K Number
K972964

Validate with FDA (Live)

Date Cleared
1997-10-10

(60 days)

Product Code
Regulation Number
876.5520
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The inflation device is intended for single use while performing gastrointestinal (GI), gastrouterine (GU) and lacrimal duct balloon catheter dilation procedures to inflate the balloon, monitor the pressure within the balloon and deflate the balloon.

Device Description

The Atrion Medical Balloon Catheter Inflation Device consists of a plastic syringe with a screw-type plunger and a locking lever and rotating palm grip that control the plunger. a manometer to measure pressure and a connecting tube. The Device has a fluid capacity of 20 cc and an operating pressure range of vacuum to between 10 and 20 atm, depending on the manometer attached.

AI/ML Overview

The provided text describes a medical device, the Atrion Medical Balloon Catheter Inflation Device, and its clearance through the 510(k) pathway. However, it does not contain information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or multi-reader multi-case studies.

This document is a "Summary of Safety and Effectiveness" and a subsequent FDA Letter, both primarily focused on establishing substantial equivalence to a predicate device (Ryder International Urological Balloon Catheter Inflation Device). The safety testing mentioned refers to material compatibility with contrast solution, not the functional performance of the inflation device itself in a clinical or simulated clinical setting.

Therefore, I cannot populate the requested table or answer most of the questions using the provided text.

Here's what can be extracted based on the limitations:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Not specified in the provided text.The document states the device is "safe for its intended use" based on material compatibility testing. No specific performance metrics (e.g., pressure accuracy, inflation time, deflation time, leakage rates) are provided.

2. Sample size used for the test set and the data provenance:
Not provided in the document. The "safety testing" mentioned refers only to material tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. No clinical or performance testing requiring expert ground truth is described.

4. Adjudication method for the test set:
Not applicable. No clinical or performance testing requiring adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a mechanical inflation device, not an AI-assisted diagnostic tool, and no MRMC studies are mentioned.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a mechanical device, not an algorithm.

7. The type of ground truth used:
For the "Summary of Safety Testing," the ground truth used was compliance with "USP guidelines" for material testing with contrast solution. No clinical ground truth is mentioned.

8. The sample size for the training set:
Not applicable. This is a medical device, not an AI model, and no training set is mentioned.

9. How the ground truth for the training set was established:
Not applicable. This is a medical device, not an AI model.

In summary: The provided FDA submission documents focus on comparing the Atrion Medical Balloon Catheter Inflation Device to a predicate device for substantial equivalence, with an emphasis on technological characteristics and material safety. They do not contain the type of performance study data, acceptance criteria, or ground truth information typical for diagnostic or AI-driven medical devices.

{0}------------------------------------------------

Atrion Medical Products, Inc. 1426 Curt Francis Road Post Office Box 564 Arab, AL 35016 Tel 205 586 1580 Fax 205 586 8529

K972964

OCT 10 1997

Atrion

SUMMARY OF SAFETY AND EFFECTIVENESS 12.

Date of Preparation:August 7, 1997
Device Name:Atrion Medical Balloon Catheter Inflation Device
Classification Name:Balloon Inflation/Deflation Device
Manufacturer:Atrion Medical Products, Inc.PO Box 564, 1426 Curt Francis RoadArab, AL 35016
Contact:Mr. Dan Clark, Atrion Medical Products, Inc.PO Box 564, 1426 Curt Francis RoadArab, AL 35016Telephone: (205) 586-1580, Fax: (205) 586-5553
Predicate:Ryder International Urological Balloon Catheter (Bard®Eagle™) Inflation Device cleared under K962611

Device Description/Intended Use:

The Atrion Medical Balloon Catheter Inflation Device consists of a plastic syringe with a screw-type plunger and a locking lever and rotating palm grip that control the plunger. a manometer to measure pressure and a connecting tube.

The Device has a fluid capacity of 20 cc and an operating pressure range of vacuum to between 10 and 20 atm, depending on the manometer attached.

The inflation device is intended for single use while performing balloon dilation procedures to inflate the balloon, monitor the pressure within the balloon and deflate the balloon.

Technological Characteristics:

The Atrion Medical Balloon Catheter Inflation Device has an operating pressure range of vacuum to between 10 and 20 atm, depending on the manometer attached, while the predicate device has a range of vacuum to 20 atm. The Atrion Medical Balloon Catheter Inflation Device has no stopcock, while the predicate device does. There are no other significant technological characteristics that distinguish the two devices, and no differences that should pose a risk to patient safety.

Summary of Safety Testing:

The materials of the device which contact the contrast solution in use have been tested using USP guidelines and the results of these studies indicate that the product is safe for its intended use.

Image /page/0/Picture/12 description: The image shows a sequence of numbers. The numbers are 000040. The numbers are black and the background is white.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure or a series of flowing lines, possibly representing people or services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 0 1997

Mr. Dan Clark Vice President, Regulatory and Ouality Atrion Medical Products, Inc. 1426 Curt Francis Road P.O. Box 564 Arab, Alabama 35016

Re: K972964

Atrion Medical Balloon Catheter Inflation Device Dated: August 7. 1997 Received: August 11, 1997 Regulatory class: II and I 21 CFR 8876.5520/Product code: 78 KOE 21 CFR §886.4350/Product code: 86 HNW

Dear Mr. Clark:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

William Yu

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known): K972964_____________________________________________________________________________________________________________________________________________

Device Name: Atrion Medical Balloon Catheter Inflation Device

Indications For Use:

The inflation device is intended for single use while performing gastrointestinal (GI), gastrouterine (GU) and lacrimal duct balloon catheter dilation procedures to inflate the balloon, monitor the pressure within the balloon and deflate the balloon.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------
Prescription Use(Per 21 CFR 801.109)OR Over-The-Counter Use
-------------------------------------------------------------------

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) NumberK972964(Optional Format-1-2-96)
--------------------------------------------------

§ 876.5520 Urethral dilator.

(a)
Identification. A urethral dilator is a device that consists of a slender hollow or solid instrument made of metal, plastic, or other suitable material in a cylindrical form and in a range of sizes and flexibilities. The device may include a mechanism to expand the portion of the device in the urethra and indicate the degree of expansion on a dial. It is used to dilate the urethra. This generic type of device includes the mechanical urethral dilator, urological bougies, metal or plastic urethral sound, urethrometer, filiform, and filiform follower.(b)
Classification. (1) Class II (special controls). Except when it is a mechanical urethral dilator, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the urethrometer, urological bougie, filiform and filiform follower, and metal or plastic urethral sound. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.