Genotypic Analysis. Am J Clin Pathol 1991; 95:517

{12}------------------------------------------------

Image /page/12/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines representing its body and wings. The eagle's head is facing left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Stephen A. Tillson, Ph.D. Vice President Scientific Affairs/Quality Assurance Ventana Medical Systems, Inc. 3865 North Business Center Drive Tucson, Arizona 85705

K973393 Re: Trade Name: ChemMate™ LCA (CD45) Regulatory Class: II Product Code: DEH Dated: June 27, 1997 Received: July 1, 1997

SEP 2 5 1997

Dear Dr. Tillson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. this response to your premarket notification Please note: submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{13}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

Page_1 of 1

510(k) Number (if known):K 973393

Device Namel LCA(leukocyte common antigen) ChemMate: includes ChemMate secondary detection kits peroxidase DAB SDK 605 & 305;ChemMate Negative Control reagent; ChemMate standard secondary buffer kit

Indications For Use:

For the LCA primary antibodyreagent:

To qualitatively aid in the identification by light microscopy of human cells of lymphoid orgin, by recognizing Leukocyte Common Antigen (LCA) in normal and pathologic paraffin embedded tissues processed in neutral buffered formalin, B5, or Bouin's fixative.

For the Ancillary Reagents:

To aid in the performance of immunohistochemical tests

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluat

Petur E. Massi

(Division Sign-Off) Division of Clinical Laborator 510(k) Number _

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter. Use_

(Optional Format 1-2-96)

O

§ 866.5550 Immunoglobulin (light chain specific) immunological test system.

(a)
Identification. An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.(b)
Classification. Class II (performance standards).